AQUACEL AG EXTRA HYDROFIBER DRESSING WITH SILVER AND STRENGTHENING FIBER
K121275 · Convatec, Inc. · FRO · Jul 25, 2012 · SU
Device Facts
| Record ID | K121275 |
| Device Name | AQUACEL AG EXTRA HYDROFIBER DRESSING WITH SILVER AND STRENGTHENING FIBER |
| Applicant | Convatec, Inc. |
| Product Code | FRO · SU |
| Decision Date | Jul 25, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Intended Use
AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber may be used for: - Abrasions - - Lacerations - - Minor cuts - Minor scalds and burns Under the supervision of a healthcare professional: AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber may be used for the management of: - Wounds as an effective barrier to bacterial penetration of the dressing as this may help reduce infection; - Partial thickness (second degree) burns; - Diabetic foot ulcers, leg ulcers, (venous stasis ulcers, arterial ulcers and leg ulcers of mixed etiology) and pressure ulcers/sores (partial & full thickness); - Surgical wounds left to heal by secondary intention such as dehisced surgical incisions; - Surgical wounds that heal by primary intent such as dermatological and surgical incisions (e.g. orthopedic and vascular); - Traumatic wounds; - Wounds that are prone to bleeding, such as wounds that have been mechanically or surgically debrided and donor sites; - Oncology wounds with exudate, such as fungoides-cutaneous tumors, fungating carcinoma, cutaneous metastasis, Kaposi's sarcoma, and angiosarcoma; - Painful wounds; - Infected wounds
Device Story
AQUACEL™ Ag EXTRA™ is a sterile, two-layer non-woven wound dressing; composed of Hydrofiber™ material and ionic silver; stitchbonded with regenerated cellulose fibers for increased absorbency and tensile strength. Dressing absorbs wound exudate; transforms into a soft gel; maintains moist wound environment; provides barrier to bacterial penetration. Used in clinical or home settings; applied by healthcare professionals or patients (for minor wounds). Gel formation supports healing; silver content provides antimicrobial barrier to reduce infection risk. Output is physical wound coverage and exudate management.
Clinical Evidence
No clinical data provided for this specific submission. Safety and effectiveness are supported by literature and clinical data previously provided for the predicate device (K080383).
Technological Characteristics
Two-layer non-woven dressing; materials include Hydrofiber™ (carboxymethylcellulose) and ionic silver; reinforced with regenerated cellulose fibers via stitchbonding. Sterile; conformable; high-absorbency gel-forming matrix. No electronic components or software.
Indications for Use
Indicated for management of minor wounds (abrasions, lacerations, cuts, scalds, burns) for OTC use. Under professional supervision, indicated for management of partial/full thickness wounds including diabetic foot, venous/arterial/mixed leg ulcers, pressure ulcers, surgical wounds (primary/secondary intention), traumatic wounds, bleeding wounds, oncology wounds, and infected wounds.
Predicate Devices
- AQUACEL™ Ag Hydrofiber™ Dressing (K080383)
Related Devices
- K221720 — Extra Silver Gelling Fiber Dressing · Winner Medical Co., Ltd. · Dec 21, 2022
- K080383 — AQUACEL HYDROFIBER WOUND DRESSING AND AG HYDROFIBER DRESSING · Convatec, A Division of E.R. Squibb & Sons, LLC · May 2, 2008
- K252167 — Enhanced Silver Gelling Fiber Dressing (OTC) Extra Enhanced Silver Gelling Fiber Dressing (Prescription) · Qingdao Bright Moon Biomedical Materials Co., Ltd. · Dec 19, 2025
- K210466 — Silver Gelling Fiber Dressing · Winner Medical Co., Ltd. · Jan 14, 2022
- K240485 — LUOFUCON® Extra Silver Gelling Fiber Dressing Plus (Prescription use),LUOFUCON® Silver Antimicrobial Gelling Fiber Dressing Plus (OTC use) · Huizhou Foryou Medical Devices Co., Ltd. · Oct 7, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
K121275
page 1/2
510(k) Premarket Notification AQUACEL™ Ag Extra™ Hydrofiber™ Dressing with Silver and Strengthening Fiber
April 25, 2012
JUL 2 5 2012
## SECTION 5: 510(K) SUMMARY
Device: AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber Applicant: ConvaTec Inc. Contact: Katrina Fiedler Associate Director, US Regulatory Affairs 908-904-2541 Fax: 908-904-2235 Email: katrina.fiedler@convatec.com Date: April 25, 2012 Trade Name: AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber Classification Name: Dressing, Wound, Drug Device Class: Unclassified Product Code: FRO
Predicate Device: AQUACEL™ Ag Hydrofiber™ Dressing, K080383
AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber is a one piece wound dressing comprised of two layers of soft, sterile non-woven material. The non-woven pads are comprised of Hydrofiber™ dressing and ionic silver stitchbonded together with regenerated cellulose fibers and designed to provide additional absorbency and tensile strength properties. This conformable and highly absorbent dressing absorbs wound fluids, creating a soft gel which maintains a moist environment and supports the body's healing process.
AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber is indicated for the management of wounds and can be used over-the-counter for minor wounds such as abrasions, lacerations, minor cuts, minor scalds and burns. Under the supervision of a healthcare professional, AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber may be used for the management of wounds as an effective barrier to bacterial penetration of the dressing as this may help reduce infection; partial thickness (second degree) burns; diabetic foot ulcers, (venous stasis
ConvaTec Inc.
14 of 435
{1}------------------------------------------------
K121275
ye 2/2
510(k) Premarket Notification AQUACEL™ Ag Extra™ Hydrofiber™ Dressing with Silver and Strengthening Fiber
April 25, 2012
ulcers, arterial ulcers and leg ulcers of mixed etiology) and pressure ulcers/sores (partial & full thickness); surgical wounds left to heal by secondary intention such as dehisced surgical incisions; surgical wounds that heal by primary intent such as dermatological and surgical incisions (e.g. orthopedic and vascular); traumatic wounds; wounds that are prone to bleeding, such as wounds that have been mechanically or surgically debrided and donor sites; oncology wounds with exudate, such as fungoids-cutaneous tumors, fungating carcinoma, cutaneous metastasis, Kaposi's sarcoma, and angiosarcoma; painful wounds and infected wounds.
Since AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber is largely based on the AQUACEL™ Ag Hydrofiber™ technology, the safety and effectiveness of AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber has been demonstrated by the literature and clinical data provided in previous 510(k)s (i.e., K080383). In summary, a careful and thorough review of the literature suggests that Hydrofiber™ dressings have been used safely and effectively in clinical trials for the management of wounds.
Thus we believe that, AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber is substantially equivalent to the previously cleared Hydrofiber™technology based products (reference K080383) and that AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber can be used safely and effectively for the management of wounds.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 25 2012
Convatec. Incorporated % Ms. Katrina Fiedler Associate Director, US Regulatory Affairs 200 Headquarters Park Drive Skillman, New Jersey 08558
Re: K121275
Trade/Device Name: AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber Regulatory Class: Unclassified Product Code: FRO Dated: April 25, 2012 Received: April 27, 2012
Dear Ms. Fiedler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Ms. Katrina Fiedler
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Milker
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Premarket Notification AQUACEL™ Ag Extra™ Hydrofiber™ Dressing with Silver and Strengthening Fiber
### SECTION 4: INDICATIONS FOR USE STATEMENT
510(k) Number (if known): Not Known
Device Name: AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber
For Over-the-Counter Use:
AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber may be used for:
- Abrasions -
- Lacerations -
- Minor cuts
- Minor scalds and burns
Under the supervision of a healthcare professional:
AQUACEL™ Ag EXTRA™ Hydrofiber™ Dressing with Silver and Strengthening Fiber may be used for the management of:
- Wounds as an effective barrier to bacterial penetration of the dressing as this may help reduce infection;
- Partial thickness (second degree) burns; l
AND/OR
- Diabetic foot ulcers, leg ulcers, (venous stasis ulcers, arterial ulcers and leg . ulcers of mixed etiology) and pressure ulcers/sores (partial & full thickness);
- Surgical wounds left to heal by secondary intention such as dehisced surgical incisions;
- Surgical wounds that heal by primary intent such as dermatological and surgical = incisions (e.g. orthopedic and vascular);
- Traumatic wounds; l
- Wounds that are prone to bleeding, such as wounds that have been mechanically or surgically debrided and donor sites;
- Oncology wounds with exudate, such as fungoides-cutaneous tumors, fungating carcinoma, cutaneous metastasis, Kaposi's sarcoma, and angiosarcoma;
- Painful wounds;
- Infected wounds
x
Prescription Use
Over-The-Counter Use _
(Part 21 CFR 801 Subpart D)
#### (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kane for MM
Division Sign Off
ConvaTec Inc.
al. Orthopedic. and Restorative Devi
13 of 435
510(k) Number K12(275