Under the supervision of a healthcare professional, Dash-Topic Plus Cream is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including atopic dermatitis, allergic contact dermatitis and radiation dermatitis. Dash-Topic Plus Cream also helps to relieve dry, waxy skin by maintaining a moist wound & skin environment, which is beneficial to the healing process.
Device Story
Dash-Topic Plus Cream is a non-sterile, topical, oil-in-water emulsion; functions as a physical barrier to maintain a moist wound and skin environment; aids in managing symptoms of dermatoses (burning, itching, pain) and dry skin. Applied topically by patient under healthcare professional supervision. Provides moisture and protective barrier; supports healing process. No complex electronics or software.
Clinical Evidence
Bench testing only. Biocompatibility testing included Agar Diffusion Cytotoxicity (ISO 10993-5:2009), Direct Primary Skin Irritation (ISO 10993-10:2010), and Kligman Maximization Sensitization (ISO 10993-10:2010). Results showed non-cytotoxic, negligible irritation, and non-sensitizing. Bench performance testing included appearance, pH, preservative content, viscosity, and package integrity. Preservative effectiveness and microbial testing performed per USP<51>, USP<61>, and USP<62>. In-use stability testing confirmed shelf-life.
Technological Characteristics
Oil-in-water emulsion; contains humectants and emollients. Non-sterile cream. Biocompatibility per ISO 10993-5 and ISO 10993-10. Stability testing per USP<51>, USP<61>, and USP<62>.
Indications for Use
Indicated for patients suffering from dermatoses, including atopic dermatitis, allergic contact dermatitis, and radiation dermatitis, requiring management of burning, itching, and pain, or relief of dry, waxy skin. Use under healthcare professional supervision.
K123678 — HYALMIX CREAM · Igi Laboratories, Inc. · Dec 16, 2013
K132329 — NIVATOPIC PLUS · Nivagen Pharmaceuticals, Inc. · Dec 20, 2013
K082089 — PRUMYX CREAM · Prugen, Inc. · Jan 13, 2009
K190342 — BIAFINE Topical Cream · Bausch Health Americas Incorporated · Oct 18, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized depiction of human profiles facing right, with three distinct faces layered on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2016
Founders Science Group, LLC Ronald M Gurge, Ph.D. Chief Science Officer 400 Highland Corporate Drive Cumberland, RI 02864
Re: K150914
Trade/Device Name: Dash-Topic Plus Cream Regulatory Class: Unclassified Product Code: FRO Dated: March 18, 2016 Received: March 21, 2016
Dear Dr. Gurge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150914
Device Name Dash-Topic Plus Cream
### Indications for Use (Describe)
Under the supervision of a healthcare professional, Dash-Topic Plus Cream is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including atopic dermatitis, allergic contact dermatitis and radiation dermatitis. Dash-Topic Plus Cream also helps to relieve dry, waxy skin by maintaining a moist wound & skin environment, which is beneficial to the healing process.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size:16px;">☑</span> </div> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="font-size:16px;">☐</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### I. Submitter
Puddingstone Pharma Regulatory Affairs 400 Highland Corporate Drive Cumberland, RI 02864 Contact Person: Robert Munroe Email: puddingstone@founderssg.com Date Prepared: March 31, 2015
#### II. Device Name
| Name of Device: | Dash-Topic Plus Cream |
|-----------------------|---------------------------|
| Common or Usual Name: | Dressing, Wound, Drug |
| Classification Name: | Dressing, Wound, Drug |
| Panel: | General & Plastic Surgery |
| CFR Number: | Unclassified |
| Product code: | FRO |
#### III. Predicate Device
HylatopicPlus Cream cleared under 510(k) K110727, from PreCision Dermatology Inc.
#### IV. Device Description
Dash-Topic Plus Cream is a non-sterile, off-white, low odor, fragrance free, topical product. Dash-Topic Plus Cream forms a physical barrier that helps to maintain a moist wound and skin environment. Dash-Topic Plus Cream is a prescription device.
#### V. Indications for Use
Under the supervision of a healthcare professional, Dash-Topic Plus Cream is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including atopic dermatitis, allergic contact dermatitis and radiation dermatitis. Dash-Topic Plus Cream also helps to relieve dry, waxy skin by maintaining a moist wound and skin environment, which is beneficial to the healing process.
#### VI. Comparison of Technological Characteristics with the Predicate Device
Both the proposed and predicate device are oil-in-water emulsions containing humectant and emollient components which donate moisture to the skin, and form a semi-permeable protective barrier. Both products are non-sterile creams containing similar preservative systems, and are used topically to relieve symptoms of various dermatoses.
{4}------------------------------------------------
#### VII. Substantial Equivalence
Dash-Topic Plus Cream is similar in function and intended use to HylatopicPlus Cream manufactured by PreCision Dermatology Inc., and includes identical components, indicated uses and operating principles.
#### VIII. Performance Data
Non-clinical testing was conducted to confirm the safe and effective performance of Dash-Topic Plus Cream. Agar Diffusion Cytotoxicity (ISO 10993-5: 2009). Direct Primary Skin Irritation (ISO 10993-10:2010) and Kligman Maximization Sensitization (ISO 10993-10:2010) Tests were performed on the proposed device. The results of the biocompatibility tests demonstrate that Dash-Topic Plus Cream is non-cytotoxic, a negligible irritant and non-sensitizing. For release and shelf life stability studies, the device product had undergone bench performance testing, the following parameters were monitored: Appearance, pH, preservative content, viscosity and package integrity. The device product had undergone Preservative Effectiveness Testing as per USP<51> (Antimicrobial Effectiveness Testing) and USP<61> <62> (Microbial Enumeration Tests and Tests for Specified Microorganisms). In-use stability testing was conducted to determine the shelf-life of the opened 450g container, the following parameters were monitored during the study: Appearance, pH, preservative content, viscosity, package integrity, USP<51> and USP<61> <62>.
#### IX. Conclusion
The bench performance data confirms the physical characteristics, stability and shelf-life of the proposed device product. The biocompatibility testing data confirms the safety of the proposed device product. Dash-Topic Plus Cream has identical indicated uses, operating principles and component composition when compared to the predicate. Therefore, Dash-Topic Plus Cream is substantially equivalent to HylatopicPlus Cream, cleared under 510(k) K110727.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.