anti-bacterial bandage

K200821 · Planet (Suzhou) Medical Products Co., Ltd. · FRO · Oct 19, 2020 · SU

Device Facts

Record IDK200821
Device Nameanti-bacterial bandage
ApplicantPlanet (Suzhou) Medical Products Co., Ltd.
Product CodeFRO · SU
Decision DateOct 19, 2020
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

Anti-bacterial bandages are to be applied topically to the skin for the management of minor cuts,minor scrapes.

Device Story

Anti-bacterial bandage; consists of fabric or polyethylene tape with an absorbent pad containing 0.1% or 0.8% benzalkonium chloride. Applied topically to skin by patient or caregiver for management of minor cuts and scrapes. Benzalkonium chloride reduces bacterial colonization within the dressing to help prevent infection. Device is EO sterilized, single-use, and intended for OTC use.

Clinical Evidence

Bench testing only. Testing included biocompatibility (cytotoxicity, intracutaneous reactivity, acute systemic toxicity, sensitization, material-mediated pyrogenicity, endotoxin), sterilization validation (ISO 11135:2014), shelf-life (ASTM F1980-16), storage/transport (ASTM 4169-16), and antibacterial efficacy (AATCC 100:2012).

Technological Characteristics

Materials: Fabric tape or polyethylene tape, absorbent pad with 0.1%/0.8% benzalkonium chloride. Sterilization: Ethylene oxide (ISO 11135:2014). Biocompatibility: ISO 10993-1, -5, -10, -11. Shelf life: 3 years (ASTM F1980-16). Packaging: Medical packing paper. Single-use.

Indications for Use

Indicated for topical application to the skin for the management of minor cuts and minor scrapes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, which features a stylized depiction of human figures. To the right of this symbol, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is in a larger, bolder font, while "U.S. FOOD & DRUG ADMINISTRATION" is in a smaller font below it. October 19, 2020 Planet (Suzhou) Medical Products Co., Ltd % Van Lee Senior Consultant LinkHope medical technology Inc No. 231 science avenue Huangpu district, Guangzhou Guangzhou, GuangDong 510700 China Re: K200821 Trade/Device Name: Anti-bacterial bandage Regulatory Class: Unclassified Product Code: FRO Dated: March 23, 2020 Received: March 30, 2020 Dear Van Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Kimberly Ferlin, Ph.D. For Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200821 Device Name Anti-bacterial Bandage Indications for Use (Describe) Anti-bacterial bandages are to be applied topically to the skin for the management of minor cuts,minor scrapes. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### 1. Submission Sponsor Planet (Suzhou) Medical Products Co., Ltd No.33 Qinglian Rd Xuguan Industrial Park, Hi-tech Development Zone, Suzhou, Jiangsu, China 215151 Contact: Frank Zhou Contact title: General Manager #### 2. Submission Correspondent LinkHope medical technology Inc. No. 231 science avenue Huangpu district, Guangzhou, China Phone: 86 -13632273583 Contact: Van Lee Title: Senior Consultant, RA #### 3. Date Prepared Mar 23, 2020 # 4. Device Identification Trade/Proprietary Name: Anti-bacterial Bandage Common/Usual Name: Wound Dressing Classification Name: unclassified Regulation Number: NA Product Code: FRO {4}------------------------------------------------ | Device Name | 510(k)<br>No. | Product<br>Code | Classification<br>Regulation | Manufacture | |-----------------------------------|---------------|-----------------|------------------------------|-----------------------------| | Curad<br>Antibacterial<br>Bandage | K113583 | FRO | Unclassified | Medline Industries,<br>Inc. | ## 5. Legally Marketed Predicate Device(s) #### 6. Indication for Use Statement Anti-bacterial bandages are to be applied topically to the skin for the management of minor cuts,minor scrapes. #### 7. Device Description 7.1.1 Anti-bacterial bandages are to be applied topically to the skin for the management of minor cuts,minor scrapes. Anti-bacterial bandage is made of Fabric tape/polyethylene tape, and absorbent pad. Absorbent pad contains 0.1%/0.8% benzalkonium chloride. 7.1.2 The Anti-bacterial bandage is EO sterilized, and is for single use only. 7.1.3 Product shelf- life time: Three years ## 8. Comparison of Technological Characteristics with predicate device | Manufacturer | Planet (Suzhou) Medical Products Co., Ltd | Medline Industries, Inc. | Comparison | | management of minor<br>cuts,minor scrapes. | minor cuts, scrapes and burns. | Page 3 of 5 | Sterile methods | EO sterile, compliance<br>with ISO 11135:2014 | EO sterile, compliance with<br>ISO 11135:2014 | Same | |----------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|------------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|----------------------------|------------------------------------------------------------|------------------------------------------------------------|------| | Trade Name | Anti-bacterial Bandage | Curad Antibacterial Bandage | NA | Model | 0.1% benzalkonium<br>chloride, fabric tape;<br>0.8% benzalkonium<br>chloride, fabric tape;<br>0.1% benzalkonium<br>chloride,polyethylene<br>tape;<br>0.8% benzalkonium<br>chloride,polyethylene tape | 0.8% benzalkonium chloride,<br>fabric tape;<br>0.8% benzalkonium<br>chloride,polyethylene tape | Similar | Single-Use | Yes | Yes | Same | | 510(k) Number | K200821 | K113583 | NA | Antimicrobial<br>agent | Benzalkonium chloride<br>0.1% and 0.8% | Benzalkonium chloride 0.8% | Similar | Shelf Life | 3 years | 3 years | Same | | Product Code | FRO | FRO | Same | Mechanism of<br>Action | Benzalkonium chloride<br>reduce bacterial<br>colonization within<br>dressing. | Benzalkonium chloride in the<br>wound pad helps prevent<br>infection. Helps prevent<br>infection and protects cuts,<br>scrapes or minor burns from<br>dirt and germs | Similar | Complies with<br>ISO 10993 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11 | Same | | Regulation Number | NA | NA | Same | Use time | No more than 24 hours | No more than one week | Different | Packaging | medical packing paper | medical packing paper | Same | | Device Classification Name | dressing, wound, drug | dressing, wound, drug | Same | Anatomical<br>Location | For use on minor cuts,<br>minor scrapes. | For use on minor cuts, scrapes,<br>and burns. | Similar | | | | | | Indications for Use | Anti-bacterial bandages are to be applied topically to the skin for the | Antibacterial bandages are to be applied topically to the skin to help prevent infection in | Similar | Material | Anti-bacterial bandage is<br>made of Fabric<br>tape/polyethylene tape,<br>and absorbent pad.<br>Absorbent pad contains<br>0.1%/0.8% benzalkonium<br>chloride. | Anti-bacterial bandage is made<br>of Fabric/polyethylene tape,<br>and absorbent pad. Absorbent<br>pad contains 0.8%<br>benzalkonium chloride. | Similar | | | | | | | | | | | | | | | | | | {5}------------------------------------------------ {6}------------------------------------------------ #### 9. Substantial Equivalence Discussion The Anti-bacterial Bandage is compared to the predicate device with respect to indications for use, mechanism of action, materials,antimicrobial agent, etc. According to the comparison information, most of the characteristics of the subject device are the same as the predicate device, some of the characteristics are similar, one is different, but none of them will cause new safety or effectiveness issues. #### 10. Non-Clinical Performance Data As part of demonstrating safety and effectiveness of Anti-bacterial Bandage and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Planet (Suzhou) Medical Products Co., Ltd completed a number of non-clinical performance tests. The Anti-bacterial Bandage meets all the requirements for overall design, sterilization, biocompatibility, and other test results confirming that the design output meets the design inputs and specifications for the device. The Anti-bacterial Bandage passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device: - Biocompatibility; . - Cytotoxicity o - Intracutaneous reactivity o - 0 Acute system toxicity - O Sensitization testing - Material mediated pyrogenicity O - endotoxin testing O {7}------------------------------------------------ - Sterilization Testing ISO 11135:2014, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. - . Shelf Life Testing ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Device - Storage and Transport Testing . ASTM 4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems - . AATCC 100: 2012, Antibacterial Finishes on Textile Materials ## 11. Conclusion Based on the above non-clinical tests and the information compared with the predicate device, the anti-bacterial bandage is as safe and as effective as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...