K202797 · Accurate Medical Therapeutics, Ltd. · DQO · Oct 23, 2020 · Cardiovascular
Device Facts
Record ID
K202797
Device Name
Drakon and Sequre Microcatheters
Applicant
Accurate Medical Therapeutics, Ltd.
Product Code
DQO · Cardiovascular
Decision Date
Oct 23, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Drakon™ and SeQure® microcatheters are intended for the infusion of contrast media into all peripheral vessels. The Drakon™ and SeQure® microcatheters are also intended for drug infusion in intra-arterial therapy and infusion of embolic materials. The Drakon™ and the SeQure® microcatheters should not be used in cerebral vessels.
Device Story
Single-use microcatheters for peripheral vascular access; facilitate infusion of contrast media, therapeutic drugs, and embolic materials. Device comprises luer lock hub, strain relief, central shaft, and distal tip with radiopaque markers for fluoroscopic visualization. Inner lumen PTFE-lined for smooth passage of guide wires and agents; distal shaft features hydrophilic coating for lubricity. Two models: Sequre® (distal side holes, two markers) and Drakon™ (no side holes, one marker). Operated by clinicians in interventional settings; used to navigate peripheral vasculature. Output is delivery of fluids/agents to target site; clinical decision-making guided by fluoroscopic imaging of markers. Benefits include precise, localized delivery of therapeutic or diagnostic agents to peripheral anatomy.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993 series, ASTM F756, ASTM F2382, USP <151>), sterilization validation (ISO 11135-1), and mechanical performance (tensile, burst pressure, torque, trackability, power injection, and leakage tests). All tests met predetermined acceptance criteria.
Technological Characteristics
Single lumen microcatheter; PTFE inner lumen; hydrophilic polymer coating. Radiopaque markers for fluoroscopic visualization. Available in 1.7 Fr to 3.0 Fr diameters and 105-155 cm lengths. Sterilized per ISO 11135-1. Mechanical testing per internal specs and standards including ISO 10993 and ASTM F756/F2382.
Indications for Use
Indicated for infusion of contrast media, drugs, and embolic materials into peripheral vessels. Contraindicated for use in cerebral vessels.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
K140329 — RENEGADE HI-FLO MICROCATHETER, MICROCATHETER KIT, FATHOM SYSTEM, FATHOM KIT · Boston Scientific · May 15, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
October 23, 2020
Accurate Medical Therapeutics Ltd. Osnat Harbater R&D and RA Manager 19 Eli Hurvitz Street Rehovot, 7608802 Israel
Re: K202797
Trade/Device Name: Drakon™ and Sequre® Microcatheters Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DQO Dated: September 17, 2020 Received: September 23, 2020
Dear Osnat Harbater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202797
Device Name Drakon™ and Sequre® microcatheters
#### Indications for Use (Describe)
The Drakon™ and SeQure® microcatheters are intended for the infusion of contrast media into all peripheral vessels. The Drakon™ and SeQure® microcatheters are also intended for drug infusion in intra-arterial therapy and infusion of embolic materials.
The Drakon™ and the SeQure® microcatheters should not be used in cerebral vessels.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Accurate Medical, a Guerbet company. The logo features a teal circle with a purple arrow pointing upwards and to the right. The text "accuratemedical" is written in purple and gray next to the circle. Below the company name, the text "a Guerbet company" is written in a smaller font.
### 510(k) Summary Drakon™ and Sequre® microcatheters 510(k) Number K202797
Date Prepared: October 19th, 2020
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:
#### I. SUBMITTER
#### Company
Accurate Medical Therapeutics Ltd. 19 Eli Hurvitz St Rehovot, Israel 7608802 +972-54-3386871
#### Contact Person
Osnat Harbater 19 Eli Hurvitz St Rehovot, Israel 7608802 +972-54-3386871 osnat(@accurmed.com
### II. DEVICE
Name of Device: Common or Usual Name: Classification Name: Regulation: Regulatory Product Code:
Drakon™ and Sequre® microcatheters Drakon™ and Sequre® microcatheters Catheter, Intravascular, Diagnostic 21 CFR 870.1200 Class: II DQO
## III. PREDICATE DEVICE
Accurate Medical Therapeutics Ltd. claims substantial equivalence to the Sequre NF and Sequre® microcatheters, cleared under K173430
### IV. DEVICE DESCRIPTION
The Drakon™ and Sequre® microcatheters are single use microcatheters primarily comprised of a luer lock hub, a strain relief cover and tube, central shaft, and a distal tip with radiopaque markers for visualization. The two models differ only in the design of the distal tip. The Sequre®'s distal end has side holes and two radiopaque markers while the Drakon™'s distal end has no side holes and one radiopaque marker. These markers allow for the fluoroscopic visualization of the distal tip of the microcatheters.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Accurate Medical, a Guerbet company. The logo features a teal circle with a purple arrow pointing upwards and to the right. The text "accurate medical" is written in purple next to the circle, and the text "a Guerbet company" is written in a smaller font below the main text.
The inner lumen is made of PTFE (polytetrafluoroethylene), which allows for the smooth passage of fluids, embolic agents and devices such as guide wires. The distal section of the shaft in both models is coated in a hydrophilic polymer laver. which ensures high lubricity when wet with saline or blood.
The Drakon™ and Sequre® microcatheters are sterile single lumen devices and are available in several different diameters (1.7 Fr., 2.4 Fr, 2.7 Fr., 2.8Fr, and 3.0 Fr.) and lengths (105 cm, 130 cm, or 150 or 155 cm).
## V. INDICATIONS FOR USE
The Drakon™ and Sequre® microcatheters are intended for the infusion of contrast media into all peripheral vessels.
The Drakon™ and Sequre® microcatheters are also intended for drug infusion in intraarterial therapy and infusion of embolic materials.
The Drakon™ and Sequre® microcatheters should not be used in cerebral vessels.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device is substantially equivalent to the predicate device with respect to indications for use, principle of operation, fundamental design principles, performance, sterilization, and packaging. The primary reason for submitting this special 510(k) is the addition of sizes that differ in outer and inner diameter and material changes. Note that the 1.7 Fr, 1.9 Fr, and 3.0 Fr. microcatheters also have a minimal addition to the length (from 150 cm to 155 cm).
These differences do not impact product performance or modify the intended use but are mainly designed to reach smaller or larger target vessels in the same anatomy. The Microcatheter's instructions for use identically indicate the sizes requirements for devices compatible with the Microcatheter outer diameter/inner diameter. Therefore, these differences do not raise any new issues of safety and effectiveness.
Although different materials were used in the additional sized models, the materials are all of the same type (e.g. polymer based hydrophilic coating, PTFE, etc). Thus, these changes also do not raise new questions of safety and effectiveness in comparison to predicate device.
## VII. PERFORMANCE DATA
Risk assessment pursuant to ISO 14971 was performed to assess the impact of the change. The following bench tests were performed to evaluate the design elements and performance characteristics of the modified Drakon™ and Sequre® microcatheters and to demonstrate substantial equivalence to the predicate device. The modified Drakon™ and Sequre® microcatheters met the predetermined acceptance criteria.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Accurate Medical, a Guerbet company. The logo features a teal circle with a purple arrow pointing upwards and to the right. To the right of the circle is the text "accurate medical" in a dark purple sans-serif font, with the words "a Guerbet company" in a smaller, lighter purple font below.
## Biocompatibility testing
biocompatibility evaluation for the modified Drakon™ and Sequre® The microcatheters was conducted in accordance with the FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process
The following testing was conducted:
- Cytotoxicity (ISO 10993-5:2009) ●
- ISO Intracutaneous Study (ISO 10993-10:2010) .
- ISO Guinea Pig Maximization Sensitization Test (ISO 10993-10:2010)
- ISO Acute Systemic Toxicity Study in Mice (ISO 10993-11:2017)
- ASTM Hemolysis Study (ASTM F756 and ISO 10993-12:2012) ●
- ASTM Partial Thromboplastin Time (ASTM F2382:2018) ●
- SC5b-9 Complement Activation Assay (ISO 10993-4:2017) .
- USP Rabbit Pyrogen Study, Material Mediated (USP General Chapter <151> Pyrogen Test: 2019 and ISO 10993-11: 2017).
- In Vivo Thromboresistance Study in Dogs - Jugular Vein ISO 10993-4:2009
### Sterilization, Packaging and Shelf Life Testing
Sterilization validation testing of the Drakon™ and Sequre® microcatheters is performed in compliance with ISO 11135-1 for a SAL 106. In addition, shelf life and packaging testing were performed to support the labeled shelf life.
### Bench Testing
The following Bench testing was completed successfully by meeting the predefined acceptance criteria:
- Bead Compatibility Bench Test ●
- Vessel Flow Dynamic Indication (Beads Reflux) Bench Test ●
- Tensile Bench Test ●
- Burst Pressure Bench Test
- Torque Strength Bench Test
- . Strain Relief Bench Test
- Guidewire & Guide Catheter Compatibility; Dimensional and Visual ● Inspection
- Bend Radius Bench Test ●
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Accurate Medical, a Guerbet company. The logo features a teal circle with a purple arrow pointing upwards and to the right. To the right of the logo is the text "accuratemedical" in purple and gray, with the words "a Guerbet company" in smaller gray font below.
- Power Injection Bench Test ●
- . Preconditioning and Injected Substances Compatibility Bench Test
- Torque Transmission Bench Test
- Acute Particulate Matter Evaluation
- Air Leakage Bench Test
- Liquid Leakage Bench Test .
- . Trackability Bench Test
### VIII. CONCLUSIONS
The modified Drakon™ and Sequre® microcatheters are substantially equivalent in intended use and indications for use, principles of operation, fundamental design, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness. In conclusion, the modified Drakon™ and Sequre® are substantially equivalent to its predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.