MEDI-TECH 4 AND 5 FRENCH NIGHTHAWK FLUSH ANGIOGRAPHIC CATHETERS
Applicant
Boston Scientific Corp
Product Code
DQO · Cardiovascular
Decision Date
Jul 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Intended Use
The Medi-tech 4 and 5 French Nighthawk Flush Angiographic Catheters are designed to provide a pathway to be used for delivering contrast media to selected sites in the vascular system during an angiographic procedure.
Device Story
The Medi-tech Nighthawk Flush Angiographic Catheter is a sterile, single-use device designed for vascular access. It functions as a conduit for delivering contrast media to specific vascular sites during angiography. The catheter features a coextruded Pebax shaft with radiopaque fillers for visualization, a polycarbonate hub, and a strain relief. It is available in 4F and 5F sizes with various tip tapers for compatibility with 0.035" or 0.038" guidewires. Some models include sideholes and an optional silicone coating to reduce friction. The device is operated by a clinician in a clinical setting. By providing a stable pathway and facilitating contrast delivery, the catheter enables visualization of the vascular system, aiding in diagnostic decision-making.
Clinical Evidence
Bench testing only. Functional testing included pressure burst, tip bond tensile, dye flow, tip coefficient of friction, force transmission, radiopacity, torque, tip straightener peel-away force, and particulate testing. Silicone coating integrity was also verified. Results confirmed the device is adequate for its intended use.
Technological Characteristics
Materials: Pebax (shaft), polycarbonate (hub), radiopaque fillers, optional silicone coating. Dimensions: 4F and 5F diameters, 65-100 cm lengths. Features: Sideholes, tapered tips for 0.035"/0.038" guidewires. Sterilization: Provided sterile. Energy source: None (mechanical).
Indications for Use
Indicated for use in patients undergoing angiographic procedures requiring the delivery of contrast media to selected sites in the vascular system. Prescription use only.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
Medi-tech 4 and 5 F Imager Tennis Racket Flush catheters
K233268 — Impress Angiographic Catheter · Merit Medical Systems, Inc. · May 22, 2024
K991673 — CORDIS TEMPO ANGIOGRAPHY CATHETER · Cordis Corp. · May 26, 1999
K123818 — URSA ANGIOGRAPHIC CATHETER · Curative Medical Technology, Inc. · Mar 4, 2013
K191608 — Impress Angiographic Catheter · Merit Medical Systems, Inc. · Jul 16, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
K9716007
510(k) Notification Medi-tech® 4 and 5 F Flush Angiographic Catheters
### Section 4
# Summary of Safety and Effectiveness
# (Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)
#### General Provisions L
JUL 2 4 1997
| Submitter's Name<br>and Address | SCIMED Life Systems, Inc.<br>One SCIMED Place<br>Maple Grove, Minnesota 55311 |
|---------------------------------|-------------------------------------------------------------------------------|
| Contact Person | Angela Raun<br>(612) 494-2456 |
| Classification Name | Diagnostic Intravascular Catheters<br>(21CFR Part 870.1200) |
| Common or Usual Name | Angiographic Intravascular Catheter |
| Proprietary Name | Medi-tech® 4 and 5 French Nighthawk™<br>Flush Angiographic Catheters |
| Name of Predicate Devices | Medi-tech® 4 and 5 F Imager™ Tennis<br>Racket™ Flush catheters |
#### III. Device Description
II.
The Medi-tech 4 and 5 French Nighthawk Flush Angiographic Catheters are available in the following dimensions and tip taper configurations. The 4 F Nighthawk catheters are tapered at the tip to be compatible with 0.035" guidewires. The 5 F Nighthawk catheters have two tapered tip configurations to be compatible with 0.038" and 0.035" guide wires.
| Model | Catheter ID (ins) | Final Tip ID (ins) | Available Lengths | Sideholes |
|---------------------------------|-------------------|--------------------|-------------------|-----------|
| 4 F Nighthawk Non-braided Flush | 0.040" | tapers to 0.038" | 65-100 cm | yes |
| 5 F Nighthawk Non-braided Flush | 0.049" | tapers to 0.038" | 65-100 cm | yes |
| 5 F Nighthawk Non-braided Flush | 0.049" | tapers to 0.041" | 65-100 cm | yes |
The Nighthawk Flush catheters utilize common biocompatible materials. The Nighthawk catheter shaft is coextruded using Pebax combined with a radiopaque filler for the inner layer and Pebax with a radiopaque filler and colorant as the outer surface layer. The shaft provides a smooth surface to allow for dye flow and easy
{1}------------------------------------------------
# Section 4
# Summary of Safety and Effectiveness, cont.
guide wire movement as well as stiffness and curve retention. The radiopaque materials are utilized in the shaft and tip to allow visualization of the catheter during the procedure. The catheter will be provided with sideholes and optional silicone coating applied to the exterior surface of the outer layer.
The catheters utilize a polycarbonate hub and Pebax strain relief. The hub is molded to the proximal end of the catheter shaft. A tip straightener is included as an accessory to the Flush line of catheters. The devices will be provided sterile and are intended for one procedure use only.
### IV. Intended Use
The Medi-tech 4 and 5 French Nighthawk Flush Angiographic Catheters are designed to provide a pathway to be used for delivering contrast media to selected sites in the vascular system during an angiographic procedure.
### v. Non-clinical Test Summary
Functional testing consisted of pressure burst, tip bond tensile, dye flow, tip coefficient of friction, force transmitted by the catheter tip, radiopacity, torque, tip straightener peel-away force, and particulate and verification of the silicone coating integrity. Test results verified that the 4 and 5 French Nighthawk Flush Angiographic Catheters are adequate for their intended use. The 4 and 5 French Nighthawk Flush Angiographic Catheters are considered substantially equivalent to angiographic catheters currently marketed by Medi-tech based on a comparison of the intended use, design and results of in-vitro testing and evaluation.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Angela Raun Senior Regulatory Affairs Specialist Boston Scientific Corporation One. SCIMED Place Maple Grove, Minnesota 55311-1566
JUL 2 4 1997
Re: K971607 Trade Name: Medi-tech® and 5 French Nighthawk™ Flush Angiographic Catheters Regulatory Class: II Product Code: 74DQQ Dated: April 30, 1997 Received: May 1, 1997
Dear Ms. Raun:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labelinq, and prohibitions against misbranding, and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP requlation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Paqe 2 - Ms. Anqela Raun
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section 2
Indications for Use
510(k) Number (if known) _____________________________________________________________________________________________________________________________________________________
Device Name: Medi-tech® 4 and 5 French Nighthawk™ Flush Angiographic Catheters ·······························································································
Indications for Use:
The Medi-tech® 4 and 5 French Nighthawk™ Flush Angiographic Catheters are designed to provide a pathway to be used for delivering contrast media to selected sites in the vascular system during an angiographic procedure.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
*OR*
Over The Counter Use
(Optional Format 1-2-96)
Lynne a. Reamer
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
1607
510(k) Number _
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.