K160662 · Surefire Medical, Inc. · DQO · Mar 11, 2016 · Cardiovascular
Device Facts
Record ID
K160662
Device Name
Surefire Infusion System
Applicant
Surefire Medical, Inc.
Product Code
DQO · Cardiovascular
Decision Date
Mar 11, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Intended Use
The Surefire Infusion System is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
Device Story
Surefire Infusion System is a 0.027" lumen coaxial microcatheter featuring a distal, funnel-shaped, expandable tip; used in peripheral vascular angiographic procedures. Operated by physicians in clinical settings. Device serves as a conduit for contrast agents, flush solutions, and embolic beads (hydrogel particles ≤700um, glass microspheres ≤190um). Tip expansion/collapse is controlled by retracting or advancing the inner microcatheter relative to an outer sheath; allows for vessel wall apposition in 4-6 mm vessels. Expanded tip improves infusion efficiency of embolic agents while maintaining antegrade blood flow. Radiopaque markers facilitate positioning under fluoroscopy. Device is single-use, sterile, and compatible with standard 0.018" guidewires and rotating hemostatic valves. Benefits include targeted delivery of therapeutic agents and improved infusion efficiency.
Clinical Evidence
No human clinical data. Evidence consists of bench testing (visual/dimensional, pull force, infusion efficiency, particulates, kink resistance, antegrade flow, torque, trackability) and an acute animal study in a simulated clinical environment. Animal study confirmed the device met user needs and performed comparably to the predicate.
Technological Characteristics
Coaxial microcatheter with Teflon inner liner and hydrophilic coating. Distal funnel-shaped expandable tip (4-6 mm). Radiopaque markers. Materials identical to predicate (K143588). Sterilized via Ethylene Oxide (EtO). Compatible with 0.018" guidewires. Biocompatibility per ISO 10993-1, -4, -5, -10, -11.
Indications for Use
Indicated for use in angiographic procedures to deliver radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
K143588 — Surefire Infusion System 021 · Surefire Medical, Inc. · Jan 12, 2015
K121677 — SUREFIRE HI-FLOW MICROCATHETER · Surefire Medical, Inc. · Jun 14, 2012
K180677 — Surefire Spark Infusion System · Surefire Medical, Inc. · Apr 3, 2018
K171355 — Surefire Precision Infusion System · Surefire Medical, Inc. · May 17, 2017
K230957 — TriSalus TriNav® LV Infusion System · Trisalus Life Sciences · May 2, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure, with three faces in profile overlapping each other. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2016
Surefire Medical, Inc. c/o Mark Job Regulatory Technical Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K160662
Trade/Device Name: Surefire Infusion System Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Dated: March 7, 2016 Received: March 9, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160662
Device Name Surefure Infusion System
Indications for Use (Describe)
The Surefire Infusion System is intended for use in angiographic procedures.
It delivers radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
Type of Use (Select one or both, as applicable)
| | <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|--|--------------------------------------------------------------------------|
| | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
#### Device Name
Surefire Infusion System
#### Manufacturer Name and Address
Surefire Medical, Inc. 6272 W. 91st Avenue Westminster, CO 80031 Owner Operator Number: 10038066
### Submitter Contact Information
Surefire Medical, Inc. 6272 W. 91st Avenue Westminster, CO 80031 Lynne Aronson, VP Regulatory Affairs and Quality Assurance Contact: Phone: 303-426-1222 -----------------------------------------------------------------------------------------------------------------------------------------------------------------Fax: 303-426-1223
### Common, Classification & Proprietary Names
| Common Name: | diagnostic intravascular catheter |
|----------------------------|-----------------------------------|
| Classification Name: | diagnostic intravascular catheter |
| Proprietary Name: | Surefire Infusion System |
| Classification: | Class II |
| Classification Panel: | Cardiovascular Devices |
| Classification Regulation: | 21 CFR 870.1200 |
| Product Code: | DQO |
### Predicate Device
- Surefire Infusion System 021 ■ (marketed as Surefire Precision Infusion System)
### Device Description
The modified Surefire Infusion System is a 0.027" lumen coaxial microcatheter with the Surefire Expandable Tip at the distal end. The Surefire Infusion System has a female luer lock hub on the proximal end of the catheter and has a catheter shaft length of 150 cm.
The modified Surefire Infusion System serves as the conduit for physician-specified agents such as contrast agents, flush solutions, and embolic beads. It is compatible with standard 0.018" guide wires, infusion syringes, rotating hemostatic valves (RHVs/Tuohy Borsts), and embolic hydrogel particles 700um or less in size and glass microspheres 190um or less in size. The Surefire Infusion System, serves as the conduit for physician-specified agents such as contrast agents, flush solutions, and embolic beads. The Surefire Infusion System has a Teflon inner liner to provide a lubricious surface for passage of physician-specified agents and other accessory devices. The device is hydrophilically coated. The soft, pliable, funnel-shaped Surefire Expandable Tip is sized for use in vessels of 4 mm to 6 mm.
K143588
The Surefire Infusion System has an outer sheath to facilitate expanding and collapsing the Surefire Expandable Tip. There are three radiopaque markers located just proximal and distal to the Surefire Expandable Tip. The Tip can be expanded or collapsed for re-positioning during an interventional procedure by simply retracting or advancing the inner microcatheter while holding the outer sheath stationary. When expanded, the Expandable Tip is designed to improve infusion efficiency of compatible embolic agents while maintaining antegrade flow in the vessel.
{4}------------------------------------------------
The Surefire Infusion System is provided sterile (EtO) for single patient use.
### Indications for Use
The Surefire Infusion System is intended for use in angiographic procedures. It delivers radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
### Biocompatability Testing
All of the patient-contact materials used to fabricate the modified Surefire Infusion System are identical to the predicate Surefire Infusion System as it was approved in 510(k) K143588 (cleared January 12, 2015) in formulation, processing, and sterilization, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.).
Biocompatibility testing of the patient-contact materials used in the construction of the modified Surefire Infusion System were previously performed in accordance with ISO 10993-1 for an external communicating device in contact with circulating blood with a limited duration of less than 24 hours. The testing was conducted in accordance with GLP by NAMSA (Northwood, OH).
As the modified Surefire Infusion System is constructed of the same materials as the predicate Surefire Infusion System, the biocompatibility test requirements for the modified Surefire Infusion System were met by leveraging the previously completed biocompatibility testing listed below.
Additionally, testing for thrombogenicity was performed on the modified Surefire Infusion System Catheter as a part of an Animal Study.
The results of the biocompatibility testing did not indicate any significant biological reaction that would affect the patient due to contact with the materials used in the device construction.
| Category | Standard | Test Method |
|-----------------------------------------------|--------------|----------------------------------------------------------------------------------------|
| Cytotoxicity | ISO 10993-5 | Cytotoxicity Study Using the ISO Elution Method - 1x Minimal Essential Media Extract |
| Sensitization | ISO 10993-10 | ISO Maximization Sensitization Study - Extract - 0.9% Sodium Chloride Solution Extract |
| | ISO 10993-10 | ISO Maximization Sensitization Study - Extract - Sesame Oil, NF Extract |
| Irritation or<br>Intracutaneous<br>Reactivity | ISO 10993-10 | ISO Intracutaneous Study - Extract - 0.9% Sodium Chloride Solution Extract |
| | ISO 10993-10 | ISO Intracutaneous Study - Extract - Sesame Oil, NF Extract |
| Systemic Toxicity | ISO 10993-11 | ISO Systemic Toxicity Study - Extract - 0.9% Sodium Chloride Solution Extract |
| | ISO 10993-11 | ISO Systemic Toxicity Study - Extract - Sesame Oil, NF Extract |
| | ISO 10993-11 | Pyrogen - Material Mediated – 0.9% Sodium Chloride Solution Extract |
| Hemocompatability | ISO 10993-4 | ASTM Hemolysis – CMF-PBS Extract |
| | ISO 10993-4 | C3a Complement Assay - Normal Human Serum Extract |
| | ISO 10993-4 | SC5b-9 Complement Assay - Normal Human Serum Extract |
| | ISO 10993-4 | Coagulation - ASTM Partial Thromboplastin Time |
{5}------------------------------------------------
## Performance Testing
The following design verification / validation tests were performed as a result of the risk analysis assessment of the dimensional modifications. The test results demonstrate that the modified Surefire Infusion System meets the same performance specifications and acceptance criteria as the predicate device.
- l Visual/Dimensional Inspections
- . Pull Force
- Infusion Efficiency
- . Particulates
- . Device Compatibility
- Kink Resistance
- Antegrade Flow
- Torque Resistance (catheter body)
- Trackability
The following testing was leveraged from previous testing of the predicate device:
- . Packaging seal strength and integrity
- . Hub Aspiration
- . Corrosion
- Pull Force (handle components) 트
- 트 Hydrophilic Coating Lubricity and Wear
- Diagnostic Infusion Agent Compatibility ■
Burst Pressure
- Embolic Agent Compatibility
- . Torque (handle components)
## Animal Testing
An animal study was performed to assess the comparative acute performance of the modified Surefire Infusion System to the predicate device, as defined by physicians in a simulated clinical environment. The modified Surefire Infusion System was found to be acceptable in all evaluated categories, met the defined user needs, and performed comparably to the predicate device.
■
### Substantial Equivalence
The modified Surefire Infusion System is substantially equivalent in intended use, design, and technology/principles of operation to the predicate.
Comparative Summary: Design / Technological Characteristics
The modified Surefire Infusion System and predicate device are coaxial microcatheters with the Surefire Expandable Tip at the distal end. Both infusion systems have an outer catheter to facilitate expanding and collapsing the Surefire Expandable Tip.
The modified Surefire Infusion System and the predicate device are constructed of the same materials utilizing similar construction and manufacturing processes.
The modified Surefire Infusion System is a larger version of the predicate device. The predicate device has an inner lumen diameter of .021", length of 120 cm, and a Surefire Expandable Tip sized for use in vessels 1.5 - 3.5 mm. The modified device has an inner lumen diameter of .027", lengths of 150 cm, and a Surefire Expandable Tip sized for use in vessels 4-6 mm.
The modified Surefire Infusion System and predicate device are provided in identical packaging, sterilized by ethylene oxide, and labeled for single use only.
{6}------------------------------------------------
Comparative Summary: Indications for Use
The modified Surefire Infusion System has the same indications for use as the predicate device. Both devices are intended for use in angiographic procedures to deliver radiopaque media and therapeutic agents to selected sites in the peripheral vascular system.
Comparative Summary: Performance
Animal and bench performance test data demonstrate that the modified Surefire Infusion System performance is comparable to the predicate device.
In summary, the modified Surefire Infusion System is substantially equivalent in intended use, design, and technology/principles of operation to the predicate device. Animal and bench performance test data demonstrate that the modified Surefire Infusion System performance is comparable to the predicate device. Differences between the devices do not raise any issues of safety or effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.