K970389 · Dade Chemistry Systems, Inc. · JIT · Feb 18, 1997 · Clinical Chemistry
Device Facts
Record ID
K970389
Device Name
THYROID CALIBRATOR
Applicant
Dade Chemistry Systems, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Feb 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The Thyroid Calibrator is intended to be used on the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module to calibrate the FT4 and TSH methods.
Device Story
Liquid bovine albumin-based calibrator; contains thyroxine and human TSH (levels 2-5); level 1 contains no detectable thyroxine or TSH. Used on Dimension® RxL clinical chemistry system with heterogeneous immunoassay module. Provides reference points for calibrating FT4 and TSH assays. Used by laboratory personnel in clinical settings to ensure accuracy of patient thyroid hormone measurements.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Liquid bovine albumin matrix; 5-level kit; 2.0 mL per vial. Designed for use with Dimension® RxL clinical chemistry system.
Indications for Use
Indicated for use as a calibrator for FT4 and TSH methods on the Dimension® RxL clinical chemistry system.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
aca® plus FT4 Calibrator
aca® plus TSH Calibrator
Related Devices
K081103 — LOCI THYROID CALIBRATOR WITH MODEL RC610 · Siemens Healthcare Diagnostics, Inc. · May 23, 2008
K100778 — DIMENSION EXL LOCI THYROID CALIBRATOR, MODEL RC 610A · Siemens Healthcare Diagnostics · May 5, 2010
K993014 — MODIFICATION TO DOCUMENT THYROID CAL.VER · Casco-Nerl Diagnostics Corporation (Maine) · Nov 26, 1999
Submission Summary (Full Text)
{0}
DADE
FEB 18 1997
K970389
Interoffice Memorandum
# Summary of Safety and Effectiveness Information
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Submitter's Name: Cathy P. Craft
Dade International Inc.
P.O. Box 6101
Newark, DE 19714-6101
Date of Preparation: 1/29/97
Name of Product: Thyroid Calibrator
FDA Classification Name: Calibrator
Predicate Device: aca® plus FT4 Calibrator and aca® plus TSH Calibrator
Device Description: The Thyroid Calibrator is a liquid bovine albumin based product. The Level 1 calibrator contains no detectable thyroxine or TSH. Levels 2 through 5 contain thyroxine and human TSH. The kit consists of ten vials; two at each of five levels.
Intended use: The Thyroid Calibrator is intended to be used on the Dimension® RxL clinical chemistry system with the heterogeneous immunoassay module to calibrate the FT4 and TSH methods.
4 000005
{1}
Comparison to Predicate
Device:
| Item | Thyroid Calibrator | aca® plus FT4 Calibrator | aca® plus TSH Calibrator |
| --- | --- | --- | --- |
| Intended Use | Calibrator | Calibrator | Calibrator |
| Analytes | thyroxine, TSH | thyroxine | TSH |
| Matrix | bovine albumin | human serum | horse serum |
| Form | liquid | lyophilized | lyophilized |
| Volume | 2.0 mL per vial | 3.0 mL per vial, reconstituted | 5.0 mL per vial, reconstituted |
| Levels | 5 levels | 3 levels | 3 levels |
Comments on Substantial
Equivalence: Both the Thyroid Calibrator and the aca® plus FT4 Calibrator are intended to be used as calibrators for free thyroxine methods. Both the Thyroid Calibrator and the aca® plus TSH Calibrator are intended to be used as calibrators for thyroid stimulating hormone methods.
Conclusion: The Thyroid Calibrator is substantially equivalent to the aca® plus FT4 Calibrator and the aca® plus TSH Calibrator for their respective analytes based on the comparison discussed above.

Cathy P. Craft
Regulatory Affairs and
Compliance Manager
Date: 1/29/97
000009
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.