UniCel DxC SYNCHRON System HDL Cholesterol reagent(HDL),Unicel DxC SYNCHRON System HDL Calibrator
K142985 · Beckman Coulter, Inc. · JIT · Feb 26, 2015 · Clinical Chemistry
Device Facts
Record ID
K142985
Device Name
UniCel DxC SYNCHRON System HDL Cholesterol reagent(HDL),Unicel DxC SYNCHRON System HDL Calibrator
Applicant
Beckman Coulter, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Feb 26, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The Olympus HDL Cholesterol Reagent is intended for the quantitative determination of high density lipoprotein cholesterol (HDL-C) in human serum and plasma on the Olympus AU400/AU600/AU2700/AU5400 and UniCel DxC 600/800 SYNCHRON systems.
Device Story
The UniCel DxC SYNCHRON Systems HDL Cholesterol reagent and calibrator are used with Beckman Coulter UniCel DxC 600/800 clinical chemistry analyzers. The system measures HDL cholesterol levels in human serum or plasma samples. The device is intended for professional use in clinical laboratory settings. By providing quantitative measurements of HDL cholesterol, the system assists clinicians in diagnosing and managing lipid disorders, atherosclerosis, and associated liver or renal conditions. The calibrator ensures accurate system performance by providing standardized calibration levels for the reagent assay.
Clinical Evidence
No clinical data provided; substantial equivalence is based on bench testing and performance characteristics of the reagent and calibrator system.
Technological Characteristics
In vitro diagnostic reagent; colorimetric assay. Wavelength: 560 nm. Measuring interval: 5-135 mg/dL. Compatible with UniCel DxC 600/800 SYNCHRON and Olympus AU series systems. Reagent bottle shape modified.
Indications for Use
Indicated for quantitative determination of HDL-C in human serum and plasma for clinical diagnostic use.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K042195 — SYNCHRON SYSTEMS HDL CHOLESTEROL (HDL), SYNCHRON SYSTEMS LIPID PLUS CALIBRATORS 1 & 2, MODELS A15625/A16747 · Beckman Coulter, Inc. · Sep 16, 2004
K040767 — SYNCHRON SYSTEMS HDL CHOLESTEROL (HDLD) REAGENT · Beckman Coulter, Inc. · Jun 7, 2004
K050823 — DIRECT HDL CHOLESTEROL AND DIRECT HDL/LDL CALIBRATOR · Teco Diagnostics · Jul 26, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The symbol is composed of three abstract human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 26, 2015
BECKMAN COULTER, INC. C/O EIMEAR CARR REGULATORY AFFAIRS SPECIALIST 250 S. KREAMER BOULEVARD BREA CA 92821
Re: K142985
Trade/Device Name: UniCel DxC SYNCHRON Systems HDL Cholesterol Reagent (HDL) Unicel Dxc SYNCHRON Systems HDL Calibrator Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT, LBS Dated: January 23, 2015 Received: January 27, 2015
Dear Eimear Carr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Katherine Serrano -A
For :Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
UniCel DxC SYNCHRON Systems HDL Cholesterol reagent (HDL) UniCel DxC SYNCHRON Systems HDL Calibrator
## Indications for Use (Describe)
Reagent: UniCel DxC SYNCHRON Systems HDL Cholesterol reagent (HDL), when used in conjunction with UniCel DxC 600/800 SYNCHRON System(s) and UniCel DxC SYNCHRON Systems HDL Calibrator, is intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis, and various liver and renal diseases.
Calibrator: The UniCel DxC SYNCHRON Systems HDL Calibrator is designed to provide suitable calibration levels for Beckman Coulter UniCel DxC 600/800 SYNCHRON Systems employing the quantitative UniCel DxC SYNCHRON Systems HDL Cholesterol reagent (HDL).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.