ACCESS ULTRASENSITIVE HGH CALIBRATOR, MODEL 33585

K061282 · Beckman Coulter, Inc. · JIT · Jun 2, 2006 · Clinical Chemistry

Device Facts

Record IDK061282
Device NameACCESS ULTRASENSITIVE HGH CALIBRATOR, MODEL 33585
ApplicantBeckman Coulter, Inc.
Product CodeJIT · Clinical Chemistry
Decision DateJun 2, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The Beckman Coulter Access Ultrasensitive hGH Calibrators are intended for use with the Access Immunoassay Systems for the calibration of the Access Ultrasensitive hGH assay.

Device Story

Access Ultrasensitive hGH Calibrators; used with Access Immunoassay Systems; calibrates hGH assay for quantitative measurement of human growth hormone in serum/plasma; modification involves re-standardization from WHO pituitary 80/505 IS to recombinant WHO 98/574 2nd IS; fundamental scientific technology remains unchanged; used in clinical laboratory settings by trained personnel; output provides calibration curves for assay performance; ensures accurate hGH quantification for clinical decision-making.

Clinical Evidence

Bench testing only. Verification and validation activities performed per design control procedures; results met predetermined acceptance criteria.

Technological Characteristics

Lyophilized calibrators containing hGH. Traceable to WHO 98/574 standard. Used with Access Immunoassay Systems (one-step immunoenzymatic sandwich assay). No changes to materials, functionality, or stability compared to predicate.

Indications for Use

Indicated for the calibration of the Access Ultrasensitive hGH assay on Access Immunoassay Systems for the quantitative determination of human growth hormone (hGH) in human serum and plasma.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Image /page/0/Picture/4 description: The image shows the text "The assigned 510(k) number is:" followed by the number "K061282". The text is written in a clear, legible font. The number is written in a handwritten style. ### Submitter's Name and Address JUN - 2 2006 Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 368-7629 Fax: (952) 368-7610 Contact: Jennifer Ruether Date Prepared: May 5, 2006 ## Device Names | Proprietary Name: | Ultrasensitive hGH Calibrators on the Access®<br>Immunoassay Systems | |-------------------|----------------------------------------------------------------------| |-------------------|----------------------------------------------------------------------| Common Name: Calibrators - Classification Name: Calibrator, Secondary (862.1150, JIT) ## Predicate Device Access Ultrasensitive hGH Calibrators Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 510(k) Number: K003089 {1}------------------------------------------------ ## Device Description The Access Ultrasensitive hGH calibrators are lyophilized calibrators to be used with the Access Ultrasensitive hGH assay to generate the hGH calibration curve on the Access Immunoassay Systems. The Access Ultrasensitive hGH calibrator kit contains six 2.0 mL vials, one for each calibrator level. The Access Immunoassay Systems utilize a one-step immunoenzymatic ("sandwich") assay for the quantitative measurement of hGH. #### Intended Use The Access Ultrasensitive hGH Calibrators are intended to calibrate the Access Ultrasensitive hGH Assay for the quantitative determination of hGH levels in human serum and plasma using the Access Immunoassay Systems. | Attribute | Access Ultrasensitive hGH Calibrators | Access Ultrasensitive hGH Calibrators (modified) | |--------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | Intended Use | Calibration for quantitative determination of hGH levels in human serum and plasma | Calibration for quantitative determination of hGH levels in human serum and plasma | | Calibrators | hGH at approximate levels of 0.1, 1.0, 10, 20, and 50 ng/mL (µg/L), lyophilized | hGH at approximate levels of 0.07, 0.7, 7, 14, or 35 ng/mL (µg/L), lyophilized | | Traceability | Traceable to WHO 80/505 | Traceable to WHO 98/574 | #### Summary of Technological Characteristics The device modification consists of restandardizing the Access Ultrasensitive hGH Calibrators from the WHO Pituitary hGH 80/505 standard to the second international standard, WHO 98/574. The restandardization involves the use of the new WHO standard as the reference for assigning the calibrator values; there is no change to calibrator materials, functionality, or stability, {2}------------------------------------------------ #### Conclusion The restandardization of the Access Ultrasensitive hGH calibrators does not change the intended use or indications for use, alter the fundamental scientific technology, or affect the safety and efficacy of the device. Performance data generated from validation testing demonstrates that the restandardized Access Ultrasensitive hGH Calibrators on the Access Immunoassay Systems is substantially equivalent to the currently commercialized Access Ultrasensitive hGH calibrators. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Jennifer Ruether Regulatory Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 JUN - 2 2006 k061282 Re: Trade/Device Name: Access Ultrasensitive hGH Calibrators on the Access® Immunoassay Systems Regulation Number: 21 CFR§862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT Dated: May 5, 2006 Received: May 8, 2006 Dear Ms. Ruether: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Gutz Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): Kd61282 Device Name: Access Ultrasensitive hGH Calibrators on the Access® Immunoassay Systems Indications For Use: The Access Ultrasensitive hGH Calibrators are intended to calibrate the Access Ultrasensitive hGH Assay for the quantitative determination of hGH levels in human serum and plasma on the Access Immunoassay Systems. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carol C Benam Division Sign-Off Cice of In Vitro Diagnostic Device Laation and Safery Page 1 of 1 K061282
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...