Elecsys Progesterone III CalSet
Device Facts
| Record ID | K152526 |
|---|---|
| Device Name | Elecsys Progesterone III CalSet |
| Applicant | Roche Diagnostics |
| Product Code | JIT · Clinical Chemistry |
| Decision Date | Oct 2, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1150 |
| Device Class | Class 2 |
Intended Use
The Progesterone III CalSet is used for calibrating the quantitative Elecsys Progesterone III assay on the Elecsys and cobas e immunoassay analyzers.
Device Story
Elecsys Progesterone III CalSet is a lyophilized, plant-derived progesterone product in a human serum matrix. It serves as a secondary calibrator for the Elecsys Progesterone III assay on Elecsys and cobas e immunoassay analyzers. The device is used in clinical laboratory settings by trained laboratory personnel. The calibrator is reconstituted with distilled or deionized water and used to establish the calibration curve for the immunoassay. By providing known target values, it enables the analyzer to quantify progesterone levels in patient samples, assisting clinicians in diagnostic decision-making related to progesterone levels. The primary change from the predicate is the addition of stability claims at -20°C.
Clinical Evidence
Bench testing only. No clinical data. Stability studies included reconstitution testing, accelerated stability (35°C for 3 weeks), and ongoing real-time stability (up to 19 months). Value assignment was validated across multiple cobas e 411 and cobas e 601/MODULAR ANALYTICS E170 analyzers. Acceptance criteria for recovery were 95-105% of reference values.
Technological Characteristics
Lyophilized plant-derived progesterone in human serum matrix. Traceable via ID-GC/MS to BCR-348R and ERM-DA347. Two-level calibrator. Requires reconstitution with 1.0 mL distilled/deionized water. Stability: 2-8°C (unopened until expiration), 24 hours (opened at 2-8°C), 31 days (opened at -20°C).
Indications for Use
Indicated for use as a calibrator for the quantitative Elecsys Progesterone III assay on Elecsys and cobas e immunoassay analyzers. No specific patient population is described as this is an in vitro diagnostic calibrator.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
- K150955 — Elecsys Progesterone III Cal Check 5 · Roche Diagnostics · May 1, 2015
- K142021 — Elecsys Estradiol III CalSet · Roche Diagnostics · Aug 12, 2014
- K163546 — Lumipulse G Progesterone-N Calibrators · Fujirebio Diagnostics,Inc. · Jan 13, 2017
- K153417 — ST AIA-PACK PROG III Calibrator Set · Tosoh Bioscience, Inc. · Dec 18, 2015
- K970984 — THE ELECSYS CALCHECK PROGESTERONE · Boehringer Mannheim Corp. · Apr 2, 1997