DIMENSION VISTA MYO REAGENT CARTRIDGE AND CALIBRATOR
K053576 · Dade Behring, Inc. · JIT · Feb 7, 2006 · Clinical Chemistry
Device Facts
Record ID
K053576
Device Name
DIMENSION VISTA MYO REAGENT CARTRIDGE AND CALIBRATOR
Applicant
Dade Behring, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Feb 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
Method For the quantitative measurement of myoglobin in human serum and plasma on the Dimension Vista™ System as an aid in the rapid diagnosis of acute myocardial infarction. Calibrator For the calibration of the myoglobin (MYO) method on the Dimension Vista™ System.
Device Story
The Dimension Vista™ MYO reagent cartridge and calibrator are used for in vitro diagnostic testing of human serum and plasma samples. The device utilizes Luminescent Oxygen Channeling Immunoassay (LOCI™) technology; a homogenous sandwich chemiluminescent immunoassay. Input samples are incubated with Chemibeads (coated with anti-myoglobin monoclonal antibody) and biotinylated anti-myoglobin monoclonal antibody. Sensibeads (coated with streptavidin and photosensitizer dye) are added to form bead-myoglobin-biotinylated antibody immunocomplexes. Illumination at 680 nm triggers singlet oxygen diffusion from Sensibeads to Chemibeads, inducing a chemiluminescent reaction measured at 612 nm. The signal intensity is directly proportional to myoglobin concentration. The system is operated by laboratory personnel in a clinical setting. Results are provided to clinicians to assist in the rapid diagnosis of acute myocardial infarction.
Clinical Evidence
No clinical studies performed. Bench testing only. Precision assessed per NCCLS EP5-A2 (N=80); linearity/reportable range (0.5-1000 ng/mL) verified; analytical sensitivity (0.11 ng/mL limit of detection); interference testing (27 substances, no interference <10%); hook effect testing (no effect up to 350,000 ng/mL). Method comparison (n=160) vs. predicate showed correlation coefficient 0.998.
Technological Characteristics
Homogenous sandwich chemiluminescent immunoassay (LOCI™). Reagents: latex beads (Sensibeads with photosensitizer, Chemibeads with chemiluminescent dye) and biotinylated monoclonal antibody. Calibrator: frozen liquid human heart myoglobin in 6% bovine albumin matrix. Detection: 680 nm excitation, 612 nm emission. System: Dimension Vista™ automated analyzer. Connectivity: Integrated system.
Indications for Use
Indicated for the quantitative measurement of myoglobin in human serum and plasma to aid in the rapid diagnosis of acute myocardial infarction.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dade Behring MYO immunoassay and calibrator (K984191 and K984193)
K063756 — DIMENSION VISTA CTNI FLEX REAGENT CARTRIDGE · Dade Behring, Inc. · Mar 19, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
K 053576
## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Submitter's Name: | George M. Plummer<br>Dade Behring Inc.<br>P.O. Box 6101<br>Newark, DE 19714-6101 |
|-----------------------------|----------------------------------------------------------------------------------|
| Date of Preparation: | December 20, 2005 |
| Name of Product(s): | Dimension Vista™ MYO reagent cartridge and<br>Dimension Vista™ MYO calibrator |
| FDA Classification Name(s): | Myoglobin, Antigen, Antiserum, Control and associated calibrator |
| FDA Guidance Documents: | Not applicable |
| Predicate Device(s): | Dade Behring MYO immunoassay and Calibrator<br>(K984191/K984193) |
### Device Description(s):
#### Method
The MYO method is a homogenous sandwich chemiluminescent immunoassay based on Luminescent Oxygen Channeling Immunoassay (1.OCI™) technology. LOCI™ reagents include two latex bead reagents and a biotinylated anti-myoglobin monoclonal antibody fragment. The first bead reagent (Sensibeads) is coated with streptavidin and contains photosensitizer dye. The second bead reagent (Chemibeads) is coated with a second anti-myoglobin monoclonal antibody and contains chemiluminescent dye. Sample is incubated with Chemibeads and biotinylated antibody to form a beadmyoglobin-biotinylated antibody sandwich. Sensibeads are added and bind to the biotin to form bead-pair immunocomplexes. Illumination of the complex by light at 680 nm generates singlet oxygen from Sensibeads which diffuses into Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 nm and is a direct function of the myoglobin concentration in the sample.
### Calibrator
The Dade Behring MYO Calibrator is a three level (A, B, C), frozen liquid product containing purified human heart myoglobin in a 6% bovine albumin matrix with stabilizers and preservatives. The kit contains 3 vials of each level (A= 2.0 mL. B= 1.0 mL. C= 1.5 mL).
#### Intended Use:
#### Method
For the quantitative measurement of myoglobin in human serum and plasma on the Dimension Vista™ System as an aid in the rapid diagnosis of acute myocardial infarction.
#### Calibrator
For the calibration of the myoglobin (MYO) method on the Dimension Vista™ System.
{1}------------------------------------------------
# Comparison to Predicate Device:
Method
A summary of the features of the Dade Behring Dimension Vista™ MYO reagent cartridge and the A summal y of the reatures of the Dade Bearing = mmunoassay (K984191/K984193) is provided in the following charts.
| Feature | Dimension® MYO | Dimension Vista™ MYO |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For the <i>in vitro</i> quantitative<br>determination of myoglobin in<br>human serum and heparinized<br>plasma as an aid in the diagnosis of<br>myocardial infarction. | For the <i>in vitro</i> quantitative<br>determination of myoglobin in human<br>serum and heparinized plasma as an<br>aid in the diagnosis of myocardial<br>infarction. |
| Assay Type<br>(detection) | photometric immunoassay | chemiluminescent immunoassay |
| Reportable Range | 1 to 1000 ng/mL | 0.5 - 1000 ng/mL |
| Antibody | Dade Behring mouse monoclonal | Dade Behring mouse monoclonal |
| Analytical<br>Sensitivity | 1 ng/mL | 0.5 ng/mL |
| Analytical<br>Specificity | There are no known cross reactive<br>materials. | There are no known cross reactive<br>materials. |
| Interferences | No significant interference from:<br>bilirubin up to 60 mg/dL,<br>hemoglobin up to 1000 mg/dL and<br>triglycerides up to 1500 mg/dL | No significant interference from:<br>bilirubin, conj. up to 60 mg/dL<br>bilirubin, unconjugated up to 60<br>mg/dL, hemoglobin up to 1000<br>mg/dL and triglycerides up to 3000<br>mg/dL |
| Hook Effect | No high dose effect<br>(up to 300,000 ng/mL) | No high dose effect<br>(up to 300,000 ng/mL) |
| Calibration Interval | Calibration curve updated for each<br>lot, using five levels and every 90<br>days, thereafter with the same<br>reagent lot. | Calibration curve updated for each<br>lot, using six levels every 30 days<br>with the same reagent lot. |
| Sample Volume | 20 uL | 2 uL |
| Feature | Dimension® MYO | Dimension Vista™ MYO |
| Intended Use | MYO method calibration | MYO method calibration |
| Analyte | Human heart myoglobin | Human heart myoglobin |
| Matrix | Bovine serum albumin | Bovine serum albumin |
| Form | Liquid | Frozen liquid |
| Volume | 10 vials, 2 at each level, 1 mL each | 9 vials, 3 at each level, 2.0, 1.0, 1.5 mL for level A, B, C, respectively |
| Levels | 5 levels (0, 35, 100, 500, and 1060 ng/mL) | 3 levels provided (0, 125 and 1050 ng/mL); 3 additional levels made on-board (36.7, 367 and 733 ng/mL) |
# Method:
{2}------------------------------------------------
# Calibrator:
## Method performance Summary:
## Analytical Results
## Method Comparison
A split sample method comparison demonstrated good agreement between the Dade Behring Dimension Vista™ MYO method and the predicate Dade Behring Dimension® MYO immunoassay with serum and heparinized plasma patient samples.
| Comparative<br>Method | Slope | Intercept<br>(ng/mL) | Correlation<br>Coefficient | n |
|-----------------------|-------|----------------------|----------------------------|-----|
| Dimension® MYO | 1.003 | 6.98 | 0.998 | 160 |
The model equation for linear regression statistics is: [results for Dimension Vista™ MYO] = slope x [comparative method results] + intercept. The range of MYO values in the correlation study was: 12 to 932 ng/mL.
## Serum/Plasma Comparison
Serum and heparin plasma matched pairs were examined on the Dimension Vista™ system. Serum samples (n=37) ranging from 28 to 600 ng/mL when compared to lithium heparin samples gave a slope of 1.05, correlation coefficient of 1.0, and an intercept of -5.12 ng/mL using linear least squares regression statistics.
A separate study was conducted to evaluate the comparison of 115 lithium and sodium heparin samples ranging from 16 ng/mL to approximately 823 ng/mL. A linear least squares regression analysis comparing the lithium to sodium heparin samples gave a slope of 1.00, a correlation coefficient of 1.0, and an intercept of -0.54 ng/mL.
{3}------------------------------------------------
## Reproducibility Typical precision observed for the Dimension Vista™ MYO method is summarized below:
| Sample | Mean (ng/mL) | Repeatability | | Within Lab | |
|--------------------------------------------|--------------|---------------|-----|------------|-----|
| | | SD (ng/mL) | %CV | SD (ng/mL) | %CV |
| Human Serum Pool | | | | | |
| Pool 1 | 110.3 | 5.4 | 4.9 | 5.5 | 5.0 |
| Pool 2 | 501.5 | 17.3 | 3.4 | 18.7 | 3.7 |
| Pool 3 | 830.8 | 23.1 | 2.8 | 27.6 | 3.3 |
| Biorad Liquichek™Cardiac Marker Control LT | | | | | |
| Level 1 | 113.5 | 2.7 | 2.4 | 4.0 | 3.6 |
* Liquichek™ is a registered trademark of Biorad Laboratories, Irvine CA.
The reproducibility testing was conducted in accordance with the NCCLS (CLSI) Approved Guideline for User Evaluation of Precision Performance of Clinical Chemistry Devices EP5-A2. For each test level, a single test from two independent cups was analyzed twice per day. The within-run and total standard deviations were calculated by the analysis of variance method.
## Calibrator
The Dimension Vista™ MYO calibrator is similar to other calibrator products associated with their assays, such as the Dimension® MYO calibrator.
# Comments on Substantial Equivalence:
Both the Dimension Vista™ MYO reagent cartridge and the Dimension® MYO immunoassays are intended for the quantitative determination of myoglobin. Comparative data for serum and human plasma samples demonstrate good analytical and clinical agreement between the methods.
# Conclusion:
The Dade Behring Dimension Vista™ MYO and the predicate Dade Behring Dimension® MYO immunoassays (K984191/K984193) are substantially equivalent based on their intended use and performance characteristics as described above. The calibrator products are also equivalent in their design and intended use with their respective assay systems.
Gmt. Plummer
George M. Plummer Regulatory Affairs and Compliance Manager December 20, 2005
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, and the words "U.S. Department of Health & Human Services - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The background is white.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB 7 2006
Mr. George M. Plummer Regulatory Affairs & Compliance Manager Dade Behring Inc. Glasgow Business Community Bldg. 500. M.S. 514 PO Box 6101 Newark, DE 19714-6101
k053576 Re:
Trade/Device Name: Dimension Vista™ MYO reagent cartridge and Dimension Vista™ MYO calibrator Regulation Number: 21 CFR§ 866.5680 Regulation Name: Myoglobin immunological test system Regulatory Class: Class II Product Code: DDR, JIT Dated: December 20, 2005 Received: December 23, 2005
Dear Mr. Plummer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
## Page 2 --
This letter will allow you to begin marketing your device as described in your Section 510(k) rms lot notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Garcia, Ph.D.
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
Image /page/6/Picture/1 description: The image shows a sequence of alphanumeric characters, specifically 'K053576'. The characters are written in a bold, sans-serif font, and they appear to be handwritten or stylized. The sequence is presented in a clear and legible manner against a white background.
510(k) Number (If Known):
Device(s) Name(s):
Dimension Vista™ MYO reagent cartridge and Dimension Vista™ MYO calibrator
Indications for Use:
Method
For the quantitative measurement of myoglobin in human serum and plasma on the Dimension Vista™ System as an aid in the rapid diagnosis of acute myocardial infarction.
## Calibrator
Prescription Use
For the calibration of the myoglobin (MYO) method on the Dimension Vista™ System.
Prescription Use
(Part 21 CFR 801 Subpart D)
and/or
Over-the-counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ============================================================================================================================================================================= ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Qim Chappie
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
K053576
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.