Dimension Vista MMB Assay

K143720 · Siemens Healthcare Diagnostics · JHY · Sep 18, 2015 · Clinical Chemistry

Device Facts

Record IDK143720
Device NameDimension Vista MMB Assay
ApplicantSiemens Healthcare Diagnostics
Product CodeJHY · Clinical Chemistry
Decision DateSep 18, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1215
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K143720 · Sep 18, 2015Dimension Vista MMB AssaySiemens Healthcare DiagnosticsNative de-identified human serum and plasma samplesThe sponsor performed a split-sample method comparison between the modified device and the predicate device using 111 native clinical samples to demonstrate substantial equivalence.Method comparison; Clinical samples; Substantial equivalence

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Split-sample method comparisonNative de-identified human serum and plasma samples; Sample Size: 111Predicate Dimension Vista 1500 Integrated System (K051087)Correlation and bias between modified and predicate systems

Indications for Use

The MMB method is an in vitro diagnostic test for the quantitative measurement of mass creatine kinase MB isoenzyme (EC 2.7.3.2) in human serum and plasma on the Dimension Vista® System for confirmation of acute myocardial infarction.

Device Story

The Dimension Vista MMB Assay is a homogeneous sandwich chemiluminescent immunoassay using LOCI technology; it measures mass creatine kinase MB isoenzyme in human serum and plasma. The assay utilizes two synthetic bead reagents: Sensibeads (streptavidin-coated, photosensitive dye) and Chemibeads (anti-mass CK-MB monoclonal antibody, chemiluminescent dye). Sample incubation with biotinylated antibody and Chemibeads forms a sandwich; Sensibeads bind to biotin to create bead-pair immunocomplexes. Illumination at 680 nm triggers singlet oxygen diffusion from Sensibeads to Chemibeads, inducing chemiluminescence measured at 612 nm. The signal intensity is proportional to CK-MB concentration. The assay runs on the Dimension Vista 1500 System, a fully automated, microprocessor-controlled floor instrument. The system performs pipetting, mixing, and spectral intensity measurement. Healthcare providers use the quantitative output to confirm acute myocardial infarction. The device is intended for clinical laboratory use.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP5-A2 (20 days, 3 instruments, 5 levels); within-lab CVs ranged 1.80%–5.99%. Linearity confirmed per CLSI EP6-A (1–300 ng/mL, r=0.999). Detection limits (LOB 0.4 ng/mL, LOD 0.8 ng/mL, LOQ 1.0 ng/mL) established per CLSI EP17-A2. Method comparison (n=111) against predicate system showed Passing Bablok slope 0.99 and intercept -0.16 ng/mL. No clinical studies performed.

Technological Characteristics

LOCI (Luminescent Oxygen Channeling Immunoassay) technology. Reagents: biotinylated monoclonal antibody, Chemibeads, Streptavidin Sensibeads. Detection: Photomultiplier tube (PMT) with multiple dynodes. System: fully automated, microprocessor-controlled, floor-model analyzer. Connectivity: integrated, non-wireless. Sterilization: N/A (reagents). Software: embedded firmware for instrument control and signal processing.

Indications for Use

Indicated for quantitative measurement of mass creatine kinase MB isoenzyme in human serum and plasma for confirmation of acute myocardial infarction. For prescription use only.

Regulatory Classification

Identification

A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k143720 B. Purpose for Submission: Modification of a previously cleared device – Change in the photomultiplier tube (PMT) for the Dimension Vista 1500 System. C. Measurand: Mass creatine kinase MB isoenzyme (MMB) D. Type of Test: Quantitative E. Applicant: Siemens Healthcare Diagnostics, Inc. F. Proprietary and Established Names: Dimension Vista MMB Assay (Mass creatine kinase MB isoenzyme) Dimension Vista 1500 System G. Regulatory Information: 1. Regulation section: 21 CFR 862.1215, Creatine phosphokinase/creatine kinase or isoenzymes test system 21 CFR 862.2160, Discrete photometric chemistry analyzer for clinical use 2. Classification: Class II Class I {1} 3. Product code: JHY JJE 4. Panel: Chemistry (75) H. Intended Use: 1. Intended use(s): See Indication(s) for use below. 2. Indication(s) for use: Dimension Vista MMB Assay: The MMB method is an in vitro diagnostic test for the quantitative measurement of mass creatine kinase MB isoenzyme (EC 2.7.3.2) in human serum and plasma on the Dimension Vista System for confirmation of acute myocardial infarction. Dimension Vista 1500 System: The Siemens Healthcare Diagnostics Dimension Vista 1500 System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric and integrated ion-selective multisensory technology for clinical use. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Performance data was obtained using the Dimension Vista 1500 System. I. Device Description: The Dimension Vista MMB Assay is available as a prepackaged Flex cartridge for use on the Dimension Vista 15001500 System. Each cartridge contains biotinylated monoclonal antibody, mass creatine kinase MB isoenzyme Chemibeads, Streptavidin Sensibeads, and 2 {2} assay buffer. The Dimension Vista 1500 System is a floor model, fully automated, microprocessor-controlled, integrated instrument system that uses prepackaged Flex reagent test cartridges to measure a variety of analytes in human body fluids. The system is a multi-functional analytical tool that processes chemical and immunochemical methodologies, utilizing photometric, turbidimetric, chemiluminescence, nephelometric, and integrated ion selective multisensory detection technologies for clinical use. The Dimension Vista 1500 SystemSystem can analyze up to 1500 tests/hour (typical, depending on the method mix) using a variety of analytical detection capabilities. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Dimension Vista MMB AssayAssay Dimension Vista 1500 Integrated System 2. Predicate 510(k) number(s): k970343 k051087 3. Comparison with predicate: | SimilaritiesSimilarities and Differences | | | | --- | --- | --- | | Item | Candidate Device Dimension Vista MMB Flex Assay on the Dimension Vista 1500 System | Predicate Dimension Vista MMB Flex Assay/Assay (k970343, k051087) | | Intended Use | The MMB method is an in vitro diagnostic test for the quantitative measurement of mass creatine kinase MB isoenzyme (EC 2.7.3.2) in human serum and plasm on the Dimension Vista System for confirmation of acute myocardial infarction. | Same | | Assay Range | 1 – 300 ng/mL | 0.5 – 300 ng/mL | | Sample Type | Human serum and plasma | Same | | Technology | LOCI technology | Same | | Sample Size | 5 μL | Same | | Reagents and antibody | Biotinylated monoclonal | Same | {3} | SimilaritiesSimilarities and Differences | | | | --- | --- | --- | | Item | Candidate Device Dimension Vista MMB Flex Assay on the Dimension Vista 1500 System | Predicate Dimension Vista MMB Flex AssayAssay (k970343, k051087) | | | antibody, mass creatine kinase MB isoenzyme Chemibeads, Streptavidin Sensibeads, assay buffer | | | Instrument | Dimension Vista 1500 System | Dimension Vista 1500 Integrated System | | | Photomultiplier tube - multiplier channel: multiple dynodes | Photomultiplier tube - multiplier channel: enhanced glass single surface tube | | | Updated component area network | | K. Standard/Guidance Document Referenced (if applicable): CLSI EP09-A2 “Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline” CLSI EP05-A2 “Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline” CLSI EP-6A “Evaluation of Linearity of Quantitative Measurement Procedures: a Statistical approach” CLSI EP17-A2 “Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition” L. Test Principle: The MMB method is a homogeneous sandwich chemiluminescent immunoassay based on LOCI technology. LOCI reagents include two synthetic bead reagents and a biotinylated anti-mass creatine kinase MB isoenzyme monoclonal antibody fragment. The first bead reagent (Sensibeads) is coated with streptavidin and contains photosensitive dye. The second bead reagent (Chemibeads) is coated with a second anti-mass creatine kinase MB isoenzyme monoclonal antibody and contains chemiluminescent dye. The samplessamplesare incubated with Chemibeads and biotinylated antibody to form a beadmass creatine kinase MB isoenzyme-biotinylated antibody sandwich. The sensibeadsssensibeadss are added and bind to the biotin to form bead-pair immunocomplexes. Illumination of the complex by light at $680~\mathrm{nm}$ generates singlet oxygen from Sensibeads which diffuses into the Chemibeads, {4} triggering a chemiluminescent reaction. The resulting signal is measured at $612\mathrm{nm}$ and is a direct function of the mass creatine kinase MB isoenzyme concentration in the sample. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Reproducibility testing was conducted in accordance with the CLSI Guideline EP5-A2. Testing included three levels of commercially available quality control material at normal, cut-off, and abnormal CK-MB concentrations and two patient plasma pools spiked with method calibrator to normal and abnormal CK-MB concentrations. For each test level, a single test from two independent cups was analyzed twice per day for 20 days with 1 reagent lot and 3 instruments. | Material | Mean ng/mL | Repeatability (within-run) | | Within-Lab Precision | | | --- | --- | --- | --- | --- | --- | | | | SD | %CV | SD | %CV | | QC1 | 8.66 | 0.15 | 1.72 | 0.22 | 2.59 | | QC2 | 24.13 | 0.36 | 1.50 | 0.56 | 2.30 | | QC3 | 70.73 | 0.91 | 1.28 | 1.27 | 1.80 | | Plasma Pool 1 | 3.26 | 0.15 | 4.71 | 0.19 | 5.99 | | Plasma Pool 2 | 6.15 | 0.15 | 2.45 | 0.21 | 3.40 | # b. Linearity/assay reportable range: Linearity across the assay range (1 to $300\mathrm{ng / mL}$ ) was confirmed according to CLSI Guideline EP6-A. A sample containing a high concentration of MMB was serially diluted with the zero level calibrator to produce 10 test samples ranging from 0 to $313.5\mathrm{ng / mL}$ . Each dilution was assayed in replicates of 5. Data were analyzed using linear regression analysis. A summary of the linearity data is presented below. | Range of Samples | Slope | Intercept | Correlation Coefficient. | n | | --- | --- | --- | --- | --- | | 0–313.5ng/mL | 1.004 | 0.170 | 0.999 | 10 | {5} c. Traceability, Stability, Expected values (controls, calibrators, or methods): Dimension Vista MMB Calibrators (liquid, bovine serum albumin based products containing human heart CK-MB isoenzyme) were previously cleared in k970336. d. Detection limit: Detection limit studies were carried out according to CLSI EP17-A2. For the limit of blank (LOB), 4 zero level calibrator samples were tested over 3 days on 2 reagent lots at 5 replicates per day for a total of 60 measurements per lot. The limit of detection (LOD) and limit of quantitation (LOQ) were determined using a patient serum sample pool diluted to 5 levels targeted to fall between 0.5 and 3.6 ng/mL. 8 replicates per day were run over 5 days with 2 lots for a total of 40 measurements per lot. The LOD and the LOQ were calculated based on within-laboratory precision profiles according to CLSI EP17-A2. The LOQ corresponds to a within-laboratory imprecision coefficient of variation of ≤ 20% at a MMB concentration of ≤ 1.0 ng/mL. Results are shown below: $$ \mathrm{LOB} = 0.4\ \mathrm{ng/mL} $$ $$ \mathrm{LOD} = 0.8\ \mathrm{ng/mL} $$ $$ \mathrm{LOQ} = 1.0\ \mathrm{ng/mL} $$ These studies support a low end for the assay measuring range of 1.0 ng/mL - . e. Analytical specificity: Specificity studies were performed as described in k970343. None of the substances tested showed significant interference with the MMB assay. Since there were no changes to the MMB assay antibody or concentration of the antibody conjugates, specificity testing was not repeated. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: {6} The Dimension MMB assay was evaluated on the proposed Dimension Vista 1500 System with new PMT and compared to the Dimension MMB assay on the predicate Dimension Vista 1500 Integrated System with the current photomultiplier tube. One hundred eleven (111) de-identified, natural human serum and plasma samples were assayed in duplicate across the assay range; however, only the first result was used in each analysis. | Method | Predicate Sample Range Analyzed (ng/mL) | Slope (95% CI) | Intercept ng/mL (95% CI) | Correlation Coefficient (standard linear regression) | n | | --- | --- | --- | --- | --- | --- | | MMB (Passing Bablok) | 1.0 – 274.0 | 0.99 (0.98 – 0.99) | -0.16 (-0.20 – -0.10) | Not applicable | 111 | | | | | | | | | MMB (linear regression) | 1.0 – 274.0 | 0.97 | 0.64 | 0.999 | 111 | b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: A literature-based clinical cut-off can be found in the labeling. 5. Expected values/Reference range: A reference range of 0.5 – 3.6 ng/mL in serum was determined in k970343 as the central 95% in a population of 257 healthy adults ages 19-75. The labeling includes the {7} following statement: Each laboratory should establish its own expected values for mass creatine MB isoenzyme as performed on the Dimension Vista System. ## N. Instrument Name: Dimension Vista 1500 System ## O. System Descriptions: 1. Modes of Operation: Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? ☐ Yes ☑ or ☐ No Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? ☐ Yes ☑ or ☐ No 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: ☐ Yes ☑ or ☐ No 3. Specimen Identification: Previously described in k051087. 4. Specimen Sampling and Handling: Previously described in k051087. 5. Calibration: Previously described in k051087. 6. Quality Control: Previously described in k051087. {8} P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: Not applicable. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...