INHIBIN A CALIBRATORS & CONTROLS ON THE ACCESS IMMUNOASSAY SYSTEMS

K070144 · Beckman Coulter, Inc. · JIT · Feb 1, 2007 · Clinical Chemistry

Device Facts

Record IDK070144
Device NameINHIBIN A CALIBRATORS & CONTROLS ON THE ACCESS IMMUNOASSAY SYSTEMS
ApplicantBeckman Coulter, Inc.
Product CodeJIT · Clinical Chemistry
Decision DateFeb 1, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The Access Inhibin A calibrators are intended to calibrate the Access Inhibin A assay for the quantitative determination of dimeric Inhibin A levels in human serum and plasma using the Access Immunoassay Systems. The Access Inhibin A QC is intended for monitoring system performance of the Access Inhibin A assay.

Device Story

Access Inhibin A Calibrators and QC are accessory materials for the Access Immunoassay System; used to calibrate and monitor system performance for quantitative measurement of dimeric Inhibin A in human serum and plasma. Calibrator kit includes seven levels; control kit includes three levels. System utilizes competitive binding immunoenzymatic method. Used in clinical laboratory settings by professional operators. Output provides calibration curve generation and system performance monitoring, enabling accurate quantitative analyte measurement for clinical decision-making.

Clinical Evidence

No clinical data. Bench testing only. Stability studies (real-time) confirmed 28-day open-vial stability at 2–30 °C and 34-day shelf life. Traceability to WHO 91/624 established.

Technological Characteristics

Liquid solution format; matrix consists of Bovine Serum Albumen; aligned to WHO international standard 91/624. Designed for use with automated competitive binding immunoenzymatic assay systems.

Indications for Use

Indicated for use as calibrators and quality controls for the quantitative determination of dimeric Inhibin A levels in human serum and plasma on Access Immunoassay Systems.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K070144 ### Access Inhibin A Calibrators & Controls 510(k) Summary #### 1.0 Submitted By: Tyler Foutch Regulatory Affairs Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 368-1653 FAX: (952) 368-7610 #### 2.0 Date Submitted January 15, 2007 #### 3.0 Device Name(s): - 3.1 Proprietary Names Access® Inhibin A Calibrators Access® Inhibin A Controls 3.2 Classification Names Calibrator (21 CFR § 862.1150) Control (21 CFR § 862.1660) #### 4.0 Legally Marketed Device The Access Inhibin A Calibrators & Controls claim substantial equivalence to the Diagnostic Systems Laboratories (DSL) ACTIVE® Inhibin A ELISA Calibrators and Controls currently in commercial distribution, FDA 510(k) Number K002128. #### 5.0 Device Description The Access Inhibin A Calibrators & Controls are designed for use with the Access Inhibin A Reagent for generation of the Inhibin A assay calibration curve on Beckman Coulter's Access immunoassay Systems. The Access Inhibin A Calibrator kit contains 7 X 2.5 mL bottles, one for each of seven calibrator levels. The Access Inhibin A Control kit contains 3 X 2.5 mL bottles, one for each of three control levels. The Access Immunoassay Systems utilize a competitive binding immunoenzymatic method for quantitative analyte measurement. FEB - 】2007 {1}------------------------------------------------ #### 6.0 Intended Use The Access Inhibin A calibrators are intended to calibrate the Access Inhibin A assay for the quantitative determination of dimeric Inhibin A levels in human serum and plasma using the Access Immunoassay Systems. The Access Inhibin A QC is intended for monitoring system performance of the Access Inhibin A assay. #### 7.0 Comparison to the Predicate The Access Inhibin A Calibrators and the predicate calibrator utilize different test methodologies, automated competitive binding immunoenzymatic method for the Access and manual ELISA method for the DSL. The Access Inhibin A calibrators and controls are provided in Bovine Serum Albumen, while the predicate calibrator and control are provided in Fetal Bovine Serum. Both are aligned to WHO international standard 91/624. Both devices come in a liquid solution format. The Access kit contains three controls while the predicate contains two controls. The measured range differs between the two devices, the Access assay range is 0-1500 pg/mL, while the predicate range is 0-1000 pg/mL #### 8.0 Summary of Performance Data The Access Inhibin A Calibrators have been demonstrated to be equivalent to the predicate product. Based on the results of the performance characteristics testing, these calibrator materials meet product claims and specifications. Performance data from validation testing supports a finding of substantial equivalence to the Diagnostic Systems Laboratories (DSL) ACTIVE® Inhibin A ELISA Calibrators and Controls already in commercial distribution. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with three curved lines representing its wings or body. The logo is black and white. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Tyler Foutch Regulatory Affairs Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 FEB - 1 2007 Re: k070144 > Trade/Device Name: Access Inhibin A Calibrators and QC Regulation Number: 21 CFR § 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT, JJX Dated: January 15, 2007 Received: January 18, 2007 Dear Mr. Foutch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please othe the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the from the (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean m. Cooper, M.S., D.V.M. Sean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): k070144 Device Name:_Access Inhibin A Calibrators and QC Indications For Use: The Access Inhibin A calibrators are intended to calibrate the Access Inhibin A assay for the quantitative determination of dimeric Inhibin A levels in human serum and plasma using the Access Immunoassay Systems. The Access Inhibin A QC is intended for monitoring system performance of the Access Inhibin A assay. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carol C Benson Vision Sign-Off Office of In Vitro Diagnostic Device Braluation and Safety Page 1 of ____________________________________________________________________________________________________________________________________________________________________ K070144
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...