DIMENSION CLINICAL CHEMISTRY SYSTEM LIPASE CALIBRATOR, MODEL DC56
K081319 · Siemens Healthcare Diagnostics, Inc. · JIT · Aug 4, 2008 · Clinical Chemistry
Device Facts
Record ID
K081319
Device Name
DIMENSION CLINICAL CHEMISTRY SYSTEM LIPASE CALIBRATOR, MODEL DC56
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Aug 4, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The LIPL Calibrator is an in vitro diagnostic product to be used to calibrate the Lipase (LIPL) method for the Dimension® clinical chemistry systems.
Device Story
The Dimension® Lipase Calibrator (LIPL CAL - DC56) is a liquid, bovine serum albumin-based product used to calibrate Lipase (LIPL) assays on Dimension® clinical chemistry systems. The kit consists of three levels of calibrators: Level 1 contains no detectable lipase; Levels 2 and 3 contain porcine pancreas lipase. The product is ready-to-use with no preparation required. It is intended for use by laboratory professionals in clinical settings. The calibrator allows the clinical chemistry system to establish a calibration curve, ensuring accurate measurement of lipase levels in patient samples. This supports clinicians in diagnosing and monitoring conditions associated with lipase activity, such as pancreatic disorders.
Clinical Evidence
No clinical data. Performance is established through bench testing, specifically verifying the assigned values of the calibrator levels against the Master Pool and ensuring traceability to an external reference system (PBS/Precical®).
Technological Characteristics
Liquid, bovine serum albumin-based matrix. Contains porcine pancreas lipase (Levels 2 and 3). Three-level kit. Traceable to Master Pool values. No specific energy source or software algorithm; functions as a chemical reagent for instrument calibration.
Indications for Use
Indicated for use as an in vitro diagnostic calibrator for the Lipase (LIPL) method on Dimension® clinical chemistry systems. No specific patient population, age, or gender restrictions are defined as this is a laboratory reagent product.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dimension Vista® Enzyme 1 Calibrator
Related Devices
K061923 — DIMENSION VISTA ENZYME 1 CALIBRATOR (ENZ 1 CAL - KC310) · Dade Behring, Inc. · Aug 25, 2006
K970905 — SYNCHRON CX SYSTEMS LIPASE (LIPA) REAGENT · Beckman Instruments, Inc. · Jun 4, 1997
K081789 — DIMENSION CLINICAL CHEMISTRY SYSTEM ENZYME I CALIBRATOR · Siemens Healthcare Diagnostics, Inc. · Sep 10, 2008
K061750 — DIMENSION VISTA SYSTEM LIPID CALIBRATOR, MODEL KC220 · Dade Behring, Inc. · Aug 11, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
| | 510(k) Summary for the | |
|--|--------------------------------------------------------|--|
| | Dimension® Clinical Chemistry System Lipase Calibrator | |
| | (LIPL CAL - DC56) | |
AUG - 4 2008
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA
1990 and 21 CFR 807.92.
May 8, 2008
| A. 510(k) Number: | 2081319 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| B. Analyte: | Lipase |
| C. Type of Test: | Calibrator Material |
| D. Applicant: | Siemens Healthcare Diagnostics Inc, P.O. Box 6101, Newark, DE 19714-6101 |
| | Victor M. Carrio, Regulatory Affairs and Compliance Manager |
| | Office: (302) 631-0376 Fax: (302) 631-6299 |
| E. Proprietary and Established Names: | Dimension® Clinical Chemistry System Lipase Calibrator |
| | (LIPL CAL - DC56) |
| F. Regulatory Information: | |
| 1. Regulation section: | 21 CFR § 862-1150 - Calibrator |
| 2. Classification: | Class II |
| 3. Product Code: | JIT - Secondary Calibrator |
| 4. Panel: | Clinical Chemistry |
| G. Standard/Guidance Document Referenced: | |
| | Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final,02/22/1999 |
| | ISO 14971:2007 Medical devices -Application of risk management to medical devices |
| H. Intended Use: | The LIPL Calibrator is an in vitro diagnostic product to be used to calibrate the Lipase (LIPL) method for the Dimension® clinical chemistry systems. |
| I. Device Description: | The LIPL calibrator is a liquid bovine serum albumin-based product. The level 1 calibrator contains no detectable lipase. Levels 2 and 3 contain porcine pancreas lipase. The kit consists of six vials, two vials of Calibrator Level 1, two vials of Calibrator Level 2, |
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
i
1
{1}------------------------------------------------
and two vials of Calibrator Level 3 which are ready for use (no preparation is required). The volume per vial is 1.0 mL.
J. Substantial Equivalence Information:
The Dimension® Clinical Chemistry System Lipase Calibrator (DC56) and the predicate device, the Dimension Vista® Enzyme 1 Calibrator were compared. The following table provides a comparison of the important similarities and differences between the device and the predicate:
| Item | Device<br>Dimension® clinical<br>chemistry system Liquid<br>Lipase calibrator | Predicate Device<br>Dimension Vista® System<br>Enzyme 1 Calibrator |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The LIPL Calibrator is an in<br>vitro diagnostic product to be<br>used to calibrate the Lipase<br>(LIPL) method for the<br>Dimension® clinical chemistry<br>systems. | The Vista® System Enzyme 1<br>Calibrator is an <i>in vitro</i><br>diagnostic product for the<br>calibration of Amylase (AMY),<br>Gamma-Glutamyl Transferase<br>(GGT), Lactate Dehydrogenase<br>(LDH), Lipase (LIP), and<br>Pseudocholinesterase (PCHE)<br>methods on the Dimension<br>Vista® System. |
| Analytes | Lipase | Amylase, Gamma-Glutamyl<br>Transferase, Lactate<br>Dehydrogenase, Lipase, and<br>Pseudocholinesterase |
| Form | Liquid | Liquid |
| Traceability | Master Pool, Dimension®<br>clinical chemistry system<br>values. | Master Pool, Dimension® clinical<br>chemistry system values. |
| Matrix | Bovine serum base with Lipase<br>(porcine pancreas). | Bovine serum base with amylase<br>(human saliva), GGT (bovine<br>kidney), LDH (chicken heart),<br>lipase (porcine pancreas), and<br>PCHE (horse serum). |
| Number of<br>Levels | Three (3) levels. | Two (2) levels. |
K. Performance Characteristics:
Traceability: The assigned values of the Dimension® LIPL Calibrator are traceable to Master Pool values, assigned on the Dimension® clinical chemistry system.
Master Pool level 1 of BSA base is assigned a value of 0 U/L. LIPL Master Pool bottle values levels 2 and 3 are assigned on multiple instruments calibrated with LIPL Anchor Pool. The LIPL Anchor Pool values are assigned using an external reference system (PBS/Precical®). A previous Master Pool lot is used as a control. Calibrators are prepared gravimetrically from porcine lipase. The concentration of each level is verified against Master Pool
{2}------------------------------------------------
| values. | | | |
|-----------------|--|--|--|
| File Attachment | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AUG - 4 2008
Siemens Healthcare Diagnostics Inc. c/o Mr. Victor M. Carrio Senior Manager of Regulatory Affairs P.O. Box 6101, Mailbox 514 Newark, DE 10714-6101
Re: k081319
Trade Name: Dimension® Lipase Calibrator (DC56) Regulation Number: 21 CFR §862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT Dated: July 11, 2008 Received: July 14, 2008
Dear Mr. Carrio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollofre and (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Jean M. Coopes, M.S., D.v.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indication for Use
510(k) Number (if known): K081319
Device Name:
Dimension® Lipase Calibrator (DC56)
Indication For Use:
The LIPL Calibrator is an in vitro diagnostic product to be used to calibrate the Lipase (LIPL) method for the Dimension® clinical chemistry systems.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Carol C. Benson
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K081319
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.