K060371 · Dade Behring, Inc. · JIT · Mar 9, 2006 · Clinical Chemistry
Device Facts
Record ID
K060371
Device Name
THE EMIT 2000 TACROLIMUS CALIBRATORS
Applicant
Dade Behring, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Mar 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The Emit® 2000 Tacrolimus Calibrators are intended for use as a reference in measuring tacrolimus in human whole blood using the Emit® 2000 Tacrolimus Assay.
Device Story
Emit® 2000 Tacrolimus Calibrators serve as reference materials for tacrolimus quantification in human whole blood. The kit contains six levels of tacrolimus (0, 2.5, 5, 10, 20, and 30 ng/mL) in preserved whole blood hemolysate. Used in clinical laboratory settings to calibrate the Emit® 2000 Tacrolimus Assay. By establishing a reference curve, the calibrators enable the assay to accurately measure tacrolimus concentrations, assisting clinicians in monitoring therapeutic drug levels in patients.
Clinical Evidence
No clinical data. Bench testing only. Stability studies (real-time) conducted over 37 months for shelf-life and 12 weeks for open-vial stability. Traceability established via gravimetric preparation of stock solution and verification against a master lot calibrated by LC/MS.
Technological Characteristics
Calibrator kit containing tacrolimus in preserved human whole blood hemolysate. Six levels: 0, 2.5, 5, 10, 20, 30 ng/mL. Materials tested for HBsAg, HIV 1/2, HCV, and Syphilis. Storage: -20°C (shelf-life) and 9°C (open-vial).
Indications for Use
Indicated for use as a reference standard for the measurement of tacrolimus in human whole blood samples. Intended for prescription use.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K060503 — DIMENSION TACR CALIBRATOR, MODEL DC107 · Dade Behring, Inc. · May 18, 2006
K060385 — EMIT 2000 TACROLIMUS ASSAY AND SAMPLE PRETREATMENT REAGENT · Dade Behring, Inc. · Apr 6, 2006
K061503 — DIMENSION CYCLOSPORINE EXTENDED RANGE CALIBRATOR (DC108A) · Dade Behring, Inc. · Jun 22, 2006
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k060371
B. Purpose for Submission:
New Device
C. Measurand:
Tacrolimus
D. Type of Test:
Calibrator materials for tacrolimus
E. Applicant:
DADE BEHRING, INC.
F. Proprietary and Established Names:
Emit® 2000 Tacrolimus Calibrators
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JIT | II | 21 CFR 862.1150 | 75 |
H. Intended Use:
1. Intended use(s):
The Emit® 2000 Tacrolimus Calibrators are intended for use as a reference in measuring tacrolimus in human whole blood using the Emit® 2000 Tacrolimus Assay.
2. Indication(s) for use:
The Emit® 2000 Tacrolimus Calibrators are intended for use as a reference in measuring tacrolimus in human whole blood using the Emit® 2000 Tacrolimus Assay
3. Special conditions for use statement(s):
{1}
For prescription use only.
4. Special instrument requirements:
Not Applicable (N/A)
I. Device Description:
The calibrators contain tacrolimus in preserved whole blood hemolysate. The calibrator kit consists of one vial of each calibrator level with target concentrations of 0, 2.5, 5, 10, 20, and 30 ng/mL of tacrolimus.
The calibrators contain human blood components. Each lot was tested and found to be non-reactive for Hepatitis B Surface Antigen (HBsAg), Antibody to Human Immunodeficiency Virus (HIV 1/2), Antibody to Hepatitis C Virus (HCV) and Antibody to Syphilis.
J. Substantial Equivalence Information:
| Predicate | Abbott IMx Tacrolimus II Calibrator, P970007 |
| --- | --- |
| | |
| Describe the item being compared | |
| The Abbott IMx Tacrolimus II Calibrators are for use in calibrating the Abbott IMx Tacrolimus II Assay. The Abbott Calibrators are prepared in preserved human whole blood with preservatives. There are 6 calibrator levels with nominal values of 0, 3, 6, 12, 20 and 30 ng/mL tacrolimus. | |
| Similarites | |
| Both devices are calibrators intended for use as a reference in measuring tacrolimus with their respective assays. | |
| Both devices contain tacrolimus in preserved whole blood hemolysate. | |
| Both devices contain 6 levels of calibrators. | |
| Differences | |
| The Emit® 2000 Tacrolimus Calibrators consist of 6 calibrator materials with concentrations of -0, 2.5, 5, 10, 20 and 30 ng/mL while the Abbott IMx® Tacrolimus II Calibrators consist of 6 calibrator materials with concentrations of -0, 3, 6, 12, 20, and 30 ng/mL. | |
K. Standard/Guidance Document Referenced (if applicable):
| GUIDANCE | | | |
| --- | --- | --- | --- |
| Document Title | Office | Division | Web Page |
| Class II Special Controls
Guidance Document:
Cyclosporine and Tacrolimus | OIVD | DCTD | http://www.fda.gov/cdrh/ode/guidance/1380.html |
{2}
3
Assays; Guidance for Industry and FDA
L. Test Principle:
N/A
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
N/A
b. Linearity/assay reportable range:
N/A
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The tacrolimus frozen liquid stock solution was prepared gravimetrically from commercially available pharmaceutical grade tacrolimus. Calibrators are prepared by directly spiking the stock solution into whole blood hemolysate to form the target concentrations. Verification occurred via recovery against a master lot. The master lot is formulated by diluting the stock into the whole blood hemolysate with the preservatives at six different levels and stored at 20°C. The potential master lot of calibrators is assigned by multiple runs using the Emit Tacrolimus Assay calibrated with the existing master lot of calibrators verified by LC/MS.
Shelf-life and open vial stability was conducted using real-time data from 3 lots of calibrators. Each level of calibrators was tested in duplicate and replicates of 6 with commercially available control material. The calibrators (stored at -20°C) were tested at 0,1,2,3,7,9,13,15,22,25,31 and 37 months. An open vial stability (stored at 9°C) study was conducted at 0, 8 and 12 weeks. Verification of the 8 week open vial stability throughout the 36 month shelf life, the open vial stability was repeated from 7 to 37 months at 6 month intervals. The study results support the sponsors 8 week open vial and 36 month shelf life stability.
d. Detection limit:
N/A
e. Analytical specificity:
{3}
N/A
f. Assay cut-off:
N/A
2. Comparison studies:
a. Method comparison with predicate device:
N/A
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
4. Clinical cut-off:
N/A
5. Expected values/Reference range:
The concentrations in the Emit Tacrolimus Calibrators have lot specific values that range from 1 to 30 ng/mL.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
{4}
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
5
{5}
6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches