A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.
Device Story
Olympus LDL Cholesterol Reagent and Calibrator; in vitro diagnostic reagents for quantitative determination of LDL cholesterol. Modification involves change in formulation and manufacturing source. Device used in clinical laboratory settings by trained personnel. Output provides quantitative LDL cholesterol levels to assist clinicians in lipid profile assessment and cardiovascular risk management. Fundamental scientific technology remains unchanged from predicate.
Clinical Evidence
No clinical data provided; substantial equivalence is based on the device's intended use and classification as a lipoprotein test system and calibrator.
Technological Characteristics
In vitro diagnostic reagent and calibrator system. Quantitative measurement of lipoprotein in serum/plasma. Class II device under 21 CFR 862.1150.
Indications for Use
Indicated for measurement of lipoprotein in serum and plasma to assist in diagnosis and treatment of lipid disorders (e.g., diabetes mellitus), atherosclerosis, and liver/renal diseases. Intended for prescription use.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification ODE Review Memorandum
To: THE FILE
RE: DOCUMENT NUMBER K040471
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Reserved Class I device. The following items are present and acceptable (delete/add items as necessary) for the OLYMPUS LDL CHOLESTEROL REAGENT, OSR6196/OSR6296; OLYMPUS LDL CHOLESTEROL CALIBRATOR, ODC0024:
1. The name and 510(k) number of OLYMPUS AMERICA, INC.'s previously cleared device, OLYMPUS LDL CHOLESTEROL REAGENT, OLYMPUS LDL CHOLESTEROL CALIBRATOR cleared under K014032. (For a preamendments device, a statement to this effect has been provided.)
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATION (S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for a modification in formulation, by a different manufacturer.
Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics.
4. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Comments – Validation Protocols were adequate to address hazards identified in the Risk analysis (FMEA)
revised:3/27/98