K203400 · Trudell Medical International · CAF · Jun 14, 2021 · Anesthesiology
Device Facts
Record ID
K203400
Device Name
Corrugated tube with mouthpiece accessory
Applicant
Trudell Medical International
Product Code
CAF · Anesthesiology
Decision Date
Jun 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The corrugated tube is an accessory to the MC 300 ® Nebulizer. It is designed to work in conjunction with the MC 300 ® Nebulizer to extend the patient interface away from the nebulizer. The accessory is a single patient is intended to be used with either mouthpiece or mask for pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician.
Device Story
Corrugated tube accessory extends patient interface away from MC 300 Nebulizer; used with mouthpiece or mask. Device is disposable, single-patient use, non-sterile. Operates via pneumatic jet nebulization (inherited from predicate). Used in hospital, clinic, or home environments by patients under physician care. Accessory supplements parent device performance without altering principle of operation. Benefits include increased flexibility in patient positioning during aerosolized medication delivery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including aerosol characterization (Total Mass, Fine Particle Fraction, MMAD, GSD) across multiple medications (Albuterol Sulfate, Budesonide, Ipratropium Bromide) and biocompatibility testing (ISO 10993-1, 3, 5, 10, 11, 12, 17, 18).
Technological Characteristics
Plastic molded corrugated tube accessory. Pneumatic jet nebulization principle. Non-sterile, disposable, single-patient use. Biocompatibility per ISO 10993. Dry gas pathway VOC and PM2.5 assessment performed.
Indications for Use
Indicated for pediatric (ages 2 years and above) and adult patients under the care of a licensed healthcare provider or physician for use with the MC 300 Nebulizer to deliver aerosolized inhalation medications.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Trudell Medical International Marianne Tanton Director, Quality and Regulatory Affairs 725 Baransway Drive London, Ontario N5V 5G4 Canada
### Re: K203400
Trade/Device Name: Corrugated tube with mouthpiece accessory Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: May 20, 2021 Received: May 21, 2021
### Dear Marianne Tanton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brandon Blakely, PhD Acting Assistant Director DHT1C: Division of ENT. Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203400
#### Device Name
Corrugated tube with mouthpiece accessory
#### Indications for Use (Describe)
The corrugated tube is an accessory to the MC 300 ® Nebulizer. It is designed to work in conjunction with the MC 300 ® Nebulizer to extend the patient interface away from the nebulizer. The accessory is a single patient is intended to be used with either mouthpiece or mask for pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> |
|----------------------------------------------|-------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div></div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Prepared: 14 June 2021
### 1. Submitter
Trudell Medical International 725 Baransway Drive London, Ontario N5V 5G4, Canada
Contact: Marianne Tanton Director, Quality and Regulatory Affairs Phone: 1-519-455-7060 Email: mtanton@trudellmed.com
### 2. Device Name
| Trade Name: | None |
|----------------------|-------------------------------------------|
| Common Name: | Corrugated tube with mouthpiece accessory |
| Classification Name: | Nebulizer 21 CFR 868.5630 |
| Regulatory Class: | II |
| Product Code: | CAF |
### 3. Predicate Device
MC300* Nebulizer - K173367 Trudell Medical International
The predicate device has not been subject to a recall.
### 4. Device Description and Principle of Operation
The corrugated tube with mouthpiece accessory is designed to work in conjunction with the cleared predicate device, MC300 * Nebulizer (K173367) to extend the patient interface away from the nebulizer.
In line with FDA's definition of a "medical device accessory", the corrugated tube with mouthpiece accessory is intended to supplement the performance of the parent device, MC300* Nebulizer (K173367).
The principle of operation of MC300* Nebulizer remains the same as cleared under K173367.
### 5. Indications for Use
The corrugated tube is an accessory to the MC 300 ® Nebulizer. It is designed to work in conjunction with the MC 300 ® Nebulizer to extent interface away from the nebulizer. The accessory is a single patient use device that is intended to be used with either
<sup>*</sup>Trademarks and registered trademarks of Trudell Medical International
{4}------------------------------------------------
mouthpiece or mask for pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician.
The MC300* Nebulizer indications for use remains the same as cleared under K173367.
### 6. Comparison to predicate device
The predicate device, MC300* Nebulizer (K173667) and the subject device (corrugated tube with mouthpiece) used with MC300* Nebulizer, are identical in purpose, function, core technology and method of operation.
The purpose of this submission is the addition of corrugated tube with mouthpiece accessory for use with the predicate device, MC300* Nebulizer (K173367). Table 1 provides a comparison of the subject and predicate devices.
| Element of<br>Comparison | Corrugated tube with<br>mouthpiece to be used<br>with MC300* Nebulizer<br>(Subject Device) | MC300* Nebulizer<br>K173367<br>(Predicate Device) | Comparison |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use | The nebulizer indications<br>for use remains the same<br>as cleared under K173367.<br>The corrugated tube is an<br>accessory to the MC 300®<br>Nebulizer. It is designed to<br>work in conjunction with the<br>MC 300® Nebulizer to<br>extend the patient interface<br>away from the nebulizer.<br>The accessory is a single<br>patient use device that is<br>intended to be used with<br>either mouthpiece or mask<br>for pediatric (ages 2 years<br>and above) and adult<br>patients, who are under the<br>care of a licensed<br>healthcare provider or<br>physician. | The nebulizer is intended to<br>be used with pediatric<br>(ages 2 years and above)<br>and adult patients, who are<br>under the care of a<br>licensed healthcare<br>provider or physician. The<br>device is designed to<br>aerosolize prescribed<br>medication for inhalation by<br>a patient in the hospital,<br>clinic or home care<br>environment. The nebulizer<br>is a single patient use<br>device. | Supplement |
| Principle of<br>Operation | The nebulizer's principle of<br>operation remains the<br>same as cleared under<br>K173367. | Pneumatic Jet Nebulizer | Supplement |
### Table 1: Comparison to Predicate Device
<sup>*</sup>Trademarks and registered trademarks of Trudell Medical International
{5}------------------------------------------------
| Element of<br>Comparison | Corrugated tube with<br>mouthpiece to be used<br>with MC300* Nebulizer<br>(Subject Device) | MC300* Nebulizer<br>K173367<br>(Predicate Device) | Comparison |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|------------|
| | The corrugated tube with<br>mouthpiece is an accessory<br>to the MC300* Nebulizer. It<br>is designed to work in<br>conjunction with the<br>MC300* Nebulizer to<br>extend the patient interface<br>away from the nebulizer. | | |
| Environment of<br>use | Hospital, Clinic or Home | | Similar |
| Patient<br>population | Adult and pediatric patients<br>(ages 2 years and above) | | Similar |
| Single Patient<br>Use | Yes | | Similar |
| Aerosolization | Continuous during inhalation and exhalation | | Similar |
| Type of device | Disposable, single patient use, prescription only, non-<br>sterile | | Similar |
| Manufacturing<br>process | Plastic molding | | Similar |
Note: Type of gas source, flow rates and maximum fill volume are nebulizer characteristics and are not listed in the comparison table.
## 7. Performance Data
### 7.1. Aerosol Characterization
Aerosol characterization testing was performed in accordance with relevant sections of the CDRH Guidance Document "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH - 1993). The test results demonstrated substantially equivalent in-vitro performance between the subject device (corrugated tube with mouthpiece or mask) used with MC300* Nebulizer, and the predicate device, MC300* Nebulizer (K173367). The results of the testing are summarized in Tables 2 to 4 below.
{6}------------------------------------------------
#### Table 2. Summary of Aerosol Characterization Test Results for Corrugated Tube with Mouthpiece (Subject Device) Used with MC300* Nebulizer and MC300* Nebulizer (K173367) (Predicate Device) at 8 L/min
| Metric | Albuterol Sulfate | | Budesonide | | Ipratropium Bromide | |
|---------------------------------------------------------------------------|-------------------|---------------------|-------------------|---------------------|---------------------|---------------------|
| | Subject Device | Predicate<br>Device | Subject<br>Device | Predicate<br>Device | Subject<br>Device | Predicate<br>Device |
| Total Mass<br>(μg) | 1248.8±57.8 | 1409.0±84.5 | 403.6±54.6 | 381.2±21.4 | 535.3±28.2 | 583.2±17.1 |
| Fine Particle<br>Fraction<br>(0.98-5.39µm<br>aerodynamic<br>diameter) (%) | 70.5±1.1 | 69.1±1.9 | 53.6±8.2 | 58.9±4.5 | 70.3±0.9 | 70.7±1.8 |
| Fine Particle<br>Mass (µg) | 905.7±51.9 | 974.4±76.7 | 212.9±9.9 | 223.9±14.5 | 376.6±22.1 | 412.7±17.2 |
| MMAD (µm) | 2.9 | 2.9 | 5.0 | 4.6 | 3.4 | 3.1 |
| GSD | 2.2 | 2.2 | 1.7 | 1.8 | 2.2 | 2.1 |
### Table 3. Summary of Aerosol Characterization Test Results for Corrugated Tube with Mouthpiece (Subject Device) Used with MC300* Nebulizer and MC300* Nebulizer (K173367) (Predicate Device) at 13 psi (4 L/min)
| Metric | Albuterol Sulfate | | Budesonide | | Ipratropium Bromide | |
|---------------------------------------------------------------------------|-------------------|---------------------|-------------------|---------------------|---------------------|---------------------|
| | Subject Device | Predicate<br>Device | Subject<br>Device | Predicate<br>Device | Subject<br>Device | Predicate<br>Device |
| Total Mass<br>(µg) | 1394.4±121.8 | 1348.8±158.9 | 358.1±16.5 | 383.3±21.8 | 513.3±40.3 | 555.4±35.8 |
| Fine Particle<br>Fraction<br>(0.98-5.39µm<br>aerodynamic<br>diameter) (%) | 71.6±3.5 | 72.0±2.5 | 59.5±3.9 | 59.4±7.0 | 74.4±1.1 | 72.1±2.4 |
| Fine Particle<br>Mass (µg) | 995.9±67.0 | 971.3±124.8 | 213.0±16.5 | 227.8±22.8 | 381.8±28.2 | 400.8±32.2 |
| MMAD (µm) | 2.8 | 2.4 | 4.7 | 4.6 | 2.4 | 2.9 |
| GSD | 2.2 | 2.2 | 1.7 | 1.9 | 2.2 | 1.9 |
{7}------------------------------------------------
### Table 4. Summary of Aerosol Characterization Test Results for MC300* Nebulizer with Supine Accessory with Mask (Subject Device with Mask) and MC300* Nebulizer with Mask (Predicate Device with Mask) at 8 L/min
| | Albuterol Sulfate | | Budesonide | | Ipratropium Bromide | |
|----------------------------------------------------------------------------|--------------------------------|----------------------------------|--------------------------------|----------------------------------|--------------------------------|----------------------------------|
| Metric | Subject<br>Device with<br>mask | Predicate<br>Device with<br>mask | Subject<br>Device with<br>mask | Predicate<br>Device with<br>mask | Subject<br>Device with<br>mask | Predicate<br>Device with<br>mask |
| Total Mass (µg) | 1396.7±124.7 | 1435.4±92.2 | 356.3±13.0 | 333.5±22.8 | 607.9±33.5 | 570.5±29.8 |
| Fine Particle<br>Fraction (0.98-<br>5.39µm<br>aerodynamic<br>diameter) (%) | 71.5±1.3 | 69.5±2.1 | 64.2±1.7 | 61.3±7.1 | 73.4±1.4 | 70.3±1.9 |
| Fine Particle<br>Mass (µg) | 999.7±103.9 | 997.5±72.5 | 228.8±9.6 | 203.6±16.6 | 446.1±26.0 | 401.3±21.2 |
| MMAD (µm) | 2.6 | 2.8 | 4.3 | 4.5 | 2.6 | 3.1 |
| GSD | 2.2 | 2.2 | 1.8 | 1.8 | 2.1 | 2.0 |
## 7.2. Biocompatibility Testing
Biological endpoints applicable to the corrugated tube with mouthpiece accessory are listed below. Materials were tested in accordance with ISO 10993-1 (2018) and the results satisfied the requirements. All in vitro and in vivo studies were performed and included the following battery of tests: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Genotoxicity, Extractables, and Chemical Characterization with a Biological Risk Assessment.
## Table 5. Summary of Biocompatibility Testing Conducted
| ISO Standard | Test/ Assessment |
|--------------|-------------------------------------------------|
| 10993-1 | Biological Risk Assessment |
| 10993-5 | Cytotoxicity Study Using the ISO Elution Method |
| 10993-3 | Genotoxicity: Bacterial Reverse Mutation Study |
| 10993-3 | Genotoxicity: Mouse Lymphoma Assay |
| 10993-11 | Acute Systemic Toxicity Study in Mice |
{8}------------------------------------------------
| ISO Standard | Test/ Assessment |
|--------------|-----------------------------------------------------------------------------------------------------|
| 10993-10 | Intracutaneous Study in Rabbits |
| 10993-10 | Guinea Pig Maximization Sensitization |
| 10993-17 | Establishment of allowable limits for leachable substances |
| 10993-12 | Solvent and Extraction Condition Verification for ISO 10993-18 Chemical<br>Characterization Program |
| 10993-18 | Chemical characterization of materials |
## 7.3. Dry Gas Pathway Testing
To support the safe use of the corrugated tube with mouthpiece accessory in dry gas conditions, a worst-case assessment of volatile organic compounds (VOCs) and fine particles (particulate matter PM2.5) was conducted. Testing results and risk assessment demonstrated that exposure during use of the corrugated tube with mouthpiece accessory is unlikely to result in toxicological effects.
## 8. Clinical Performance Summary
Not applicable. The determination of substantial equivalence is not based on Clinical Performance data.
### 9. Conclusion
The non-clinical data demonstrate that the use of the corrugated tube with mouthpiece accessory used with MC300* Nebulizer (K173367) is safe and effective for use in patients for the delivery of aerosolized inhalation medications and is therefore substantially equivalent to the predicate device, MC300* Nebulizer, cleared under K173367.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.