K964078 · Medical Industries America, Inc. · CAF · Jan 10, 1997 · Anesthesiology
Device Facts
Record ID
K964078
Device Name
3050 SPORT-NEB
Applicant
Medical Industries America, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Jan 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The model 3050 "Sport-Neb" is a small air compressor designed to provide sufficient air pressure and flow to power a hand held nebulizer such as the Salter #8900.
Device Story
Small, lightweight air compressor; provides air pressure/flow to power hand-held nebulizers (e.g., Salter #8900). Device comprises motor compressor assembly, wiring, switch, power cord, and ABS plastic housing. Used in home or clinical settings; operated by patients or caregivers. Provides pneumatic power to nebulizer; enables aerosol medication delivery. Portable design; smaller form factor than predicate AEROMAX.
Clinical Evidence
Bench testing only; no clinical data provided. Device equivalence established through design comparison and functional similarity to predicate.
Technological Characteristics
ABS plastic housing; dimensions 7.5" x 7.5" x 4.5"; AC-powered motor compressor assembly; pneumatic output via air outlet barb; manual switch operation.
Indications for Use
Indicated for patients requiring aerosolized medication delivery via a hand-held nebulizer powered by an external air compressor.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
K992643 — MODEL IRC 1001 AEROSOL COMPRESSOR · Invacare Corp. · Oct 27, 1999
K970035 — JOHN BUNN NEB-U-LITE II MEDICATION COMPRESSOR WITH DISPOSABLE NEBULIZER, MODEL #JBO-112-009 · Graham-Field, Inc. · Apr 1, 1997
K020797 — PULMO-MIST COMPRESSOR MODEL # 4323 · Nidek Medical Products, Inc. · May 3, 2002
K050183 — SILFAB SUPER 2 MAX COMPRESSOR, N28 SERIES · Silvestrin Fabris S.R.L. · Aug 16, 2005
Submission Summary (Full Text)
{0}
MEDICAL INDUSTRIES AMERICA INC.
2879 R AVENUE
ADEL, IOWA
U.S.A. 50003-8055
TEL. (515) 993-5001
1-800-759-3038
FAX (515) 993-4172
K964078
JAN. 10, 1997
ENCLOSURE 'B'
EXECUTIVE SUMMARY OF THE MODEL 3050 "SPORT-NEB"
The model 3050 "Sport-Neb" is a small air compressor designed to provide sufficient air pressure and flow to power a hand held nebulizer such as the Salter #8900.
Nebulizer compressors are very basic devices comprising of a motor compressor assembly, wiring, switch, power cord and case. Nebulizer compressors have been on the market for many years. "Medical Industries America, Inc." is currently marketing two other nebulizer compressors - the "Aerosol Two" and the "AEROMAX". The "Sport-Neb" is virtually the same as the "AEROMAX" with the exception of the case which is smaller and more portable.
## DEVICE DESCRIPTION:
The Model 3050 "sport-Neb" is a small, lightweight, nebulizer compressor. It measures 7 1/2" X 7 1/2" X 4 1/2". The housing is grey in color and moulded from ABS plastic. External components include a switch, filter housing, AC cord and the air outlet barb. An optional carry bag will be offered to store the patients hand held nebulizer.
The handle is built into the housing
The operating components are located internally. The compressor, some minor wiring and the exhaust/intake plumbing are located inside.
## MODIFICATIONS:
At this time, there are no intentions to modify the device.
26
{1}
MEDICAL INDUSTRIES AMERICA INC.
2879 R AVENUE
ADEL, IOWA
U.S.A. 50003-8055
TEL. (515) 993-5001
1-800-759-3038
FAX (515) 993-4172
ENCLOSURE 'B'
EXECUTIVE SUMMARY OF THE MODEL 3050 "SPORT-NEB" COMPRESSOR (CONTINUED)
DESCRIPTION OF THE MODEL 3002 "AEROMAX" (PREDICATE DEVICE)
The Model 3002 "AEROMAX" manufactured by "Medical Industries America, Inc. is also a small, lightweight nebulizer compressor. It measures approximately 12 X 8 1/2 X 4 1/2. It is grey in color.
In the front of the device is the lid. Under the lid are the following components - switch, AC cord, filter housing and air outlet barb.
The "AEROMAX" has the operating components located internally also. The compressor, the intake/exhaust plumbing and the wiring system are located here.
INTENDED USE:
The Model 3002 "AEROMAX" has the same intended use as the Model 3050.
Page 3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.