SALTER AIRE COMPRESSOR

K992285 · Salter Labs · CAF · Jan 7, 2000 · Anesthesiology

Device Facts

Record IDK992285
Device NameSALTER AIRE COMPRESSOR
ApplicantSalter Labs
Product CodeCAF · Anesthesiology
Decision DateJan 7, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Salter Aire Compressor is an AC-powered, piston type air compressor that generates pressurized air for use with a small volume nebulizer to produce an aerosol. This product is intended to be used primarily by patients in the home care market, although it may also be used by trained hospital staff personnel as well. The Salter Aire Compressor is a prescription use, non-sterile, durable device.

Device Story

Salter Aire Compressor is an AC-powered, piston-type device generating pressurized air for small volume nebulizers to produce aerosols. Used in home care and hospital settings by patients or trained staff. Device is portable, lightweight, and durable. Operation involves connecting to a nebulizer to deliver medication; healthcare providers use the output to facilitate respiratory treatment. Benefits include reliable, portable aerosol delivery for patients. No software or complex electronics; mechanical operation only.

Clinical Evidence

No clinical data. Bench testing only, including verification and validation to design specifications, back pressure, flow rate, temperature testing, shipping/drop testing, and UL 1431 compliance testing.

Technological Characteristics

Piston-type air compressor; AC-powered (115/220 VAC); plastic housing; includes replaceable foam filter for air cleanliness; portable form factor; over-temperature protection; conforms to UL 1431 and IEC 601-1 standards.

Indications for Use

Indicated for patients requiring aerosol therapy via a small volume nebulizer in home care or hospital settings. Prescription use only.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K992285 # 13. Salter Aire Compressor 510(k) Summary: In accordance with 21 CFR section 807.92, Salter Labs is submitting the following safety and effectiveness summary. #### 1) Submitter Information Salter Labs 100 W. Sycamore Road Arvin, CA 93203 (805) 854-3166 #### 2) Name of Device Proprietary Name: Salter Labs Salter Aire Compressor Common Name is Portable Air Compressor Classification Name: Portable Air Compressor ### 3) Substantially equivalent to: Medical Industry America Sport-Neb Compressor. #### 4) Device Description and System Overview: The Salter Labs Salter Aire Compressor is an AC powered, piston-type compressor that generates pressurized air for a nebulizer. It is designed to be small, light weight and portable. The device is manufactured by Salter Labs to Salter Labs design specifications. The Salter Aire Compressor will be available in both 115 and 220 VAC models. The units have been designed and built according to applicable requirements of UL 1431 for USA models and IEC 601-1 for European models. All USA versions of the device will be 115VAC. The Salter Aire Compressor is a prescription device and is designed to be a durable, reusable device that is intended to be used by hospital, clinic and home care markets. It has a 5 year warranty. #### System Similarities and Differences to Currently Marketed Devices The Salter Aire Compressor and the predicate devices upon which equivalence is sought share the same purpose: to provide the source of compressed air used to provide pressurized airflow in a nebulizer for the production of aerosol. Like the predicate devices reviewed, the Salter Aire Compressor is intended for the home care market and the hospital market. {1}------------------------------------------------ All comparative products are essentially equivalent, and many share similar components and have similar output specifications. Performance characteristics, as tested, are essentially equivalent to predicate devices, and all products share the same basic size and weight. The Salter Aire Compressor has been designed and tested to comply with all appropriate sections of the UL 1431 Standard for Personal Hygiene and Health Care Appliances, and complies with IEC 601-1 for European 220 volt versions. Other than basic configuration and dimensional differences between the Salter Aire and predicate devices, there are no appreciable differences between units in terms of function, performance or utility. #### 5) Statement of Intended Use: The Salter Aire Compressor is an AC-powered, piston type air compressor that generates pressurized air for use with a small volume nebulizer to produce an aerosol. This product is intended to be used primarily by patients in the home care market, although it may also be used by trained hospital personnel. The Salter Aire Compressor is a prescription use, non-sterile, durable device. | Test Reading: | Salter Aire Compressor | MIA Sport-Neb | |------------------------------------|------------------------|------------------| | Internal Temp. | 92.3 to 128.5° F | 80.5 to 134.8° F | | Dead Head<br>Pressure in<br>p.s.i. | 32 to 38.5 psi | 28 to 40 psi | | Flow with<br>Nebulizer | 6.7 to 7.5 LPM | 6.2 to 7.1 LPM | | Free Flow Rate<br>in SLPM | 11.3 to 12.8 LPM | 7.3 to 12.1 LPM | | Time to<br>Nebulize 3 cc<br>H2O: | 8 minutes | 8.5 minutes | #### 6) Comparative data concerning the Salter Aire Compressor and the defined benchmark predicate device: {2}------------------------------------------------ # b. Predicate Devices: | Product: | Manufacturer: | 510(k) Number: | |-----------|--------------------------|----------------| | Sport-Neb | Medical Industry America | K964078 | | Aeromax | Medical Industry America | K942444 | Table 3.1 Predicate Devices # c. Comparative Product Matrix The chart below compares Salter Aire Compressor with predicate devices: | item: | Characteristic: | Salter Labs<br>Compressor | Medical Industries<br>America Sport-<br>Neb Compressor | Medical Industries<br>America AeroMax<br>Compressor | | |-------|---------------------------------------|------------------------------------------|--------------------------------------------------------|-------------------------------------------------------|--------| | 1 | Type of compressor: | piston | piston | piston | | | 2 | Input power: | 110V 60Hz, 2.5<br>amps | 110V 60Hz, 1.8<br>amps | 110V 60Hz, 2.5<br>amps | | | 3 | Maximum output pressure at nebulizer: | 6.5 lpm | 6 lpm | approx. 6 lpm | | | 4 | Maximum flow: | 4 | 11 lpm | 11 lpm | 13 lpm | | 5 | Operating Environment: | home care and<br>hospital | home care | home care | | | 6 | Maximum operating temperature: | 128.5° F | 134.8° F | 130 ° F | | | 7 | Designed to conform to UL 1431 | Yes | unknown | unknown | | | 8 | Operating Life: | 5 years | 5 years | 5 years | | | 9 | Over-temperature Protection? | Yes | Yes | Yes | | | 10 | Portable? | Yes | Yes | Yes | | | 11 | Case material: | Plastic | Plastic | Plastic | | | 12 | Warranty: | 1 year | 1 year | 1 year | | | 13 | Electromagnetic Compatibility: | conforms to UL<br>1431, and<br>IEC 601-1 | conforms to<br>UL 1431 (personal<br>hygiene products) | conforms to<br>UL 1431 (personal<br>hygiene products) | | ## d. Brief Description of Non-Clinical Tests Identified in the Premarket Notification: - Verification and Validation testing to design specifications. . - . Back pressure, flow rate and temperature testing and comparative product testing. - Shipping test, including multiple container drops from a height of 4 feet onto a . concrete floor. - UL testing to UL 1431 Standard. . {3}------------------------------------------------ - e. Clinical Tests submitted: None. #### f. Conclusions of all Testing: The Salter Labs Salter Aire Compressor met all design requirements and passed all validation test requirements. The device functioned properly and exhibited a predictable flow rate and pressure at an acceptable internal case temperature. The device was demonstrated to possess equivalent performance to predicate devices tested. - g. Software Validation: Not applicable: there is no software in this product. - h. Sterilization Validation: Not applicable: this product is sold and used as a nonsterile product. - Biocompatibility: There are no biocompatible issues. No direct patient contact. All i. air lines are clean and oil-free. A replaceable foam filter maintains air cleanliness. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three wavy lines, which is the symbol of the department. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN - 7 2000 Mr. Gus Bock Salter Labs 100 W. Sycamore Road Arvin, CA 93203-2300 Re: K992285 Salter Aire Compressor Regulatory Class: II (two) Product Code: 73 CAF Dated: November 9, 1999 Received: November 12, 1999 Dear Mr. Bock: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {5}------------------------------------------------ Page 2 - Mr. Gus Bock This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Jeanne A Waterhouse for Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # SALTER LABS ### 3. Intended Use The Salter Aire Compressor is an AC-powered, piston type air compressor that generates pressurized air for use with a small volume nebulizer to produce an aerosol. This product is intended to be used primarily by patients in the home care market, although it may also be used by trained hospital staff personnel as well. The Salter Aire Compressor is a prescription use, non-sterile, durable device. Please refer to section 7 of this submission for product use, testing and results. to Hunter (Division Sign Division of Cardie and Neurologica 510(k) Numbe prescription use Salter Labs 510(k) Submission for Salter Aire Compressor
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...