SMARTTOUCH

K133951 · Nexus6, Ltd. · CAF · Apr 25, 2014 · Anesthesiology

Device Facts

Record IDK133951
Device NameSMARTTOUCH
ApplicantNexus6, Ltd.
Product CodeCAF · Anesthesiology
Decision DateApr 25, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2

Indications for Use

The SmartTouch is intended for single-patient use as an electronic data capture accessory for recording actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their trial MDI . medication; - In clinical practice, where specialists, general practitioners, nurses, and educators need to know if . a patient has actuated their prescribed MDI medication. - In self-management, where patients need to track their medication use as part of their . management plan. The SmartTouch is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function.

Device Story

SmartTouch is a clip-on electronic accessory for Metered Dose Inhalers (MDI). It uses optical and pressure sensors to detect and log the date and time of MDI actuations. The device features an LED indicator and a control button for status checks and manual communication initiation. Data is transmitted wirelessly via Bluetooth to a communications device for remote review by clinicians or researchers. Used in clinical trials, clinical practice, and patient self-management to monitor medication adherence. It does not measure remaining medication or count doses. The device benefits patients and providers by providing objective, time-stamped compliance data to support treatment management.

Clinical Evidence

No clinical data. Safety and effectiveness were established through bench testing, including functional verification of the MDI presence and actuation sensor system, software verification, and compliance with IEC 60601 series standards (electrical safety, EMC, environmental performance) and FCC radio frequency regulations.

Technological Characteristics

Clip-on MDI accessory; microprocessor-controlled; internal battery power; optical and pressure sensors for actuation detection; Bluetooth wireless connectivity; software-based data logging. Complies with IEC 60601 series and FCC standards.

Indications for Use

Indicated for single-patient use to record actuations of prescribed Metered Dose Inhalers (MDI). Applicable for patients in clinical trials, clinical practice (monitored by specialists, GPs, nurses, educators), and self-management plans. Not for dose counting or indicating remaining medication quantity.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 133951 Image /page/0/Picture/1 description: The image shows the date APR 2 5 2014 at the top. Below the date is the logo for nexus6 LTD. The logo is black and white. Nexus 6 Limited Suite 205, 8 Commerce Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 20 December 2013 ### 510(k) SUMMARY This summary of 510(k) sofety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92: #### (a)(1) Submitter Information | Company Details: | Refer to information above | |------------------|-------------------------------------------| | Contact Person: | Garth Sutherland, Chief Executive Officer | #### (a)(2) Name of the Device | Model Number / Name: | SmartTouch | |----------------------|------------------------------------------------------| | Classification Name: | Nebulizer (Direct Patient Interface) | | | Anesthesiology Devices 21 CFR 868 5630 Class II, CAF | #### (a)(3) Identification of Legally Marketed Devices K091803, SmartTrack System, Nexus6 Limited #### (a)(4) Description of the Device SmartTouch is used to provide a compliance monitoring for use of a Metered Dose Inhaler (MDI). SmartTouch is a clip-on device that attaches externally around the enclosure of an MDI. Optical and pressure sensor methods are used to detect MDI presence and actuation, and the device logs the usage history of the MDI. The device includes an LED indicator and control button to check device status and manually initiate communications functions. The SmartTouch has a Bluetooth interface to wirelessly exchange data with a communications device. # (a){5) Statement of the Intended Use The SmartTouch is intended for single-patient use as an electronic data capture accessory for recording actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their trial MDI . medication; - In clinical practice, where specialists, general practitioners, nurses, and educators need to know if . a patient has actuated their prescribed MDI medication. {1}------------------------------------------------ - In self-management, where patients need to track their medication use as part of their management plan. The SmartTouch is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function. #### (a)(6) Technological Characteristics Summary Technological characteristics of the SmartTouch are largely equivalent to the predicate device listed above. Equivalent features between the devices include: configuration in attaching to the outside of an MDI enclosure: microprocessor control and use of an internal clock, to log date and time of MDI actuations; power supply from an internal battery; interface to a communications device to upload MDI usage data; and capability to provide compliance data for further analysis using remote review software. The sensor technology used to detect MDI actuation is different from the predicate device, as is the communications technology. These aspects of the device have been verified by non-clinical testing to establish equivalent performance to the predicate device. #### (b)(1)_Discussion of the Non-Clinical Tests Non-clinical testing of the SmartTouch has been carried out to cover functional verification and device performance. This included completion of software and device verification procedures, with performance testing of the MDI presence and actuation sensor system to ensure data is logged accurately for MDI usage. This established correct functionality of the SmartTouch according to requirements. Review and testing of the SmartTouch for compliance to IEC 60601 series standards for electrical safety, electromagnetic compatibility, and environmental performance; and FCC regulations for radio frequency devices; has been completed by external laboratories. #### (b)(2) Discussion of the Clinical Tests Clinical testing was not required to demonstrate the safety and effectiveness of the SmartTouch. The product functionality has been adequately assessed by bench testing as above. ### (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance Finished device testing carried out for the SmartTouch indicates it meets design and performance functional requirements. Software verification demonstrates that device features are effective, and functions equivalently to the predicate device meets standard requirements for wireless communications, electrical safety, electromagnetic compatibility, and environmental performance. This information indicates that the SmartTouch is equivalent to the predicate device in terms of device safety and effectiveness. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three wavy lines extending from its body, all in black. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 25, 2014 Nexus6, Ltd. Mr. Garth Sutherland Chief Executive Officer Suite 205 8 Commerce Street Auckland, 1010 New Zeland Re: K133951 > Trade/Device Name: SmartTouch Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer (Nebulizer) Regulatory Class: Class H Product Code: CAF Dated: April 22, 2014 Received: April 23, 2014 Dear Mr. Sutherland, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. Garth Sutherland Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Image /page/3/Picture/7 description: The image shows a logo or emblem with some text. The text reads "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID". The logo is a stylized design with some intricate details. The text is positioned to the right of the logo. Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Nexus6 LTD. The logo consists of a plus sign made up of four squares, with the center square being white and the other four squares being gray. To the right of the plus sign is the text "nexus6 LTD" in a sans-serif font. Nexus6 Limited Suite 205, 8 Commerce Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 20 December 2013 ### INDICATIONS FOR USE KI33951 510(k) Number: SmartTouch Device Name: Indications for Use: The SmartTouch is intended for single-patient use as an electronic data capture accessory for recording actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their trial MDI . medication; - In clinical practice, where specialists, general practitioners, nurses, and educators need to know if . a patient has actuated their prescribed MDI medication. - In self-management, where patients need to track their medication use as part of their . management plan. The SmartTouch is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) Deepika A*Lakh 2014.04.25
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...