K960854 · Pacesetter, Inc. · CAF · Dec 23, 1996 · Anesthesiology
Device Facts
Record ID
K960854
Device Name
SIEMENS SERVO ULTRA NEBULIZER 345
Applicant
Pacesetter, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Dec 23, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Siemens Model 345 Ultra Nebulizer is to nebulize liquid medication for introduction into the patient’s airway via an endotracheal tube. It can be used on any patient for whom the use of the Siemens Servo Ventilator 300 is indicated, in any ventilatory mode appropriate for the patient.
Device Story
Accessory to Servo Ventilator 300; nebulizes liquid medication for delivery into patient airway via endotracheal tube. Uses ultrasonic energy (2.4 MHz) to aerosolize 10ml medication cup contents. Nebulization rate 0.3ml/min at 0.5l/min gas flow; mass median particle diameter 4.0µm. Integrated into patient circuit; adds no gas volume, ensuring ventilator readings/settings remain unaffected. Used in clinical settings by healthcare providers. Benefits patient by enabling medication delivery during mechanical ventilation.
Clinical Evidence
Bench testing only. Validation included unit and system-level testing for particle size and nebulization rates across various medications and ventilator flow rates. Electrical safety and electromagnetic compatibility testing performed per EN 60601 standards.
Technological Characteristics
Ultrasonic nebulizer; 2.4 MHz excitation frequency; 10ml medication cup capacity; 4.0µm mass median particle diameter; 0.3ml/min nebulization rate at 0.5l/min gas flow. Designed as an accessory to the Servo Ventilator 300. Tested to EN 60601 electrical safety and EMC standards.
Indications for Use
Indicated for patients requiring nebulized liquid medication delivery via endotracheal tube during mechanical ventilation using the Siemens Servo Ventilator 300.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Microsonic Ultrasonic Nebulizer
Omron NE-U07 Ultra-Air
Siemens Servo Nebulizer 945
Related Devices
K020357 — 800 SERIES EASYNEB NEBULIZER · Nellcor Puritan Bennett, Inc. · Aug 7, 2002
K041082 — NEWPORT NEBSONIC N550 ULTRASONIC NEBULIZER · Newport Medical Instruments, Inc. · Sep 7, 2004
K990897 — NELLCOR PURITAN BENNETT 700 SERIES VENTILATOR SYSTEM, AND EASYNEB NEBULIZER · Puritan-Bennett Ireland, Ltd. · Mar 15, 2000
K980024 — SHINMED MODEL SW-966 ULTRASONIC NEBULIZER · Shining World Health Care Co., Ltd. · Sep 3, 1999
K082541 — HEALTHLINE MEDICATOR / REDINEB AEROSOL DRUG DELIVERY SYSTEM · Medi Nuclear Corp., Inc. · Nov 26, 2008
Submission Summary (Full Text)
{0}
K960854
510(k) Summary for Siemens Servo Ultra Nebulizer 345
DEC 23 1996
1. Date this summary was prepared
February 28, 1996
2. Submitter's Name and Address
Siemens-Elema AB
Röntgenvägen 2
S-171 95 Solna
Sweden
3. Contact Person
Mr. Anders Lodin
Telephone 011-46 8 730 7228
Telefax 011-46 8 98 63 05
4. Device Name
Trade/Proprietary Name: Servo Ultra Nebulizer 345
Common Name: Ultrasonic Nebulizer Accessory to Servo Ventilator 300
Classification Names: (1) Nebulizer
(2) (Accessories to) Ventilator, Continuous (Respirator)
5. Predicate Devices
The legally marketed devices to which equivalence is being claimed are:
- Microsonic Ultrasonic Nebulizer, marketed by Medisonic U.S.A. Inc.
- Omron NE-U07 Ultra-Air, marketed by Omron Healthcare, Inc.
- Siemens Servo Nebulizer 945
Siemens Ultra Nebulizer 345 510(k)
2/28/96
{1}
# 6. Device Description
The Servo Ultra Nebulizer 345 (SUN 345) is an accessory to the Servo Ventilator 300 that is used to nebulize medication during ventilator treatment. An ultrasonic nebulizer similar to those commonly marketed as a component of home “medication inhaler” devices has been adapted for use in the patient circuit of the Servo Ventilator. The ultrasonic nebulizer does not add any gas volume to the circuit, so that readings and settings on the ventilator are unaffected. The SUN 345 medication cup holds $10\mathrm{ml}$ of liquid, which is nebulized at a rate of $0.3\mathrm{ml/min}$ at $0.5\mathrm{l/min}$ gas flow. The mass median diameter of particles is $4.0\mu\mathrm{m}$.
# 7. Intended Use
The intended use of the Siemens Model 345 Ultra Nebulizer is to nebulize liquid medication for introduction into the patient’s airway via an endotracheal tube. It can be used on any patient for whom the use of the Siemens Servo Ventilator 300 is indicated, in any ventilatory mode appropriate for the patient.
# 8. Comparison of Technological Characteristics
The Servo Ultra Nebulizer 345 works by exciting the medication with ultrasonic energy at a frequency of $2.4\mathrm{Mhz}$, as do the Medisonic and Omron products. The Servo Nebulizer 945 uses the jet principal.
The SUN 345 produces particles with a mass median diameter of $4\mu\mathrm{m}$. The Medisonic has particle size of $3-5\mu\mathrm{m}$, and the Omron $1-5\mu\mathrm{m}$.
The SUN 345 has a nebulization rate of $0.5\mathrm{ml}$ water/min at 1 l/sec gas flow, and $0.3\mathrm{ml}$ water/min at $0.5\mathrm{l/sec}$ gas flow. This compares to up to $2\mathrm{ml/min}$ for the Medisonic and $0.4\mathrm{ml/min}$ for the Omron.
# 9. Nonclinical Tests Used in Determination of Substantial Equivalence
The design of the Siemens Model 345 Ultra Nebulizer has been thoroughly validated at the unit and system level. Non-clinical tests were conducted to determine particle size and nebulization rate for various medications and ventilator flow rates. The complete system consisting of the SUN 345 and the Servo Ventilator 300 has been tested to the electrical safety requirements and the Electromagnetic Compatibility requirements of EN 60601.
Siemens Ultra Nebulizer 345 510(k)
2/28/96
Page E-2
{2}
0. Conclusions From Nonclinical Testing
The testing of the SUN 345 demonstrates that the performance is substantially equivalent to predicate devices cited above.
Siemens Ultra Nebulizer 345 510(k)
2/28/96
Page E-3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.