SMDIA-1000

K101953 · Southmedic, Inc. · CAF · Jan 7, 2011 · Anesthesiology

Device Facts

Record IDK101953
Device NameSMDIA-1000
ApplicantSouthmedic, Inc.
Product CodeCAF · Anesthesiology
Decision DateJan 7, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SMDIA-1000 MDI adaptor is intended to be used with an MDI for intermittent delivery of prescribed aerosol medications dispensed in MDIs. It is intended for use only when connected to ventilator tubing or tracheal tubes. The SMDIA-1000 MDI adaptor is single patient use, disposable.

Device Story

SMDIA-1000 is a disposable, single-patient-use adaptor acting as an actuator for meter dose inhalers (MDI). Device connects to ventilator tubing or endotracheal tubes; provides a port for MDI insertion to deliver aerosolized medication into the breathing circuit. Features a tee design with 15mm/22mm fittings and a cap for the MDI port when not in use. Used in critical care, ICU, sub-acute centers, and home settings by clinicians or caregivers. Facilitates intermittent drug delivery for ventilator-dependent patients. Output is the aerosolized medication delivered into the patient's airway via the ventilator circuit.

Clinical Evidence

Bench testing only. Performance validated via comparative particle characterization using a cascade impactor with three drug types. Additional testing included simulated life cycle, environmental, operational, drop, canister activation, and connection trials. Device met all pass/fail criteria and demonstrated equivalence to predicate performance.

Technological Characteristics

Tee-style adaptor with 15mm/22mm fittings; MDI port with cap. Single-patient use, disposable. Materials in gas/fluid pathway tested per ISO 10993 for cytotoxicity, irritation, and sensitization. No electronic components or software.

Indications for Use

Indicated for ventilator-dependent patients requiring intermittent delivery of prescribed aerosol medications via MDI. Intended for use only when connected to ventilator tubing or tracheal tubes. Prescription use only.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K101953 JAN - 7 2011 | Southmedic, Inc.<br>50 Alliance Blvd.<br>Barrie, ONT | Tel – (705) 720-1902 x 342<br>Fax – (705) 728-9537 | |------------------------------------------------------|-----------------------------------------------------------------------| | Official Contact: | Tish Anger - VP Quality & Regulatory Assurance | | Proprietary or Trade Name: | SMDIA-1000 | | Common/Usual Name: | Accessory to a nebulizer | | Classification Name/Code: | CAF – nebulizer<br>CFR 868.5630 | | Device: | Southmedic SMDIA 1000 | | Predicate Devices: | Instrumentation Industries<br>K991355 - RTC-22-D<br>K091111 - RTC-24V | #### Device Description: The Southmedic SMDIA-1000 is an adaptor which acts as an actuator for intermittent delivery of prescribed aerosol medications dispensed by meter dose inhalers (MDI). The SMDIA-1000 is connected to ventilator tubing and the MDI connected when drug delivery into the circuit is required. The SMDIA-1000 adaptor may also be connected to the endotracheal tube. #### Indications for Use: The SMDIA-1000 MDI adaptor is intended to be used with an MDI for intermittent delivery of prescribed aerosol medications dispensed in MDI's. They are intended for use only when connected to ventilator tubing or tracheal tubes. #### Environment of Use: The expected clinical environment is critical care and / or long or short term ventilation, which can occur in hospitals, ICU, sub-acute centers, and home settings. #### Patient Population: The SMDIA-1000 is intended to be prescribed for any patient who is ventilator dependent and to whom a MDI has been prescribed. {1}------------------------------------------------ | 510(k) Summary<br>Page 2 of 3<br>24-Dec-10 | Predicate<br>K991355 - R | |--------------------------------------------|--------------------------| | A-1000 | | | Summary of substantial equivalence | Southmedic SMDIA-1000 | Predicate - Instrumentation Industries<br>K991355-RTC-22-D & K091111 - RTC-24V | |---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The SMDIA-1000 MDI adaptor is intended to be used with an<br>MDI for intermittent delivery of prescribed aerosol<br>medications dispensed in MDIs. They are intended for use<br>only when connected to ventilator tubing or tracheal tubes. | The RTC series MDI adapters are actuators for intermittent<br>delivery of prescribed aerosol medications dispensed in meter<br>dose inhalers. The RTC series MDI adapters are intended for<br>use only when connected to ventilator tubing or tracheal tubes. | | Environment of use | The expected clinical environment is critical care and /<br>or long or short term ventilation, which can occur in<br>hospitals, ICU, sub-acute centers, and home settings. | The expected clinical environment is critical care and /<br>or long or short term ventilation. | | Patient Population | For any patient who is ventilator dependent and to<br>whom a MDI has been prescribed. | For any patient who is ventilator dependent and to whom<br>a MDI has been prescribed. | | Prescriptive | Yes | Yes | | Design | | | | Placed in the breathing circuit | Yes | Yes | | May attach to a tracheal tube | Yes | Yes | | Tee with standard 15 mm / 22<br>mm fittings | Yes | Yes | | Fitting for adaption to MDI | Yes | Yes | | Cap when not in use | Yes | Yes | | Single patient use, disposable | Yes | Yes | | Performance Testing | | | | Particle characterization via<br>Cascade Impactor | 3 samples with 3 different drugs<br>MDI burst | 3 drug types<br>1 sample each | | Simulated life cycle and | Same sample testing for intra-sample variability<br>Environment, operational, drop, canister activation and | Not listed | {2}------------------------------------------------ ### 510(k) Summary Page 3 of 3 24-Dec-10 In summarizing the above comparative table the SMDIA-1000 is viewed as substantially equivalent to the predicate devices because: #### Indications - - Intended to be used with an MDI for intermittent delivery of prescribed aerosol medications . dispensed in MDIs. They are intended for use only when connected to ventilator tubing or tracheal tubes. These indications for use are identical to the predicate - K091111 - RTC-24V #### Technology - - The tee design with a port which is capped when not in used is identical to the predicate -. K091111 - RTC-24V #### Materials - - . The materials in the gas or fluid pathway have been tested per ISO 10993 for Cytotoxicity, Irritation, and Sensitization #### Environment of Use - - The expected clinical environment is critical care and / or long or short term ventilation, � which can occur in hospitals, ICU, sub-acute centers, and home settings in substantially equivalent to the predicate - K091111 -- RTC-24V #### Comparative Performance - - . Comparative Particle characterization testing via Cascade Impactor demonstrated that the performance of the proposed device and the predicate were equivalent - Additional performance testing included Simulated life cycle, environmental, operational, . drop, canister activation and connection trial testing, inter- and intra-sample testing for variability. For which the proposed device met its pass / fail criteria or was found substantially equivalent to the predicate. #### Differences - There are no significant differences between the proposed device and the predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MID 20993-0002 Southmedic, Incorporated C/O Mr. Paul E. Dryden President Promedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134 JAN - 7 2011 Re: K101953 Trade/Device Name: SMDIA-1000 Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: December 24, 2010 Received: December 27, 2010 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. . {4}------------------------------------------------ Page 2- Mr. Dryden Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Nh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ JAN - 7 2011 ## Indications for Use Statement Page 1 of 1 510(k) Number: K101953 Device Name: SMDIA - 1000 Indications for Use: The SMDIA-1000 MDI adaptor is intended to be used with an MDI for intermittent delivery of prescribed aerosol medications dispensed in MDIs. It is intended for use only when connected to ventilator tubing or tracheal tubes. The SMDIA-1000 MDI adaptor is single patient use, disposable. Prescription Use XX (Part 21 CFR 801 Subpart D) or Over-the-counter use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schultheis (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K101953
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...