SALTER LABS NEBULIZER EXHALATION AEROSOL FILTER

K983403 · Salter Labs · CAF · Dec 17, 1998 · Anesthesiology

Device Facts

Record IDK983403
Device NameSALTER LABS NEBULIZER EXHALATION AEROSOL FILTER
ApplicantSalter Labs
Product CodeCAF · Anesthesiology
Decision DateDec 17, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2

Indications for Use

The Salter Labs Nebulizer Exhalation Aerosol Filter is indicated for all circumstances where the reduction of medical aerosols during nebulizer use is required or recommended. The Salter Labs Nebulizer Exhalation Aerosol Filter comes in two configurations: a disposable device, which has a filter contained within a permanently sealed plastic case, and a reusable device designed to be used with Salter Labs nebulizer product. The Salter Labs Exhalation Aerosol Filter is intended to be used to reduce the amount of unused medical aerosols exhaled by the patient using the Salter Labs nebulizer system. The product specification calls for at least a 95% effectiveness at aerosol removal. No claims of bacterial filtration effectiveness are made or intended.

Device Story

The Salter Labs Nebulizer Exhalation Aerosol Filter is an accessory designed to attach to the exhalation valve of Salter Labs nebulizer mouthpieces. It utilizes a hydrophobic 3M Filtrete material (0.3 micron) housed in an injection-molded plastic casing to capture exhaled medical aerosols. The device is available in disposable and reusable (up to 1 year) configurations. It is used in home care and hospital settings by patients undergoing nebulizer therapy. By filtering exhaled air, the device reduces the amount of unused medical aerosol released into the environment, potentially benefiting caregivers and others in the vicinity by minimizing exposure to medication. The device is non-sterile and features a non-standard connector size specific to the Salter Labs product line.

Clinical Evidence

No clinical data. Bench testing only. Performance validation confirmed >95% aerosol removal efficiency for particles in the 1 to 10 micron range, with average efficiency exceeding 99%. Aerosol sizes monitored ranged from 0.01 to 9.9 microns. Exhalation resistance was measured at ≤ 0.8 cm/H2O.

Technological Characteristics

Hydrophobic 3M Filtrete material (0.3 micron); injection-molded plastic housing; rectangular form factor; surface area ≥ 6.0 sq. inches; non-standard 0.794" connector; non-sterile; disposable or reusable (1-year lifespan); exhalation resistance ≤ 0.8 cm/H2O.

Indications for Use

Indicated for patients using a Salter Labs nebulizer system when a physician or healthcare professional wishes to minimize the amount of medical aerosol exhaled into the air.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### Salter Labs Nebulizer Aerosol Expiration Filter 510(k) Summary 9.2 In accordance with 21 CFR section 807.92 Salter Labs is submitting the following safety and effectiveness summary. 1) Submitter Information Salter Labs 100 W. Sycamore Road Arvin, CA 93203 (805) 854-3166 2) Name of Device Proprietary Name: Salter Labs Nebulizer Exhalation Aerosol Filter Common Name is Exhalation Filter Classification Name: Breathing Circuit Bacterial Filter 3) Substantially equivalent to Pari Exhalation Filter, K926455. 4) Device Description and System Overview: The Salter Labs Aerosol Filter is designed to connect to the exhalation valve of the Salter Labs nebulizer mouthpiece. Since the exhalation valve of the Salter Labs Nebulizer product is designed with a non-standard diameter, the Salter Labs Nebulizer Exhalation Aerosol Filter is configured in such a way that it only fits the Salter Labs nebulizer product line: this device is not intended to be used with any other product. Product labeling clearly defines this product as an accessory to the Salter Labs nebulizer assembly. The aerosol filter is a rectangular component that has been preformed to fit into an injection molded plastic housing. There are two design configurations: disposable and reusable. The reusable configuration allows the filter to be easily inserted and removed after each use. The filter is specified as a bacterial filter, although the intended use of the device is to remove aerosol particles, only. The surface area of the filter is ≥ 6.0 square inches, resulting in a specified resistance (exhalation force) of ≤ 2,0 cm Ho0 from the patient. Product design performance has been validated using product design specifications according to Salter Labs procedures and validation requirements. There are two defined product configurations for this device: disposable and reusable. - 5) Statement of Intended Use: The Salter Labs Nebulizer Exhalation Aerosol Filter is indicated for all circumstances where the reduction of medical aerosols during nebulizer use is required or The Salter Labs Nebulizer Exhalation Aerosol Filter comes in two recommended. configurations: a disposable device, which has a filter contained within a permanently sealed plastic case, and a reusable device designed to be used with Salter Labs nebulizer product. The Salter Labs Exhalation Aerosol Filter is intended to be used to reduce the amount of {1}------------------------------------------------ unused medical aerosols exhaled by the patient using the Salter Labs nebulizer system. The product specification calls for at least a 95% effectiveness at aerosol removal. No claims of bacterial filtration effectiveness are made or intended. - 6) Comparative data concerning the Salter Labs Nebulizer Aerosol Filter and competitive products follows: - a) Technological Characteristics and Comparison: | | Salter Labs:<br>Disposable Filter | Salter Labs:<br>Reusable Filter | Pari | Marquest | King Systems | |------------------------------------|----------------------------------------------------|-------------------------------------------------------|----------------------------------------------------------|----------------------------------------------|----------------------------------------------| | Brand Name: | Exhalation<br>Aerosol Filter | Exhalation<br>Aerosol Filter | Exhalation<br>Filter | Respirguard II<br>Filter | Virobac II Filter | | Model #: | TBD | TBD | unknown | 303 | 20800 | | Filter Type: | Hydrophobic<br>Filter | Hydrophobic<br>Filter | Hydrophobic<br>Filter | Hydrophobic<br>Filter | Hydrophobic<br>Filter | | Filter<br>Manufacturer: | 3M | 3M | 3M | 3M | 3M | | Filter Material: | 0.3 micron<br>Filtrete Material | 0.3 micron<br>Filtrete<br>Material | 0.3 micron<br>Filtrete<br>Material | 0.3 micron | 0.3 micron | | Total Filter<br>Area: | ≥ 6.0 sq. inches<br>(rectangular) | ≥ 6.0 sq.<br>inches<br>(rectangular) | approximately<br>4 sq. inches<br>(circular) | approximately<br>4 sq. inches<br>(circular) | approximately<br>4 sq. inches<br>(circular) | | Single-Use/<br>Reusable: | Disposable | Reusable<br>(1 year) | Reusable<br>(1 year) | Disposable | Disposable | | Reuse Claims: | none | 1 year | 1 year | none | none | | Sterile/Non-<br>Sterile: | Non-Sterile | Non-Sterile | Non-Sterile | Non-Sterile | Non-Sterile | | Intended Use: | Nebulizer | Nebulizer | Nebulizer | Nebulizer,<br>humidifier, O₂<br>Concentrator | Nebulizer,<br>humidifier, O₂<br>Concentrator | | Target<br>Population: | Home Care &<br>Hospital | Home Care &<br>Hospital | Home Care &<br>Hospital | Home Care &<br>Hospital | Home Care &<br>Hospital | | Aerosol<br>Filtration? | Yes | Yes | Yes | Yes | Yes | | Bacterial<br>Filtration? | No | No | Yes (but used<br>primarily as<br>aerosol filter) | Yes | Yes | | Exhalation<br>Resistance<br>Force: | 0.8 cm/H₂O** | 0.8 cm/H₂O** | 0.8 cm/H₂O<br>(no valve)<br>7.0 cm/H₂O<br>(with valve)** | 1.2 cm/H₂O | unknown | | Connector<br>Size: | .794" (non-<br>standard Salter<br>sized connector) | .794" (non-<br>standard Salter<br>sized<br>connector) | 22mm | 22mm | 22mm | {2}------------------------------------------------ - b) Brief Description of Non-Clinical Tests Identified in the Premarket Notification: - performance validation to design specification 0 - 0 aerosol removal testing and validation - c) Clinical Tests submitted: None. - The Salter Labs Nebulizer Expiratory Aerosol Filter d) Conclusions of all Testing: (in both the disposable and reusable configurations) met all design requirements and passed all validation test requirements. It was shown to have a > 95% rate of aerosol removal efficiency, with aerosols in the 1 to 10 micron range, with an average efficiency rating exceeding 99%. Aerosol sizes monitored varied from .01 microns to 9.9 microns. Typical aerosol sizes generated during nebulization range from 1 micron to 5 microns. ## Cleaning and Reuse: Only the Reusable Aerosol Filter is intended for reuse. The Salter Labs Reusable Nebulizer Exhalation Aerosol Filter is designed to be cleaned prior to each use. Cleaning instructions are contained in product literature and include directions for washing, chemical cleaning and short term autoclaving. Since this device is incorporated in the exhalation path of the Salter Labs Nebulizer family, this device is provided and intended to be used in a non-sterile condition. ### Product Warranty: Standard product warranties apply to the disposable product. The reusable product is warranted for one year. All products are warranted to be free of defects at time of receipt. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird figure. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 17 1998 Mr. Duane Kazal Salter Labs Inc. 100 W. Sycamore Road Arvin, CA 93203 Re : K983403 Salter Labs Nebulizer Exhalation Aerosol Filter Regulatory Class: II (two) Product Code: 73 CAF Dated: September 15, 1998 September 28, 1998 Received: Dear Mr. Kazal: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major regulations affecting your device additional controls. can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Mr. Duane Kazal This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### K983403 510(k) Number: Unknown -C Device Name: Salter Labs Nebulizer Exhalation Aerosol Filter. Indications For Use: This product is indicated for use whenever the physician or healthcare professional administering or prescribing medical aerosol products to a patient with a Salter Labs nebulizer wishes to minimize the amount of medical aerosol exhaled into the air. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Usen ﺖ (Per 21 CFR 801.109) OR Over-The Counter-Use (Optional Format 1-2-96) Zank Medoo 12-16-98 (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number ...
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...