FAMILIES OF MASTER NEB, PRIMO NEB, SC03, NEB AID, 4.2.NEB, Q03, FJ03, AND WALKIE-NEB
K062014 · Flaem Nuova S.P.A. · CAF · Mar 22, 2007 · Anesthesiology
Device Facts
Record ID
K062014
Device Name
FAMILIES OF MASTER NEB, PRIMO NEB, SC03, NEB AID, 4.2.NEB, Q03, FJ03, AND WALKIE-NEB
Applicant
Flaem Nuova S.P.A.
Product Code
CAF · Anesthesiology
Decision Date
Mar 22, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
The nebulizer compressors Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, and Q03 are intended to provide a source of compressed air for medical purposes, to be used in home health care and hospital use. The Walkie-Neb model is intended to be used only in home health care. These devices are indicated for use with all commercially available small volume pneumatic nebulizers, to produce a fine aerosol mist of medication for respiratory therapy, for both adult and pediatric patients who have been prescribed inhalation therapy or medication.
Device Story
Family of nebulizer compressors (AC-powered: Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, Q03; DC-powered: Walkie-Neb) acting as compressed air generators for pneumatic nebulizers. Devices connect to small volume nebulizers to convert liquid medication into aerosol mist for inhalation. Used in home health care and hospital settings by patients or caregivers. AC models utilize electrical motors to drive pneumatic output; Walkie-Neb includes rechargeable battery pack and external charger for portability. Healthcare providers prescribe the therapy; patients operate the device to administer medication. Output is a fine aerosol mist; clinical benefit is delivery of respiratory medication to the lungs.
Clinical Evidence
Bench testing only. Performance tests conducted per 1993 FDA reviewer guidance for nebulizers demonstrated equivalent operating flow rate and pressure compared to predicate devices. No clinical data required.
Technological Characteristics
Compressor-based pneumatic air generator. AC models (F700, F1000, F1500 series) and DC model (Walkie-Neb) with rechargeable battery. Plastic housing. Operates via electrical motor to provide compressed air. Standalone device.
Indications for Use
Indicated for adult and pediatric patients prescribed inhalation therapy or medication; used with small volume pneumatic nebulizers to produce aerosol mist for respiratory therapy.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
K072163 — HAPPYNEB II, HAPPYNEB III, SPEEDY, NEBBY PLUS AND MYNEB, MODEL 5 · 3A Health Care S.R.L. · Mar 6, 2008
K053490 — MERITS HEALTH PRODUCTS PORTABLE NEBULIZER · Merits Health Products Co., Ltd. · Jul 24, 2006
K032170 — TRANSNEB COMPRESSOR W/NEBULIZER, MODEL 5000 · Medical Industries America, Inc. · May 28, 2004
K983736 — ICEL EVOLUSONIC PORTABLE ULTRASONIC NEBULIZER · Brasamerica Medical Equipment, Inc. · Nov 10, 1998
K192633 — GUS831 Compressor Nebulizer · Globalcare Medical Technology Co., Ltd. · May 1, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
K062014
MAR 2 2 2007
# 510(k) Summary for the nebulizer compressors Master Neb, Primo Neb, SC03 , Neb Aid, 4.2.Neb, Q03, FJ03, and Walkie-Neb
This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
### 1. General Information
Submitter:
Contact Person:
FLAEM NUOVA Via Colli Storici 221, 223, 225 S. Martino della Battaglia Brescia 25010 Italy
> Maureen O'Connell O'Connell Regulatory Consultants. Inc. 5 Timber Lane North Reading, MA 01864 Telephone: 978-207-1245 Fax: 978-207-1246
Summary Preparation Date:
January 30, 2007
2. Names
Device Name:
The families of Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, Q03, FJ03, and the Walkie-Neb.
Classification Name:
Portable air compressor Regulation number: 868.6250 Product Code: CAF
## 3. Predicate Devices
The nebulizer compressors Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, Q03 and Walkie-Neb are substantially equivalent to a combination of the following devices:
- > Respironics, nebulizer compressor MISTER NEB - K013027,
- V Healthdyne Inc NEBULIZER SYSTEM K922623,
- Invacare PRO, COMPACT e PORTABLE DESKTOP Nebulizer -K042483, and
- V Salter Labs SALTER AIRE COMPRESSOR - K992285.
{1}------------------------------------------------
### 4. Device Description
The nebulizer compressors Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb. FJ03, Q03 and Walkie-Neb are a family of nebulizer compressors intended to be used as compressed air generators for pneumatic nebulizers used by patients for aerosol therapy. There are 7 model families with an AC powered motor (Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, and Q03) and one model with a DC low voltage motor (Walkie-Neb), with a rechargeable battery pack and an external charger/power supplier.
The 7 AC model families have different plastic housings, which define the model name of the product. Inside they have the same compressor but with different electrical motors, to provide different pneumatic performances.
### 5. Indications for Use
The nebulizer compressors Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, and Q03 are intended to provide a source of compressed air for medical purposes, to be used in home health care and hospital use. The Walkie-Neb model is intended to be used only in home health care.
These devices are indicated to be used with all commercially available small volume pneumatic nebulizers, to produce a fine aerosol mist of medication for respiratory therapy, for both adult and pediatric patients who have been prescribed inhalation therapy or medication.
#### 6. Performance Data
The performance tests performed by Flaem Nuova, as requested in "Reviewer guidance for nebulizers, metered dose inhalers, spacers and actuators" issued in October 1993, demonstrate that Flaem Nuova nebulizer compressors have the same effectiveness as their predicate devices because they have equivalent operating flow rate and pressure. Clinical data were not required.
#### 7. Comparison to Predicate Devices
The Flaem Nuova compressors (Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, Q03 and Walkie-Neb) and the predicate devices are indicated for the same intended use (Salter Labs, Salter Aire Compressor, K992285). They have the same AC and DC power supply and meet the same performance, safety and EMC requirements. The AC Flaem Nuova compressors, series F700, have the same compressor, electrical motor and thus the identical operating pressure and flow ranges as the Respironics Mister Neb compressor (K013027). The AC Flaem Nuova compressors series F1000 and F1500 have the same compressor, electrical motor and thus the identical operating pressure and flow ranges as the Healthdyne Inc. NEBULIZER SYSTEM compressor (K922623). The Walkie Neb Model has
{2}------------------------------------------------
mechanical configuration, batteries, charging system and accessories which are substantially equivalent to the configuration of the Invacare Portable Desktop (K042483). Concerning the operating pressure and flow ranges, the Walkie Neb has the equivalent performance to the Respironics Mister Neb compressor (K013027).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three figures instead of snakes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Flaem Nuova S.P.A. C/O Ms. Maureen O' Connell Regulatory Consultant O' Connell Regulatory Consultants, Incorporated 5 Timber Lane North Reading, Massachusetts 01864
MAR 2 2 2007
Re: K062014
Trade/Device Name: Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, Q03 and WalkieNeb Regulation Number: 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: March 5, 2007 Received: March 7, 2007
Dear Ms. O' Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Ms. O' Connell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, Q03 and Walkie-Neb
Indications for Use:
The nebulizer compressors Master Neb, Primo Neb, SC03, Neb Aid, 4.2.Neb, FJ03, and Q03 are intended to provide a source of compressed air for medical purposes, to be used in home health care and hospital use. The Walkie-Neb model is intended to be used only in home health care.
These devices are indicated for use with all commercially available small volume pneumatic nebulizers, to produce a fine aerosol mist of medication for respiratory therapy, for both adult and pediatric patients who have been prescribed inhalation therapy or medication.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 and of simmel Hossellist, Comici, Land Streetses
510(K) FILE
Page viii
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.