The SmartTrack System is intended for single-patient use as an electronic data capture accessory for recording and monitoring actuations of prescribed MDI usage. This may be used in the following applications: - In clinical trials, where researchers need to know when a patient has actuated their trial MDI medication; - In clinical practice, where specialists, general practitioners, and nurse educators need to know if a patient has actuated their prescribed MDI medication. The SmartTrack is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function.
Device Story
SmartTrack System is a compliance monitoring accessory for Metered Dose Inhalers (MDI). It consists of a clip-on module, an optical docking station, and PC-based analysis software. The clip-on module attaches to the MDI exterior and uses an optical sensor to detect and log the date and time of MDI actuations. Data is transferred via the docking station to a PC. The Respiratory Analyzer software allows clinicians, researchers, or nurse educators to review, export, and analyze usage history in table or graph formats. The device provides objective data on medication adherence to inform clinical decision-making and support clinical trial monitoring. It does not count doses or indicate remaining medication quantity.
Clinical Evidence
Bench testing only. No clinical data was required or provided. Functional verification and performance testing of the MDI actuation sensor system confirmed accurate logging of MDI usage. Compliance with IEC 60601 (safety/EMC) and IEC 60068 (environmental) standards was performed.
Technological Characteristics
Clip-on module for MDI; optical sensor for actuation detection; microprocessor-controlled; internal battery power; USB-connected optical docking station; PC-based offline software for data reporting. Complies with IEC 60601 and IEC 60068 series standards.
Indications for Use
Indicated for single-patient use to record and monitor MDI actuations in clinical trials or clinical practice settings for patients prescribed MDI medication.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
K091803
Image /page/0/Picture/1 description: The image shows the logo for Nexus 6. The logo consists of a plus sign made up of four squares, followed by the word "nexus6" in lowercase letters. The "6" is slightly raised and has a small "tm" symbol next to it, indicating a trademark.
OCT - 9 2009
Nexus 6 Limited Level 2, 109 Cook Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 Fax +64 9 307 2773
6 July 2009
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
(a)(1) Refer to information above and concluding this summary.
## (a)(2) Name of the Device
| Model Number / Name: | SmartTrack System |
|----------------------|-------------------------------------------------------------------------------------------------|
| Classification Name: | Nebulizer (Direct Patient Interface)<br>Anesthesiology Devices, 21 CFR §868.5630, Class II, CAF |
## (a)(3} Identification of Legally Marketed Devices
| K990185 | MDILog Model MDC-512 | Medtrac Technologies Inc |
|---------|----------------------|--------------------------|
| K970344 | MDILog Model MDC-511 | Medtrac Technologies Inc |
| K935955 | Doser MDI Counter | Newmed Corporation |
#### (a)(4) Description of the Device
The SmartTrack System consists of three modules used to provide a compliance monitoring function for use of a Metered Dose Inhaler (MDI).
- · SmartTrack is a clip-on module that attaches externally around the enclosure of an MDI, using an optical sensor to detect MDI actuation, to log the usage history of the MDI. The device includes an LCD display and control buttons to review basic data and adjust device settings. Various device status conditions are indicated to the user, and logged to provide potentially relevant context for later review of MDI usage history data.
- SmartTrack Docking Station is an optical interface cradle used to download data stored on the SmartTrack module, via USB cable connection to a PC, and allow for interaction by the offline software with device settings and status monitoring.
- Respiratory Analyzer is a software application for use on a PC that provides for offline review of the MDI compliance data logged by the SmartTrack module. Data is stored in database files with patient, medication, and prescribed dose information, and is available for review and export in various table and graph formats to highlight MDI usage characteristics.
{1}------------------------------------------------
#### (a)(5) Statement of the Intended Use
The SmartTrack System is intended for single-patient use as an electronic data capture accessory for recording and monitoring actuations of prescribed MDI usage. This may be used in the following applications:
- In clinical trials, where researchers need to know when a patient has actuated their trial MDI medication;
- . In clinical practice, where specialists, general practitioners, and nurse educators need to know if a patient has actuated their prescribed MDI medication.
The SmartTrack is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function.
## (a)(6) Technological Characteristics Summary
Technological characteristics of the SmartTrack system are largely equivalent to the predicate devices listed above. Equivalent features between the devices include: configuration in attaching to the outside of an MDI enclosure; microprocessor control and use of an internal clock, to log date and time of MDI actuations; power supply from an internal battery; interface to a personal computer via an optical docking station to download MDI usage data; and offline review software providing for data handling, device interaction, and patient data reporting functions.
The sensor technology used to detect MDI actuation is different from the predicate devices, and this aspect of the device has been verified by non-clinical testing to establish equivalent performance to the predicate devices.
#### (b)(1) Discussion of the Non-Clinical Tests
Non-clinical testing of the SmartTrack system has been carried out to cover functional verification and device performance. This included completion of software verification procedures, with performance testing of the MDI actuation sensor system to ensure data is logged accurately for MDI usage. This established correct functionality of the SmartTrack System according to requirements.
Third party testing of the SmartTrack System for compliance to IEC 60601 series standards for general safety and electromagnetic compatibility, and IEC 60068 series standards for environmental testing, will be completed by an accredited laboratory before marketing of the device.
## (b)(2) Discussion of the Clinical Tests
Clinical testing was not required to demonstrate the safety and effectiveness of the SmartTrack System. The product functionality has been adequately assessed by bench testing as above.
#### (b)(3) Conclusions Demonstrating Safety, Effectiveness and Performance
Hardware testing carried out for the SmartTrack System indicates it meets design and performance functional requirements. Software verification demonstrates that device features are effective, and the system configuration functions equivalently to the predicate devices. The device will meet standard requirements for electrical safety and electromagnetic compatibility before marketing.
This information indicates that the SmartTrack System is equivalent to the predicate devices in terms of device safety and effectiveness.
Garth Sutherland Chief Operating Officer Nexus6 Ltd
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing left. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the circumference of the circle. The text is in all capital letters and is arranged to follow the curve of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002
Nexus6 Limited C/O Mr. Morten Simon Christensen Responsible Third Party Official Underwriters Laboratories, Incorporated 455 East Trimble Road San Jose, California 95131-1230
OCT - 9 2009
Re: K091803
Trade/Device Name: SmartTrack System Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: September 22, 2009 Received: September 24, 2009
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Mr. Christensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
h for
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nexus6 LTD. The logo consists of a plus sign made up of four squares, with a white square in the center. To the right of the plus sign is the word "nexus6" in lowercase letters, with "LTD" in superscript.
Nexus6 Limited Level 2, 109 Cook Street, Auckland 1010, New Zealand PO Box 106-612, Auckland 1143, New Zealand Phone +64 9 307 2771 Fax +64 9 307 2773
9 October 2009
# INDICATIONS FOR USE
510(k) Number: K091803
SmartTrack System Device Name:
Indications for Use:
The SmartTrack System is intended for single-patient use as an electronic data capture accessory for recording and monitoring actuations of prescribed MDI usage. This may be used in the following applications:
- In clinical trials, where researchers need to know when a patient has actuated their . trial MDI medication;
- In clinical practice, where specialists, general practitioners, and nurse educators need . to know if a patient has actuated their prescribed MDI medication.
The SmartTrack is not intended to indicate remaining quantity of medication in an MDI and does not include a dose counting function.
Prescription Use ✓ (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulten
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091803
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.