HIOX80 AEROSOL ADAPTER

K032516 · Sensor Medics Corp. · CAF · Mar 4, 2004 · Anesthesiology

Device Facts

Record IDK032516
Device NameHIOX80 AEROSOL ADAPTER
ApplicantSensor Medics Corp.
Product CodeCAF · Anesthesiology
Decision DateMar 4, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Hi-Ox® High FiO2 Mask with the Aerosol Adapter is intended as an accessory for compatible nebulizers. The mask with adaptor is -intended to enable the delivery of nebulized drugs to patients who also require elevated inspired oxygen concentrations.

Device Story

Accessory for compatible small-volume nebulizers (cup type, gas-driven, 18mm OD or 22mm ID ports); enables concurrent nebulized drug delivery and high-concentration oxygen therapy. Device consists of a right-angle Santoprene thermoplastic elastomer adapter; snaps into existing port on Hi-Ox mask between inhalation/exhalation valves. Operates by allowing nebulizer gas flow to supplement oxygen supply; excess flow exits via mask exhalation pathway. Used in clinical settings; operated by healthcare practitioners. Practitioner monitors patient respiration and valve function during/after nebulization to ensure no sticky residue buildup from drugs interferes with valve performance. Benefits patient by maintaining elevated FiO2 during aerosol treatments without impacting particle delivery.

Clinical Evidence

Bench testing only. Evaluated valve flow resistance (typical drop 1.07 cmH2O at 60 lpm) and material compatibility with common nebulized aqueous solutions (Albuterol Sulfate, Ipratropium Bromide, Cromolyn Sodium) over 2.5 hours of continuous use. Results indicate no material impact on valve performance.

Technological Characteristics

Material: Santoprene (medical grade thermoplastic elastomer). Form factor: Right-angle fitting. Connectivity: Mechanical interface for 18mm OD or 22mm ID nebulizer ports. Principle: Passive fluid/gas pathway. No energy source. No software.

Indications for Use

Indicated for patients requiring nebulized drug delivery while simultaneously receiving elevated inspired oxygen concentrations.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### SMDA Summary: ର) ### Hi-Ox® High FiO2 Oxygen Mask with the Aerosol Adapter - Summary of Safety and Effectivenesss | Company: | SensorMedics Corporation | |------------|--------------------------| | Address: | 22705 Savi Ranch Parkway | | | Yorba Linda, CA 92887 | | Telephone: | 714 283-2228 | | Fax: | 714 283-8426 | ### Contact: Yvette Lloyd #### Proprietary Name: Hi-Ox® High FiO2 Oxygen Mask with the Aerosol Adapter #### Common Name: Oxygen Mask with Nebulizer Adapter #### Intended Use: The Hi-Ox® High FiO2 Mask with the Aerosol Adapter is intended as an accessory for compatible nebulizers. The mask with adaptor is -intended to enable the delivery of nebulized drugs to patients who also require elevated inspired oxygen concentrations. ### Description of the Device: The Aerosol Adapter for the Hi-Ox® is an adapter that enables standard small-volume nebulizers (use only cup type, gas driven nebulizers without tee-fittings or air entrainment ports) with either 18 mm OD or 22 mm ID outlet ports to be adapted to the Hi-Ox® . This will enable patients to be treated with nebulized products while receiving high concentrations of oxygen at moderate flow rates of 8 -10 lpm. The valves in the Hi-Ox® have been tested and found to function normally after nebulization of some commonly nebulized drugs in aqueous formulations. However, the valves could malfunction if the prescribed drug creates a sticky residue on the valves. The practitioner should observe the patient's respiration and the function of the Hi-Ox valves during and after nebulization of drugs via the Hi-Ox80 The Aerosol Adapter is a simple right angle fitting made of Santoprene, a medical grade thermoplastic elastomer material. The adapter can be snapped into a normally plugged port on the Hi-Ox® just below the mask and between the inhalation-exhalation valves. This location for {1}------------------------------------------------ the adapter places nothing between the nebulizer and the patient's airway so as to not impact any of the particles produced by the nebulizer. The qas driving the nebulizer can supplement the normal oxygen supply for the Hi-Ox®, and any flow in excess of what is required by the patient can exit the mask via the normal exhalation pathway. There is no risk of pressure buildup beyond that of the cracking pressure and flow resistance of the exhalation valve (~ 1 cmH2O/L/sec). This enables the patient to receive their nebulizer treatments while maintaining an elevated FiO2. # Clinical and Non-Clinical Tests of Equivalency: Most aerosol adapters, which are considered as Class I devices, connect a nebulizer either directly to a disposable aerosol mask or via a right-angle adapter to a mouthpiece. The Aerosol Adapter for the Hi-Ox 0 is equivalent to a right-angle adapter that connects a nebulizer to a mask. The Aerosol Adapter for the Hi-Ox80 would have the same effects on nebulizer performance as any other tee-fitting for nebulizer connections. Flow resistance of all the one-way valves in the Hi-Ox® employed have been found to be well below the design target of <1.5 cmH2O at flows of 60 lpm, (i.e. 0.025 cmH2O per lpm): "Typical pressure drops are in the range of 1.07 cmH2O at 60 lpm. Testing of up to 2.5 hours of continuous nebulization of commonly nebulized aqueous solutions (i.e., Albuterol Sulfate, Ipratropium Bromide and Cromolyn Sodium) indicates that nebulization of these drugs have no material impact on the performance of the valves within the Hi-Ox® and therefore imposes no increase in risk to the patient. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling human figures or waves, arranged in a cascading manner. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR - 4 2004 Mr. Paul Kittinger Manager, Regulatory Affairs SensorMedics Corporation 22705 Savi Ranch Parkway Yorba Linda, California 92887-4645 Re: K032516 Trade/Device Name: Hi-OX 80 Aerosol Adapter Regulation Number: 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: December 9, 2003 Received: December 11, 2003 Dear Mr. Kittinger: We have reviewed your Section 510(k) premarket notification of intent to market the device we mayor and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Interside commerce proces that have been reclassified in accordance with the provisions of Amendments, of to are , and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls provided by annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Mr), it may of badyers in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device A may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Kittinger Please be advised that F.DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. hiu Lin, Ph Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Ko32516 510{k) Number (if known): Hi-Ox® High FiO2 Mask with the Aerosol Adapter Device Name: Indications For Use: The Hi-Ox® High FiO2 Mask with the Aerosol Adapter is intended as an accessory for compatible nebulizers. The mask with adaptor is intended to enable the delivery of nebulized drugs to patients who also require elevated inspired oxygen concentrations. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subcar C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) CAWh (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices Page 1 of 1 __________________________________________________________________________________________________________________________________________________________________ 510(k) Number: K032516
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...