MED2000 SPA NEBULIZER COMPRESSORS, MODELS P1 AND P2, WITH NEBULIZER, AND MED2000 SPA NEBULIZER, MODEL A1/C (ANDYFLOW)
Applicant
Med2000 S.R.L.
Product Code
CAF · Anesthesiology
Decision Date
Jan 14, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Med2000 SpA Nebulizer Compressors, Models P1 and P2, are AC-powered air compressors, with adapter, intended to provide a source of compressed air for medical purposes for use in home health care. These devices are provided with the Model A1/C (AndyFlow) pneumatic nebulizer and should only be used with this nebulizer to produce a fine aerosol mist of medication for respiratory therapy, for inhalation by a patient for treatment of respiratory disorders such as allergies, asthma, cystic fibrosis, COPD, etc. It can be used with adult or pediatric patients. The Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories, is a pneumatic nebulizer which, when driven by a compatible air compressor, nebulizes specific inhalable drugs for inhalation by a patient. It can be used with adult or pediatric patients.
Device Story
System comprises AC-powered piston compressor (Models P1/P2) and pneumatic nebulizer (Model A1/C). Compressor provides compressed air via Venturi principle to nebulizer; nebulizer converts liquid medication into fine aerosol mist for patient inhalation. Used in home health care settings by patients or caregivers. Device lacks microprocessors. Healthcare provider prescribes medication; patient operates device to self-administer therapy. Benefits include targeted delivery of respiratory medication to lungs for treatment of chronic/acute respiratory conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including particle characterization, biocompatibility (ISO 10993), electrical safety, EMC, and environmental stress testing (vibration, impact, temperature, humidity).
Technological Characteristics
Piston-style air compressor; polypropylene nebulizer; Venturi-based aerosolization. Dimensions: 6.0" x 4.7" x 2.4"; weight: 1.1 lbs. Power: 115 VAC/16 VAC adapter. No microprocessors. Biocompatibility per ISO 10993.
Indications for Use
Indicated for adult and pediatric patients requiring respiratory therapy for disorders including allergies, asthma, cystic fibrosis, and COPD. Used to deliver aerosolized medication via inhalation.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Mister Neb™ Nebulizer Compressor, Model HS123 (K013027)
Salter Labs Ultramist Nebulizer, 8960 Series (K961476)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten text string that appears to be a combination of letters and numbers. The string is oriented diagonally, running from the lower-left to the upper-right corner of the image. The characters are written in a simple, somewhat childlike style, with thick, uneven lines. The string appears to read "K031908".
JAN 1 4 2004
Attachment #2
## 510(K) SUMMARY
This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:________________________________________________________________________________________________________________________________________________
#### Submitter's Identification: 1.
Med2000 SpA Via dell'Artigianato, 23/25 25080 Padenghe Sul Garda, (BS) ITALY
Contact:
Mr. Sandro Rossi CEO
Date Summary Prepared:
January 9, 2004
#### 2. Name of the Device:
- 1. Med2000 Nebulizer Compressor, Model P1, with Nebulizer
- 2. Med2000 Nebulizer Compressor, Model P2, with Nebulizer
- 3. Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories
#### Predicate Device Information: 3.
- 1. K#013027, Mister Neb ™ Nebulizer Compressor, with Nebulizer, Model HS123, Respironics Healthscan, Inc., Cedar Grove, NJ
- 2. K#961476, Salter Labs Ultramist Nebulizer, 8960 Series, Salter Labs, Arvin, CA
- 3. K#014056, Salter Labs Modified Ultramist Nebulizer, Salter Labs, Arvin, CA
- 4. K#021742, Galemed Neb-Easy Nebulizer, Galemed Corp., Temecula, CA
{1}------------------------------------------------
#### 4. Device Description:
#### Med2000 SpA Nebulizer Compressors, Models P1 and P2: .
The two models (P1 and P2) differ only in minor cosmetic aspects of their shape that have no effect on performance or safety. These piston compressors are housed in a compact, irregularly-shaped white-plastic housing. Dimensions are: 60" x 4.7" x 2.4" for both compressors, and weigh 1.1 lbs. Neither models contain microprocessors. A "power cube" transformer (Med2000 Class 2 power unit) plugs into a wall outlet (2-blade plug) and provides 16 VAC to the motor via a thin 2-conductor cord that plugs into the compressor. The compressors are supplied with compatible patient tubing (2 meters), and will be marketed with a compatible medication nebulizer and respective instruction manuals.
#### Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories: .
The Nebulizer, Model A1/C (AndyFlow), is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing.
This device operates on the Venturi principle and is refillable. It is manufactured using polypropylene plastics, injection molded and is intended for single patient use.
The Nebulizer, Model A1/C (AndyFlow), has several attachments that are used with the nebulizer during the nebulization process. The attachments include a mouthpiece, a 2 meter connecting tube (which connects to the lower part of the nebulizer opposite the mouthpiece), a pediatric breathing mask and an adult breathing mask. These are standard accessories for nebulizer devices.
The Nebulizer, Model A1/C (AndyFlow), is a hand-held pneumatically powered nebulizer that consists of a nebulizer top that is screwed into a nebulizer cup. The bottom of the cup has a fitting to accept a source of nebulizing gas. The device consists of a nebulizer portion (which forms and conveys an aerosol) and a reservoir portion (which contains non-aerosolized drug).
#### 5. Intended Use:
The Med2000 SpA Nebulizer Compressors, Models P1 and P2, are AC-powered air compressors, with adapter, intended to provide a source of compressed air for medical purposes for use in home health care. These devices are provided with
{2}------------------------------------------------
the Model A1/C (AndyFlow) pneumatic nebulizer and should only be used with this nebulizer to produce a fine aerosol mist. of medication for respiratory therapy, for inhalation by a patient for treatment of respiratory disorders such as allergies, asthma, cystic fibrosis, COPD, etc. It can be used with adult or pediatric patients.
The Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories, is a pneumatic nebulizer which, when driven by a compatible air compressor, nebulizes specific inhalable drugs for inhalation by a patient. It can be used with adult or pediatric patients.
#### Comparison to Predicate Devices: 6.
### Nebulizer Compressors:
The Med2000 SpA Nebulizer Compressors, Models P1 and P2 (difference between models is body shape only) is substantially equivalent to the Mister Neb™ Nebulizer Compressor, with Nebulizer, Model HS123, K#013027, Respironics Healthscan, Inc., Cedar Grove, NJ. This predicate device was cleared with the same indications for use as our device.
Table 1, below, outlines similarities and differences between our nebulizer compressor and the predicate device, as follows:
| CHARACTERISTICS | MED 2000 SpA<br>NEBULIZER<br>COMPRESSORS WITH<br>NEBULIZER | MISTER NEB™<br>NEBULIZER<br>COMPRESSOR WITH<br>NEBULIZER |
|-------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------|
| Model | P1 and P2 | HS123 |
| Dimensions | 6.0" x 4.7" x 2.4" * | 7.48" x 7.08" x 4.3" |
| Weight | 2.3 lbs. | 3.2.lbs |
| Electrical Requirements | 115 VAC/16 VAC (with<br>AC/AC adapter) 60 Hz | 115 VAC/ 60 Hz |
| Power Consumption | 23 watts | 81 watts |
| Maximum Compressor<br>Pressure | 29 psig | 29 psig |
| Maximum Flow | 15 Ipm | Not Available |
| Typical Performance with<br>Nebulizer (Average<br>Compressor Flow Rate) | 9.0 lpm @ 6 psig | 8.0 lpm @ 10 psig |
| Power Indicator | NO | NO |
| Intensity Control | NO | NO |
| Intensity Indicator | NO | NO |
# TABLE 1: NEBULIZER COMPRESSOR COMPARISON CHART
{3}------------------------------------------------
| On/Off Switch | YES | YES |
|-----------------------|-----|-----|
| Without AC/AC Adapter | | |
## Nebulizer:
The Med200 SpA Nebulizer, and Accessories, Model A1/C (AndyFlow), is substantially equivalent to the Salter Labs Ultramist Nebulilzer, Model 9860 Series, K#961476 and K#014056, Salter Labs, Aruin, Ca, and, the Galemed Neh-Easy Nebulizer, K#021742, Galemed Corp, Temecula, CA. These predicate devices were cleared with the same indications for use as our device.
### Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:
The following performance testing was conducted:
- EPA PM 10 Testing .
- Nebulizer Particle Characterization Studies .
- ISO 10993 Biocompatibility Testing .
- Maximum pressure and flow under all combinations of the following: .
- Temperatures of +5º C and +20º C, and 40º C with 90% RH
- Line voltage of 95, 115 and 132 V
- Storage at -20º C and at +60º C
- Fluid Spill Resistance C.
- Maximum Surface and Air Temperatures ರ.
- Sinusoidal Vibration e.
- Impact (drop) Resistance ت
- Leakage Current and Dielectric Withstand (Electrical Safety) ದಿ.
- EMC Testing .
b.
#### Discussion of Clinical Tests Performed: 8.
Not Applicable
{4}------------------------------------------------
#### 9. Conclusions:
The Med2000 SpA Nebulizer Compressors, Models P1 and P2, with Nebulizer, and, a separate Nebulizer, to be known as the Med2000 SpA Nebulizer, Model A1/C soparate NoballED), to sories have the same intended use and similar characteristics as the predicate. Moreover, bench testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of any allettiveness. Thus, the Med2000 SpA Nebulizer Compressors, Models P1 and P2, with Nebulizer, and, a separate Nebulizer, to be known as the Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling human figures or flowing lines. The logo is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 4 2004
Mcd2000 SpA c/o Ms. Susan D. Goldstein-Falk mdi Consultants 55 Northern Blvd Suite 200 Great Neck. New York 11021
K031908 Re:
Trade Name: Med2000 Nebulizer Compressor, Model P1, with Nebulizer Med2000 Nebulizer Compressor, Model P2, with Nebulizer Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories Regulation Number: 21 CFR 868.5630 Regulation Namc: Nebulizer (Direct Patient Interface) Regulatory Class: II Product Code: CAF, BTI Dated: October 22, 2003 Received: October 23, 2003
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device micrate commerce prior to was as that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket നിം Federal I vou, Drug, und Commay, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls providents of an annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 MA), it may be subject to the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device ear be roundsh further announcements concerning your device in the Ecderal Register.
{6}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA' s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements mean that 227 mas made statutes and regulations administered by other Federal agencies. or the Act of any 1 each a the Act' s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and ibuing (= = = = = forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), rt you deems specifical of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Attachment #1
Page __ 1 __ of __ 1
510(k) Number (if known): _(03(409)
Device Name: Med2000 SpA Nebulizer Compressors, Models P1 and P2, with Nebulizer, and, Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories
### Indications For Use:
The Med2000 SpA Nebulizer Compressors, Models P1 and P2, are AC-powered air compressors, with adapter, intended to provide a source of compressed air for medical purposes for use in home health care. These devices are provided with the Model A1/C (AndyFlow) pneumatic nebulizer and should only be used with this nebulizer to produce a fine aerosol mist of medication for respiratory therapy, for inhalation by a patient for treatment of respiratory disorders such as allergies, asthma, cystic fibrosis. COPD. etc. It can be used with adult or pediatric patients.
The Med2000 SpA Nebulizer, Model A1/C (AndyFlow) with Accessories, is a pneumatic nebulizer which, when driven by a compatible air compressor, nebulizes specific inhalable drugs for inhalation by a patient. It can be used with adult or pediatric patients.
Prescription Use Use (Per 21 CFR 801 Subpart D) Over-The Counter
OR (21 CFT 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
((Division Sign-On)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number.*_*_
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.