BASIS SPINAL SYSTEM

K051039 · Medtronic Sofamor Danek USA, Inc. · KWQ · May 12, 2005 · Orthopedic

Device Facts

Record IDK051039
Device NameBASIS SPINAL SYSTEM
ApplicantMedtronic Sofamor Danek USA, Inc.
Product CodeKWQ · Orthopedic
Decision DateMay 12, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The BASIS™ anterior cervical plates are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. The BASIS™ anterior Thoracic and Lumbar plates are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. When used as an anterolateral thoracic/lumbar, or anterior cervical system, BASIS™ components are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion.

Device Story

BASIS™ Spinal System consists of titanium alloy screws, bolts, nuts, and plates; used for spinal intervertebral body fixation. Components assembled into tailor-made, rigidly locked constructs for individual patient anatomy. Used by surgeons in clinical settings to stabilize spine segments. Provides mechanical support to facilitate fusion or stabilization in patients with spinal pathologies. Contraindicated to mix stainless steel and titanium components in same construct.

Clinical Evidence

No clinical data provided; substantial equivalence based on design and material characteristics.

Technological Characteristics

Materials: Titanium alloy. Components: Screws, bolts, nuts, plates. Fixation: Rigidly locked constructs. Form factor: Modular spinal fixation system. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring spinal intervertebral body fixation for spinal stenosis, spondylolisthesis, spinal deformities (scoliosis, kyphosis, lordosis), fracture, pseudarthrosis, tumor resection, or failed previous fusion. Applicable to anterior cervical and anterolateral thoracic/lumbar regions.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAY 1 2 2005 ## BASISTM Spinal System Summary of Safety and Effectiveness May 2005 - Medtronic Sofamor Danek, Inc. USA Company: I. 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 Richard W. Treharne, PhD Contact: Senior Vice President Regulatory Affairs - Proposed Proprietary Trade Name: BASIS™ Spinal System II. - Classification Name(s)/Product Code(s): III. Classification Name: Spinal Intervertebral Body Fixation Orthosis (per 21 CFR Section 888.3060) Product Codes: KWQ #### Product Description IV. The BASIS™ Spinal System components included in this submission consist of screws, bolts, nuts and plates as well as ancillary instrument sets. The BASIS™ implant components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The BASIS™ Spinal System implant components are made from titanium alloy. No warranties express, or implied, are made. Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded. See the MSD Catalog for further information about warranties and limitations of liability. Never use stainless steel and titanium implant components in the same construct. {1}------------------------------------------------ #### Indications V. ### BASIS™ Cervical Plates The BASIS™ anterior cervical plates are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. ### BASIS™ Thoracic/Lumbar Plates The BASIS™ anterior Thoracic and Lumbar plates are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. ### BASIS™ Components When used as an anterolateral thoracic/lumbar, or anterior cervical system, BASIS™ components are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. #### Substantial Equivalence VI. Documentation was provided which demonstrated the BASIS™ Spinal System to be substantially equivalent to the following systems: SYNTHES® Cervical Vertebrae Plates (K792352), SYNTHES® Spinal Implant System (K802462), ORION® Anterior Cervical Plate System (K973854), ZPLATE-ATL™ Anterior Fixation System and ZPLATE-ATL™ Anterior Fixation System ZPLATE-II (K922543 and K990603) and the VANTAGE™ Anterior Fixation System Plate (K023797). {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, placed within a circle. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged around the upper portion of the circle. MAY 1 2 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Richard W. Treharne, PhD Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132 Re: K051039 Trade/Device Name: BASIS™ Spinal System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: KWQ Dated: April 22, 2005 Received: April 25, 2005 Dear Dr. Treharne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and in your of substantial equivalence of your device to a legally premits to the time in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acvies in 1 3 % (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general mionnairen val and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Stupt Clurdon A Miriam Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # KOSIO34 CONFIDENTIAl Page 1 of of 1 l April 2005 510(k) Number (if known): Device Name: _______________________________________ BASISTM Spinal System Indications For Use #### BASIS™ Cervical Plates The BASIS™ anterior cervical plates are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. #### BASIS™ Thoracic/Lumbar Plates The BASIS™ anterior Thoracic and Lumbar plates are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. #### BASIS™ Components When used as an anterolateral thoracic/lumbar, or anterior cervical system, BASIS™ components are intended for the following indications: (1) spinal stenosis, (2) spondylolisthesis, (3) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (4) fracture, (5) pseudarthrosis, (6) tumor resection, and/or (7) failed previous fusion. | Prescription Use | X | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | AND/OR | Over-The-Counter Use | | |------------------------|--| | (21 CFR 807 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Hurt Chords (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number ................................................................................................................................................................ 000000 ( Image /page/4/Picture/18 description: The image shows a black and white abstract pattern. The pattern appears to be a cluster of dark, irregular shapes against a white background. The shapes are densely packed together, creating a textured effect. The overall impression is one of randomness and organic form.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...