POLARIS SPINAL SYSTEM
K103393 · Biomet Spine · KWQ · Feb 2, 2011 · Orthopedic
Device Facts
| Record ID | K103393 |
| Device Name | POLARIS SPINAL SYSTEM |
| Applicant | Biomet Spine |
| Product Code | KWQ · Orthopedic |
| Decision Date | Feb 2, 2011 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3060 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion. The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The Accu Vision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
Device Story
Polaris Spinal System is a non-cervical spinal fixation system; line extension adds 6.35mm titanium uniplanar screws. System provides immobilization and stabilization as an adjunct to fusion via pedicle, hook, or sacral/iliac screw fixation. Used in posterior or anterolateral spinal surgery; supports percutaneous (Ballista instruments) and minimally invasive (Accu Vision instruments) approaches. Dominos allow connection to other Biomet spinal systems (Altius, Array, Omega21, Synergy) for multi-level fixation. Operated by surgeons in clinical settings. Provides mechanical stabilization to the spine; aids in fusion process; benefits patients by correcting deformities or stabilizing unstable spinal segments.
Clinical Evidence
Bench testing only. Mechanical testing conducted per ASTM F1717-04, including static compression bending, static torsion, and dynamic compression bending fatigue in a vertebrectomy model. Testing demonstrated the subject device is substantially equivalent to predicate spinal systems.
Technological Characteristics
Titanium alloy components; uniplanar screw design. Mechanical testing performed per ASTM F1717-04. System includes cannulated screws and percutaneous/minimally invasive instrumentation. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring non-cervical spinal fixation as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), deformity (scoliosis, kyphosis, lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion. Pedicle screw fixation requires autograft.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
- Polaris Spinal System (K090523)
- Polaris Spinal System (K011437)
- Polaris Spinal System (K041449)
- Polaris Spinal System (K081952)
- Polaris Spinal System (K100220)
Reference Devices
- Altius Spinal System
- Array Spinal System
- Biomet Omega21 Spinal System
- Synergy Spinal System
Related Devices
- K111957 — POLARIS SPINAL SYSTEM · Biomet Spine (Ebi, LLC) · Aug 19, 2011
- K100438 — POLARIS SPINAL SYSTEM, POLARIS/SYNERGY STRUCTURE ANTERIOR FIXATION SYSTEM · Ebi, L.P. · Jun 23, 2010
- K100409 — POLARIS SPINAL SYSTEM, POLARIS/SYNERGY STRUCTUE ANTERIOR FIXATION SYSTEM · Biomet Spine (Aka Ebi, LLC) · Jun 23, 2010
- K133746 — POLARIS SPINAL SYSTEM · Biomet Spine (Aka Ebi, LLC) · Mar 12, 2014
- K123549 — POLARIS SPINAL SYSTEM - BALLISTA II PERCUTANEOUS SCREW PLACEMENT SYSTEM · Biomet Spine (Aka Ebi, LLC) · Feb 25, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
page 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Biomet Spine. The logo is in black and white and features the word "BIOMET" in large, bold letters. Below the word "BIOMET" is the word "SPINE" in smaller letters. The logo is simple and modern.
# FEB - 2 2011
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
| Preparation Date: | November 9, 2010 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant/Sponsor: | Biomet Spine<br>100 Interpace Parkway<br>Parsippany, NJ 07054 |
| Contact Person: | Vivian Kelly<br>Phone: 973-299-9300 x2214<br>Fax: 973-257-0232 |
| Trade name: | Polaris Spinal System |
| Common Name: | Non-cervical spinal fixation system |
| Classification Name<br>(Product Code): | Posterior, noncervical, nonpedicle use (KWP)<br>Anterior/anterolateral noncervical use (KWQ)<br>Noncervical pedicle applications (MNI, MNH and NKB) |
| Device Panel - Regulation No.: | Orthopedic - 21 CFR 888.3050, 888.3060 and 888.3070 |
## Device Description:
1103393
This submission is a line extension to Polaris Spinal System to add 6.35 titanium uniplanar screws to the system.
#### Indications for Use:
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion.
The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The Accu Vision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration
{1}------------------------------------------------
page 2 of 2
of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
#### Summary of Technologies:
The technological characteristics of the new components are the same as, or similar to, the predicate devices.
#### Performance Data:
K103293
Mechanical testing was conducted in accordance with FDA's Guidance for Industry and FDA Staff -Spinal System 510(k)s dated May 3, 2004. Per the guidance document, the following testing was conducted: static compression bending, static torsion and dynamic compression bending fatigue per ASTM 1717-04, Standard Test Methods for Static and Fatigue for Spinal Implant Constructs in a Vertebrectomy Model. The mechanical testing verifies that the subject device is substantially equivalent to other spinal systems currently on the market and has met all mechanical test requirements based on the worst-case construct testing.
### Substantial Equivalence:
The 6.35 titanium uniplanar screws in the Polaris Spinal System are substantially equivalent to the other screws in Polaris Spinal System cleared in K090523, K011437, K041449 K081952 and K 100220. These new screws are substantially equivalent to the predicate screws with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness.
#### Conclusion:
The subject device is substantially equivalent to its predicates when used as a spinal fixation device. The indications for use and fundamental technology of the device remain unchanged. Furthermore, mechanical testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject device to the other components in the Polaris Spinal System. Based on this information, the subject device does not raise any new issues regarding the safety or efficacy.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three wavy lines, which are meant to represent the human form.
Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet Spine % Ms. Vivian Kelly 100 Interpace Parkway Parsippany, New Jersey 07054
Re: K103393
FEB - 2 201
Trade/Device Name: Polaris Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, KWP, KWQ Dated: January 05, 2011 Received: January 06, 2011
Dear Ms. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 – Ms. Vivian Kelly
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
ACy B.D.h
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use K10339 510(k) Number (if known):
Device Name: Polaris Spinal System
Indications for Use:
. )
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and fordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion.
The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The AccuVision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------|--|
| Division of Surgical. Orthopedic, | |
| and Restorative Devices | |
| 510(k) Number | K103393 |
|---------------|---------|
|---------------|---------|
Page 1 of 1
Page 4-3