XTEND ANTERIOR CERVICAL PLATE SYSTEM

K092146 · Globus Medical, Inc. · KWQ · Nov 16, 2009 · Orthopedic

Device Facts

Record IDK092146
Device NameXTEND ANTERIOR CERVICAL PLATE SYSTEM
ApplicantGlobus Medical, Inc.
Product CodeKWQ · Orthopedic
Decision DateNov 16, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The XTEND™ Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis.

Device Story

XTEND™ Anterior Cervical Plate System provides anterior stabilization of cervical spine (C2-C7). System comprises standard, extender, and universal extender plates; used with variable or fixed angle screws. Extender plates facilitate revision surgery by attaching to existing implants. Implants are surgically attached to anterior vertebral body. Device used in clinical/OR setting by surgeons to stabilize spine; aids fusion process; benefits patients by correcting deformity or providing stability in trauma/degenerative cases.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Titanium alloy construction per ASTM F136 and F1295. System includes plates of various lengths and variable/fixed angle screws. Mechanical fixation principle. No software or electronic components.

Indications for Use

Indicated for patients requiring anterior cervical spine fixation (C2-C7) due to degenerative disc disease, trauma (fractures), tumors, deformity (kyphosis, lordosis, scoliosis), pseudarthrosis, failed fusion, spondylolisthesis, or spinal stenosis.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092146 NOV 1 6 2009 # 510(K) Summary: XTEND™ Anterior Cervical Plate System | Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>(610) 415-9000 | |-----------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Kelly J. Baker, Ph.D<br>Director, Clinical Affairs & Regulatory | | Device Name: | XTEND™ Anterior Cervical Plate System | | Classification: | Per 21 CFR as follows:<br>§888.3060 Spinal Intervertebral Body Fixation Orthosis<br>Product Code KWQ.<br>Regulatory Class II, Panel Code 87. | | Predicate(s): | VIP® K081391 (SE date July 3, 2008)<br>ASSURE® K040721 (SE date June 17, 2004)<br>PROVIDENCE® K070775 (SE date April 19, 2007) | ## Device Description: The XTEND™ Anterior Cervical Plate System consists of standard plates, Extender plates and Universal Extender plates. Extender plates may be used for revision surgery in which additional stabilization is required. Extender plates are attached to an adjacent XTEND™ plate, and Universal Extender plates are inserted adjacent to other plates. XTEND™ plates are available in various lengths to be used with either variable angle screws or fixed angle screws. Each XTEND™ plate is attached to the anterior portion of the vertebral body of the cervical spine (levels C2-C7). The XTEND™ Anterior Cervical Plate System implants are composed of titanium allov, as specified in ASTM F136 and F1295. #### Intended Use: The XTEND™ Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis. #### Basis of Substantial Equivalence: XTEND™ Anterior Cervical Plate System is similar to the predicate systems with respect to technical characteristics, performance, and intended use. Mechanical testing in accordance with the "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s", May 3, 2004 is presented. {1}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Globus Medical. Inc. % Kelly J. Baker, Ph.D. 2560 General Armistead Avenue, Valley Forge Audubon, Pennsylvania 19403 NOV 1 6 2009 Re: K092146 Trade/Device Name: XTEND Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Fixation Orthosis Regulatory Class: Class II Product Code: KWO Dated: November 10, 2009 Received: November 12, 2009 Dear Dr. Baker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the incications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Foond, Or to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misburoni, in and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your dovine can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Kelly J. Baker. Ph.D. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm fou the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. .fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm Sincerely vours. Mark A. Millmann Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Statement Kog 2146 510(k) Number: XTEND™ Anterior Cervical Plate System Device Name: ## INDICATIONS: The XTEND™ Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine C2-C7 for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis. Prescription Use × (Per 21 CFR §801.109) OR Over-The-Counter Use (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark A. Milliken (Division Sign-Off) (Division Sign-gical, Orthopedic, and Restorative Devices 510(k) Number K092144
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...