Anterior Cervical Plate System

K202972 · Nutech Spine and Biologics · KWQ · Nov 24, 2020 · Orthopedic

Device Facts

Record IDK202972
Device NameAnterior Cervical Plate System
ApplicantNutech Spine and Biologics
Product CodeKWQ · Orthopedic
Decision DateNov 24, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Nutech Spine and Biologics Anterior Cervical Plate System is interior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.

Device Story

Anterior Cervical Plate System; provides spinal intervertebral body fixation. System comprises titanium bone plates and screws in various sizes/lengths for patient-specific configurations. Triangular plate design facilitates adjacent level fusion without requiring removal of prior plates. Used in surgical settings by physicians for cervical spine stabilization. Implants provide mechanical fixation to support fusion; clinical benefit includes stabilization of cervical spine segments affected by degenerative disease, trauma, or deformity.

Clinical Evidence

Bench testing only. Static compression bending, static torsion, static tension bending, and dynamic compression bending testing performed per ASTM F1717-18.

Technological Characteristics

Materials: Ti 6Al4V ELI (ASTM F136). Components: Titanium bone plates and screws. Design: Triangular plate shape for adjacent level fusion. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for anterior screw fixation to the cervical spine (C2-T1) in patients with degenerative disc disease, trauma (fractures), tumors, deformity (kyphosis, lordosis, scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, or spinal stenosis.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 24, 2020 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' in blue, with 'ADMINISTRATION' written below in a smaller font. Nutech Spine and Biologics % Mr. Daniel Lanois Consultant SurgOp Support 101 Lamond Ct Prosper, Texas 75078 Re: K202972 Trade/Device Name: Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: March 17, 2020 Received: September 30, 2020 Dear Mr. Lanois: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) ## K202972 Device Name Anterior Cervical Plate System #### Indications for Use (Describe) The Nutech Spine and Biologics Anterior Cervical Plate System is interior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | Date Prepared | November 18, 2020 | | | |-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|--------------| | Manufacturer | Nutech Spine and Biologics | | | | Address | 600 Luckie Drive, Suite 424, Birmingham, Alabama 35223 | | | | Telephone | 205-329-7285 | | | | Fax | 888-475-7793 | | | | Contact Person | Daniel Lanois | | | | Address | Consultant | | | | | SurgOp Support LLC | | | | | 101 Lamond Ct, Prosper, TX 75078 | | | | Telephone | 678-371-3605 | | | | Fax | | | | | Email | daniel@surgopsupport.com | | | | Trade Name | Anterior Cervical Plate System | | | | Common Name | Anterior Cervical Plate System | | | | Panel Code | Orthopedic/87 | | | | Classification Name | Spinal Intervertebral Body Fixation Orthosis | | | | Class | Class II | | | | Regulation Number | 21 CFR 888.3060 | | | | Product Code | KWQ | | | | | Name of Primary Predicate Device | 510(k) # | Manufacturer | | | Diamond Anterior Cervical Plate System | K100265 | Amendia | | Description | The Anterior Cervical Plate System consists of single level titanium bone<br>plates of various sizes and titanium bone screws in various diameters and<br>lengths to allow for patient specific configurations. The triangular shape of<br>the plate allows for adjacent level fusion without removal of prior Nutech<br>Cervical Plates. Implants are made from Ti 6AI4V ELI, per ATSM F136. The<br>instrumentation allows for simple preparation and placement of the implant<br>construct. | | | | Indications and<br>Intended Use | The Nutech Spine and Biologics Anterior Cervical Plate System is intended<br>for anterior screw fixation to the cervical spine (C2-T1) for the following<br>indications: degenerative disc disease (as defined by neck pain of discogenic<br>origin with degeneration of the disc confirmed by patient history and<br>radiographic studies), trauma (including fractures), tumors, deformity<br>(kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions,<br>spondylolisthesis, and spinal stenosis. | | | | Technological<br>Characteristics/<br>Substantial<br>Equivalence | Documentation was provided to demonstrate that the Subject device, is<br>substantially equivalent to the Predicate. The Subject device is substantially<br>equivalent to the Predicate device in intended use, indications for use,<br>materials, technological characteristics, and labeling. | | | | Performance<br>Data | Static Compression Bending, Static Torsion, Static Tension Bending, and<br>Dynamic Compression Bending testing (per ASTM F1717-18) confirmed that<br>the Subject device performed as intended. | | | | Conclusion | Based on the intended use, indications for use, technological characteristics,<br>materials, and comparison to Predicate devices, the Subject device has been | | | # 510(k) Summary (as required by 21 CFR 807.92)
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%