Codman Certas Plus Electronic Tool Kit

K181902 · Integra Lifesciences Corp. · JXG · Oct 25, 2018 · Neurology

Device Facts

Record IDK181902
Device NameCodman Certas Plus Electronic Tool Kit
ApplicantIntegra Lifesciences Corp.
Product CodeJXG · Neurology
Decision DateOct 25, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5550
Device ClassClass 2

Indications for Use

The Codman Certas Plus Electronic Tool Kit allows the non-invasive reading or adjustment of the Codman Certas and Certas Plus Programmable Valves.

Device Story

The Codman Certas Plus Electronic Tool Kit is a non-invasive system for reading and adjusting implanted Codman Certas and Certas Plus Programmable Valves. It consists of a Locator/Indicator Tool, an Adjustment Tool, and an X-Ray Overlay Tool. The Locator/Indicator Tool uses magnetic field sensing to determine the valve setting, displaying the result on an integrated LED screen. The Adjustment Tool uses magnetic coupling to rotate the valve's internal construct to one of eight settings. The device is used in a clinical setting by healthcare providers. The LED screen provides visual feedback for valve location and setting, aiding the clinician in accurate adjustment. This system improves upon the predicate by integrating location and indication functions into a single battery-operated tool with digital feedback, facilitating precise valve management and potentially improving patient outcomes by ensuring correct shunt performance.

Clinical Evidence

No clinical studies were performed. Evidence consists of bench testing, including design verification (mechanical performance, electrical safety per IEC 60601-1, environmental, transit, and disinfection compatibility), biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993-1), and simulated-use validation with clinicians. Simulated-use testing confirmed system safety and clinical acceptability, including successful performance in a simulated head model with angular variability.

Technological Characteristics

Materials include PC/ABS, polycarbonate, MuMetal (ASTM-A753 Alloy 4), brass, stainless steel, and neodymium magnets. Operates via magnetic coupling for adjustment and magnetic field sensing for indication. Battery-operated (CR 123A) with LED display. IP4X ingress protection. Non-sterile. Software is used for screen iconography and feedback.

Indications for Use

Indicated for patients with hydrocephalus requiring non-invasive reading or adjustment of Codman Certas and Certas Plus Programmable Valves.

Regulatory Classification

Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products. October 25, 2018 Integra LifeSciences Corp. Nancv MacDonald Regulatory Affairs Manager 11 Cabot Blvd. Mansfield, Massachusetts 02048 Re: K181902 Trade/Device Name: Codman Certas Plus Electronic Tool Kit Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: September 25, 2018 Received: September 26, 2018 Dear Nancy MacDonald: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Xiaolin Zheng -S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K181902 Device Name Codman Certas Plus Electronic Tool Kit #### Indications for Use (Describe) The Codman Certas Plus Electronic Tool Kit allows the non-invasive reading or adjustment of the Codman Certas and Certas Plus Programmable Valves. Type of Use (Select one or both, as applicable) | <span style="font-size: 100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |--------------------------------------------------------------------------------------| | <span style="font-size: 100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary I. Submitter Integra LifeSciences Corp. 11 Cabot Blvd Mansfield, MA 02048 > Primary Contact: Nancy MacDonald Phone: (781) 971-5664 > Secondary Contact: Jocelyn Raposo Phone: (781) 971-5688 Date of Preparation: September 25, 2018 #### II. Device | 510(k) Number | K181902 | |---------------------|------------------------------------------------------------------------| | Name of Device | Codman Certas Plus Electronic Tool Kit | | Common Name | Hydrocephalus Shunt | | Classification Name | Central Nervous System Fluid Shunt and Components<br>(21 CFR 882.5550) | | Regulatory Class | II | | Product Code | JXG | III. Predicate The predicate device for this submission is: Device | 510(k) Number | Date Cleared | Title | |---------------|-------------------|------------------------| | K143111 | February 19, 2015 | Codman Certas Tool Kit | IV. Device The Certas Plus Electronic Tool Kit is used to non-invasively read or adjust the Description setting of a Certas or Certas Plus Programmable Valve before and after implantation in the treatment of hydrocephalus. The Certas Plus Electronic Tool Kit consists of the following tools: Locator/Indicator Tool, Adjustment Tool and X-Ray Overlay Tool. The Locator/Indicator Tool facilitates correct placement of the Adjustment Tool over a pre-implanted valve in the packaging or a post-implanted valve, and measures the valve's magnetic field to display the valve setting. The Adjustment Tool adjusts the valve to one of the 8 valve settings. The magnets in the Adjustment Tool couple with the magnets in the rotating construct of the valve, causing the rotating construct to lift and follow the Adjustment Tool as it is rotated to one of 8 valve settings. {4}------------------------------------------------ #### The Indications for Use statement for the Certas Plus Electronic Tool Kit is V. Indications for Use identical to the predicate Certas Tool Kit: The Codman Certas Plus Electronic Tool Kit allows the non-invasive reading or adjustment of the valve setting for the Codman Certas and Certas Plus Programmable Valves. VI. Compared to the predicate device, the proposed Certas Plus ElectronicTool Kit Comparison includes the modifications listed in Table 1. to Predicate Device | Table 1: Similarities and Differences – Certas Tool Kit and Certas Plus Electronic Tool Kit | | | |---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Certas Took Kit<br>(Predicate) | Certas Plus Electronic Tool Kit<br>(this submission) | | 510(k) Number | K143111 | K181902 | | Indications for Use | The Codman Certas Tool Kit allows the<br>non-invasive reading or adjustment of the<br>valve setting. | The Codman Certas Plus Electronic Tool<br>Kit allows the non-invasive reading or<br>adjustment of the valve setting for the<br>Codman Certas and Certas Plus<br>Programmable Valves. | | Handheld Tools | Locator tool (low profile + adjustable<br>height)<br>Indicator tool<br>Adjustment tool | Locator/Indicator tool<br>Adjustment tool | | Operating<br>Principle | Locator Tool: Facilitates correct<br>placement of the Adjustment Tool and<br>Indicator Tool over an implanted valve. | The Locator and Indicator Tools have<br>been combined into one tool.<br>Locator/Indicator Tool facilitates correct<br>placement of the Adjustment Tool over an<br>implanted valve. | | | Indicator Tool: Incorporates a detection<br>wheel (indication dial) that aligns the<br>valve's magnetic field to display the valve<br>setting. | The Locator/Indicator Tool incorporates a<br>detection wheel (indication dial) that<br>measures the valve's magnetic field to<br>display the valve setting. | | | Adjustment Tool: The magnets in the<br>Adjustment Tool couple with the magnets<br>in the rotating construct of the valve,<br>causing the rotating construct to lift and<br>follow the Adjustment Tool as it is rotated<br>to one of 8 valve settings. | Adjustment Tool: The magnets in the<br>Adjustment Tool couple with the magnets<br>in the rotating construct of the valve,<br>causing the rotating construct to lift and<br>follow the Adjustment Tool as it is rotated<br>to one of 8 valve settings. | | Energy Type | Mechanical, magnetic energy, not battery<br>operated | Locator/Indicator: Battery operated<br>Adjustment Tool: Magnetic energy | | Table 1: Similarities and Differences – Certas Tool Kit and Certas Plus Electronic Tool Kit | | | | | Certas Took Kit<br>(Predicate) | Certas Plus Electronic Tool Kit<br>(this submission) | | Location/Indication<br>Feedback | Location: No location feedback; the user is<br>required to manually palpate to determine<br>the valve location<br><br>Indication: display a mechanical indication<br>dial for the determination of valve<br>performance setting | Location: LED screen is used to display<br>location feedback to aid in the alignment<br>of the Locator/Indicator Tool with the<br>valve mechanism center<br><br>Indication: LED screen is used to display<br>an indication dial for the determination of<br>valve performance settings. | | Materials - Patient<br>Contacting | • Polycarbonate<br>• Polycarbonate, Aluminum with<br>anodized surface, Decal<br>• Acetal Copolymer | • Polycarbonate and acrylonitrile<br>butadiene styrene (PC/ABS)<br>• Pad Printing Ink (blue)<br>• Pad Printing Ink (White) | | Materials - Not<br>Patient Contacting<br>(Locator Tool) | There are various Locator and Indicator<br>Tool materials including, but not limited<br>to:<br>• Polycarbonate<br>• Acetal (Delrin) | Locator/Indicator Tool:<br>• PC/ABS<br>• MuMetal ASTM-A753 Alloy 4<br>• Brass<br>• Polycarbonate<br>• ROHS Compliant Electronics<br>• Nickel plated Stainless Contact and<br>Copper Wire<br>• Polyimide<br>• Insulated Copper Wire<br>• CR 123A, Lithium Battery<br>• Thermal Transfer Polyester<br>• Polyester Film | | Materials - Not<br>Patient Contacting<br>(Adjustment Tool) | There are various Adjustment Tool<br>materials including, but not limited to:<br>• Polycarbonate<br>• Acetal (Delrin) | Adjustment Tool:<br>• PC/ABS TPE<br>• PC/ABS<br>• Neodymium Magnets<br>• Stainless Steel<br>• Thermal Transfer Polyester | | Protective<br>Carrying Case | Yes | Yes | | Accessory tool | X-ray overlay | X-ray overlay | | Sterilization | NA, non-sterile | NA, non-sterile | {5}------------------------------------------------ VII. The following performance data is provided in support of the substantial Performance equivalence determination for this submission. Data ## Verification Testing The following table summarizes the design verification activities. All samples in design verification testing met predefined acceptance criteria, and the proposed device passed design verification test activities. These test results, in conjunction with the similarities to the predicate device, demonstrate that the Certas Plus Electronic Tool Kit is substantially equivalent to the predicate device. {6}------------------------------------------------ | Table 2: Design Verification Activities Codman Certas Plus Electronic Tool Kit | | | |--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Test | Test Method Summary | Results | | Comprehensive<br>Design<br>Verification Study | Testing verified that design outputs adequately met<br>the specified designed inputs including but not limited<br>to the following:<br>Product requirements verified against drawing<br>specifications for predicate Performed to verify expected performance of<br>indication and adjustment features Verify tool screen sequence and iconography<br>features Verify mechanical performance of assembly and<br>carrying case Ensure that a successful indication is provided<br>across the range of tissue thickness Verify that electronic components are<br>incorporated such that their rated maximums are<br>not exceeded during normal use. This includes<br>component thermal rating, operating temperature,<br>electrical voltage withstanding, electrical current<br>capacity and other parameters applicable to<br>specific components | PASS | | Electrical Safety<br>and Ingress<br>Testing | Protect against ingress with a IP4X rating. Per IEC<br>60601-1 (Edition 3.1) | PASS | | Force Tests | Device capable performing intended function after<br>minimum 2700 cycles | PASS | | User Feedback<br>Verification | Verify user interaction requirements associated<br>with screen and user feature design and function. Measure the time to receive an indication from<br>the device. Confirm that users receive appropriate feedback<br>device during adjustment Verify system failure messages. | PASS | | Environmental | Environmental chamber testing durability across<br>typical operational conditions. Demonstrate that<br>environmental storage conditions experienced by the<br>tool kit will not affect its functionality. | PASS | | Transit Testing | Confirm that device and case are secure during<br>transit and that case shall prevent damage to the<br>product in normal conditions of transit, handling,<br>and storage in accordance with ISTA 3A. Vibration, drop, and environmental conditioning (Packaged-Products for Parcel Delivery<br>System Shipments 70kg (150 lb) or Less<br>(standard, small, flat or elongated). | PASS | | Disinfection<br>Compatibility | Demonstrate efficacy of the reprocessing<br>procedure used to disinfect reusable components<br>of the device to achieve low level disinfection<br>using 70% Isopropyl Alcohol Surface Wipes. Confirm no appreciable signs of deterioration of<br>the enclosure after 1000 disinfection cycles. | PASS | {7}------------------------------------------------ ## Validation Testing ## Biocompatibility Testing Codman performed a biocompatibility evaluation in accordance with ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process and FDA guidance, Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a risk management process', published June 16, 2016. The Certas Plus Electronic Tool Kit is considered a surface device with limited skin contact (≤ 24 hours). Biocompatibility testing performed on the subject device, the Certas Plus Electronic Tool Kit, with successful results, confirming that no new issues of safety with successful results, confirming that no new issues of safety and effectiveness are raised. The following tests in Table 3 were conducted on the patient contacting components: cytotoxicity, sensitization and irritation. The device met all predefined acceptance criteria and passed biocompatibility testing. | Table 3: Biocompatibility Testing for the Locator/Indicator Bottom Housing of<br>the Proposed Device | | | |------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------| | Biocompatibility<br>Test | Summary of<br>Results | Conclusion | | Cytotoxicity | No cytotoxicity or cell lysis was<br>noted in any of the test wells. No pH<br>shift was observed at 48 hours. | Non-cytotoxic | | Sensitization | The test article extracts showed no<br>evidence of causing delayed dermal<br>contact sensitization in the guinea<br>pig. The test article was not<br>considered a sensitizer in the guinea<br>pig test. | Not considered a<br>sensitizer | | Irritation | There was no erythema and no edema<br>observed on the skin of the animals<br>treated with the test article extract.<br>The Primary Irritation Index for the<br>test article extracts of sodium<br>chloride and sesame oil were both<br>calculated to be 0.0. The irritation<br>responses to the test article extracts of<br>sodium chloride and sesame was<br>categorized as negligible. | Non-Irritant | All tests results are acceptable and conform to the acceptance criteria outlined in the applicable ISO 10993 standards. The test results demonstrate that the materials in the proposed device are non-cytotoxic, non-sensitizing, and nonirritating and thus biocompatible per ISO 10993-1. Therefore, the proposed Certas Plus Electronic Tool Kit is substantially equivalent to the predicate device with respect to biocompatibility. {8}------------------------------------------------ ## Simulated Post-Implantation Use Codman conducted validation testing with clinicians which included a system safety study and qualitative assessment of clinical acceptability. All acceptance criteria were met. | Table 4: Design Validation – Certas Plus Electronic Tool Kit | | | |--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Test | Test Method Summary | Results | | System Safety Study | Clinicians evaluated the maximum acceptable rate<br>for false positives and maximum success rate for<br>completing a full procedure. Testing conducted on<br>a simulated head model with clinically relevant<br>implant locations to create angular variability. | PASS | | Qualitative<br>Assessment of<br>Clinical Acceptability | After completing simulated use testing with the<br>Certas Plus Electronic Tool Kit, clinicians were<br>asked to:<br>• Provide a qualitative assessment of:<br>◦ Patient comfort (anticipated)<br>◦ Clarity of information on screen<br>◦ acceptability for use in retro-auricular<br>placement<br>◦ Ease of use<br>◦ Clinical acceptability<br>• Identify images of low battery, critical low<br>battery and graphics system failure screen and<br>asked to identify the intent. | PASS | Design Validation testing of the Certas Plus Electronic Tool Kit confirmed that the design outputs meet the customer requirements and confirm various user needs for the individual tools. These results demonstrate the safe and effective use of the Certas Plus Valve and the Certas Plus Electronic Tool Kit as a system for post-implant adjustment of the valve. ## Animal Studies No animal studies were required as appropriate verification and validation of the modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench and biocompatibility testing. ## Clinical Studies No clinical studies were required as appropriate verification and validation of the modifications was achieved based on the similarities of the proposed device to the predicate device, and from results of bench and biocompatibility testing. {9}------------------------------------------------ The Certas Plus Electronic Tool Kit passed the verification and validation testing, demonstrating that it performs similarly to the predicate device, the Certas Tool Kit, with respect to design verification and validation testing. VIII. Based upon the intended use, design, materials, function, comparison to the Conclusion predicate device, and testing performed by Codman, it is concluded that the Certas Plus Electronic Tool Kit is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...