Codman Certas Plus Programmable Valve; Codman Certas Toll Kit
K143111 · Medos International SARL · JXG · Feb 19, 2015 · Neurology
Device Facts
Record ID
K143111
Device Name
Codman Certas Plus Programmable Valve; Codman Certas Toll Kit
Applicant
Medos International SARL
Product Code
JXG · Neurology
Decision Date
Feb 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Codman Certas Plus Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. The Codman Certas Tool Kit allows the noninvasive reading or adjustment of the valve setting.
Device Story
The Codman Certas Plus Programmable Valve is a sterile, single-use implantable device for hydrocephalus management. It utilizes a ruby ball-in-cone valve seat design with a pressure-inducing spring to regulate CSF flow; as differential pressure increases, the ball displaces to increase flow. The valve offers 8 performance settings (25-215 mmH2O, plus a 'virtually off' setting at 400 mmH2O). The Codman Certas Tool Kit, used by clinicians, enables noninvasive reading and adjustment of valve settings post-implantation via magnetic force. The kit includes locator tools for valve positioning and an indicator tool for setting verification. The device is MR-Conditional at 3.0 Tesla. By allowing precise, noninvasive pressure adjustments, the system helps clinicians manage CSF drainage without surgical intervention, potentially improving patient outcomes in hydrocephalus treatment.
Clinical Evidence
No clinical studies were performed. Substantial equivalence was established through bench testing, including structural testing, pressure-flow testing per ISO 7197:2009 and ASTM F647:1994, MRI safety testing per ASTM standards (F2052-14, F2182-11a, F2119-07, F2213-06), and biocompatibility testing per ISO 10993. Validation included pre- and post-implantation simulation, valve patency, and tool kit functionality.
Technological Characteristics
Implantable pressure-regulating valve; ruby ball-in-cone mechanism with spring. Materials: Synthetic ruby. Dimensions: 8 pressure settings (25-215 mmH2O, 400 mmH2O). Energy: Magnetic force (for adjustment). Connectivity: None. Sterilization: Sterile, single-use. Standards: ISO 7197:2009, ASTM F647:1994, ASTM F2052-14, ASTM F2182-11a, ASTM F2119-07, ASTM F2213-06, ISO 10993.
Indications for Use
Indicated for patients requiring management of hydrocephalus via constant intraventricular pressure and drainage of cerebrospinal fluid (CSF).
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Codman Certas Programmable Valve System and Codman Certas Therapy Device Management System (K112156)
Related Devices
K182265 — CodMan Certas Plus Programmable Valve · Integra Lifesciences Corp. · Nov 20, 2018
K152152 — Codman Certas Plus Programmable Valve · Medos International SARL · Oct 27, 2015
K223330 — Certas Plus Programmable Valves · Integra Lifesciences Production Corporation · Feb 1, 2023
K112156 — CODMAN CERTAS PROGRAMMABLE VALVE SYSTEM, CODMAN CERTAS THERAPY MANAGEMENT SYSTEM · Medos International SARL · Oct 17, 2011
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles layered on top of each other. The profiles are facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 19, 2015
Medos International Sarl % Michelle Godin Project Manager, Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767
Re: K143111
> Trade/Device Name: Codman Certas Plus Programmable Valve: Codman Certas Tool Kit Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt And Components Regulatory Class: Class II Product Code: JXG Dated: October 28, 2014 Received: October 29, 2014
Dear Ms. Godin,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143111
Device Name
Codman Certas Plus Programmable Valve; Codman Certas Tool Kit
Indications for Use (Describe)
The Codman Certas Plus Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
The Codman Certas Tool Kit allows the noninvasive reading or adjustment of the valve setting.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------|
| <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 6. 510(k) Summary
#### I. Submitter Codman & Shurtleff, Inc. 325 Paramount Drive Raynham, MA 02767
On behalf of: Medos International SARL Chemin-Blanc 38 CH 2400 LeLocle, Switzerland
Phone: 508-828-3312 508-977-6979 Fax:
Contact Person: Michelle Godin Date of Submission: October 28, 2014
#### II. Device
| Name of Device | Codman Certas Plus Programmable Valve<br>Codman Certas Tool Kit |
|---------------------|------------------------------------------------------------------------|
| Common Name | Hydrocephalus Shunt |
| Classification Name | Central Nervous System Fluid Shunt and Components<br>(21 CFR 882.5550) |
| Regulatory Class | II |
| Product Code | JXG |
III. Predicate Codman Certas Programmable Valve System and Codman Certas Therapy Device Management System (K112156)
IV. Device The Codman Certas Plus Programmable Valve is a sterile, single use, Description implantable device designed for shunting cerebrospinal fluid (CSF) for the treatment of hydrocephalus.
> The Codman Certas Plus Programmable Valves are pressure-regulating valves utilizing the ruby ball-in-cone principle with a pressure inducing spring design. Intraventricular pressure is maintained by the ball and cone valve seat design. As the differential pressure across the shunt increases, the ball further displaces from the cone, through which CSF flows, thereby increasing flow and re-establishing the selected pressure. The ball is manufactured of synthetic ruby, as is the matching cone. Together these components provide a precise fit for regulating the flow of CSF through the valve.
{4}------------------------------------------------
| IV. Device<br>Description<br>(Cont.) | The valve is available with 8 different performance settings for constant<br>intraventricular pressure and drainage of CSF. Seven (7) of the settings<br>provide for a change in operating pressure, with a range of 25 to 215 mmH2O.<br>The eighth setting provides a minimum opening pressure of '400' mmH20,<br>thus allowing a physician to turn the valve “virtually off” without the need to<br>surgically remove the valve to limit flow. The pressure of the valve is set<br>preoperatively and can be noninvasively changed post-implantation by using<br>the Codman Certas Tool Kit, which employs magnetic force to select one of<br>the 8 settings. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| V. Indications<br>for Use | The Indications for Use statements are identical to the predicate device. |
| The Codman Certas Plus Programmable Valve is an implantable device that<br>provides constant intraventricular pressure and drainage of CSF for the<br>management of hydrocephalus. | |
| The Codman Certas Tool Kit allows the noninvasive reading or adjustment of<br>the valve setting. | |
| VI.<br>Comparison to<br>Predicate | Compared to the predicate device, the Codman Certas Plus Programmable<br>Valve and Codman Certas Tool Kit include the modifications listed in Table |
| Predicate |
|-----------|
| Device |
l
| Table 1. Modifications Proposed for the Codman Certas Plus Programmable Valve and<br>Codman Certas Tool Kit | | |
|-------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component | Modification | Rationale |
| Codman Certas<br>Plus<br>Programmable<br>Valve | Changes to components contained in the<br>Adjustable Mechanism of the programmable<br>valve<br>Minor dimensional changes to the profile<br>of the Cam of the Rotating Construct and<br>dimensional changes to the Cam/Ball Arm<br>Assembly Addition of Ruby Bearing (same ruby<br>material currently used in the predicate<br>device, K112156) to through-hole of the<br>Rotating Construct | Minor modifications improve the overall<br>performance of the programmable valve<br>without changing the fundamental<br>scientific technology or intended use of<br>the predicate device. |
{5}------------------------------------------------
# 510(k) Summary (Cont)
| Table 1. Modifications Proposed for the Codman Certas Plus Programmable Valve and<br>Codman Certas Tool Kit | | |
|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Component | Modification | Rationale |
| Codman Certas<br>Tool Kit | Locator Tool: Includes 2 Locator Tools: The<br>Low Profile Locator Tool which allows the<br>clinician to get as close to the valve as<br>possible in cases of edema and an Adjustable<br>Height Locator Tool which allows the<br>clinician to rotate the center ring to change the<br>height of the locator when valve protrusion<br>exists in a patient. In addition, the petals will<br>be slightly wider for better grip on both tools.<br>Minor enhancements to the tools include<br>machined black centering lines versus pad<br>printed black lines to improve precision for<br>location and the addition of a v-groove to<br>contain the 360° stop alignment pin for<br>increased pin retention strength. | Modification of ergonomic and user<br>interface; no change to fundamental<br>scientific technology or intended use of<br>predicate device. |
| | Indicator Tool: Removed the black button<br>and improved the pivot-bearing design to<br>reduce toggle time to indicating a valve<br>setting. Modified background of setting<br>number and added a contoured handle. | |
| | Carrying Case: Increased in size to<br>accommodate 4 tools rather than 3 tools and<br>added a bump feature for easier removal of the<br>Adjustment Tool. | |
| | Accessory Tool: X-ray Overlay Tool added to<br>verify the valve setting on x-ray. | |
| Labeling | Updated IFU and product labeling for name<br>change, clarification, and/or additional steps. | Labeling update to ensure proper use of<br>device, no change to indications for use or<br>intended use. |
The following performance data were provided in support of the substantial VII. Performance equivalence determination. Data
#### Bench Testing
Bench testing on the proposed device, Codman Certas Plus Programmable Valve and Certas Tool Kit, included the following:
- Structural testing
- . Pressure flow testing in accordance with ISO 7197:2009 Neurosurgical Implants – Sterile, Single-Use Hydrocephalus Shunts and Components and ASTM F647:1994 (R2006) Standard Practice
{6}------------------------------------------------
for Evaluating and Specifying Implantable Shunt Assemblies for Neurosurgical Application
- MRI testing ●
- Indication and Adjustment testing ●
- Shelf Life testing including structural testing, pressure flow testing, ● and MRI testing.
Validation testing included:
- Pre-implantation Use in-package indication and programming ●
- Post-implantation Use indication and adjustments post-implantation ● using simulated skin thicknesses
- Valve patency
- . Tool Kit validation
- Identifying the valve setting on x-ray with and without using the X-● Ray Overlay Tool
- Manometer IFU language ●
Test results demonstrated that the acceptance criteria were met, therefore, the Codman Certas Plus Programmable Valve and the Codman Certas Tool Kit conform to expected device performance and intended use. Results of verification and validation testing have demonstrated that the proposed Codman Certas Plus Programmable Valve and the Codman Certas Tool Kit are substantially equivalent to the predicate Codman Certas Programmable Valve and Codman Therapy Management System, and that the modifications do not impact the safety or effectiveness of the proposed device.
#### Magnetic Resonance (MR) Testing
The safety test requirements of the ASTM MR standards for the proposed Certas Plus Programmable Valve have been met through testing (ASTM F2052-14 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment; ASTM F2182-11a Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging; ASTM F2119-07 (R2013) Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants; ASTM F2213-06 (R2011) Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment).
The Codman Certas Plus Programmable Valve is MR-Conditional at 3.0 Tesla per the ASTM standards.
#### Biocompatibility Testing
The biocompatibility evaluations for the Codman Certas Plus Programmable Valve and the Codman Certas Tool Kit were conducted in accordance with the FDA Blue Book Memorandum #G95-1 Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and
{7}------------------------------------------------
Testing May 1, 1995, and International Standard ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process as recognized by FDA.
### Codman Certas Plus Programmable Valve
The Codman Certas Plus Programmable Valve is considered a permanent contact implant device. The only new component in the proposed device is the inclusion of a ruby bearing. Since the same ruby material is already included in the predicate Codman Certas Programmable Valve, additional biocompatibility testing was not required. However, as part of a cleaning validation for this component, cytotoxicity testing was completed with passing results.
# Codman Certas Tool Kit
The Codman Certas Tool Kit is considered a surface device with limited contact (≤24 hours) with breached or compromised surfaces. Biocompatibility testing was completed with passing results. The following tests were conducted:
- Cytotoxicity ●
- Sensitization ●
- Irritation
- . USP Limits Testing for USP <661> for Plastics, Physiochemical Testing
# Animal Studies
No animal studies were required as appropriate verification and validation of the design modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
# Clinical Studies
No clinical studies were required as appropriate verification and validation of the design modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
Based upon the intended use, design, materials, function, comparison to VIII. Conclusion currently marketed devices, and testing performed by Codman & Shurtleff, Inc., it is concluded that the Codman Certas Plus Programmable Valve and Codman Certas Tool Kit was found to have a safety and effectiveness profile that is similar to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.