Phasor EVAC™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system is intended for drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The EVAC system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma.
Device Story
EVAC system provides access to subdural space for evacuation of cranial subacute/chronic hematomas or hygromas. System includes battery-powered drill, cranial bolt, and drainage reservoir connected via tubing. Used in clinical settings by neurosurgeons to drain fluid accumulations to external suction reservoir; promotes gradual brain re-expansion via low homogeneous negative pressure. Side port allows for occlusion relief or additional aspiration without tubing puncture. Device is sterile, single-use, and disposable. Benefits include minimally invasive access and controlled drainage of subdural space.
Clinical Evidence
Bench testing only. No clinical data performed. Testing included biocompatibility, packaging, sterility, safety, and performance validation. Closed system integrity and bone purchase were verified.
Technological Characteristics
Sterile, single-use, disposable system. Components: battery-powered drill, cranial bolt, 100cc reservoir, silicon tubing. Features side port for aspiration. Sterilization: EO (kit/reservoir) and gamma (drill/tubing/bolt).
Indications for Use
Indicated for patients requiring access to and evacuation of cranial subacute or chronic hematomas or hygromas, and for drainage of air and fluids from the subdural space following craniotomy for hematoma removal.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Subdural Evacuation Port System (SEPS) Cranial Access Kit (K042359)
K241458 — Borvo EVAC System (Ergo); Borvo EVAC System (Classic) · Borvo Medical, Inc. · Aug 20, 2024
K042359 — SUBDURAL EVACUATING PORT SYSTEM CRANIAL ACCESS KIT (SEPS-CA), (SEPS-CAL), (SEPS-CAND),MODELS 11-0401 CA, 11-0401-CAL · Medical Designs, LLC · Nov 17, 2004
K974726 — SUBDURAL DRAINAGE CATHETER KIT, SUBDURAL CATHETER · Elekta Implants, SA · Mar 18, 1998
K162437 — Codman EDS3 CSF External Drainage System · Codman & Shurtleff, Inc. · Oct 28, 2016
Submission Summary (Full Text)
{0}
K243205 - Ray King
Page 1
FDA U.S. FOOD & DRUG ADMINISTRATION
June 3, 2025
Phasor Health, LLC
Ray King
Chief Executive Officer
8944 Kirby Drive
Houston, Texas 77054
Re: K243205
Trade/Device Name: EVAC
Regulation Number: 21 CFR 882.5550
Regulation Name: Central nervous system fluid shunt and components
Regulatory Class: Class II
Product Code: JXG
Dated: December 19, 2024
Received: December 19, 2024
Dear Ray King:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}
K243205 - Ray King
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K243205 - Ray King
Page 3
Sincerely,
Adam D. Pierce -S
Digitally signed by
Adam D. Pierce -S
Date: 2025.06.03
14:15:55 -04'00'
Adam D. Pierce, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243205
Device Name
EVAC
Indications for Use (Describe)
Phasor EVAC™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system is intended for drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The EVAC system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K243205
# 510(k) Summary
A. Device Description
| Category | Comments |
| --- | --- |
| Sponsor | Phasor Health, LLC, Ray King, 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234 |
| Correspondent Contact Information | Ray King 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234 |
| Device Common Name | Central nervous system fluid shunt and components |
| Device Regulation & Name | 21 CFR 882.5550 |
| Classification & Product Code | Class II, JXG |
| Device Proprietary Name | EVAC |
Predicate Device(s) Information:
| Predicate Device(s) | (1) Subdural Evacuation Port System (SEPS) Cranial Access Kit; (2) Phasor Drill |
| --- | --- |
| Predicate Device Manufacturer(s) | (1) Medtronic, Inc., (2) Phasor Health, LLC; respectively |
| Predicate Device Common Name | (1) Central nervous system fluid shunt and components [subdural fluid drainage system]; (2) Powered simple cranial drills, burrs, trephines, and their accessories; respectively |
| Predicate Device Notification # | (1) K042359; (2) K161704 |
| Predicate Device Classification & Name | (1) 21 CFR 882.5550; (2) 21 CFR 882.4310; respectively |
| Predicate Device Classification & Product Code: | Class II for each of (1) JXG & (2) HBE, respectively |
B. Date Summary Prepared: May 21, 2025
C. Description of Device
Phasor EVAC™ subdural evacuation system is indicated when access to the subdural space and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system consists of surgical instruments and accessories used for draining subdural
Page 1 of 5
{5}
K243205
fluid accumulations such as hygromas and liquid-state subdural hematomas to an external suction reservoir without touching the brain. Utilizing a minimally invasive technique, the EVAC system components are designed to promote gradual brain re-expansion by creating a low homogeneous negative pressure throughout the subdural space as fluid is drained to an external suction reservoir.
# D. Indications for Use
Phasor EVAC™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system is intended for drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The EVAC system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma."
# E. Comparison of the Technological Characteristics
| | Application Device | Predicate Device | Predicate Device | Impact on Substantial Equivalence |
| --- | --- | --- | --- | --- |
| Company | Phasor Health, LLC | Medtronic, Inc. | Biotex, Inc. (now Phasor Health, LLC) | N/A |
| Regulation Number | 21 CFR 882.5550 | 21 CFR 882.5550 | 21 CFR 882.4310 | includes both predicates |
| Product Code | JXG (Class II) | JXG (Class II) | HBE (Class II) | together, comparable to both devices |
| Intended Use & Indications for Use | EVAC™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The EVAC system is intended for drainage of subdural fluid accumulations such as | SEPS™ is indicated when access to and evacuation of a cranial subacute or chronic hematoma or hygroma is necessary. The Subdural Evacuating Port System Cranial Access Kit is intended for | The Phasor™ Drill is a sterile, single-use, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone. | Taken together, comparable to both devices |
{6}
K243205
| | hygromas and chronic or subacute hematomas to an external suction reservoir. The EVAC system is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma." | drainage of subdural fluid accumulations such as hygromas and chronic or subacute hematomas to an external suction reservoir. The Subdural Evacuating Port System Cranial Access Kit is also intended for draining air and fluids from the subdural space immediately following craniotomy procedures performed to remove a chronic or subacute subdural hematoma." | | |
| --- | --- | --- | --- | --- |
| Technology | | | | |
| Drill power source | Batteries | Manual | Batteries | comparable when considered together |
| Bolt | screws into skull at threaded end | screws into skull at threaded end | N/A | comparable bolts |
| Drain reservoir connected to Bolt via Tubing | 100cc connected with silicon tubing | 100cc connected with silicon tubing | N/A | comparable |
Page 3 of 5
{7}
K243205
| Occlusion relief/additional aspiration | Side port | puncture tubing via needle aspiration | N/A | no impact |
| --- | --- | --- | --- | --- |
## F. Summary of Supporting Data
EVAC™ has the same intended use, patient population, and classification as the predicate. At a high level, the subject and predicate devices are based upon the following technological elements, when the two predicates are taken together:
* Sterile packaged for single use, and disposable;
* Phasor Drill operates on battery power, as per the predicate (same mechanism as itself);
* Bolt screws into the skull in the same manner, with comparable inner diameter;
* Reservoir connects via similar tubing as the first predicate
The following technical characteristics of the subject device differ from the predicate device:
* Taken by itself, SEPS utilizes a disposable single-use sterile manually operated drill, whereas EVAC™ utilizes a disposable single-use sterile battery-powered drill (with same characteristics as the second predicate, the Phasor Drill);
* for relief of occlusion or need for additional aspiration, SEPS utilizes a needle placed through tubing (which punctures the tubing), whereas EVAC™ accomplishes similar feature and goal via a side port precluding need for puncture of tubing;
* SEPS™ is sterilized using EO, whereas EVAC™ is sterilized with EO (kit and reservoir) and gamma (Phasor Drill, tubing, and bolt) with individual packages placed together into the same shelf carton after respective sterilization.
The proposed device does not differ from the predicates taken together in any manner which would negatively impact the safety or effectiveness of the device.
## G. Discussion of Performance Testing
Various tests including for biocompatibility, packaging, sterility, safety, and performance were conducted and passed successfully. The Phasor™ Drill has been previously cleared, and any other components were justified as comparable and/or supported by documentation with adequate performance. Specifically, testing for a closed system with firm purchase into bone was established and verified by testing. No clinical testing was needed or performed otherwise.
## H. Conclusion
The performance and design validation testing conducted on the EVAC™ device on the bench demonstrated that it performs equivalent to the stated predicates that are currently
Page 4 of 5
{8}
K243205
marketed individually for the same intended use. The proposed device, comprised of the elements noted above, demonstrates that it should perform as safely and effectively as the predicate device(s).
Page 5 of 5
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.