K251598 · Brainspace, Inc. · JXG · Dec 10, 2025 · Neurology
Device Facts
Record ID
K251598
Device Name
Intellidrop
Applicant
Brainspace, Inc.
Product Code
JXG · Neurology
Decision Date
Dec 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intellidrop system is indicated for use to provide external drainage of cerebrospinal fluid (CSF) and/or monitoring of CSF drainage and intracranial pressure (ICP) for ventricular or lumbar use.
Device Story
Intellidrop system provides continuous intracranial/spinal pressure monitoring and automated external CSF drainage. Input: pressure data from wearable sensor array; drainage volume data. Operation: closed-loop, gravity-based drainage; actuator-controlled tube pinching mechanism in console adjusts flow to user-selected pressure or volume targets. Output: real-time ICP waveform, historical trend data, audio-visual alarms. Used in clinical settings; operated by healthcare professionals. Features: continuous drift detection technology; AC/battery power; real-time streaming to patient monitor. Healthcare providers use output to manage CSF drainage and monitor ICP, facilitating clinical decision-making regarding patient fluid management and pressure regulation.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1), sterility (ISO 11135), and sterile barrier performance (ISO 11607). Functional performance, accuracy, and stability were validated for up to 14 days per AAMI NS28. Human factors usability validation was conducted per IEC 62366-1, confirming successful execution of critical tasks by intended users. MR compatibility was assessed per ISO TS 10974.
Technological Characteristics
System includes a wearable sensor array, console with actuator-controlled tube pinching mechanism, and sterile disposable tubing. Powered by AC or rechargeable battery. Connectivity includes real-time data streaming via cable. Sterilization of patient interface via ethylene oxide. Software-controlled console manages ICP targets, alarms, and data storage. MR Conditional per ISO TS 10974. Compliant with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients requiring external drainage of cerebrospinal fluid (CSF) and/or monitoring of intracranial pressure (ICP) via ventricular or lumbar access.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
K161605 — Smart External Drain (SED) System · Aqueduct Critical Care, Inc. · Oct 27, 2016
K200807 — IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit · Irras USA, Inc. · Apr 29, 2020
K192289 — IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit · Irras USA, Ltd. · Nov 15, 2019
K121248 — LIQUOGUARD CSF LIQUOGUARD CSF TUBING SET (1500MM) LIQUOGUARD CSF TUBING SET (2000MM) SAFETY CUSHION FOR TUBING SET ATTAC · Moller Medical GmbH · Dec 13, 2013
K171586 — Smart External Drain (SED) System · Aqueduct Critical Care, Inc. · Jun 30, 2017
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 10, 2025
BrainSpace, Inc.
Caitlin Morse
CEO
22121 17th Avenue SE, Suite 112
Bothell, Washington 98021
Re: K251598
Trade/Device Name: Intellidrop
Regulation Number: 21 CFR 882.5550
Regulation Name: Central Nervous System Fluid Shunt and Components
Regulatory Class: Class II
Product Code: JXG, GWM
Dated: November 12, 2025
Received: November 12, 2025
Dear Caitlin Morse:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251598 - Caitlin Morse
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the QS regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K251598 - Caitlin Morse
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
XIAOLIN
ZHENG -S
For Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251598 | |
| Device Name | |
| Intellidrop | |
| Indications for Use (Describe) | |
| The Intellidrop system is indicated for use to provide external drainage of cerebrospinal fluid (CSF) and/or monitoring of CSF drainage and intracranial pressure (ICP) for ventricular or lumbar use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) SUMMARY (K251598)
General Information:
Date of Summary: December 9, 2025
Name and Address of Manufacturer: BrainSpace Inc.
22121 17th Ave SE, Suite 112
Bothell, WA 98021, USA
425.725.5008
Contact Person: Caitlin Morse, CEO
Trade Name: Intellidrop
Common Name: External CSF Drainage System
Regulation Number, Primary: 21 CFR 882.5550
Product Code, Primary: JXG
Regulation Description: Central Nervous System Fluid Shunt & Components
Regulatory Class, submission: Class II, Tradition 510(k)
Classification Panel: Neurology
Additional Associated Product Code: GWM
Device Description:
The Intellidrop system continuously monitors intracranial or spinal pressure and manages external drainage of cerebrospinal fluid (CSF) and other body fluids to a user-selected pressure target or volume target.
Key features of the Intellidrop include:
- Wearable sensor array facilitating position-agnostic pressure monitoring
- Closed-loop, gravity-based drainage to a pressure or volume target
- Real-time display of intracranial pressure (ICP) waveform
- Historical trend data of daily, hourly, and minute-by-minute ICP and drainage volume
- User selectable audio-visual alarm thresholds for both pressure and volume
- Patented Continuous Drift Detection Technology
- Continuous operation on AC power and up to 4 hours of continuous operation on rechargeable battery power
- Real-time data streaming of ICP to a patient monitor via cable
- Proximal sampling port and optional distal sampling port
Indications for Use:
The Intellidrop system is indicated for use to provide external drainage of cerebrospinal fluid (CSF) and/or monitoring of CSF drainage and intracranial pressure (ICP) for ventricular or lumbar use.
{5}
# Predicate Device:
- Smart External Drain (SED) System by Aqueduct Critical Care Inc. (K172759)
The BrainSpace Intellidrop is substantially similar to the predicate. The intended use and patient populations are identical and the minor differences in technical characteristics do not generate new questions of safety or effectiveness. Refer to Table 1 for details.
Table 1: Comparative Summary of the BrainSpace Intellidrop to Predicate Device
| Description | Predicate: SED System by Aqueduct Critical Care, Inc. (K172759) | Subject Device: BrainSpace Intellidrop |
| --- | --- | --- |
| Device Trade Name | Smart External Drain (SED) System | Intellidrop |
| Intended Use | External ventricular or lumbar drainage of cerebrospinal fluid (CSF) and monitoring of CSF drainage and/or pressure. | Same as predicate |
| Attaches to separate commercially available catheter | Yes | Same as predicate |
| Sterile disposable tubing set | Yes | Same as predicate |
| CSF drainage bag | Yes | Same as predicate |
| Gravity drainage of CSF | Yes | Same as predicate |
| Method to control gravity drainage of CSF | Yes – automated adjustment based on user settings via an actuator-controlled tube pinching mechanism in the console to either compress or release the compliant drainage tubing contained within the cartridge. | Same as predicate |
| Pressure transducer for ICP measurement | Yes | Same as predicate |
| Pressure Anatomical Alignment Method | Disposable includes wearable reference at level of EAM (external auditory meatus). | Disposable includes wearable transducer at level of EAM (external auditory meatus). |
| Software-based powered console for user interface, ICP target and alarm settings, data storage and display, and alarms | Yes | Same as predicate |
| Pressure Target Range | -5 to 40 cm H2O (ventricular only) | -20 to 50 cm H2O (ventricular only) |
| Hourly Drainage Volume Range | 0 to 45 mL/hr | 0 to 60 mL/hr |
| Displayed ICP | Yes | Same as predicate |
| Battery backup | Yes | Same as predicate |
# Testing Summary:
The testing presented confirm substantial equivalence to the predicate in terms of biocompatibility, electrical safety, resistance to electromagnetic interference, packaging, sterilization, shelf-life and performance to specifications and key consensus standards. Therefore, the sum of the presented test data is sufficient to demonstrate intended device performance. Refer to Table 2 for details.
{6}
Table 2: Summary Table for BrainSpace Intellidrop Testing
| Test | Test Summary | Result / Conclusion |
| --- | --- | --- |
| Electrical Safety | Basic safety and essential performance of the system was completed in accordance with IEC 60601-1: Edition 3.2, 2020, and relevant sub-parts. | Pass – Fully compliant |
| Storage and Transit Simulation | Accelerated shelf-life conditioning and transit simulation were completed as applicable and the system confirmed fully functional. | Pass – All requirements met for the duration of labeled shelf-life |
| Sterility | The patient interface disposable of the Intellidrop system was sterilized via ethylene oxide (EO) sterilization and assessed for cycle effectiveness in accordance with ISO 11135: 2014/A1: 2018. | Pass - Fully compliant Disposable confirmed sterile |
| Sterile Barrier Performance | The sterile barrier of the patient interface disposable was assessed and confirmed appropriate to maintain sterility in accordance with ISO11607: 2019. | Pass – All requirements verified Disposable packaging confirmed to maintain sterility |
| Biocompatibility | Biocompatibility testing was performed based on risk assessment and in accordance with ISO 10993-1:2018 and associated sub-parts. | Pass – System is biocompatible and non-pyrogenic |
| Cleanability | The reusable console was assessed for cleanability. | Pass – All requirements verified |
| Magnetic Resonance (MR) Compatibility | MRI compatibility was assessed and the system confirmed as MR Conditional in accordance with ISO TS 10974: Ed. 2: 2018. | MR Conditional |
| Basic Safety | The Intellidrop system was tested for basic safety and maintenance of essential performance in accordance with IEC 60601-1: Edition 3.2: 2020. | Pass – Fully compliant |
| Electromagnetic Interference and Immunity | The essential performance of the Intellidrop system was assessed against the risk of electromagnetic interference in accordance with IEC 60601-1-2: 2014 and IEC TR 60601-4-2:2016. | Pass – Fully compliant |
{7}
| Test | Test Summary | Result / Conclusion |
| --- | --- | --- |
| Software Verification and Validation | Software code verification, unit testing and system-based verification and validation testing for software controlled or impacted functions was completed, including user-interface review. | Pass – All requirements verified or validated |
| Functional Performance | Verification tests were performed on the Intellidrop system and sub-systems to confirm functional performance. | Pass – All requirements verified |
| Accuracy, Stability and Performance | The system was assessed for accuracy, stability, and durability of performance for up to 14 days (the longest expected per-patient use of the Intellidrop system) in accordance with AAMI NS28: 1988/R: 2015. | Pass – Fully compliant |
| Human Factors Usability Validation | Usability of the system and the effectiveness of provided information was assessed through summative human factors validation on select groups of users in accordance with IEC 62366-1:2015+AMDI:2020. | Pass – All requirements validated; all critical tasks successfully executed.
System confirmed usable by the intended users. |
Conclusion (Statement of Equivalence):
The data and information presented within this submission support a determination of substantial equivalence to the predicate.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.