NEURON THERAPEUTICS, INC., 14-GAUGE LUMBAR PUNCTURE SET, MODEL PN1000735
Applicant
Neuron Therapeutics, Inc.
Product Code
JXG · Neurology
Decision Date
Jul 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Neuron Therapeutics Inc., 14-Gauge Lumbar Puncture Set is indicated for temporary access to the lumbar subarachnoid region to drain cerebrospinal fluid (CSF) and other fluids of similar physical characteristics.
Device Story
The 14-Gauge Lumbar Puncture Set provides temporary access to the lumbar subarachnoid space for CSF drainage. The set includes a 14-gauge polyurethane catheter (30 cm length, 1.25 mm ID, 1.74 mm OD) featuring 35 fenestrations to reduce flow resistance and occlusion risk, a 14-gauge Extra-Thin Wall Touhy needle, a 45 cm stainless steel J-tip guidewire, a 6-French Touhy-Borst connector, and accessories (dressing, anchor, drape). The physician inserts the catheter into the subarachnoid space using the needle and guidewire. Catheter markings at 1 cm intervals assist in distal tip placement. The device is intended for single-use. The output is the facilitated drainage of CSF, which is collected or monitored via external equipment connected to the Touhy-Borst connector. This procedure aids in managing CSF pressure or fluid removal, benefiting patients requiring temporary subarachnoid access.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Polyurethane catheter; 14-gauge Extra-Thin Wall Touhy needle; stainless steel J-tip guidewire; 6-French Touhy-Borst connector. Features 35 fenestrations and 13 location markings at 1 cm intervals. Sterile, single-use, tamper-evident packaging.
Indications for Use
Indicated for temporary access to the lumbar subarachnoid region for drainage of cerebrospinal fluid (CSF) or fluids with similar physical characteristics in patients requiring such access.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
CODMAN Lumbar External Drainage Catheter with Touhy Needle (K964923)
Related Devices
K964923 — CODMAN LUMBAR DRAINAGE CATHETER KIT · Johnson & Johnson Professionals, Inc. · Jun 25, 1997
K081942 — NEUROCATH AG · Vygon Neuro · Aug 12, 2008
K962097 — VENTRICULAR CATHETER SET · Cook, Inc. · Aug 29, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 1 2002
Neuron Therapeutics, Inc. 14-Gauge Lumbar Puncture Set 510(k)
12 April 2002
021196
# SECTION 9: 510 (k) SUMMARY (FFD&C ACT 513(b)(3)(3)(3)(3)(3)(3)(3)(4))
Image /page/0/Picture/5 description: The image shows the logo for Neuron Therapeutics, Inc. The word "neuron" is written in a bold, sans-serif font. Below the word "neuron" is the text "Therapeutics, Inc." in a smaller font. To the left of the word "neuron" is a graphic that appears to be a stylized representation of a neuron.
## Summary of Safety and Effectiveness Data NEURON THERAPEUTICS INC., 14-GAUGE LUMBAR PUNCTURE SET 12 April 2002
Neuron Therapeutics, Inc. 81 Great Valley Parkway Malvern, PA 19355
## Contact Person
Judy P. Ways, Ph.D. Vice President, Regulatory Affairs and Quality Assurance Phone: (610) 578-9494 x112 Fax: (610) 578-9205
## Name of Device_
| Proprietary Name: | Neuron Therapeutics, Inc., 14-Gauge Lumbar Puncture Set |
|----------------------|---------------------------------------------------------|
| Trade Name: | 14-Gauge Lumbar Puncture Set |
| Common Name: | Sterile, 14-Gauge Lumbar Puncture Set |
| Classification Name: | Central Nervous System Fluid Shunt and Components |
| Regulatory Class: | II (21CFR § 882.5550) |
| Product Code: | 84 JXG |
Establishment Registration Number: 9001269
{1}------------------------------------------------
## Device Classification
This device has been placed in a Class II for Central Nervous System shunt and components per 21 CFR § 882.5550.
## Statement of Substantial Equivalence
The 14-Gauge Lumbar Catheter is the component of the14-Gauge Lumbar Puncture Set subject to 510(k) approval. Neuron Therapeutic Inc.'s (NT) 14-Gauge Lumbar Catheter is substantially equivalent to the CODMAN Lumbar External Drainage Catheter with Touhy Needle that is currently marketed in the United States under 510(k) No. K964923. NTI's 14-Gauge Lumbar Catheter and 14-Gauge Lumbar Puncture Set have the same intended use, design characteristics, and similar methods of manufacture as the CODMAN Lumbar External Drainage Catheter with Touhy Needle.
#### Indications for Use
The Neuron Therapeutics Inc., 14-Gauge Lumbar Puncture Set has an substantially equivalent indication for use as the Codman Lumbar Drainage Catheter Kit. Both are indicated for temporary access to the lumbar subarachnoid region to drain cerebrospinal fluid (CSF) and other fluids of similar physical characteristics.
#### Physical Description
The Neuron Therapeutics 14-Gauge Lumbar Puncture Set, part number PN1000735, consists of a 14-Gauge Lumbar Catheter Kit, a Transparent Tegaderm™ Dressing, a Suture-less Catheter Anchor, and a Sterile Drape. These individual, sterile packaged components are provided in sealed tamper-evident packaging for single-use.
The 14-Gauge Lumbar Catheter Kit, part number PN1000737, contains a single 30 cm coiled lumbar catheter with an inner diameter of 1.25 mm, an outer diameter of 1.74 mm and 35 fenestrations. The shorter length, larger diameter and higher number of fenestrations of the catheter allow for decreased flow resistance and higher liquid flow rates. The higher number of fenestrations decreases the possibility of occlusion of the catheter during use. The catheter is constructed of polyurethane to provide resistance to kinking or collapsing. The catheter has13 location markings positioned at 1 cm intervals to assist the physician in identifying the location of the distal end of the catheter in the subarachnoid space. Special
{2}------------------------------------------------
markings are provided to indicate the location of the catheter tip, the proximal fenestration and the coiled shaft as they exit from the needle tip.
The 14-Gauge Lumbar Catheter Kit also contains a 15.0 cm, 14-gauge Extra-Thin Wall Touhy Needle and a 45 cm stainless steel flexible J-tip guidewire for inserting and guiding the catheter into the lumbar subarachnoid space. A 6-French Touhy-Borst Connector is also included for connection to equipment for patient monitoring or fluid collection.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, repeated three times to create a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD. 20850
์ 1 2002 JUL
Neuron Therapeutics, Inc. Judy P. Ways, Ph.D. Vice President, Regulatory Affairs and Quality Assurance 81 Great Valley Parkway Malvern, Pennsylvania 19355
Re: K021196
Trade Name: Neuron Therapeutics, Inc., 14-Gauge Lumbar Puncture Set Regulation Number: 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: II Product Code: JXG Dated: April 12, 2002 Received: April 16, 2002
Dear Dr. Ways:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Dr. Judy Ways
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Kacelia M. Witten, Ph.D., M.D.
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 7: STATEMENT OF INDICATIONS FOR USE
Device Name: Neuron Therapeutics, Inc., 14-Gauge Lumbar Puncture Set
Indications for Use: The Neuron Therapeutics Inc., 14-Gauge Lumbar Puncture Set is indicated for temporary access to the lumbar subarachnoid region to drain cerebrospinal fluid (CSF) and other fluids of similar physical characteristics.
K021196 510(k) Number (if known):
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) CONCURRENCE OF CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR § 801.109) OR
Over-the-Counter Use
Stupt Purdu
(Optional Format 1-2-96)
(Division Sign-Off) Division of General. Restorative and Neurological Devices
Confidential
510(k) Number_KOZ | 196
13
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.