Bundled Neurological Shunts and Accessories Product Families Labeling Modification to Support MR Conditional Labeling
Applicant
Integra LifeSciences Corporation
Product Code
JXG · Neurology
Decision Date
Mar 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Standard-LPV II and Mini-LPV II Valves, utilized in the treatment of hydrocephalic patients in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the brain into either the right atrium of the heart or the peritoneum. The Mini-LPV II Valve can be used in (but is not restricted to) situations where skin erosion may be a problem, as with older patients. The Novus and Novus Mini Valves, utilized in the treatment of hydrocephalic patients, are components in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the peritoneum. The Novus Mini Valve can be used in (but is not restricted to) situations where skin erosion may be a problem, as with older patients. Valves with a Physiological Flow Device are intended to reduce the hazard of negative intraventricular pressure (with respect to atmospheric pressure) when the patient is sitting, standing or semi-recumbent. The Multi-Purpose Valve, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the perioneum. Valves with an Anti-Siphon Device are intended to reduce the hazard of negative intraventricular pressure when the patient is sitting, semi-recumbent or standing. The Mishler Dual Chamber Flushing Valve, Flat Bottom Design, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum. Valves with an Anti-Siphon Device are intended to reduce the hazard of negative intraventricular pressure when the patient is sitting, semi-recumbent or standing. The Pudenz Flushing Valve, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the brain into either the right atrium of the heart or the peritoneum. Valves with an Anti-Siphon Device are intended to reduce the intraventricular pressure when the patient is sitting, semi-recumbent or standing. The Ultra VS In-Line Valves, utilized in the treatment of hydrocephalic patients, are components in systems designed to shunt cerebrospinal fluid (CSF) from the lateral ventricles into either the peritoneal cavity or the right atrium of the heart. The in-line and burr-hole systems are designed to shunt cerebrospinal fluid from the lateral ventricles into the peritoneal cavity. A ventriculoperitoneal shunting system may be indicated to avoid the cardiovascular complications of an atrial shunt or for a hydrocephalic patient in whom an atrial shunt is contraindicated. The Small and Neonate Models can be used in (but are not restricted to) situations where skin erosion may be a problem, as with premature infants, pediatric patients and older patients. The Pudenz Cardiac Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart. The Infant Cardiac Catheter is utilized when the common facial and/or internal jugular veins are too small to accommodate the larger cardiac catheter. The Pudenz Ventricular Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the peritoneum. The Pudenz Peritoneal Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventriculoperitoneum. A ventriculoperitoneal shunting system may be indicated to avoid the cardiovascular complications of an atrial shunt or for a hydrocephalic patient in whom an atrial shunt is contraindicated. The Peritoneal Reflux Control Catheter and Peritoneal Open-Ended Catheter with Slits, utilized in the treatment of hydrocephalic patients, are components for systems designed to shunt cerebrospinal fluid from the lateral ventricles into the peritoneum. A ventriculoperitoneal shunting system may be indicated to avoid the cardiovascular complications of an atrial shunt or for a hydrocephalic patient in whom an atrial shunt is contraindicated. The Portnoy Ventricular Catheter, utilized in the treatments, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum. The Neuroview Ventricular Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum. The Integra CSF Reservoir provides access to the lateral cerebral ventricles via hypodermic puncture for sampling and/or injection of fluids. It is useful in obtaining CSF samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The Convertible Integra CSF Reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart or the peritoneum. The CSF Control Valve, utilized in the treatments, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the peritoneum. The Essential Shunt Kit – Flat Bottom Design, utilized in the treatment of hydrocephalic patients, is designed to shunt cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum. The Essential Shunt Kit – Flat Bottom Design, can be used in (but is not restricted to) situations where skin erosion may be a problem, as with older patients. Integra connectors are utilized principally in the treatment of hydrocephalic patients, as components in systems designed to shunt cerebrospinal fluid from the lateral cerebral ventricles of the right atrium of the heart or the peritoneum. The On-Off Flushing Reservoir, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the peritoneum. Reservoirs with an Anti-Siphon Device are intended to reduce the hazard of negative intraventricular pressure when the patient is sitting, semi-recumbent or standing. The Braden Flushing Reservoir, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum. The Foltz Flushing Reservoir, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum. The Anti-Siphon Device, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the heart or the peritoneum. The device is designed to reduce the potential hazards of excessive lowering of intraventricular pressure (with respect to atmospheric pressure) when the patient is in a sitting, standing or erect position.
Device Story
Hydrocephalus treatment via CSF shunting; system creates flow pathway from cerebral ventricles to peritoneal space or right atrium. Components include catheters (silicone elastomer), valves (diaphragm or miter mechanisms), reservoirs, and connectors. Valves control drainage rate; some include anti-siphon mechanisms to prevent over-drainage in upright positions. Reservoirs allow ventricular access for sampling, pressure monitoring, or drainage. Clinician-implanted; used in neurosurgical settings. Device modification adds MRI safety labeling (MR Conditional).
Clinical Evidence
Bench testing only. No clinical data. Testing performed to support MR Conditional labeling: Magnetically Induced Displacement Force (ASTM F2052-06e1), Magnetically Induced Torque (ASTM F2213-06), RF Heating (ASTM F2182-09), and Image Artifact (ASTM F2119-07).
Technological Characteristics
Materials: Silicone elastomer (catheters), nylon/silicone (connectors). Valve types: Diaphragm (umbrella component) and miter (duckbill flaps). MR Conditional. No metallic components in valve mechanisms.
Indications for Use
Indicated for hydrocephalic patients requiring cerebrospinal fluid (CSF) shunting from lateral ventricles to the right atrium of the heart or the peritoneum. Includes neonates, pediatric, and older patients; specific models indicated for skin erosion prevention. Anti-siphon devices indicated to reduce negative intraventricular pressure in sitting/standing positions.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
K172022 — CODMAN Hydrocephalus Valves, Catheters & Accessories (Bundled) · Codman & Shurtleff, Inc. · Oct 3, 2017
K140722 — INTEGRA FLOW REGULATING VALVE SYSTEMS OSV II MODELS AND INTEGRA FLOW REGULATING VALVE LOW FLOW MODELS · Integra LifeSciences Corporation · Jan 6, 2015
K212641 — StrataMR II Valves and Shunts · Medtronic Neurosurgery · Sep 16, 2021
K020667 — CODMAN HAKIM SHUNT SYSTEMS · Codman & Shurtleff, Inc. · Mar 28, 2002
K974708 — GEMINI STANDARD AND MINI VALVES · Neurocare Group · Mar 10, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing to the right, arranged in a way that they resemble a bird-like shape. The profiles are black, and they are positioned within a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002
March 24, 2016
Integra LifeSciences Corporation Mr. Timothy Connors Senior Regulatory Affairs Specialist 311 Enterprise Drive Plainsboro, New Jersey 08536
Re: K153041
Trade/Device Name: Integra Neurological Shunts and Accessories Products (LPV II Valves and Kits, Novus Valves and Kits, Multi-Purpose Valve, Mishler Dual Chamber Valve with Integral Connectors, Pudenz Flushing Valve with Integral Connectors, Ultra VS In-Line Valve System, Pudenz Cardiac and Infant Catheter, Pudenz Ventricular Catheter, Pudenz Peritoneal Catheter, Peritoneal Reflux Control Catheter and Peritoneal Open-Ended Catheter With Slits, Portnoy Ventricular Catheter, Neuroview Endoscopic Ventricular Catheter, Integra CSF Reservoir with Integral Connectors, Essential Shunt Kit Burr Hole Design, Essential Shunt Kit Flat Bottom Design. Connectors for Neurosurgical Use, On-Off Flushing Reservoirs, Braden Flushing Reservoir, Foltz Flushing Reservoir, Anti-Siphon Device) Regulation Number: 21 CFR 882.5550
Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: December 14, 2015 Received: December 15, 2015
Dear Mr. Connors:
This letter corrects our substantially equivalent letter of March 14, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153041
#### Device Name
Integra Neurological Shunts and Accessories Products
# Indications for Use (Describe)
LPV II Valves and Kits
The Standard-LPV II and Mini-LPV II Valves, utilized in the treatment of hydrocephalic patients in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the brain into either the right atrium of the heart or the peritoneum. The Mini-LPV II Valve can be used in (but is not restricted to) situations where skin erosion may be a problem, as with older patients.
#### Novus Valves and Kits
The Novus and Novus Mini Valves, utilized in the treatment of hydrocephalic patients, are components in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the peritoneum. The Novus Mini Valve can be used in (but is not restricted to) situations where skin erosion may be a problem, as with older patients. Valves with a Physiological Flow Device are intended to reduce the hazard of negative intraventricular pressure (with respect to atmospheric pressure) when the patient is sitting, standing or semi-recumbent.
#### Multi-Purpose Valve
The Multi-Purpose Valve, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the perioneum. Valves with an Anti-Siphon Device are intended to reduce the hazard of negative intraventricular pressure when the patient is sitting, semi-recumbent or standing.
#### Mishler Dual Chamber Valve with Integral Connectors
The Mishler Dual Chamber Flushing Valve, Flat Bottom Design, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum. Valves with an Anti-Siphon Device are intended to reduce the hazard of negative intraventricular pressure when the patient is sitting, semi-recumbent or standing.
#### Pudenz Flushing Valve with Integral Connectors
The Pudenz Flushing Valve, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the brain into either the right atrium of the heart or the peritoneum. Valves with an Anti-Siphon Device are intended to reduce the intraventricular pressure when the patient is sitting, semi-recumbent or standing.
#### Ultra VS In-Line Valve System
The Ultra VS In-Line Valves, utilized in the treatment of hydrocephalic patients, are components in systems designed to shunt cerebrospinal fluid (CSF) from the lateral ventricles into either the peritoneal cavity or the right atrium of the heart. The in-line and burr-hole systems are designed to shunt cerebrospinal fluid from the lateral ventricles into the peritoneal cavity. A ventriculoperitoneal shunting system may be indicated to avoid the cardiovascular complications of an atrial shunt or for a hydrocephalic patient in whom an atrial shunt is contraindicated. The Small and Neonate Models can be used in (but are not restricted to) situations where skin erosion may be a problem, as with premature infants, pediatric patients and older patients.
#### Pudenz Cardiac and Infant Catheter
The Pudenz Cardiac Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart. The Infant Cardiac
{3}------------------------------------------------
Catheter is utilized when the common facial and/or internal jugular veins are too small to accommodate the larger cardiac catheter.
## Pudenz Ventricular Catheter
The Pudenz Ventricular Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the peritoneum.
## Pudenz Peritoneal Catheter
The Pudenz Peritoneal Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventriculoperitoneum. A ventriculoperitoneal shunting system may be indicated to avoid the cardiovascular complications of an atrial shunt or for a hydrocephalic patient in whom an atrial shunt is contraindicated.
## Peritoneal Reflux Control Catheter and Peritoneal Open-Ended Catheter With Slits
The Peritoneal Reflux Control Catheter and Peritoneal Open-Ended Catheter with Slits, utilized in the treatment of hydrocephalic patients, are components for systems designed to shunt cerebrospinal fluid from the lateral ventricles into the peritoneum. A ventriculoperitoneal shunting system may be indicated to avoid the cardiovascular complications of an atrial shunt or for a hydrocephalic patient in whom an atrial shunt is contraindicated.
### Portnoy Ventricular Catheter
The Portnoy Ventricular Catheter, utilized in the treatments, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum.
## Neuroview Endoscopic Ventricular Catheter
The Neuroview Ventricular Catheter, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum.
# Integra CSF Reservoir with Integral Connectors
The Integra CSF Reservoir provides access to the lateral cerebral ventricles via hypodermic puncture for sampling and/or injection of fluids. It is useful in obtaining CSF samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The Convertible Integra CSF Reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart or the peritoneum.
### Essential Shunt Kit Burr Hole Design
The CSF Control Valve, utilized in the treatments, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the peritoneum.
### Essential Shunt Kit Flat Bottom Design
The Essential Shunt Kit – Flat Bottom Design, utilized in the treatment of hydrocephalic patients, is designed to shunt cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum. The Essential Shunt Kit – Flat Bottom Design, can be used in (but is not restricted to) situations where skin erosion may be a problem, as with older patients.
### Connectors for Neurosurgical Use
Integra connectors are utilized principally in the treatment of hydrocephalic patients, as components in systems designed to shunt cerebrospinal fluid from the lateral cerebral ventricles of the right atrium of the heart or the peritoneum.
### On-Off Flushing Reservoirs
The On-Off Flushing Reservoir, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the peritoneum. Reservoirs with an Anti-Siphon Device are intended to reduce the hazard of negative intraventricular pressure when the patient is sitting, semi-recumbent or standing.
{4}------------------------------------------------
## Braden Flushing Reservoir
The Braden Flushing Reservoir, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum.
#### Foltz Flushing Reservoir
The Foltz Flushing Reservoir, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles into either the right atrium of the heart or the peritoneum.
#### Anti-Siphon Device
The Anti-Siphon Device, utilized in the treatment of hydrocephalic patients, is a component in systems designed to shunt cerebrospinal fluid from the lateral ventricles of the right atrium of the heart or the heart or the peritoneum. The device is designed to reduce the potential hazards of excessive lowering of intraventricular pressure (with respect to atmospheric pressure) when the patient is in a sitting, standing or erect position.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="10" width="10"> <rect height="10" style="fill:none;stroke-width:1;stroke:rgb(0,0,0)" width="10"></rect> <line style="stroke:rgb(0,0,0);stroke-width:1" x1="0" x2="10" y1="0" y2="10"></line> <line style="stroke:rgb(0,0,0);stroke-width:1" x1="0" x2="10" y1="10" y2="0"></line> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="10" width="10"> <rect height="10" style="fill:none;stroke-width:1;stroke:rgb(0,0,0)" width="10"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) SUMMARY
| 807.92(a)(1) – Submitter information | |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Integra LifeSciences Corporation |
| Address | 311 Enterprise Drive<br>Plainsboro, New Jersey 08536 |
| Phone Number | (609) 936-531 |
| Establishment<br>Registration Number | 9004007 |
| Name of Contact Person | Timothy Connors |
| Date Prepared | October 16, 2015 |
| 807.92(a)(2) – Name of device | |
| Trade or Propriety Names | Integra Neurological Shunts and Accessories Products<br>(LPV II Valves and Kits, Novus Valves and Kits,<br>Multi-Purpose Valve,<br>Mishler Dual Chamber Valve with Integral Connectors,<br>Pudenz Flushing Valve with Integral Connectors,<br>Ultra VS In-Line Valve System,<br>Pudenz Cardiac and Infant Catheter,<br>Pudenz Ventricular Catheter,<br>Pudenz Peritoneal Catheter,<br>Peritoneal Reflux Control Catheter,<br>Peritoneal Open-Ended Catheter With Slits,<br>Portnoy Ventricular Catheter,<br>Neuroview Endoscopic Ventricular Catheter,<br>Integra CSF Reservoir with Integral Connectors,<br>Essential Shunt Kit Burr Hole Design,<br>Essential Shunt Kit Flat Bottom Design,<br>Connectors for Neurosurgical Use,<br>On-Off Flushing Reservoirs,<br>Braden Flushing Reservoir,<br>Foltz Flushing Reservoir,<br>Anti-Siphon Device) |
| Common or Usual Name | Hydrocephalus Shunt Systems and Components |
| Classification Name | Central Nervous System Fluid Shunt and Components |
| Classification Panel | Neurology |
| Regulation | Class II, under 21 CFR 882.5550 |
| Product Code(s) | JXG |
| 807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed | |
| Equivalence is claimed to current Integra Neurological Shunts and Accessories Products,<br>as identified below: | |
LPV II Valves and Kits - K974708
{6}------------------------------------------------
Novus Valves and Kits - K961859 Multi-Purpose Valve - Pre-amendment device Mishler Dual Chamber Valve with Integral Connectors - K760784 Pudenz Flushing Valve with Integral Connectors - K760501, K760502 Ultra VS In-Line Valve System - K894072 Pudenz Cardiac and Infant Catheter - Pre-amendment device Pudenz Ventricular Catheter - Pre-amendment device Pudenz Peritoneal Catheter - Pre-amendment device Peritoneal Reflux Control Catheter - K894072 Peritoneal Open-Ended Catheter With Slits – K894072 Portnoy Ventricular Catheter - Pre-amendment device Neuroview Endoscopic Ventricular Catheter - K971617 Integra CSF Reservoir with Integral Connectors - Pre-amendment device Essential Shunt Kit Burr Hole Design - K973525 Essential Shunt Kit Flat Bottom Design - K973525 Connectors for Neurosurgical Use - Pre-amendment device On-Off Flushing Reservoirs - Pre-amendment device Braden Flushing Reservoir - Pre-amendment device Foltz Flushing Reservoir - Pre-amendment device Anti-Siphon Device - K760785
# 807.92(a)(4) - Device description
Integra Neurological Shunts and Accessories are used in the treatment of hydrocephalus. Hydrocephalus is commonly treated by creating a CSF flow pathway from a cerebral ventricle to the peritoneal spaces in the abdomen or to the right atrium of the heart. This is commonly referred to as "shunting". Integra markets a full line of products for CSF shunting procedures including catheters, valves, reservoir devices, connectors and accessories to aid in implantation.
A shunt system may comprise of a catheter, valve, reservoir and connectors, depending on clinician preference and use. In practice, a catheter is implanted into the space where CSF drainage is necessary (ventricles of the brain or lumbar subarachnoid space) and connected to a valve. The valve is used to control the CSF drainage rate. The other side of the valve is connected to a catheter which is placed to allow drainage to the desired site, typically the right atrium of the heart or the peritoneal cavity. As needed, a clinician may also incorporate a reservoir, for a closed ventricular access site, and/or additional connectors into the shunt system.
Catheters are silicone elastomer and some models are made from high durometer silicone elastomer. Valve mechanisms are categorized as diaphragm and miter. For diaphragm valves, the mechanism is an umbrella shaped component oriented at right angles to the flow path. Miter valves incorporates two silicone flaps in the shape of a duckbill. The flaps part in response to a pressure differential to allow flow. Some vales are available with low, medium or high closing pressure ranges and some contain an anti-siphon component. Reservoirs are available as standard or side-inlet or convertible and in various sizes; some have an on-off flushing feature. A variety of connectors, made of
{7}------------------------------------------------
| nylon or silicone elastomer material, are also available. | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 807.92(a)(5) – Intended use of the device | |
| Indications for Use | LPV II Valves and Kits |
| | The Standard-LPV II and Mini-LPV II Valves, utilized in<br>the treatment of hydrocephalic patients, are components in<br>systems designed to shunt cerebrospinal fluid from the<br>lateral ventricles of the brain into either the right atrium of<br>the heart or the peritoneum. |
| | The Mini-LPV II Valve can be used in (but is not restricted<br>to) situations where skin erosion may be a problem, as with<br>older patients. |
| | Novus Valves and Kits |
| | The Novus and Novus Mini Valves, utilized in the treatment<br>of hydrocephalic patients, are components in systems<br>designed to shunt cerebrospinal fluid from the lateral<br>ventricles of the brain into either the right atrium of the heart<br>or the peritoneum. |
| | The Novus Mini Valve can be used in (but is not restricted<br>to) situations where skin erosion may be a problem, as with<br>older patients. |
| | Valves with a Physiological Flow Device are intended to<br>reduce the hazard of negative intraventricular pressure (with<br>respect to atmospheric pressure) when the patient is sitting,<br>standing or semi-recumbent. |
| | Multi-Purpose Valve |
| | The Multi-Purpose Valve, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles into<br>either the right atrium of the heart or the peritoneum. |
| | Valves with an Anti-Siphon Device are intended to reduce<br>the hazard of negative intraventricular pressure when the<br>patient is sitting, semi-recumbent or standing. |
| | Mishler Dual Chamber Valve with Integral Connectors |
| | The Mishler Dual Chamber Flushing Valve, Flat Bottom<br>Design, utilized in the treatment of hydrocephalic patients, is<br>a component in systems designed to shunt cerebrospinal<br>fluid from the lateral ventricles into either the right atrium of<br>the heart or the peritoneum. |
| Valves with an Anti-Siphon Device are intended to reduce<br>the hazard of negative intraventricular pressure when the<br>patient is sitting, semi-recumbent or standing. | |
| <b>Pudenz Flushing Valve with Integral Connectors</b><br>The Pudenz Flushing Valve, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles of the<br>brain into either the right atrium of the heart or the<br>peritoneum. | |
| Valves with an Anti-Siphon Device are intended to reduce<br>the hazard of negative intraventricular pressure when the<br>patient is sitting, semi-recumbent or standing. | |
| <b>Ultra VS In-Line Valve System</b><br>The Ultra VS In-Line Valves, utilized in the treatment of<br>hydrocephalic patients, are components in systems designed<br>to shunt cerebrospinal fluid (CSF) from the lateral ventricles<br>into either the peritoneal cavity or the right atrium of the<br>heart. The in-line and burr-hole systems are designed to<br>shunt cerebrospinal fluid from the lateral ventricles into the<br>peritoneal cavity. A ventriculoperitoneal shunting system<br>may be indicated to avoid the cardiovascular complications<br>of an atrial shunt or for a hydrocephalic patient in whom an<br>atrial shunt is contraindicated. | |
| The Small and Neonate Models can be used in (but are not<br>restricted to) situations where skin erosion may be a<br>problem, as with premature infants, pediatric patients and<br>older patients. | |
| <b>Pudenz Cardiac and Infant Catheter</b><br>The Pudenz Cardiac Catheter, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles of the<br>brain into the right atrium of the heart. The Infant Cardiac<br>Catheter is utilized when the common facial and/or internal<br>jugular veins are too small to accommodate the larger<br>cardiac catheter. | |
| <b>Pudenz Ventricular Catheter</b><br>The Pudenz Ventricular Catheter, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles of the | |
| brain into either the right atrium of the heart or the<br>peritoneum. | |
| <b>Pudenz Peritoneal Catheter</b><br>The Pudenz Peritoneal Catheter, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles into<br>the peritoneum. | |
| A ventriculoperitoneal shunting system may be indicated to<br>avoid the cardio vascular<br>complications of an atrial shunt or for a hydrocephalic<br>patient in whom an atrial shunt is contraindicated. | |
| <b>Peritoneal Reflux Control Catheter and Peritoneal<br/>Open-Ended Catheter With Slits</b><br>The Peritoneal Reflux Control Catheter and Peritoneal<br>Open-Ended Catheter with Slits, utilized in the treatment of<br>hydrocephalic patients, are components for systems<br>designed to shunt cerebrospinal fluid from the lateral<br>ventricles into the peritoneum. | |
| A ventriculoperitoneal shunting system may be indicated to<br>avoid the cardiovascular complications of an atrial shunt or<br>for a hydrocephalic patient in whom an atrial shunt is<br>contraindicated. | |
| <b>Portnoy Ventricular Catheter</b><br>The Portnoy Ventricular Catheter, utilized in the treatment<br>of hydrocephalic patients, is a component in systems<br>designed to shunt cerebrospinal fluid from the lateral<br>ventricles into either the right atrium of the heart or the<br>peritoneum. | |
| <b>Neuroview Endoscopic Ventricular Catheter</b><br>The Neuroview Ventricular Catheter, utilized in the<br>treatment of hydrocephalic patients, is a component in<br>systems designed to shunt cerebrospinal fluid from the<br>lateral ventricles into either the right atrium of the heart or<br>the peritoneum. | |
| <b>Integra CSF Reservoir with Integral Connectors</b><br>The Integra CSF Reservoir provides access to the lateral<br>cerebral ventricles via hypodermic puncture for sampling<br>and/or injection of fluids. It is useful in obtaining CSF<br>samples for cytological and chemical studies, for monitoring | |
| ventricular fluid pressure and for ventricular drainage. | |
| The Convertible Integra CSF Reservoir may be utilized in<br>hydrocephalic patients as a component in systems designed<br>to shunt CSF from the lateral ventricles into either the right<br>atrium of the heart or the peritoneum. | |
| Essential Shunt Kit Burr Hole Design<br>The CSF Control Valve, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles of the<br>brain into either the right atrium of the heart or the<br>peritoneum. | |
| Essential Shunt Kit Flat Bottom Design<br>The Essential Shunt Kit - Flat Bottom Design, utilized in the<br>treatment of hydrocephalic patients, is designed to shunt<br>cerebrospinal fluid from the lateral ventricles of the brain<br>into the peritoneum. | |
| The Essential Shunt Kit - Flat Bottom Design, can be used<br>in (but is not restricted to) situations where skin erosion may<br>be a problem, as with older patients. | |
| Connectors for Neurosurgical Use<br>Integra connectors are utilized principally in the treatment of<br>hydrocephalic patients, as components in systems designed<br>to shunt cerebrospinal fluid from the lateral cerebral<br>ventricles of the brain into either the right atrium of the heart<br>or the peritoneum. | |
| On-Off Flushing Reservoirs<br>The On-Off Flushing Reservoir, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles into<br>either the right atrium of the heart or the peritoneum. | |
| Reservoirs with an Anti-Siphon Device are intended to<br>reduce the hazard of negative intraventricular pressure when<br>the patient is sitting, semi-recumbent or standing. | |
| Braden Flushing Reservoir<br>The Braden Flushing Reservoir, utilized in the treatment of<br>hydrocephalic patients, is a component in systems designed<br>to shunt cerebrospinal fluid from the lateral ventricles into<br>either the right atrium of the heart or the peritoneum. |…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.