External CSF Drainage Ventricular Catheters, External CSF Drainage Lumbar Catheters

K242974 · Sophysa · JXG · Jun 20, 2025 · Neurology

Device Facts

Record IDK242974
Device NameExternal CSF Drainage Ventricular Catheters, External CSF Drainage Lumbar Catheters
ApplicantSophysa
Product CodeJXG · Neurology
Decision DateJun 20, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days (up to 29 days maximum).

Device Story

External CSF drainage catheters (ventricular or lumbar) facilitate temporary diversion of cerebrospinal fluid; catheters inserted into ventricular or subarachnoid space and secured via suture tabs; flow diverted into external collection systems; used in clinical settings by healthcare professionals; provides pressure relief and volume reduction; benefits include management of intracranial pressure; device provided sterile for single use.

Clinical Evidence

No clinical data. Bench testing performed: visual inspection, dimensional analysis, radiopacity, pressure/flow characteristics, blockage/hemorrhagic CSF exposure, air/water tightness, dynamic breaking strength, catheter/guidewire tensile/bending strength, kink resistance, catheter fixation, corrosion resistance, and needle/hub connection strength.

Technological Characteristics

Catheters constructed from biocompatible materials; ventricular (29-35cm) and lumbar (90cm) form factors; includes stylets, trocars, guidewires, Tuohy needles, and suture tabs; sterilization via Ethylene Oxide; compliant with ISO 7197, ISO 20697, ISO 20698, and ISO 11070; lumbar version is MR Conditional.

Indications for Use

Indicated for temporary drainage of Cerebrospinal Fluid (CSF) in patients requiring ventricular or lumbar access for up to 29 days.

Regulatory Classification

Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 20, 2025 Sophysa Zheng XUE Consultant 5 Rue Guy Moquet Orsay, 91400 France Re: K242974 Trade/Device Name: External CSF drainage Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: Class II Product Code: JXG Dated: May 21, 2025 Received: May 21, 2025 Dear Zheng XUE: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K242974 - Zheng XUE Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K242974 - Zheng XUE Page 3 Sincerely, Adam D. Pierce -S Digitally signed by Adam D. Pierce -S Date: 2025.06.20 15:27:32 -04'00' Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K242974 | | | Device Name External CSF Drainage | | | Indications for Use (Describe) The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days (up to 29 days maximum). | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K242974, Page 1 of 3 510(k) Summary 510(k) Number: K242974 Sponsor Information Applicant Name: Sophysa Address: 5 Rue Guy Moquet ORSAY, FRANCE, 91400 Phone and Fax Number: +33 (0)1 69 35 35 00 (phone) +33 (0)1 69 35 36 90 (fax) Sophysa Person: Julie LOPEZ Regulatory affairs director Jlopez@sophysa.com Submission contact Person: Zheng XUE Consultant: Zheng.xuecazenave@gmail.com Date Prepared: 05-June-2025 Device Name and Classification Trade/Proprietary Name: External CSF Drainage Common Name: Central nervous system fluid shunt and components. Classification Name: Shunt, Central Nervous System and Components Regulation: 21 CFR 882.5550 Device Class: 2 Product Code: JXG Predicate Device Predicate: Phoenix Neuro Catheters FVPC2 and FVPC4, Sophysa (K853365) Lumbar Drainage Catheter Kit, Codman (K964923) Device Description The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days. There are two types of external CSF drainage catheters: - Ventricular catheters - Lumbar catheters The catheter is provided sterile and with accessories to facilitate the surgical procedures and connectors. The external CSF drainage catheters need to be connected to collection systems. Indications for Use The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days (up to 29 days maximum). Summary of Equivalence to Predicate Device The External CSF drainage is substantially equivalent to Phoenix Neuro Catheters FVPC2 and FVPC4, Sophysa (K853365) and Lumbar Drainage Catheter Kit, Codman (K964923). {5} K242974, Page 2 of 3 Table 1 : Predicate Devices Comparison | | Predicates | Proposed Devices | Rational of why no new issues of safety and effectiveness | | --- | --- | --- | --- | | Clinical Equivalence | | | | | Indications for use | K853365: Temporary, external drainage of CSF following placement in the ventricles.K964923: The Lumbar Drainage Catheter Kit is indicated for temporary access to the lumbar subarachnoid region and, when used with other Codman devices, is designed to drain cerebrospinal fluid (CSF) as a means of reducing increased intracranial volume and pressure | The Sophysa external CSF drainage catheters are intended to temporarily drain the Cerebrospinal Fluid (CSF), for less than 30 days (up to 29 days maximum). | Substantially Equivalent | | Single use | Single use | Single use | Same | | Sterility status | Supplied sterile | Supply sterile | Same | | Re-sterilization status | Cannot be re-sterilized | Cannot be re-sterilized | Same | | Technical Equivalence | | | | | Principle mode of action | K853365: The catheter is inserted into the ventricular space and fixed under the scalp by suturing. CSF flow is diverted into a compatible collection system.K964923: The catheter is inserted into the subarachnoid space and fixed to the subcutaneous tissue by suturing. CSF flow is diverted into a compatible collection system. | Same | Same | | Shape/Dimension | K853365: 23.5 cm long, 1.5 mm ID.K964923: 80 cm long, 0.76 mm ID. | Ventricular: 29-35cm long, 1.5- 2.3 mm IDLumbar: 90 cm long, 0.76 mm ID | Substantially Equivalent | | Accessories | K853365: Stylet, luer lock connector, compression hub,K964923: Guidewire, Tuohy needle, luer adapter, suture tab | Ventricular: Stylet, trocar, luer-lock connector, suture tabLumbar: Guidewire, Tuohy needle, suture tab, compression hub | Substantially Equivalent | | Performance Standards | - | ISO 7197, ISO 20697, ISO 20698, ISO 11070 | Substantially Equivalent | {6} K242974, Page 3 of 3 | | Predicates | Proposed Devices | Rational of why no new issues of safety and effectiveness | | --- | --- | --- | --- | | MRI compatibility | MRI Safe | Ventricular: MR Safe Lumbar: MR Conditional due to the presence of metallic parts within the Compression Hub which connects to an external CSF drainage collection system. | Substantially Equivalent | | Shelf life | - | 5 years | Target shelf life | | Biological Safety/Compatibility | | | | | Biocompatibility | Biocompatible materials | Biocompatible materials established by a combination of testing and rationale for the following: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Subacute/Subchronic Toxicity, Genotoxicity, Implantation (brain, 4 weeks, muscle 4 weeks), Indirect Hemolysis, Neurotoxicity | Substantially Equivalent | | Sterilization method | Ethylene Oxide | Ethylene Oxide | Same | | Primary packaging system | Tyvek pouch | Double peel-off packaging Tyvek pouch | Substantially Equivalent | ## Summary of Non-clinical testing The following bench testing has been performed on the representative samples of the External CSF drainage product line: Visual Inspection, Dimensional Analysis, Radiopacity, Pressure / Flow Characteristics, Blockage / Hemorrhagic CSF Exposure, Air and Water Tightness, Dynamic Breaking Strength, Catheter Tensile Strength, Kink Resistance, Catheter Fixation, Corrosion Resistance, Guidewire Tensile Strength, Guidewire Bending Strength, Guidewire Breakage, Tuohy Needle/Hub Connection Strength. ## Statement of Substantial Equivalence The information summarized above demonstrates that External CSF drainage is substantially equivalent to and is as safe and as effective as the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...