K161605 · Aqueduct Critical Care, Inc. · JXG · Oct 27, 2016 · Neurology
Device Facts
Record ID
K161605
Device Name
Smart External Drain (SED) System
Applicant
Aqueduct Critical Care, Inc.
Product Code
JXG · Neurology
Decision Date
Oct 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Draining and monitoring of CSF flow from the lateral ventricles is indicated in selected patients to: - Reduce intracranial pressure (ICP), e.g., pre-, intra- or postoperative. - Monitor CSF chemistry, cytology, and physiology. - Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. Monitoring of intracranial pressure (ICP) is indicated in selected patients with: - Severe head injury - Subarachnoid hemorrhage graded III, IV, or V preoperatively - Reye's syndrome or similar encephalopathies - Hydrocephalus - Intracranial hemorrhage - Miscellaneous problems when drainage is to be used as a therapeutic maneuver Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions.
Device Story
SED System automates CSF drainage and ICP monitoring; replaces manual gravity-based systems. Inputs: pressure signals from transducers in drain/reference lines. Operation: electromechanical console uses stepper-motor to pinch/release compliant tubing in disposable cartridge; maintains user-set ICP range; compensates for patient movement. Outputs: real-time ICP display, CSF flow volume, and alarms for pressure/flow excursions. Used in clinical settings (hospital); operated by healthcare staff. Benefits: eliminates manual adjustments, allows patient mobility via battery backup, provides automated alerts.
Electromechanical system with software-embedded console and sterile disposable cartridge. Uses pressure transducers for ICP measurement and stepper-motor for automated tubing occlusion. Powered by AC/battery. Dimensions/materials not specified beyond compliant tubing. Sterilization: Gamma radiation (min 25 kGy). Connectivity: Standalone console. Standards: AAMI/ANSI/ISO 11137, 11737, ES60601-1, IEC 60601-1-2, 60601-1-8.
Indications for Use
Indicated for patients requiring CSF drainage and ICP monitoring, including those with severe head injury, subarachnoid hemorrhage (grade III-V), Reye's syndrome/encephalopathies, hydrocephalus, intracranial hemorrhage, or space-occupying lesions.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Medtronic DUET External Drainage and Monitoring System (K984053)
Related Devices
K172759 — Smart External Drain (SED) System · Aqueduct Critical Care, Inc. · Nov 14, 2017
K181301 — Smart External Drain (SED) System · Aqueduct Critical Care, Inc. · Jun 14, 2018
K171586 — Smart External Drain (SED) System · Aqueduct Critical Care, Inc. · Jun 30, 2017
K192289 — IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit · Irras USA, Ltd. · Nov 15, 2019
K200807 — IRRAflow Catheter, IRRAflow Tube Set, IRRAflow Control Unit · Irras USA, Inc. · Apr 29, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2016
Aqueduct Critical Care, Inc. Mr. Thomas Clement President & CEO 11822 North Creek Parkway North, Suite 110 Bothell, Washington 98011
Re: K161605
Trade/Device Name: Smart External Drain (SED) System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: September 23, 2016 Received: September 26, 2016
Dear Mr. Clement:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena - S/A
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161605
Device Name
Smart External Drain (SED) System
#### Indications for Use (Describe)
Draining and monitoring of CSF flow from the lateral ventricles is indicated in selected patients to:
- Reduce intracranial pressure (ICP), e.g., pre-, intra- or postoperative.
- Monitor CSF chemistry, cytology, and physiology.
- Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts.
Monitoring of intracranial pressure (ICP) is indicated in selected patients with:
- Severe head injury
- Subarachnoid hemorrhage graded III, IV, or V preoperatively
- Reye's syndrome or similar encephalopathies
- Hydrocephalus
- Intracranial hemorrhage
- Miscellaneous problems when drainage is to be used as a therapeutic maneuver
Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="font-size:20px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="font-size:20px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### General Information:
| Date of Summary Preparation: | October 21, 2016 |
|-----------------------------------|---------------------------------------------------------------------------------------------------|
| Name and Address of Manufacturer: | Aqueduct Critical Care, Inc.<br>11822 North Creek Parkway North<br>Suite 110<br>Bothell, WA 98011 |
| Contact Person: | Tom Clement, President and CEO |
| | Phone: 425-985-1571<br>Fax: 425-278-9377 |
| Trade Name: | Smart External Drain (SED) System |
| Common Name: | External CSF Drainage System |
| Regulation Number: | Primary: § 21 CFR 882.5550<br>Secondary: § 21 CFR 882.1620 |
| Regulation Description: | Central Nervous System Fluid Shunt and Components<br>Intracranial Pressure Monitor |
| Regulatory Class: | Class II |
| Classification Panel: | Neurology |
| Product Code: | Primary: JXG<br>Secondary: GWM |
{4}------------------------------------------------
Device Description: The SED System is based upon traditional gravity-based drainage systems, but is designed to allow for the automated regulation of ICP without the need for continuous manual measurements, adjustments and interventions. The SED System consists of an electromechanical software embedded SED Console and a sterile, disposable SED Cartridge, which includes all components necessary to attach to the external ventricular drainage catheter via a luer-lock connector and to a drainage bag that collects the drained fluid.
The SED System is mounted on an IV pole, with the SED Console positioned by the user at an easy-to-view height, while the drainage bag is positioned below the lowest possible patient head position, which is considered to be below the height of a hospital bed.
The SED System automatically maintains a set ICP using pressure sensors (transducers) and an automated stepper-motor pinching mechanism that compresses or releases the system's compliant drain tubing in order to control the degree of CSF flow (i.e., equivalent to the alteration of CSF flow that happens when a traditional gravity drain is manually raised or lowered). The SED System displays the measured ICP information and also calculates and displays CSF flow volume output, while additionally incorporating multiple alarms given its automated functionality.
The SED System can thus automatically compensate for patient movement, allow far greater mobility (via a battery backup) and also alert hospital staff if ever the ICP exceeds the set maximum or minimum levels for a particular patient, as well as if low or high CSF drainage values (as also set by the user) are ever exceeded.
Indications for Use: Draining and monitoring of CSF flow from the lateral ventricles is indicated in selected patients to:
- . Reduce intracranial pressure (ICP), e.g., pre-, intra- or postoperative.
- . Monitor CSF chemistry, cytology, and physiology.
- Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. .
Monitoring of intracranial pressure (ICP) is indicated in selected patients with:
- Severe head injury ●
- Subarachnoid hemorrhage graded III, IV, or V preoperatively ●
- . Reye's syndrome or similar encephalopathies
- . Hydrocephalus
- Intracranial hemorrhage
- Miscellaneous problems when drainage is to be used as a therapeutic maneuver ●
Monitoring can also be used to evaluate the status pre- and postoperatively for space-occupying lesions.
{5}------------------------------------------------
Predicate Device: Aqueduct Critical Care, Inc. cites the following as the predicate device.
Predicate Device Duet External Drainage and Monitoring System K984053
Table 1 below provides a summary of the technological characteristics of the SED System in comparison to the predicate device.
| Detail or Technological<br>Characteristic | Subject Aqueduct Critical Care, Inc.<br>SED System | Predicate Medtronic Duet External<br>Drainage and Monitoring System |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name | Smart External Drain (SED) System | Medtronic DUETTM External Drainage and<br>Monitoring System |
| Intended Use | Same | To attach to an implanted, external drainage<br>catheter in order to externally drain<br>cerebrospinal fluid (CSF) and monitor both<br>CSF drainage and intracranial pressure<br>(ICP) |
| Attaches to separate,<br>commercially available EVD<br>Catheter | Yes | Yes |
| Sterile Disposable tubing set | Yes | Yes |
| CSF Drainage Bag | Yes | Yes |
| Gravity drainage of CSF | Yes | Yes |
| Method to control gravity<br>drainage of CSF | Automated adjustment based on user settings<br>(for max/min ICP) via a stepper-motor<br>controlled, tube-pinching mechanism to either<br>compress or release the compliant drainage<br>tubing contained within the sterile, disposable<br>SED Cartridge. | Manual adjustment of the drip chamber<br>either up or down the IV Pole, relative to<br>the patient's head position and ventricular<br>catheter location. |
| Pressure Transducer for ICP<br>Measurement | Yes<br>(The SED System integrates transducers into<br>its design for drain line and reference line<br>pressure measurements that are used to<br>calculate and regulate ICP.) | Yes<br>(The Duet System includes both a design<br>and instructions for attachment of a<br>transducer to the main system stopcock that<br>then allows for measurement and visual<br>display of ICP via a connected monitor.) |
| Software-based, Powered<br>Console for User Interface, User<br>ICP and Alarm Setting<br>Adjustments, Data Storage and<br>Display, and Alarms for ICP<br>Monitoring | Yes | No<br>(Note, however, that a transducer can be<br>attached (see above) and then used in<br>conjunction with an ICP Monitor to<br>provide a display and user interface for<br>ICP monitoring.) |
| Method to account for location<br>of ventricles via patient head<br>position | Reference shroud attached to patient behind<br>the ear (at the level of the EMA) to account<br>for patient's head positioning. | Laser level must be attached to the system<br>and leveled appropriately and then the<br>system "zeroed"; adjustment thereafter may<br>be needed. |
| Measured Pressure Range | -5 to 40 cm H2O (set ICP range) | 0 - 31 cm H2O (set ICP range) |
| Displayed ICP | Yes<br>(via SED Console display) | Yes<br>(via drip chamber pressure indicator<br>window or via connected transducer to<br>patient monitor display) |
| Battery Back-up | Yes | No<br>(Manual, non-powered system) |
Table 1: Comparative Summary of the SED System and Predicate Duet System
{6}------------------------------------------------
Testing Summary: To demonstrate intended device performance, as well as to support the substantial equivalence of the subject SED System, the technological and performance characteristics were evaluated as summarized in Table 2 that follows.
The results from these testing activities:
- demonstrate that the technological and performance characteristics of the subject SED . System are comparable to the predicate external drainage management system, and
- . ensure the SED System can perform in a manner equivalent to the predicate device with the same intended use as an external drainage and monitoring system.
Conclusion (Statement of Equivalence): The data and information presented within this submission support a determination of substantial equivalence, and therefore market clearance of the subject SED System via this 510(k) Premarket Notification.
{7}------------------------------------------------
| Testing | Testing Summary | Results/Conclusions |
|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| SED Console Performance Testing | | |
| SED Console<br>Verification | Simulated use test cases completed with all requirements<br>tested to verify SED Console performance. | PASS – All requirements verified. |
| SED Cartridge Performance Testing | | |
| SED Cartridge Pinch<br>Tubing Force and<br>Fatigue | SED Cartridges were tested to show that the pinch<br>tubing met requirements, including resistance to pinch<br>tubing fatigue and shut off / pinch force. | PASS - All requirements verified. |
| SED Cartridge Button<br>Verification | SED Cartridges were tested to show that the manual<br>pinch button on the cartridge met the following<br>requirements: activation force, leak resistance, and<br>fatigue life. | PASS – All requirements verified. |
| MRI Compatibility | The SED Cartridge was tested in an MRI environment of<br>1.5-Tesla and 3.0-Tesla. | PASS – All requirements verified. |
| Tubing Kink and Crush<br>Verification | SED Cartridges were tested in a bend radius to assess<br>kink resistance. For crush resistance, fluid flow<br>exhibited no change when a force was applied to the<br>tubing compared to no force applied. | PASS – All requirements verified. |
| Dimensional<br>Verification | SED Cartridges were dimensionally measured. | PASS - Measurements confirmed that all dimensional requirements were<br>met. |
| Drainage Bag<br>Verification | SED Cartridges were tested to show that drainage bags<br>met requirements for: capacity, venting, and wetting. | PASS – All requirements verified. |
| Bond Joint Testing | SED Cartridges were used to destructively pull test the<br>bond joint configurations. | PASS - All bond joints met or exceeded the minimum pull strength<br>specification. |
| SED System Performance Testing | | |
| SED System<br>Mechanical<br>Verification | The SED System was tested to ensure that all applicable<br>mechanical requirements were met. | PASS - The SED Console was shown to function properly following the<br>maximum anticipated number of latching and unlatching cycles. |
| Verification of SED<br>System Environmental<br>Conditions | The SED System was subjected to the extremes of the<br>specified environmental conditions (including<br>temperature, humidity, and altitude) to confirm intended<br>system functionality post-conditioning. | PASS - SED System functioned as intended post-conditioning. |
| Long-Term Stability of<br>SED System | SED Systems were tested under clinical-use simulation<br>for the longest expected use duration of the SED<br>System | PASS - SED System met all requirements under clinical-use simulation. |
| Testing | Testing Summary | Results/Conclusions |
| Verification of the<br>Pressure Measurement<br>Range and Response<br>Time of the SED<br>System | The SED System was tested to the extremes of the<br>pressure measurement range, as well as to measure the<br>response time of the system. | PASS - All applicable requirements verified. |
| Packaging Validation Testing | | |
| Packaging Validation<br>Testing | The packaging configurations for the reusable SED<br>Console and sterile SED Cartridge and Drain Bag were<br>subjected to applicable distribution simulation per<br>ASTM D4169, followed by testing:<br><br>The SED Console was tested to ensure it maintained<br>proper function via simulated clinical-use.<br>The sterile SED Cartridge and Drain Bag packaging was<br>subjected to seal strength, dye leak testing, and visual<br>inspection, while the SED Cartridge and Drain Bag were<br>subjected to functional testing to confirm proper device<br>function. | PASS - Package integrity and product functionality met all applicable<br>requirements. |
| Sterilization Validation Testing | | |
| Sterilization Validation | Sterilization validation testing was completed to validate<br>that the gamma radiation (min 25 kGy) sterilization<br>process provides a Sterility Assurance Level (SAL) of<br>10-6, using the following applicable standards and<br>sterilization validation methods.<br><br>• AAMI / ANSI / ISO 11737-1:2006 (R) 2011<br>• AAMI / ANSI / ISO 11737-2:2009/(R) 2014<br>• AAMI / ANSI / ISO 11137-1:2006/(R) 2010,<br>[Including: Amendment 1 (2013)]<br>• AAMI / ANSI / ISO 11137-2:2013 | PASS – The gamma radiation sterilization process provides an SAL of<br>10-6. |
| Shelf-Life Testing | | |
| Shelf-Life Testing | Accelerated age testing was completed on the SED<br>Cartridge and SED Drain bag components and their<br>respective packaging to support a 6-month shelf life. | PASS - All packaging and functional requirements were met following 6<br>months of accelerated aging. |
| Testing | Testing Summary | Results/Conclusions |
| Biocompatibility Testing | | |
| Biocompatibility<br>Testing | Biocompatibility testing completed on all direct (skin) and indirect (potential limited and brief periods of<br>retrograde drainage fluid flow) contact portions of the<br>SED System. This included cytotoxicity, sensitization,<br>intracutaneous reactivity, LAL pyrogenicity and acute<br>systemic toxicity. | PASS - SED System met all applicable biocompatibility requirements for<br>its intended use. |
| Software Testing | | |
| Software Verification<br>and Validation | Code verification and software verification and<br>validation testing was performed. | PASS - SED System met all acceptance criteria for verification and<br>validation |
| Electrical Safety and Electromagnetic Compatibility Testing | | |
| Verify the Electrical<br>Safety and<br>Electromagnetic<br>Compatibility (EMC)<br>of the SED System | Applicable electrical safety and EMC testing (including<br>emissions and immunity) was completed on the SED<br>System in accordance with the following standards.<br>• AAMI / ANSI ES60601-1:2005/(R)2012 and<br>A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012<br>(Consolidated Text) (IEC 60601-1:2005, MOD)<br>• IEC 60601-1-2 Edition 3: 2007-03<br>• IEC 60601-1-8 Edition 2.0 2006-10 | PASS - SED System met all applicable electrical safety and EMC<br>requirements. |
# Table 2: Summary Table for SED System Testing
{8}------------------------------------------------
{9}------------------------------------------------
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.