K984053 · Medtronic PS Medical · JXG · Feb 4, 1999 · Neurology
Device Facts
Record ID
K984053
Device Name
BECKER EXTERNAL DRAINAGE AND MONITORING SYSTEM
Applicant
Medtronic PS Medical
Product Code
JXG · Neurology
Decision Date
Feb 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to reduce intracranial pressure (ICP), e.g. preintra- or postoperative; monitor CSF chemistry, cytology and physiology; provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. The monitoring of the intracranial pressure (ICP) is indicated in selected patients with severe head injury; subarachnoid hemormage graded III, IV or V preoperatively; Reyes syndrome or similar encephalopathies; hydrocephalus; intracranial hemorrhage or miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre-and postoperatively for space occupying lesions.
Device Story
Becker EDMS is a closed system for draining and monitoring cerebrospinal fluid (CSF) from lateral ventricles or lumbar subarachnoid space. System components include a non-distensible patient connection line, stopcock, mounting panel, injection sites, and a removable drainage bag with microbial filter. A graduated 50 cc flow chamber with drip former and conical bottom allows for volumetric measurement of CSF. Used in clinical settings by neurosurgeons and clinical staff to manage intracranial pressure (ICP) and monitor CSF chemistry/physiology. By providing temporary drainage and pressure monitoring, the device assists in therapeutic management of neurological conditions and post-operative status. Output is visual (fluid level in chamber/bag) and allows clinicians to assess patient intracranial status and adjust drainage therapy accordingly.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material, and performance characteristic comparisons to the predicate device.
Technological Characteristics
System includes non-distensible patient connection line, stopcock, mounting panel, non-latex injection sites, and drainage bag with microbial filter. Features a graduated 50 cc flow chamber with drip former and conical bottom. Materials are consistent with currently marketed devices. Non-electronic, mechanical system.
Indications for Use
Indicated for patients requiring CSF drainage or ICP monitoring, including those with severe head injury, subarachnoid hemorrhage (grade III-V), Reyes syndrome, encephalopathies, hydrocephalus, intracranial hemorrhage, space-occupying lesions, or infected CSF shunts.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
{0}------------------------------------------------
#### Safety and Effectiveness Summary ll.
984as3
## Determination of substantial equivalence
The ICP monitoring device and method were first described in an article published in 1973. Over the years the device has come into widespread use.
Medtronic PS Medical has, with input from neurosurgeons, developed a device design, which is considered to be equivalent to the predicate Codman External Drainage System II (K902257, 19249383 The Medtronic PS Medical Becker EDMS is a complete closed system indicated for the draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space.
K910938 - listed incorrectly throughout Medtronic PS Medical considers the Medtronic PS Medical Becker EDMS to be substantially equivalent to the Codman External Drainage System II (K902257, K920938), currently in commercial distribution. The predicate device is used for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing and controlling increased intracranial volume and pressure when the insertion of a permanent internal shunt is not needed. A comparison of the Codman External Drainage System II and the Medtronic PS Medical Becker EDMS is included in Tables 1 and 2.
Medtronic PS Medical considers the Becker EDMS to be safe, effective and substantially equivalent in indications, physical characteristics and performance to the noted predicate device. Evidence of this is provided in predicate device product labeling (Attachment 2).
#### Device Name
Medtronic PS Medical Becker External Drainage and Monitoring System (Becker EDMS).
## Predicate Device(s)
Codman External Drainage System II (K902257, K920938)
#### Device Description
The Medtronic PS Medical Becker EDMS is indicated for the draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space. The Becker EDMS is substantially equivalent in indications, materials of fabrication, performance characteristics and design specifications as compared with the predicate device, Codman External Drainage System II.
The materials utilized in the Becker EDMS are not substantially different to those used in currently marketed products. The Becker EDMS is comprised of a nondistensible blue striped (proximal end) patient connection line, patient line stopcock, mounting panel/main system section, three non-latex iniection sites, and a removable drainage bag with approximate volumetric graduations and microbial filter. A graduated 50 cc
36
page 2 of 1
{1}------------------------------------------------
flow chamber with drip former and conical bottom and locking bracket is included on the main system section.
A comparison of the Becker EDMS and predicate device is provided in Table 1.
# Intended Use
The Medtronic PS Medical Becker EDMS will be indicated for use as follows: "Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to reduce intracranial pressure (ICP), e.g. preintra- or postoperative; monitor CSF chemistry, cytology and physiology; provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. The monitoring of the intracranial pressure (ICP) is indicated in selected patients with severe head injury; subarachnoid hemormage graded III, IV or V preoperatively; Reyes syndrome or similar encephalopathies; hydrocephalus; intracranial hemorrhage or miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre-and postoperatively for space occupying lesions."
Draft Medtronic PS Medical product labeling and product literature for the Medtronic PS Medical Becker EDMS is included (Attachment 1).
## Intended Use predicate device
The predicate device Codman External Drainage System II is indicated for use as follows:
"Use of the Codman External Drainage System II is indicated for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing and controlling increased intracranial volume and pressure when the insertion of a permanent internal shunt is not indicated."
Product literature for the predicate Codman External Drainage System II is included (Attachment 2).
37
page 3 of
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three curved lines that suggest feathers or wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1999 FEB
Mr. Jeffrey Henderson Vice President, Quality Medtronic PS Medical 125 Cremona Drive Goleta, California 93117-5500
Re: K984053 Becker External Drainage and Monitoring System Regulatory Class: II Product Code: JXG November 3 1998 Dated: Received: November 6, 1998
#### Dear Mr. Henderson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Failure to Administration (FDA) will verify such assumptions. comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Mr. Jeffrey Henderson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Russell S. Page
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Device Name: Becker External Drainage and Monitoring System (EDMS)
510(k) Number (if known):
984053
Indications for Use:
"Draining and monitoring of CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to reduce intracranial pressure (ICP), e.g. preintra- or postoperative; monitor CSF chemistry, cytology and physiology; provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. The monitoring of the intracranial pressure (ICP) is indicated in selected patients with severe head injury; subarachnoid hemorthage graded III, IV or V preoperatively; Reyes syndrome or similar encephalopathies; hydrocephalus; intracranial hemorrhage or miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre-and postoperatively for space occupying lesions."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Nino Panzera for J2P
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number_ 3 3 40 3
Over the Counter Use: or Prescription Use: (Per 21 CFR 801.109)
(optional format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.