Medtronic External Drainage and Monitoring System, Medtronic External Drainage and Monitoring System, Becker Replacement Drainage Bags
Applicant
Medtronic, Inc.
Product Code
GWM · Neurology
Decision Date
Apr 9, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Draining CSF and monitoring CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to: 1. Reduce intracranial pressure (ICP), e.g. pre-, intra- or postoperative. 2. Monitor CSF chemistry, cytology and physiology. 3. Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts. The monitoring of the intracranial pressure (ICP) is indicated in selected patients with: 1. Severe head injury 2. Subarachnoid hemorrhage graded III, IV or V preoperatively 3. Reye's syndrome or similar encephalopathies 4. Hydrocephalus 5. Intracranial hemorrhage 6. Miscellaneous problems when drainage is to be used as a therapeutic maneuver. Monitoring can also be used to evaluate the status pre- and postoperative for space-occupying lesions.
Device Story
Medtronic EDMS is a closed system for temporary drainage and monitoring of cerebrospinal fluid (CSF) from lateral ventricles or lumbar subarachnoid space. System components include patient line with stopcock, graduated chamber, hanging bracket, drainage bag with braided cord, and pressure scale tape. Device operates via gravity and intracerebral pressure; drainage rate determined by system height relative to patient anatomy. Used in clinical settings (e.g., hospital/OR) by healthcare professionals to manage intracranial pressure, monitor CSF chemistry/physiology, or treat infected shunts. Output is visual monitoring of CSF flow and pressure via graduated chamber and scale tape, enabling clinical decision-making regarding therapeutic drainage. Benefits include controlled pressure reduction and diagnostic access to CSF.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing only, including visual/dimensional inspection, leakage testing, flow initiation pressure, seal weld integrity, and tensile strength testing.
Technological Characteristics
Closed system for CSF drainage/monitoring. Components: patient line with latex-free injection site, graduated chamber, drainage bag, pressure scale tape. Sterilization: Ethylene Oxide. Shelf life: 2 years. Non-pyrogenic. Operates via gravity/intracerebral pressure. No electronic or software components.
Indications for Use
Indicated for patients requiring CSF drainage or ICP monitoring due to severe head injury, subarachnoid hemorrhage (grade III-V), Reye's syndrome/encephalopathies, hydrocephalus, intracranial hemorrhage, or space-occupying lesions.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 9, 2020
Medtronic, Inc. Jen Correa Manager, Regulatory Affairs 5290 California Avenue Irvine, California 92617
Re: K200630
Trade/Device Name: Medtronic External Drainage and Monitoring System (EDMS) Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM, HCA Dated: March 9, 2020 Received: March 10, 2020
Dear Jen Correa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K200630
#### Device Name
Medtronic External Drainage and Monitoring System (EDMS)
Indications for Use (Describe)
Draining CSF and monitoring CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to:
- 1. Reduce intracranial pressure (ICP), e.g. pre-, intra- or postoperative.
- 2. Monitor CSF chemistry, cytology and physiology.
- 3. Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts.
The monitoring of the intracranial pressure (ICP) is indicated in selected patients with:
- 1. Severe head injury
- 2. Subarachnoid hemorrhage graded III, IV or V preoperatively
- 3. Reye's syndrome or similar encephalopathies
- 4. Hydrocephalus
- 5. Intracranial hemorrhage
- 6. Miscellaneous problems when drainage is to be used as a therapeutic maneuver.
Monitoring can also be used to evaluate the status pre- and postoperative for space-occupying lesions.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
# Medtronic External Drainage and Monitoring System (EDMS)
### 510(k) Summary
April 6, 2020
- I. Company: Medtronic Neurosurgery 5290 California Ave Irvine, CA 92617 USA Contact: Jen Correa Manager, Regulatory Affairs Jennifer.L.Correa@medtronic.com Telephone Number: 949-297-5494
#### II. Establishment Registration Number: 3015531529
- III. Proprietary Trade Name: Medtronic External Drainage and Monitoring System (EDMS)
- IV. Regulatory Class: II
#### V. Primary Classification
Name: Intracranial pressure monitoring device Product Code: GWM Regulation: 21 CFR 882.1620 Secondary Classification: Name: Ventricular catheter Product Code: HCA Regulation: 21 CFR 882.4100
- VI. Product Description: The Medtronic External Drainage and Monitoring System EDMS is provided as a complete closed system for the drainage and monitoring of cerebrospinal fluid (CSF) flow from the lateral ventricles or the lumbar subarachnoid space. The system is offered in various kit configurations for various clinical applications. The EDMS Drainage Assembly is supplied pre-assembled, sterile and non-pyrogenic in a double wrap package system. A drainage bag with braided cord is also included with each EDMS kit.
The EDMS and components are intended for single (one time) use only and is not designed or intended to be re-used, re-processed, or re-sterilized. Some of the basic features include the following:
- . a patient line stopcock with latex-free injection site and non-distensible patient connection line:
- a graduated chamber and hanging bracket for I.V. pole suspension; ●
- a drainage bag connection line with two slide clamps and latex-free injection site;
{4}------------------------------------------------
- a removable vented drainage bag with approximate volumetric graduations and drainage port; ●
- pressure scale tape.
#### VII. Indications for Use:
Draining CSF and monitoring CSF flow from the lateral ventricles or lumbar subarachnoid space is indicated in selected patients to:
- Reduce intracranial pressure (ICP), e.g. pre-, intra- or postoperative. 1.
- 2. Monitor CSF chemistry, cytology and physiology.
- 3. Provide temporary CSF drainage in patients with infected cerebrospinal fluid shunts.
The monitoring of the intracranial pressure (ICP) is indicated in selected patients with:
- 1. Severe head injury
- 2. Subarachnoid hemorrhage graded III, IV or V preoperatively
- Reye's syndrome or similar encephalopathies 3.
- 4. Hydrocephalus
- 5. Intracranial hemorrhage
- 6. Miscellaneous problems when drainage is to be used as a therapeutic maneuver.
Monitoring can also be used to evaluate the status pre- and postoperative for spaceoccupying lesions.
- VIII. Summary of the Technological Characteristics: The proposed subject drainage bag shares the same technological characteristics as the original predicate drainage bag. The rest of the EDMS device has the same technological characteristics as the predicate device.
| | Subject Device | Predicate Device<br>K802100 |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>use/Indications<br>for Use | Draining CSF and monitoring CSF flow<br>from the lateral ventricles or lumbar<br>subarachnoid space is indicated in<br>selected patients to:<br>1. Reduce intracranial pressure (ICP), e.g.<br>pre-, intra- or postoperative.<br>2. Monitor CSF chemistry, cytology and<br>physiology.<br>3. Provide temporary CSF drainage in<br>patients with infected cerebrospinal fluid<br>shunts.<br>The monitoring of the intracranial pressure<br>(ICP) is indicated in selected patients with:<br>1. Severe head injury | The monitoring of intracranial pressure<br>(ICP) is indicated on selected patients<br>with:<br>1. Severe head injury.<br>2. Status pre- and postoperative for<br>space-occupying intracranial lesions.<br>3. Subarachnoid hemorrhage graded IV<br>or V preoperatively.<br>4. Reyes syndrome or similar<br>encephalopathies.<br>5. Hydrocephalus.<br>6. Intracranial hemorrhage.<br>7. Miscellaneous problems when<br>ventricular drainage is to be used as a<br>therapeutic maneuver. |
{5}------------------------------------------------
| | Subject Device | Predicate Device |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| | | K802100 |
| | 2. Subarachnoid hemorrhage graded III, IV<br>or V preoperatively | |
| | 3. Reye's syndrome or similar<br>encephalopathies | |
| | 4. Hydrocephalus | |
| | 5. Intracranial hemorrhage | |
| | 6. Miscellaneous problems when drainage is<br>to be used as a therapeutic maneuver. | |
| | Monitoring can also be used to evaluate the<br>status pre- and postoperative for space-<br>occupying lesions. | |
| Operating<br>Principle | External drainage is temporary drainage of<br>cerebrospinal fluid (CSF) from the lateral<br>ventricles of the brain, or the lumbar space of<br>the spine, into an external collection bag. The<br>Medtronic EDMS drains CSF by using a<br>combination of gravity and intracerebral<br>pressure. The drainage rate depends on the<br>height at which the system is placed relative<br>to the patient's anatomy. | Same |
| Anatomical<br>Sites | Drainage and monitoring of CSF flow from<br>the lateral ventricles or lumbar subarachnoid<br>space. | Same |
| Biocompatibility | No direct patient contact<br>In contact with CSF only | Same |
| Sterilization<br>Method | Ethylene Oxide | Same |
| Pyrogenicity | Non pyrogenic | Same |
| Shelf life | 2 yr. | Same |
- IX. Identification of Legally Marketing Devices: The Intracranial Pressure Monitoring System, Becker Design, K802100.
- X. Discussion of the Performance Testing: In accordance with the risk assessment of the change it was determined that dimensional verification, and design verification testing of the bag was necessary. The successful results of the testing demonstrated that the changes do not raise questions of safety and effectiveness, supporting the substantial equivalence to the predicate device.
{6}------------------------------------------------
### Performance Data - Bench
| Test | Test Method Summary / Purpose | Results |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Visual and<br>Dimensional<br>Inspection | Visual and dimensional inspection demonstrates that<br>the printed graduations meet volumetric capacity.<br>Volumetric graduations are approximate. | The EDMS device met the<br>acceptance criteria for visual and<br>dimensional inspection. |
| Leakage of<br>Drainage Bag | The drainage bag must withstand being inverted<br>without leaking. | The EDMS device met the<br>acceptance criteria for drainage<br>bag leakage. |
| Flow Initiation<br>Pressure | Record pressure at which flow initiates, for each<br>drainage bag. | The EDMS device met the<br>acceptance criteria for flow<br>initiation pressure. |
| Drainage Bag<br>Seal Weld | Ensure there are no leaks in the drainage bag. | The EDMS device met the<br>acceptance criteria for drainage<br>bag seal weld. |
| Tensile Strength<br>of Drainage Bag<br>Inlet Port | Evaluate the tensile strength of the drainage bag<br>inlet port to failure. | The EDMS device met the<br>acceptance criteria for tensile<br>strength of drainage bag inlet<br>port. |
# Performance Data – Animal
The risk assessment of the proposed modifications to the disposable bag did not require animal testing. Determination of substantial equivalence for the design change is based upon the design verification bench testing.
# Performance Data - Clinical
The risk assessment of the proposed modifications to the disposable bag did not require clinical testing. Determination of substantial equivalence for the design change is based upon the design verification bench testing.
- XI. Conclusions: The information provided in this submission demonstrates that the subject device, the Medtronic EDMS, has the same intended use/indications for use as the predicate device and the differences in technological characteristics introduced by the proposed changes to the drainage bag do not raise any questions of safety and effectiveness. Based on the information provided in this submission the subject Medtronic EDMS device is considered substantially equivalent to the previously cleared predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.