AESCULAP - MEITHKE PROGAV 2.0 ADJUSTABLE SHUNT SYSTEM

K141687 · Aesculap, Inc. · JXG · Apr 2, 2015 · Neurology

Device Facts

Record IDK141687
Device NameAESCULAP - MEITHKE PROGAV 2.0 ADJUSTABLE SHUNT SYSTEM
ApplicantAesculap, Inc.
Product CodeJXG · Neurology
Decision DateApr 2, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV 2.0 shunt can be verified by using the proGAV 2.0 Compass but must be confirmed by radiograph (X-ray).

Device Story

Adjustable differential pressure valve for hydrocephalus treatment; shunts CSF from lateral ventricles to peritoneum. Device comprises titanium housing, leaf spring, and ball mechanism; pressure settings controlled via internal magnets. Physician uses external manual tools to locate, verify, and adjust pressure settings pre- and post-operatively; Compass tool provides verification, but X-ray confirmation required. Device provides audible/tactile feedback during adjustment. Used in clinical settings by neurosurgeons. Benefits include precise, non-invasive pressure adjustment to manage intracranial pressure.

Clinical Evidence

Bench testing only. Testing performed per ISO 7197:2009 included resistance to leakage, pressure-flow performance, overpressure integrity, dynamic break strength, bursting pressure, reflux performance, long-term stability (28 days), and adjustability/readability. MRI safety evaluated per ASTM F2119, F2182, F2213, and F2052. All samples passed acceptance criteria.

Technological Characteristics

Adjustable differential pressure valve. Materials: Titanium Alloy Ti6Al4V, Neodym Ferrite Boron, Alpha Sapphire. Dimensions: 17mm diameter, 4.5mm height. 4 internal magnets. Audible/tactile feedback mechanism. Sterilization: Steam. MR Conditional (3-Tesla).

Indications for Use

Indicated for patients requiring cerebrospinal fluid (CSF) shunting from the lateral ventricles of the brain into the peritoneum to treat hydrocephalus.

Regulatory Classification

Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three overlapping profiles facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 2, 2015 Aesculap, Inc. Ms. Kathy Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034 Re: K141687 Trade/Device Name: Aesculap - Miethke ProGAV 2.0 Adjustable Shunt System Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: March 5, 2015 Received: March 6, 2015 Dear Ms. Kathy Racosky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carlos L. Pena -S 同公 Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K141687 Device Name Aesculap Miethke proGAV 2.0 Adjustable Shunt System Indications for Use (Describe) The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV 2.0 shunt can be verified by using the proGAV 2.0 Compass but must be confirmed by radiograph (X-ray). | | Type of Use (Select one or both, as applicable) | |--|-------------------------------------------------| |--|-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 510(k) SUMMARY (as required by 21 CFR 807.92) #### Miethke proGAV 2.0 Adjustable Shunt System March 31, 2015 | COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 | |----------------------|--------------------------------------------------------------------------------------------------------------------| | CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com | | TRADE NAME: | Aesculap Miethke proGAV 2.0 Adjustable Shunt System | | COMMON NAME: | Hydrocephalus Shunt System | | CLASSIFICATION: | Class II | | CLASSIFICATION NAME: | Shunt, Central Nervous System and Components | | REGULATION NUMBER: | 882.5550 | PRODUCT CODE: JXG B. ### SUBSTANTIAL EQUIVALENCE Aesculap. Inc. believes that the Aesculap Miethke proGAV 2.0 Adjustable Shunt System is substantially equivalent to the Miethke proGAV Programmable Shunt System (K062009 / K103003). The proGAV 2.0 has some differences in technological features in comparison to the predicate device. There is no difference between the subject device and the predicate with respect to indications for use. ### DEVICE DESCRIPTION proGAV 2.0 is an adjustable differential pressure valve that can be set for a range of pressures. The proGAV 2.0 valve is comprised of a titanium housing that contains a leaf spring and ball mechanism that is mechanically controlled by internal magnets. Manual devices are available to locate, verify the pressure setting and to set or re-set the pressure pre and postoperatively. These manual accessories are for external use by the physician. Once verified using the Compass the setting must be confirmed with an X-ray. The device will be distributed by itself or in combination with the ShuntAssistant valve or proSA valve. The proGAV 2.0 adjustable differential pressure valve includes the same legally marketed accessories that are available with the Miethke Shunt Systems. {5}------------------------------------------------ The Miethke proGAV 2.0 Adjustable Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Adjustments of the proGAV 2.0 shunt can be verified by using the proGAV 2.0 Compass but must be confirmed by radiograph (X-ray). ## TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s)) The Aesculap Miethke proGAV 2.0 Adjustable Shunt System is substantially equivalent to the predicate Miethke proGAV Programmable Shunt System. The subject device is shown to be substantially equivalent and has the same performance characteristic to its predicate device through comparison in design, principles of operation, intended use, and materials. The proGAV and proGAV 2.0 device characteristics are summarized below. | | New Device<br>Miethke Shunt System<br>proGAV 2.0 Valve | Predicate<br>Miethke Shunt System<br>proGAV Valve<br>K103003/K062009 | |-----------------------------|--------------------------------------------------------|----------------------------------------------------------------------| | Adjustable: | Yes | Yes | | Valve Type: | Adjustable differential pressure | Adjustable differential pressure | | Pressure levels: | 0 - 20 cmH2O | 0 - 20 cmH2O | | Materials: | Titanium Alloy Ti6Al4V | Yes | | | Neodym Ferrite Boron | Yes | | | Alpha Sapphire | Yes | | Design: | Circular | Circular | | Housing: | | | | Inner | smooth | N/A | | Outer | thin curved wave profile | smooth | | Audible or tactile feedback | Yes | No | | Dimensions: | | | | Height | 4.5 mm | 4.4 mm | | Diameter | 17 mm | 18 mm | | # of Magnets | 4 | 2 | | Sterilization: | Steam | Steam | | Packaging: | Double Peel Pouch | Double Peel Pouch | | Manual Instruments: | Yes | Yes | | Tool settings and readings | 0 – 20 cmH2O | 0 - 20 cmH2O | | Packaging: | PE case or wooden box (set) | PE case | # PERFORMANCE DATA The below preclinical testing was performed to demonstrate that the Aesculap Miethke proGAV 2.0 Adjustable Shunt System performs as intended and is substantially equivalent to the predicate device. Testing was conducted in accordance with ISO 7197:2009. {6}------------------------------------------------ | Test | Test Method Summary | Results | |---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Resistance to Leakage | 100 cm of air applied to the subject<br>device submerged in water. No<br>leakage is allowed with a differential<br>pressure from the inside to outside<br>of 100 cm water column within 5<br>min. | No leakage detectable within 5 min<br>under 100cm applied overpressure.<br>Samples passed the acceptance<br>criteria | | Pressure-Flow | Pressure-flow-performance tested<br>between the flow range of 5 to 50<br>ml/h. The measured pressure has to<br>remain inside manufacturers<br>declaration. | The pressure flow performance of<br>each tested valve remains inside<br>manufacturers declaration.<br>Samples passed the acceptance<br>criteria | | Overpressure | Function and integrity of the subject<br>device shall withstand a positive<br>pressure of 1 m water column<br>applied to the open shunt. | Valve function/performance was<br>verified due to an additional<br>measurement of the pressure flow<br>performance after applying the<br>overpressure. Integrity was verified<br>by passing the dynamic braking<br>force.<br>Samples passed the acceptance<br>criteria | | Dynamic Break Strength | Using a frequency of 1 Hz ±0.2,<br>tension is applied in flow direction<br>and should lead to an elongation of<br>the subject device of 10% or a<br>maximum force of 5 N. Testing is<br>carried out for 100,000 cycles. | No sample had a rupture or crack<br>after applying 100 000 cycles.<br>Samples passed the acceptance<br>criteria | | Bursting Pressure | Subject device must withstand a<br>positive pressure of 2 m of water<br>column inside the subject device<br>without any change within a<br>tolerance of ± 10% (no later than 2<br>hours after the burst pressure<br>application). | No later than 2 hours after applying<br>the positive pressure of 2m water<br>column the valve pressure flow<br>performance was inside<br>manufacturers declaration.<br>Samples passed the acceptance<br>criteria | | Reflux performance | To verify resistance a water bath<br>was used for the 500 mm of water<br>column against the flow direction of<br>the subject device. A maximum<br>flow of 0.04 ml/min is allowed to be<br>drained back. | All tested items had a measured<br>reflux lower that allowed.<br>Samples passed the acceptance<br>criteria | | Long Term Stability | The subject device was immersed in<br>distilled water and kept at 36°C ±5<br>while pumping distilled water at an<br>average flow rate of 20 ml/h through<br>the valve for at 28 days. Flow rate<br>was check 3 times a day. Patient<br>position was simulated (14 days/<br>horizontal and 14 days/ vertical) | All items remained inside the<br>manufacturers declaration over the<br>period of 28 days.<br>Samples passed the acceptance<br>criteria | | Valve adjustability/readability | Different materials with varying<br>thicknesses were placed between the<br>valve and the adjustment/verification<br>tools to verify the<br>adjustability/verification of the valve | The adjustability and verification of<br>all items remained inside the<br>manufacturers declaration. Samples<br>passed the acceptance criteria | Results of the above testing demonstrates that the device is substantially equivalent to the predicate device. {7}------------------------------------------------ In addition testing was performed according to the following MRI standards: - ASTM F2119 Evaluation of MR Image Artifacts ● - ASTM F2182 Measurement of Radio Frequency Induced Heating During Magnetic ● Resonance Imaging - ASTM F2213 Qualitative Measurement of Magnetically Induced Torque in the . Magnetic Resonance Environment - ASTM F2052 Measurement of Magnetically Induced Displacement Force on the in . the Magnetic Resonance Environment The results and evaluation conclude that the device is MR Conditional in 3-Tesla Magnetic Resonance Imaging (MRI) systems according to ASTM F 2503 and is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...