EAGLE EYE PLATINUM CATHETER, EAGLE EYE PLATINUM SHORT TIP CATHETER, VISIONS PV .014 PLATINUM

K143701 · Volcano Corporation · OBJ · Aug 26, 2015 · Cardiovascular

Device Facts

Record IDK143701
Device NameEAGLE EYE PLATINUM CATHETER, EAGLE EYE PLATINUM SHORT TIP CATHETER, VISIONS PV .014 PLATINUM
ApplicantVolcano Corporation
Product CodeOBJ · Cardiovascular
Decision DateAug 26, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1200
Device ClassClass 2

Intended Use

The Eagle Eye® Platinum (EEP) Digital IVUS Catheter Line, including the Eagle Eye Platinum catheter, Eagle Eye Platinum ST catheter and the PV.014P catheter, are designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. These devices are not currently indicated for use in the cerebral vessels. The Eagle Eye® Platinum (EEP) Digital IVUS Catheter Line is as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures

Device Story

Digital IVUS catheter line; incorporates distal cylindrical ultrasound transducer array; radiates acoustic energy; detects ultrasonic echoes; generates real-time cross-sectional images of coronary/peripheral vessels. Used in clinical settings; operated by physicians during angiographic procedures. Tracks over 0.014" guidewire; introduced percutaneously or via surgical cut down. Compatible with Volcano imaging systems (In-Vision, s5, s5i, CORE Mobile, CORE). Output displayed to clinician to visualize vessel lumen/wall structures; aids clinical decision-making during vascular interventions.

Clinical Evidence

Bench testing only. Evaluated friction force, coating adhesion, and particulate generation. Comprehensive biocompatibility testing performed, including cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, hemolysis, thromboresistance, complement activation, and genotoxicity.

Technological Characteristics

Cylindrical ultrasound transducer array; 0.014" guidewire compatibility; hydrophilic coating. Operates with Volcano imaging systems using VH IVUS software v1.2+. Biocompatibility per ISO/ASTM standards (e.g., ASTM Hemolysis).

Indications for Use

Indicated for evaluation of vascular morphology in coronary and peripheral vasculature; adjunct to angiography for imaging vessel lumen and wall structures. Not indicated for cerebral vessels.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Volcano Corporation Elaine Alan Senior Regulatory Affairs Specialist 1 Fortune Drive Billerica, Massachusetts 01821 Re: K143701 Trade/Device Name: Eagle Eye Platinum (EEP) Digital IVUS Catheter Line Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, ITX Dated: June 23, 2015 Received: July 16, 2015 Dear Elaine Alan, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Arthell Stein forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143701 Device Name Eagle Eye Platinum (EEP) Digital IVUS Catheter Line #### Indications for Use (Describe) The Eagle Eye® Platinum (EEP) Digital IVUS Catheter Line, including the Eagle Eye Platinum catheter, Eagle Eye Platinum ST catheter and the PV.014P catheter, are designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. These devices are not currently indicated for use in the cerebral vessels. The Eagle Eye® Platinum (EEP) Digital IVUS Catheter Line is as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Volcano. The logo consists of a gray square with a white triangle inside, and the word "VOLCANO" in large, bold, blue letters. The triangle inside the square is pointing downwards. The word "VOLCANO" is to the right of the square. Volcano Corporation 1 Fortune Drive Billerica, MA 01821 Main: +1 978-262-0049 Fax : +1 978-262-0035 www.volcanocorp.com ## 510(K) SUMMARY | SPONSOR: | Volcano Corporation<br>3721 Valley Centre Drive, Suite 500<br>San Diego, CA 92130 | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT/SUBMITTER: | Elaine Alan<br>Senior Regulatory Affairs Specialist<br>Volcano Corporation<br>1 Fortune Drive<br>Billerica, MA 01821<br>Tel: (858) 764-1281 | | DATE OF SUBMISSION: | December 21, 2014 | | DEVICE: | Eagle Eye Platinum (EEP) Digital IVUS Catheter Line | | Trade Name: | Eagle Eye Platinum (EEP) Digital IVUS Catheter Line | | Common Name: | Diagnostic Intravascular Catheter | Classification and Product Codes: | CFR Number | Class | Product Code | |---------------------------------------------------------|-------|--------------| | 21 CFR 870.1200<br>Diagnostic Intravascular<br>Catheter | II | OBJ | | 21 CFR 892.1570<br>Diagnostic Ultrasound<br>Transducer | II | ITX | PREDICATE DEVICE: Eagle Eye Platinum (EEP) Digital IVUS Catheter (K092596) Eagle Eye Platinum Short Tip (EEP-ST) Catheter (K120697) {4}------------------------------------------------ ## DEVICE DESCRIPTION: The EEP line of digital IVUS catheters consists of the following products: - Eagle Eye Platinum Catheter, Part # 809746001, Model 85900P ● - Eagle Eye Platinum Short Tip Catheter, Part # 400-0200.141, Model 85900PST - Visions PV .014 Platinum, Part # 400-0200.233, Model 85910P The EEP line of catheters incorporates a cylindrical ultrasound transducer array located near the distal tip of the catheter. The array radiates acoustic energy into the surrounding tissue and detects the subsequent ultrasonic echoes. The information from the echoes is used to generate real-time images of the coronary or peripheral vessels. All of the EEP line of catheters utilizes an internal lumen that allows the catheters to track over a 0.014" (0.36 mm) guide wire. The guide wire exits from the guide wire lumen approximately 24 cm proximal to the catheter tip. The EEP line catheters are introduced either percutaneously or via surgical cut down into the vascular system. The EEP line of catheters may be used with the In-Vision Imaging System, Volcano s5™, Volcano s5i™, Volcano CORE Mobile, and Volcano CORE imaging systems. The catheters are designed to work with Volcano VH IVUS system software v1.2 or higher. This catheter will not operate if connected to any other imaging system. # INDICATIONS FOR USE: The Eagle Eye® Platinum (EEP) Digital IVUS Catheter Line, including the Eagle Eye Platinum catheter, Eagle Eye Platinum ST catheter and the PV.014P catheter, are designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. These devices are not currently indicated for use in the cerebral vessels. The Eagle Eye® Platinum (EEP) Digital IVUS Catheter Line is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. # COMPARISON OF TECHNOLOGICAL CHARACTERISICS: The modification made to the EEP line of catheters (change in hydrophilic coating) does not affect the intended use of the device or technologies included as part of the device product line and it does not alter the fundamental scientific technologies. The indications for use are identical to those of the currently marketed devices (EEP; K092596 and EEP-ST K120697). The modified catheters are substantially equivalently to currently marketed predicate devices. {5}------------------------------------------------ ## PERFORMANCE DATA: Non-clinical device testing was conducted to confirm the performance of the modified device. Bench testing was conducted against known standards or product specification and evaluated the following: - Friction Force - Coating Adhesion - . Particulate Generation All bench testing was successfully completed. Biocompatibility testing was conducted on the device and the following tests were successfully completed: - . Cytotoxicity - Sensitization - Intracutaneous Reactivity - Systemic Toxicity - . Pyrogenicity - ASTM Hemolysis - Partial Thromboplastin Time - In vivo Thromboresistance - C3a Complement Activation - SC5-b Complement Activation - Bacterial Endotoxins (LAL) - LEAP Latex Elisa for Antigenic Protein ● - Platelet and Leukocyte Counts ● - . Genotoxicity The successful completion of performance testing and biocompatibility testing concluded that the modified Eagle Eye Platinum catheters are substantially equivalent to the currently marketed Eagle Eye Platinum catheters.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...