Visions® PV .014P RX Digital IVUS Catheter ; Visions® PV .018 Digital IVUS Catheter

K253399 · Volcano Corporation (Dba Philips Image Guided Therapy Device · OBJ · Oct 28, 2025 · Cardiovascular

Device Facts

Record IDK253399
Device NameVisions® PV .014P RX Digital IVUS Catheter ; Visions® PV .018 Digital IVUS Catheter
ApplicantVolcano Corporation (Dba Philips Image Guided Therapy Device
Product CodeOBJ · Cardiovascular
Decision DateOct 28, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1200
Device ClassClass 2

Intended Use

Visions® PV.014P RX Digital IVUS Catheter: The Visions® PV.014P RX Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in the cerebral vessels. The Visions® PV.014P RX Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. Visions® PV .018 Digital IVUS Catheter: The Visions® PV .018 Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in cerebral vessels. The Visions® PV .018 Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Device Story

Intravascular ultrasound (IVUS) catheters; cylindrical ultrasound transducer array at distal tip; radiates acoustic energy and detects ultrasonic echoes; generates real-time cross-sectional images of coronary/peripheral vessels. Used in clinical settings (cath lab) by physicians; tracks over 0.018" guidewire; introduced percutaneously or via surgical cut down. Requires connection to specific Volcano imaging systems (s5, s5i, CORE Mobile, CORE); standalone operation with system. Output provides visual vessel morphology to assist clinical decision-making during angiographic procedures; benefits patient by providing detailed lumen/wall structure visualization.

Clinical Evidence

No clinical data or performance testing was completed to support this 510(k) submission.

Technological Characteristics

Cylindrical ultrasound transducer array; internal lumen for 0.018" guidewire tracking; compatible with Volcano imaging systems. No changes to materials or design compared to predicates.

Indications for Use

Indicated for evaluation of vascular morphology in coronary and peripheral blood vessels via cross-sectional imaging. Adjunct to conventional angiography for imaging vessel lumen and wall structures. Contraindicated for use in cerebral vessels.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 28, 2025 Volcano Corporation (dba Philips Image Guided Therapy Device) Aaron Niklaus Sr. Regulatory Affairs Specialist 3721 Valley Centre Drive, Suite 500 San Diego, California 92130 Re: K253399 Trade/Device Name: Visions® PV .014P RX Digital IVUS Catheter; Visions® PV .018 Digital IVUS Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, ITX Dated: September 30, 2025 Received: September 30, 2025 Dear Aaron Niklaus: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K253399 - Aaron Niklaus Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253399 - Aaron Niklaus Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MARCO CANNELLA -S for - Aneesh Deoras - Assistant Director - Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices - Office of Cardiovascular Devices - Office of Product Evaluation and Quality - Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253399 | | | Device Name Visions® PV .014P RX Digital IVUS Catheter; Visions® PV .018 Digital IVUS Catheter | | | Indications for Use (Describe) Visions® PV.014P RX Digital IVUS Catheter: The Visions® PV.014P RX Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in the cerebral vessels. The Visions® PV.014P RX Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. | | | Visions® PV .018 Digital IVUS Catheter: The Visions® PV .018 Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in cerebral vessels. The Visions® PV .018 Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} PHILIPS K253399 # PV Digital IVUS Catheter 510(k) Summary K253399 Sponsor: Volcano Corporation (dba Philips Image Guided Therapy Devices) Valley Centre Dr #500, San Diego, CA 92130 Contact/Submitter: Aaron Niklaus Senior Regulatory Affairs Specialist Valley Centre Dr #500, aaron.niklaus@philips.com 651-968-7960 Date of Submission: 09-30-2025 ## Subject Device 1 Trade Name: Visions® PV.014P RX Digital IVUS Catheter Common Name: Diagnostic Intravascular Catheter Diagnostic Ultrasound Transducer Class: 2 CFR Numbers: 21 CFR 870.1200 Diagnostic Intravascular Catheter 21 CFR 892.1570 Diagnostic Ultrasound Transducer Product Codes: OBJ, ITX Panel: Cardiovascular ## Subject Device 2 Trade Name: Visions PV .018 Digital IVUS Catheter Common Name: Diagnostic Intravascular Catheter Diagnostic Ultrasound Transducer Class: 2 CFR Numbers: 21 CFR 870.1200 Diagnostic Intravascular Catheter 21 CFR 892.1570 Diagnostic Ultrasound Transducer Product Codes: OBJ, ITX Panel: Cardiovascular Page 1 of 4 {5} PHILIPS K253399 # Predicate Device 1: Trade Name Visions® PV .014P RX Digital IVUS Catheter Common Name Diagnostic Intravascular Catheter Diagnostic Ultrasound Transducer Class 2 CFR Numbers 21 CFR 870.1200 Diagnostic Intravascular Catheter 21 CFR 892.1570 Diagnostic Ultrasound Transducer Product Codes OBJ, ITX Panel: Cardiovascular 510(k) Number K152829 Clearance Date 11/19/2015 # Predicate Device 2: Trade Name Visions PV .018 Digital IVUS Catheter Common Name Diagnostic Intravascular Catheter Diagnostic Ultrasound Transducer Class 2 CFR Numbers 21 CFR 870.1200 Diagnostic Intravascular Catheter 21 CFR 892.1570 Diagnostic Ultrasound Transducer Product Codes OBJ, ITX Panel: Cardiovascular 510(k) Number K150442 Clearance Date 09/04/2015 # Device Description: The Visions® PV .014P RX Digital IVUS Catheter and the Visions® PV .018 PV .018 Digital IVUS Catheter are being bundled into one submission per FDA Guidance for Industry: Bundling Multiple Devices or Multiple Indications in a Single Submission, June 2007 (FDA-2003-D-0376), as the two catheters share the same indications for use, FDA classification product codes, and technological characteristics, and part of the same product line of catheters. Page 2 of 4 {6} PHILIPS K253399 # Visions® PV .014P RX Digital IVUS Catheter The Visions® PV .014P RX Digital IVUS Catheter incorporates a cylindrical ultrasound transducer array located near the distal tip of the catheter. The array radiates acoustic energy into the surrounding tissue and detects the subsequent ultrasonic echoes. The information from the echoes is used to generate real-time images of the coronary or peripheral vessels. The Visions® PV .014P RX Digital IVUS Catheter utilizes an internal lumen that allows the catheters to track over a 0.018" (0.46 mm) guide wire. The guide wire exits from the guide wire lumen approximately 31 cm proximal to the catheter tip. The PV .018 catheters are introduced either percutaneously or via surgical cut down into the vascular system. The Visions® PV .014P RX Digital IVUS Catheter may only be used with Volcano imaging systems, such as the Volcano s5™, Volcano s5i™, Volcano CORE Mobile, and Volcano CORE imaging systems. The catheter will not operate if connected to any other imaging system. # Visions® PV .018 Digital IVUS Catheter The Visions® PV .018 Digital IVUS Catheter incorporates a cylindrical ultrasound transducer array located near the distal tip of the catheter. The array radiates acoustic energy into the surrounding tissue and detects the subsequent ultrasonic echoes. The information from the echoes is used to generate real-time images of the coronary or peripheral vessels. The Visions® PV .018 Digital IVUS Catheter utilizes an internal lumen that allows the catheters to track over a 0.018" (0.46 mm) guide wire. The guide wire exits from the guide wire lumen approximately 31 cm proximal to the catheter tip. The PV .018 catheters are introduced either percutaneously or via surgical cut down into the vascular system. The Visions® PV .018 Digital IVUS Catheter may only be used with Volcano imaging systems, such as the Volcano s5™, Volcano s5i™, Volcano CORE Mobile, and Volcano CORE imaging systems. The catheter will not operate if connected to any other imaging system. # Indications For Use: ## Visions® PV .014P RX Digital IVUS Catheter The Visions® PV.014P RX Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a cross-sectional image of such vessels. This device is not currently indicated for use in the cerebral vessels. The Visions® PV.014P RX Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. ## Visions® PV .018 Digital IVUS Catheter The Visions® PV .018 Digital IVUS Catheter is designed for use in the evaluation of vascular morphology in blood vessels of the coronary and peripheral vasculature by providing a crosssectional image of such vessels. This device is not currently indicated for use in cerebral vessels. The Visions® PV .018 Digital IVUS Catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. Page 3 of 4 {7} PHILIPS K253399 # Substantial Equivalence ## Intended Use The Intended Use of the Subject Devices are the same as their respective Predicate Devices. ## Technological Characteristics The Subject Devices have the same technological characteristics as their respective Predicate Devices with no significant updates to device design, specification, or performance. The Instructions for Use (IFU) have been updated to add a new Precaution. This Precaution is the result of cases where physicians, during "radial-to-peripheral" procedures (radial access), have used a Digital IVUS Catheter without the appropriate sheath and/or guide catheter; this inappropriate use could lead to failure to provide adequate support for maneuvering the guidewire and catheter to peripheral vessels. The Precaution is intended to ensure that physicians use a guide sheath of appropriate length to prevent inadequate support for maneuvering the guidewire and catheter to peripheral vessels. The modifications made to the Visions® PV .014P RX Digital IVUS Catheter and Visions® PV .018 Digital IVUS Catheter (do not affect the intended use of the device or technological characteristics. The indications for use are identical to those of the currently marketed predicate devices (Visions® PV .014P RX Digital IVUS Catheter and Visions PV.018 Digital IVUS Catheter cleared under K150442). The Subject Devices are substantially equivalent to their respective legally marketed Predicate Devices. ## Performance Data No performance testing was completed to support this 510(k) Submission. ## Conclusion Both Subject Devices have the same Intended Use and Technological Characteristics as their respective Predicate Devices, demonstrating Substantial Equivalence. Page 4 of 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...