SOUNDSTAR 3D DIAGNOSTIC CATHETER, MODEL M-5723-05

K091352 · Biosense Webster, Inc. · OBJ · Dec 4, 2009 · Cardiovascular

Device Facts

Record IDK091352
Device NameSOUNDSTAR 3D DIAGNOSTIC CATHETER, MODEL M-5723-05
ApplicantBiosense Webster, Inc.
Product CodeOBJ · Cardiovascular
Decision DateDec 4, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1200
Device ClassClass 2

Intended Use

The Biosense Webster SOUNDSTAR™ 3D Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® EP Navigation Systems, the SOUNDSTAR™ 3D Catheter provides location information.

Device Story

The SOUNDSTAR™ 3D Ultrasound Catheter is a 10F, 90 cm intracardiac echo (ICE) catheter featuring an acoustic array and a location sensor at the tip. It connects to both an ultrasound system (via Swiftlink cable) and a CARTO® EP Navigation System (via Hypertronic connector). The device acquires real-time ultrasound images while simultaneously providing location data to the navigation system. This enables the integration of ultrasound imagery with electromagnetic-acquired cardiac maps. Used by clinicians in a cardiac setting, the device assists in visualizing anatomy and other devices within the heart, facilitating clinical decision-making during electrophysiology procedures. The modification in this submission solely updates labeling to support compatibility with additional CARTO® EP Navigation Systems; the physical device remains identical to the predicate.

Clinical Evidence

No clinical data provided. Bench testing only; previous testing for the predicate device remains applicable as the device design and performance are unchanged.

Technological Characteristics

10F, 90 cm ICE catheter; acoustic array identical to ACUSON AcuNav™; integrated location sensor for electromagnetic navigation; bifurcated tail with SOUNDSTAR tab and CARTO® Hypertronic connectors. Connectivity: compatible with Sequoia/Cypress Ultrasound Systems and CARTO® EP Navigation Systems.

Indications for Use

Indicated for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy, physiology, and other devices within the heart. No specific age or gender restrictions provided.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # DEC - 4 zuug Image /page/0/Picture/1 description: The image shows handwritten text. The text at the top reads "K091352". Below that, the text reads "pg. 1 of 3". The text appears to be a page number or identifier within a larger document. # 11 510(k) Summary ### 11.1 SPONSOR'S NAME & ADDRESS Biosense Webster, Inc. 3333 Diamond Canyon Road Diamond Bar, CA 91765 ## 11.2 Official Correspondent Balaka Das Senior Specialist, Regulatory Affairs Phone: 909-839-8599 Fax: 909-839-8804 Email: bdas2@its.jnj.com - 11.3 SUBMISSION DATE May 6, 2009 ## 11.4 TRADE NAME SOUNDSTAR™ 3D Ultrasound Catheter ### 11.5 COMMON NAME Electrophysiology Mapping/Ultrasound Catheter - CLASSIFICATION NAME/PRODUCT CODE 11.6 Intravascular Ultrasound Catheter/OBJ ## 11.7 CLASSIFICATION Class II - 11.8 PREDICATE DEVICE SOUNDSTAR™ 3D Ultrasound Catheter (K070242) cleared on May 15, 2007. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows handwritten text. The first line reads 'K0913552'. The second line reads 'pg 2 of 3'. The text appears to be a page number or identifier within a larger document. #### 11.9 DESCRIPTION OF DEVICE The Biosense Webster SOUNDSTAR" 3D Ultrasound Catheter is a 90 cm 10F IntraCardiac Echo (ICE) Catheter with an acoustic array identical to the ACUSON AcuNav" 10F Diagnostic Ultrasound Catheter. The catheter has a location sensor (providing location information to CARTO® EP Navigation Systems with ultrasound capability) and an ultrasound transducer (acquiring real time ultrasound images) embedded in the tip. The SOUNDSTAR™ 3D Ultrasound Catheter has a bifurcated 'tail' originating from its handle. One leg terminates in the SOUNDSTAR tab connector, which connects via a Swiftlink" cable to an ultrasound system. The other leg terminates in the CARTO® Hypertronic connector, which connects via a Patient Interface Unit (PIU) extension cable to the CARTO® Navigation System. The SOUNDSTAR™ 3D Ultrasound Catheter, when connected to a CARTO® EP Navigation System with ultrasound capability, and the Sequoia or the Cypress Ultrasound Systems, will provide real-time integration of ultrasound images with CARTO electromagnetic acquired maps. #### 11.10 INDICATIONS FOR USE The Biosense Webster SOUNDSTARTM 3D Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® EP Navigation Systems, the SOUNDSTAR" 3D Catheter provides location information. #### DESCRIPTION OF MODIFICATION 11.11 The modified SOUNDSTAR" 3D Ultrasound Catheter is physically identical to the predicate device in terms of design, manufacturing methods, materials and performance. There are absolutely no changes to the device whatsoever. The only modification is to the labeling for the device to allow compatibility with multiple CARTO® EP Navigation Systems with ultrasound capability. #### SUMMARY OF NONCLINICAL TESTS 11.12 All testing previously submitted for the SOUNDSTAR™ 3D Ultrasound Catheter still applies to the modified device as there were no changes to the design, materials, manufacturing methods or performance of the {2}------------------------------------------------ K091352 pg 3 of 3 device. Testing for compatibility with the CARTO® 3 EP Navigation Systems was submitted in the CARTO® 3 System 510(k). Therefore, no additional testing is submitted in this Premarket Notification. #### 11.13 SUBSTANTIAL EQUIVALENCE The modified SOUNDSTAR 3D Ultrasound Catheter is identical to the previously cleared SOUNDSTAR™ 3D Ultrasound Catheter in that the devices: - have the same intended use, . - use the same operating principle, . - use the same fundamental scientific technology, ◆ - incorporate the same design, ● - incorporate the same materials and construction, . - have the same shelf life, and . - are packaged using the same materials and processes. . In summary, the SOUNDSTAR™ 3D Ultrasound Catheter described in this submission is substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird symbol, composed of three curved lines, representing the department's mission of promoting health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. The logo is in black and white. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 # DEC 0 4 2009 Biosense Webster, Inc. Balaka Das, Senior Specialist, Regulatory Affairs 3333 Diamond Canyon Rd. Diamond Bar, CA 91765 Re: K091352 Trade/Device Name: SoundStar 3D Ultrasound Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ Dated: November 3, 2009 Received: November 5, 2009 Dear Ms. Das: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Ms. Balaka Das device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K091352 Device Name: Soundstar 3D Ultrasound Catheter Indications For Use: The Biosense Webster SOUNDSTAR™ 3D Diagnostic Ultrasound Catheter and related accessory devices are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible CARTO® EP Navigation Systems, the SOUNDSTAR™ 3D Catheter provides location information. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRN Office of Device Evaluation (ODE) | | |--------------------------------------------------------|-------------| | (Division Sign-Off) Division of Cardiovascular Devices | Page 1 of 1 | | 510(k) Number | K091352 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...