Disposable Blood Pressure Cuff

K222420 · Wenzhou Xikang Medical Instruments Co., Ltd. · DXQ · Nov 21, 2022 · Cardiovascular

Device Facts

Record IDK222420
Device NameDisposable Blood Pressure Cuff
ApplicantWenzhou Xikang Medical Instruments Co., Ltd.
Product CodeDXQ · Cardiovascular
Decision DateNov 21, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1120
Device ClassClass 2
AttributesPediatric

Intended Use

The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Device Story

Disposable blood pressure cuff; accessory for noninvasive blood pressure measurement systems. Consists of cuff with bladder and air hose (single or double tube). Wrapped around patient limb; secured via hook and loop closure. Tubing connects to external blood pressure monitor. Used in clinical settings; operated by healthcare professionals. Provides pneumatic interface for blood pressure measurement; facilitates accurate pressure transmission from monitor to patient limb. Single-patient use; non-sterile. Benefits include infection control via disposability and compatibility with standard monitoring equipment.

Clinical Evidence

No clinical data. Bench testing only. Compliance with ISO 81060-1 (non-invasive sphygmomanometers) and ISO 10993 (biocompatibility: cytotoxicity, sensitization, irritation) demonstrated.

Technological Characteristics

Materials: PVC film, non-woven fabric, nylon hook/loop tape. Sensing: Pneumatic bladder. Dimensions: 22 models covering neonatal to thigh sizes. Connectivity: Mechanical interface (single/double tube) to non-invasive blood pressure monitors. Sterilization: Non-sterile. Standards: ISO 81060-1, ISO 10993-1, ISO 10993-5, ISO 10993-10.

Indications for Use

Indicated for use as an accessory with noninvasive blood pressure measurement systems for neonate, infant, child, and adult patient populations.

Regulatory Classification

Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable. November 21, 2022 Wenzhou Xikang Medical Instruments Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161,East Lujiazui Rd.,Pudong Shanghai, Shanghai 200120 China Re: K222420 Trade/Device Name: Disposable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: October 25, 2022 Received: October 25, 2022 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Stephen C. Browning -S LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K222420 Device Name Disposable Blood Pressure Cuff #### Indications for Use (Describe) The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. Type of Use (Select one or both, as applicable) | <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) | |---------------|----------------------------------------------| | <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92. ### 1.0 Submitter's Information | Name: | Wenzhou Xikang Medical Instruments Co., Ltd. | |----------|-------------------------------------------------------------------------------------------| | Address: | No. 1478 Haining Road, Haibin Street, Longwan District<br>Wenzhou, Zhejiang 325024, China | | Tel: | +86-577-86876969 | | Contact: | Wang Tiance | #### Designated Submission Correspondent | Contact: | Mr. Boyle Wang | |----------|------------------------------------------------------------------------| | Name: | Shanghai Truthful Information Technology Co., Ltd. | | Address: | Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai,<br>200120 China | | Tel: | +86-21-50313932 | | Email: | Info@truthful.com.cn | Date Submitted: Aug.4th,2022 #### 2.0 Device Information | Trade/Device name: | Disposable Blood Pressure Cuff | |------------------------------|----------------------------------| | Common name: | Non-invasive blood pressure cuff | | Classification name: | Blood Pressure Cuff | | Classification Product Code: | DXQ | | Regulation number: | 870.1120 | | Classification: | Class II | | Panel: | Cardiovascular | #### 3.0 Predicate Device Information ### Predicate Device: | Manufacturer: | Shenzhen Caremed Medical Technology Co., Ltd. | |--------------------|---------------------------------------------------------------------------------| | Trade/Device Name: | Caremed Reusable Blood Pressure Cuff, Caremed<br>Disposable Blood Pressure Cuff | {4}------------------------------------------------ 510(k) number: K182433 ## 4.0 Device Description The subject device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes. The subject device contains Cuff with bladder and Air Hose. Air hose has single tube and double tube. Wrapped the cuff around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. The device is single use device, and which is made of non-woven fabrics & PVC (Cuff) and PVC (Air Hose), thereinto, the non-woven fabrics & PVC (Cuff) is the material used to contacting with the patient. The disposable blood pressure cuff has 22 models with different size for different population with different arm size, and the 11 models of 22 are use single tube of air hose and other 11 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system. ### 5.0 Indication for Use Statement The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. {5}------------------------------------------------ ## 6.0 Technological Characteristic Comparison Table | Item | Subject Device | Predicate Device | Comparison<br>Analysis | |----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | Product Name | Disposable Blood<br>Pressure Cuff | Caremed Disposable<br>Blood Pressure Cuff | ---- | | 510(k) No. | Pending | K182433 | ---- | | Product Code | DXQ | DXQ | Same | | Regulation No. | 21 CFR 870.1120 | 21 CFR 870.1120 | Same | | Class | II | II | Same | | Intended<br>Use/Indication<br>for Use | The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. | The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.<br>The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. | Same with the<br>Indication for<br>Use of the<br>Disposable<br>Blood Pressure<br>Cuff | | Material<br>Composition | Cuff (Patient contacted): PVC film<br>stick non-woven fabric;<br>Bladder:PVC<br>Air Hose:PVC<br>Hook tape: Nylon<br>Loop: Nylon | Non-woven (Patient contacted);<br>nylon;<br>PVC | Same | | Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | Same | | Tube Number | One or two | One or two | Same | | Limb<br>Circumference<br>(Range in cm) | Conform to AHA bladder sizes<br>recommendations | Conform to AHA bladder sizes<br>recommendations | | | | Neonatal 1 (3-6 cm) | Neonatal 1 (3-6 cm) | Different | | | Neonatal 2 (4-8 cm) | Neonatal 2 (4-8 cm) | | | | Neonatal 3 (6-11 cm) | Neonatal 3 (6-11 cm) | | | | Neonatal 4 (7-13 cm) | Neonatal 4 (7-13 cm) | | | | Neonatal 5 (8-15 cm) | Neonatal 5 (8-15 cm) | | | | Infant (9-15 cm) | Infant (9-14.8 cm) | | | | Child (13-20 cm) | Child (13.8-21.5 cm) | | | | Small Adult (18-26 cm) | Small Adult (20.5-28.5 cm) | | | | Adult (25-35 cm) | Adult (27.5-36.5 cm) | | | | Large Adult (32-42 cm) | Adult Long (27.5-36.5/46.5 cm) | | | | Thigh (42-50 cm) | Large Adult (35.5-46 cm) | | | | | Large Adult Long (35.5-46 cm) | | | Pressure Range | 0-300mmHg | 0-300mmHg | Same | | Max. Pressure | 400mmHg | 400mmHg | Same | | Sterility | Non-sterile | Non-sterile | Same | | Biocompatibility | Conform with ISO10993-1<br>(ISO10993-5, ISO10993-10) | No potential cytotoxicity;<br>No sensitization observed (test sample score 0);<br>Negligible (no observed primary irritation, test sample score 0) | Same | {6}------------------------------------------------ The Subject Device has the same Intended Use/Indication for Use, basic construction, and technology specification as the predicated device. Both devices are wrapped the patient's arm or leg and secured by a hook and loop fastener commonly called Velcro. The materials of both devices are all conformed to ISO 10993. Both devices are available in the similar size and range and are intended for the same patient populations, and the size for Infant, Child, Small Adult and Large Adult are little different. Based on the performance testing in this submission, the slight difference on the range of these blood pressure cuffs does not raise any safety or effectiveness issue. #### 7.0 Summary of Non-clinical Testing Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007. ISO 10993-5: 2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity. ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for {7}------------------------------------------------ Irritation and Skin Sensitization. #### Biocompatibility testing The biocompatibility evaluation for the Disposable Blood Pressure Cuffs was conducted in accordance with the FDA Biocompatibility guidance, 2016 (Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process") and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: Cytotoxicity Sensitization Irritation ### 8.0 _Summary of Clinical Testing No clinical study is included in this submission. ### 9.0 Conclusion The conclusions drawn from the comparison and analysis above demonstrate that the proposed device is as safe, as effective, and performs as well as the legally marketed predicated device in K182433 and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...