The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
Device Story
Caremed Blood Pressure Cuffs (reusable and disposable) function as accessories for noninvasive blood pressure measurement systems. The device consists of a cuff with an internal bladder and air hose (single or double tube). Operation involves wrapping the cuff around a patient's limb and securing it via hook-and-loop closure; the air hose connects to a blood pressure monitor. The monitor inflates the bladder to occlude the artery and measures pressure oscillations to determine blood pressure. Used in clinical settings by healthcare providers. The output (blood pressure readings) is displayed on the host monitor to assist clinicians in patient hemodynamic assessment and monitoring. Benefits include non-invasive, reliable blood pressure measurement across various patient sizes.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-5 and ISO 10993-10) and performance verification per ISO 81060-1 (non-invasive sphygmomanometer requirements). No clinical data was required or provided.
Technological Characteristics
Materials: Reusable (PU/Synthetic Leather, TPU bladder, PVC tubing, Nylon hook/loop); Disposable (Non-woven fabric, PVC, Nylon). Sensing: Pneumatic bladder inflation/oscillation. Connectivity: Mechanical connection to host monitor via air hose. Sterilization: Non-sterile. Dimensions: Various sizes (neonate to adult) conforming to AHA bladder recommendations. Pressure range: 0-300 mmHg.
Indications for Use
Indicated for use as an accessory with noninvasive blood pressure measurement systems for neonate, infant, child, and adult patient populations.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
Predicate Devices
Unimed Medical Supplies Inc. Reusable Blood Pressure Cuff (K112544)
Unimed Medical Supplies Inc. Disposable Blood Pressure Cuff (K120364)
Related Devices
K140874 — GH DISPOSABLE AND GH REUSABLE BLOOD PRESSURE CUFF · Shanghai Goahui Rubber and Plastic Company · Sep 2, 2014
K090664 — DEVON ACCUBP CUFF · Devon Medical, Inc. · Mar 23, 2009
K231961 — NIBPCuff · Shenzhen Sino-K Medical Technology Co.,Ltd · Aug 30, 2023
K233276 — Reusable Blood Pressure Cuff · Suzhou Minhua Medical Apparatus Supplies Co., Ltd. · Dec 21, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two distinct elements: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.
December 17, 2018
Shenzhen Caremed Medical Technology Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A. Nanshan Medical Devices Industrial Park Nanshan District Shenzhen, 518067 Cn
Re: K182433
Trade/Device Name: Caremed Reusable Blood Pressure Cuff, Caremed Disposable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: November 26, 2018 Received: November 26, 2018
Dear Kevin Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Stephen C. Browning -S5
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K182433
### Device Name
Caremed Reusable Blood Pressure Cuff, Caremed Disposable Blood Pressure Cuff
### Indications for Use (Describe)
The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 5px;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
### Prepared in accordance with the requirements of 21 CFR Part 807.92
### Prepared Date: 2018/08/31
#### 1. Submission sponsor
Name: Shenzhen Caremed Medical Technology Co., Ltd. Address: East Side, 3/F, C Building, Kelunte Low-Carbon Industries, Gaofeng Community, Dalang Office, Longhua District, Shenzhen, Guangdong, China Contact person: Alan Xie Title: Quality Manager E-mail: cm003@szcaremed.com Tel: +86-755-36560977-889
#### 2. Submission correspondent
Name: Chonconn Medical Device Consulting Co., Ltd.
Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen, Guangdong, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin(@chonconn.com Tel: +86-755 33941160
| Trade/Device Name | Caremed Reusable Blood Pressure Cuff,<br>Caremed Disposable Blood Pressure Cuff |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | Caremed Reusable Blood Pressure Cuff: N1A-PS, N1N-PS, N1N-PD, N1I-PS,<br>N1I-PD, N1C-PS, N1C-PD, N1S-PS, N1S-PD, N1A-PD, N1AL-PS, N1AL-PD,<br>N1L-PS, N1L-PD, N1T-PS, N1T-PD. |
| | Caremed Disposable Blood Pressure Cuff: N1A-TS, N1LL-TS, N1LL-TD, N1L-<br>TS, N1L-TD, N1AL-TS, N1AL-TD, N1A-TD, N1S-TS, N1S-TD, N1C-TS,<br>N1C-TD, N1I-TS, N1I-TD, N1N1-TS, N1N1-TD, N1N2-TS, N1N2-TD, N1N3-<br>TS, N1N3-TD, N1N4-TS, N1N4-TD, N1N5-TS, N1N5-TD |
| Common Name | Non-invasive Blood Pressure Cuff |
| Regulatory Class | Class II |
| Classification | 21CFR 870.1120 / Blood pressure cuff / DXQ |
| Submission type | Traditional 510(K) |
#### 3. Subject Device Information
#### 4. Predicate Device
By submission of the Traditional 510(k), Shenzhen Caremed Medical Technology Co., Ltd. is requesting clearance for Reusable Blood Pressure Cuff and Disposable Blood Pressure Cuff. It is comparable to the following legally marketed system:
{4}------------------------------------------------
- 1. Unimed Medical Supplies Inc. Reusable Blood Pressure Cuff under K112544.
- 2. Unimed Medical Supplies Inc. Disposable Blood Pressure Cuff under K120364.
The subject device has same intended use, same target patient population, same performance effectiveness, performance safety as the predicate devices and no question is raised regarding to effectiveness and safety. So, the conclusion is that the subject device is substantial equivalent to the predicate.
#### 5. Device Description
The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes.
The proposed device includes disposable blood pressure cuff and reusable blood pressure cuff, both disposable and reusable blood pressure cuff has same structure, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube.
The disposable blood pressure cuff is single use device, and which is made of non-woven fabrics & PVC (Cuff) and PVC (Air Hose), the non-woven fabrics & PVC (Cuff) is the material used to contacting with the patient.
The disposable blood pressure cuff has 24 models with different population with different arm size, and the 12 models of 24 are use single tube of air hose and other 12 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.
The reusable blood pressure cuff is reusable device, and which is made of PU & PVC (Cuff) and PVC (Air Hose), thereinto, the PU & PVC (Cuff) is the material used to contacting with the patient.
The reusable blood pressure cuff has 16 models with different population with different arm size, and the 8 models of 16 are use single tube of air hose and other 8 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.
#### Intended use & Indication for use 6.
The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.
#### 7. Comparison to the Predicate Device
The following tabulation indicates the detailed differences between the subject devices and the predicate devices.
{5}------------------------------------------------
| Features | Subject Device<br>Caremed Reusable Blood<br>Pressure Cuff | Predicate Device K112544<br>Unimed Reusable Blood<br>Pressure Cuff | Remark | | |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|------------------------------|------|
| | Applicant | Shenzhen Caremed Medical<br>Technology Co., Ltd. | | Unimed Medical supplies Inc. | / |
| | Classification<br>Regulation | 21CRF 870.1120 | | 21CRF 870.1120 | Same |
| Classification<br>and Code | Class II,<br>DXQ | Class II,<br>DXQ | Same | | |
| Common<br>name | Non-invasive Blood Pressure<br>Cuff | Non-invasive Blood Pressure<br>Cuff | Same | | |
| Intended use | The reusable blood pressure cuff is<br>an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and may be reused. It is<br>available in neonate, infant, child<br>and adult sizes. The cuff is not<br>designed, sold, or intended for use<br>except as indicated. | The Unimed Blood Pressure Cuff<br>is an accessory used in<br>conjunction with non-invasive<br>blood pressure measurement<br>systems. The cuff is non-sterile<br>and may be reused. It is available<br>in pediatric and adult sizes. The<br>cuff is not designed, sold, or<br>intended for use except as<br>indicated. | Different<br>(1) | | |
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | Same | | |
| Tube Number | One or two | One or two | Same | | |
| Principles of<br>Operation | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems | Bladder is wrapped around the<br>patient's limb and secured by<br>hook and loop closure Air hose is<br>connected to the noninvasive<br>blood pressure measurement<br>systems | Same | | |
| Limb<br>Circumference<br>(Range in cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11cm)<br>Infant (10-19cm)<br>Child (18-26cm)<br>Small Adult (20-28cm)<br>Adult (25-35cm)<br>Adult Long (25-35cm)<br>Large Adult (33-47cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11cm)<br>Infant (10-19cm)<br>Pediatric (18-26cm)<br>Small Adult (20-28cm)<br>Adult (25-35cm)<br>Adult Long (25-35cm)<br>Large Adult (33-47cm) | Different<br>(2) | | |
| Features | Subject Device | Predicate Device K112544 | Remark | | |
| | Caremed Reusable Blood<br>Pressure Cuff | Unimed Reusable Blood<br>Pressure Cuff | | | |
| | Adult Thigh (44-66cm) | Large Adult Long (33-47cm)<br>Adult Thigh (44-66cm) | | | |
| Pressure Range | 0-300 mmHg | 0-300 mmHg | Same | | |
| Sterility | Non-sterile | Non-sterile | Same | | |
| Max. Leakage | < 4mm Hg/ min. | < 4mm Hg/ min. | Same | | |
| Material | Cuff (Patient contacted): PU<br>Synthetic Leather<br>Bladder: Transparent Polyurethane<br>(TPU Film)<br>Tubing: PVC<br>Hook: Molded Nylon<br>Loop: Nylon | Cuff (Patient contacted): PU<br>Synthetic Leather<br>Bladder: Transparent<br>Polyurethane (TPU Film)<br>Tubing: PVC<br>Hook: Molded Nylon<br>Loop: Nylon | Same | | |
| Biocompatibility | No potential cytotoxicity;<br>No sensitization observed (test<br>sample score 0);<br>Negligible (no observed primary<br>irritation, test sample score 0) | Comply with ISO 10993-5;<br>Comply with ISO 10993-10 | Same | | |
Comparison to Primary Predicate Device Unimed Reusable Blood Pressure Cuff:
{6}------------------------------------------------
K182433
Page 4 of 7
# Comparison to Reference Device Unimed Disposable Blood Pressure Cuff:
| Features | Subject Device<br>Caremed Disposable Blood<br>Pressure Cuff | Predicate Device K120364<br>Unimed Disposable Blood<br>Pressure Cuff | Remark |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Applicant | Shenzhen Caremed Medical<br>Technology Co., Ltd. | Unimed Medical supplies Inc. | / |
| Intended use | The disposable blood pressure cuff<br>is an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and for single-patient<br>use. It is available in neonate,<br>infant, child and adult sizes. The<br>cuff is not designed, sold, or<br>intended for use except as<br>indicated. | The disposable blood pressure<br>cuff is an accessory used in<br>conjunction with noninvasive<br>blood pressure measurement<br>systems. The cuff is non-sterile<br>and for single-patient use. It is<br>available in neonatal, pediatric<br>and adult sizes. The cuff is not<br>designed, sold, or intended for<br>use except as indicated. | Different<br>(3) |
| Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | Same |
| Tube Number | One or two | One or two | Same |
{7}------------------------------------------------
| Features | Subject Device<br>Caremed Disposable Blood<br>Pressure Cuff | Predicate Device K120364<br>Unimed Disposable Blood<br>Pressure Cuff | Remark |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Limb<br>Circumference<br>(Range in cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal 1 (3-6 cm)<br>Neonatal 2 (4-8 cm)<br>Neonatal 3 (6-11 cm)<br>Neonatal 4 (7-13 cm)<br>Neonatal 5 (8-15 cm)<br>Infant (9-14.8 cm)<br>Child (13.8-21.5 cm)<br>Small Adult (20.5-28.5 cm)<br>Adult (27.5-36.5 cm)<br>Adult Long (27.5-36.5/46.5 cm)<br>Large Adult (35.5-46 cm)<br>Large Adult Long (35.5-46 cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal 1 (3-6 cm)<br>Neonatal 2 (4-8 cm)<br>Neonatal 3 (6-11 cm)<br>Neonatal 4 (7-13 cm)<br>Neonatal 5 (8-15 cm)<br>Infant (9-14.8 cm)<br>Child (13.8-21.5 cm)<br>Small Adult (20.5-28.5 cm)<br>Adult (27.5-36.5 cm)<br>Adult Long (27.5-36.5 cm)<br>Large Adult (35.5-46 cm)<br>Large Adult Long (35.5-46 cm)<br>Thigh (45-56 cm) | Different<br>(4) |
| Pressure Range | 0-300mmHg | 0-300mmHg | Same |
| Max. Pressure | 400mmHg | 400mmHg | Same |
| Sterility | Non-sterile | Non-sterile | Same |
| Material | Non-woven (Patient contacted);<br>nylon;<br>PVC | Non-woven (Patient contacted);<br>nylon;<br>PVC | Same |
| Biocompatibility | No potential cytotoxicity;<br>No sensitization observed (test<br>sample score 0);<br>Negligible (no observed primary<br>irritation, test sample score 0) | Comply with ISO 10993-5;<br>Comply with ISO 10993-10 | Same |
### Justification of differences:
Justifications for differences between Caremed Blood Pressure Cuff and the predicate device are shown as below:
Different (1): We claim the applicable population in more detail. Neonatal, infant and child are subsets of the pediatric patient population. Thus, this descriptive difference does not raise any safety or effectiveness issue.
Different (2): The limb circumference of subject device and predicate device is same. The subject device is less than predicate device with Large Adult Long. This difference limits use within the subject device and does not raise any safety or effectiveness issue.
Different (3): We claim the applicable population in more detail. Neonatal, infant and child are subsets of the pediatric patient population. Thus, this descriptive difference does not raise any safety or effectiveness
{8}------------------------------------------------
issue.
Different (4): The limb circumference of subject device and predicate device is same. The subject device is less than predicate device with Thigh. This difference limits use within the subject device and does not raise any safety or effectiveness issue.
### Conclusion:
The proposed devices share the same indications for use, device operation, overall technical and functional capabilities, meet the same standards and requirements and therefore, are substantially equivalent to the predicate device(s).
#### 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the Caremed Blood Pressure Cuffs was conducted in accordance with the FDA Biocompatibility guidance, 2016 (Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process") and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity .
- Sensitization .
- Irritation .
The Caremed Blood Pressure Cuffs are considered surface contacting for a duration of exceed 24 hours but not 30 days.
## Non-clinical data
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- . ISO 81060-1. Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007.
- . ISO 10993-5: 2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity.
- . ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
The test was selected to show substantial equivalence between the subject device and the predicate.
{9}------------------------------------------------
K182433 Page 7 of 7
#### Conclusion 9.
It has been shown in this 510(k) submission that the difference between the proposed devices and the predicate devices do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.