ALL PURPOSE CUFF, MODEL 222APC

K051904 · Suntech Medical, Inc. · DXQ · Sep 9, 2005 · Cardiovascular

Device Facts

Record IDK051904
Device NameALL PURPOSE CUFF, MODEL 222APC
ApplicantSuntech Medical, Inc.
Product CodeDXQ · Cardiovascular
Decision DateSep 9, 2005
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.1120
Device ClassClass 2
AttributesPediatric

Intended Use

The All Purpose Cuff is an accessory that is intended to be used with manual or automated noninvasive sphygmomanometers. It makes no diagnosis. The All Purpose Cuff is non-sterile and may be reused. It is available in pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated..

Device Story

All Purpose Cuff is a non-invasive blood pressure (NIBP) accessory; consists of connector, tubing, air-tight bladder, and inelastic sleeve. Used with manual or automated sphygmomanometers to measure blood pressure in pediatric and adult patients. Operated by clinicians in clinical settings. Device wraps around patient limb; secured via hook and loop fastener. Provides pneumatic interface between patient limb and blood pressure monitor; does not perform diagnosis. Reusable, non-sterile design.

Clinical Evidence

Bench testing only. The device was tested against AAMI SP10:2002, AAMI SP10:1994, AAMI SP9:1994, EN1060-1:1996, EN1060-2:1996, EN1060-3:1997, IEC60601-2-30:1999, and IEC60601-2-50:1999. All tests indicated passing results.

Technological Characteristics

Inelastic sleeve with internal air-tight bladder, tubing, and connector. Available in various sizes marked for limb circumference. Hook and loop fastener. Non-sterile, reusable. Complies with AAMI SP9/SP10, EN1060, and IEC60601-2-30/50 standards.

Indications for Use

Indicated for use with manual or automatic non-invasive sphygmomanometers to determine blood pressure parameters in pediatric and adult patients. Not for use on neonates. Intended for use with prescription (clinical grade) blood pressure monitors.

Regulatory Classification

Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SunTech Medical Section B K 051904 SEP – 9 2005 Section B 510(k) Summary {1}------------------------------------------------ SunTech Medical Page 2 of 4 Section B KDS 1904 SEP - 9 2005 SunTech Medical, Inc. SunTech Medical, Inc. Abbreviated 510(k) Submission All Purpose Cuff 510(k) Summary 27 May 2005 #### (1) Submitter information Name: Address: 507 Airport Boulevard Suite 117 Morrisville, North Carolina 27560-8200 Telephone: FAX: David Gallick (Official Correspondent). Contact person: > SunTech Medical, Inc. 507 Airport Boulevard Suite 117 Morrisville, North Carolina 27560-8200 Tel: 919-654-2332 Fax: 919-654-2301 Date prepared: 20 May 2005 919.654.2332 919.654.2301 #### (2) Name of Device Trade Name: Common Name: Classification name: All Purpose Cuff Blood Pressure Cuff Cuff, Blood Pressure, DXQ, 870.1120 {2}------------------------------------------------ #### (3) Legally-marketed predicate devices Sun Tech Medical has identified two predicate devices for the All Purpose Cuff: CRITIKON DURA-CUF® Blood Pressure Cuff, Critikon company, LLC. Cuff accessory for the Oscar 2 Oscillometric Ambulatory BP Recorder, SunTech Medical, Inc. The All Purpose Cuff is substantially equivalent to these devices. # (4) Description The All Purpose Cuff is an accessory for noninvasive blood pressure systems. It is comprised of a connector attached to tubing attached to an air tight bladder which is enclosed within an inelastic sleeve. The All Purpose Cuff comes in a range of sizes. Each is marked with the appropriate limb circumference for which the cuff is intended. # (5) Intended Use The All Purpose Cuff is an accessory that is intended to be used with manual or automated noninvasive sphygmomanometers. It makes no diagnosis. The All Purpose Cuff is non-sterile and may be reused. It is available in pediatric and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.. ### (6) Comparison to Predicate Devices The All Purpose Cuff has the same basic construction as the predicate devices. All three devices are wrapped around the patients limb and secured by means of a hook and loop type fastener. The All Purpose cuff is made from the same material as the Ocsar2 cuffs. The All Purpose cuff is available in the same size/ranges as the CRITIKON cuff and is intended for the same patient populations. {3}------------------------------------------------ # (7) Testing and Validations The All Purpose Cuff has been tested to the applicable requirements of the The All I urpose Curr has quirements documents. These tests have indicated passing results. - AAMI SP10: 2002 . - AAMI SP10:1994 . - AAMI SP9:1994 . - EN1060-1: 1996 ● - EN1060-2:1996 . - EN1060-3:1997 ● - IEC60601-2-30:1999 . - ILC00001-2 50:1999 Marketing Specification, (SunTech document # 97-0037-XX-FS), version 11 ● #### (8) Conclusion The All Purpose Cuff is equivalent in safety and efficacy to the legally-marketed predicate devices {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the caduceus. SEP - 9 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SunTech Medical, Inc. c/o Mr. David Gallick Vice President - Engineering 507 Airport Blvd., Suite 117 Morrisville, NC 27560-2301 Re: K051904 R051701 Trade Name: All Purpose Cuff, Model 222APC Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II (two) Product Code: DXQ Dated: July 12, 2005 Received: July 13, 2005 Dear Mr. Gallick: We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 510(x) premaited is substantially equivalent (for the indications referenced above and have determined the device is subsets devices marketed in referenced above and have determined the devices marketed in interstate for use stated in the enclosure) to legally marketed predicate devices Amendments for use stated in the enclosure) to regally manced produced Device Amendments, or to commerce prior to May 28, 1976, the enactment of the Federal Food. Drug. commerce prior to May 28, 1970, the chacmical with the provisions of the Federal Food, Drug, devices that have been reclassified in accordance who as provisions of the Act. The and Cosmetic Act (Act) that do not require approvate species provisions of the Act. The You may, therefore, market the device, subject to the general registration, listing You may, therefore, market the device, sucject to use grouper misbrandon, listing of general controls provisions of the Act include required it' ititing against misbranding a general controls provisions of the Act merate requirements misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mo cities and regulations affecting your device can may be subject to such additional controls. Existing major regulations affective may be subject to such additional controlis: Existing major regalato 898. In addition, FDA may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In be found in the Couc of Federal regenance.s, concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. David Gallick Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of a backer complies with other requirements of the Act that FDA has made a determination that your device complies with other requirem that FDA has made a decertification that your are four and agencies. You must or any Federal statutes and regulations and limited to: registration and listing (21 comply with an the Act 3 requirements, morealing practice requirements as set CFR Part 807), labeling (21 CFR Part 820); and if applicable, the electronic forum in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Decloils of Persons of Perior as described in your Section 510(k) I his letter will anow you to oegin maneting your antial equivalence of your device to a legally premarket nothication. The PDA miding of basification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac now on our and one and the regulation entitled, contact the Office of Compliance at (240) 276-0295. Also, please note on Comaci the Office of Compilance at (210) +16 cation" (21CFR Part 807.97) you may obtain. "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obta Other general information on your responsibilities under the Act may be obtained from the Other general information on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bhimmanfor Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): K051904 Device Name:_All Purpose Cuff; Model 222APC Indications For Use: The All Purpose Cuff is intended to be used with a manual or automatic non-invasive s and adult The All Purpose Cuff is intended to be used with a managers on pediatin and adult sphygmomanometer to determine blood pressure paremeters on pediating and adult sphygmomanometer to determine blood pressures . They are intended to be used patients. They are not intended to be used on neonates. They are intended to be used with patients. They are not intended to be used on neonators: They also could be used with with prescription (clinical grade) blood pressure monitors; they also could be used wi with prescription (Clinical grade) blood pressure monitors. Prescription Use __ X (Part 21 CFR 801 Subpart D) (AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bhumma (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...