Polaris X Unidirectional Steerable Diagnostic Catheter
Applicant
Boston Scientific Corporation
Product Code
DRF · Cardiovascular
Decision Date
Sep 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Intended Use
The catheter is intended for temporary use in electrophysiological studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
Device Story
Polaris X is a sterile, single-use, unidirectional steerable diagnostic catheter; used in electrophysiological studies for intracardiac pacing and recording of electrical potentials. Device comprises distal electrode segment (10 electrodes), shaft (6F diameter, 105cm length), and handle with thumb-slide steering control. Operator (physician) inserts catheter into patient; uses handle to actuate distal segment into 270° curve. Catheter connects to external recording/mapping systems via separate cable. Output consists of electrical signals captured from cardiac tissue; displayed on external systems for physician interpretation. Enables diagnosis of cardiac arrhythmias; facilitates clinical decision-making regarding treatment. Benefits patient through precise cardiac mapping and stimulation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1), ethylene oxide residuals (ISO 10993-7), bacterial endotoxin (ANSI/AAMI ST72), bioburden (ISO 11737-1), and design verification (mechanical integrity, steering lifecycle, and twisting).
Technological Characteristics
6F diameter, 105cm length, decapolar (10 electrode) configuration. Materials evaluated per ISO 10993-1. Thumb-slide steering mechanism for 270° curve. Sterilized via ethylene oxide. Compatible with standard electrophysiological recording/mapping systems via external cable. No software or electronic processing components.
Indications for Use
Indicated for patients requiring temporary electrophysiological stimulation (pacing) and/or recording of electrical potentials during electrophysiological studies.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
K091370 — RADIA STEERABLE DIAGNOSTIC CATHETER · C.R. Bard, Inc. · Jun 5, 2009
K051220 — REPROCESSED DEFLECTABLE ELECTROPHYSIOLOGY DIAGNOSTIC CATHETERS · Sterilmed, Inc. · Jun 14, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 14, 2021
Boston Scientific Corporation Donall Hosna Regulatory Affairs Senior Manager 4100 Hamline Ave North St. Paul, Minnesota 55112
Re: K211494
Trade/Device Name: Polaris X™ Unidirectional Steerable Diagnostic Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode recording catheter or electrode recording probe Regulatory Class: Class II Product Code: DRF Dated: August 10, 2021 Received: August 11, 2021
Dear Donall Hosna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh Deoras Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211494
### Device Name
Polaris XTM Unidirectional Steerable Diagnostic Catheter
Indications for Use (Describe)
The catheter is intended for temporary use in electrophysiological stimulation (pacing) and/or recording of electrical potentials.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## for
# K211494, Polaris X Unidirectional Steerable Diagnostic Catheter
#### I. Submitter
Boston Scientific Corporation Electrophysiology Division 4100 Hamline Ave North St. Paul, MN 55112
Contact: Donall Hosna, MS Regulatory Affairs Senior Manager Phone: 651-582-6323 E-mail: donall.hosna@bsci.com
Date Prepared: August 9, 2021
#### II. Device
Name of Device: Polaris XTM Unidirectional Steerable Diagnostic Catheter Common or Usual Name: Polaris X Diagnostic Catheter Classification Name: Electrode recording catheter or electrode recording probe (21 CFR 870.1220) Regulatory Class: II Product Code: DRF
#### III. Predicate Device
Polaris XTM Unidirectional Steerable Diagnostic Catheter, K003452 This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. Device Description
The Polaris X Unidirectional Steerable Diagnostic Catheter is a sterile, single use, unidirectional, steerable, diagnostic catheter intended for temporary use in electrophysiological studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
The catheter consists of a distal electrode segment, a shaft and a handle. The distal electrode segment consists of 10 electrodes (decapolar); 1 tip electrode that is 2 mm in length and 9 ring electrodes in a
{4}------------------------------------------------
variety of electrode spacings, as detailed in Table 1. The distal electrode segment is capable of forming a 270° staK21ndard curve configuration.
| UPN | Tip<br>Length | Tip<br>Size | Shaft<br>Size | Number of<br>Electrodes | Electrode<br>Spacing | Useable<br>Shaft<br>Length | Curve<br>Style |
|--------------|---------------|-------------|---------------|-------------------------|----------------------|----------------------------|------------------|
| M004 7000D 0 | 2 mm | 6F | 6F | 10 | 2.5 mm | 105 cm | 270°<br>Standard |
| M004 7001D 0 | | | | | 5 mm | | |
| M004 7003D 0 | | | | | 2.5/5/2.5 mm | | |
| M004 7004D 0 | | | | | 2/8/2 mm | | |
| M004 7005D 0 | | | | | 2/10/2 mm | | |
Table 1: Polaris X Catheter UPNs and Technical Descriptions
The shaft of the Polaris X catheter has a 6 French (F) outer diameter and a useable length of 105 cm. The proximal end of the shaft is connected to a handle, which contains a patented thumb-slide steering control that actuates the curve.
This catheter is compatible with most commercially available recording and mapping systems and connects to these systems via a cable, which is available separately.
#### V. Indications for Use
The catheter is intended for temporary use in electrophysiological studies for intracardiac stimulation (pacing) and/or recording of electrical potentials.
#### VI. Comparison of Technological Characteristics with the Predicate Device
The Polaris X Unidirectional Steerable Diagnostic Catheter incorporates substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device. An adhesive material change is the primary technological difference between the subject and predicate devices.
#### Performance Data VII.
The following performance data were provided in support of the substantial equivalence determination:
## Biocompatibility Testing
Biological testing was performed per Good Laboratory Practices (21 CFR. Part 58) with appropriate consideration of ISO 10993-1: 2018 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process and FDA Guidance Use of International Standard ISO 10993-1 (June 16, 2016) for devices categorized as externally communicating, limited contact, circulating blood contacting. The battery of testing included:
{5}------------------------------------------------
- Cytotoxicity .
- Intracutaneous Reactivity
- Pyrogen Testing .
- Genotoxicity, Mouse Lymphoma
- Hemolysis, Direct Contact ●
- . Platelet and Leukocyte Count
- Sensitization .
- . Acute Systemic Toxicity
- Genotoxicity, Bacterial Reverse . Mutation Study
- . Thrombogenicity, Partial Thromboplastin Time
- Hemolysis, Extract Method ●
- Complement Activation, SC5b-9 ●
# Ethylene Oxide Residual Testing
Ethylene oxide residual testing was conducted to verify acceptable residual limits per ISO 10993-7, Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals.
# Bacterial Endotoxin and Bioburden Testing
Bacterial endotoxin testing, per ANSI/AAMI ST72, Bacterial Endotoxins - Test Methods, Routine Monitoring, and Alternatives to Batch Testing, and bioburden testing, per ISO 11737-1, Microbiological Methods - Part 1: Determination of a Population of Microorganisms on Products, were performed to confirm endotoxin and bioburden levels of the catheter were not adversely affected.
# Design Verification Testing
Product specifications tested were:
- Reliability Cycle: Mechanical Integrity after Introduction / Withdrawal ●
- Reliability Cycle: Mechanical Integrity after Steering Lifecycle ●
- Twisting
# VIII. Conclusion
The subject Polaris X Unidirectional Steerable Diagnostic Catheter is substantially equivalent in indications for use, design, product function, materials, and sterility to the predicate Polaris X Unidirectional Steerable Diagnostic Catheter. The performance data, including biocompatibility, sterilization and design verification testing supports that the subject catheter is as safe and effective as its predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.