LASSO NAV DUO LOOP CATHETER

K130602 · Biosense Webster, Inc. · DRF · Aug 14, 2013 · Cardiovascular

Device Facts

Record IDK130602
Device NameLASSO NAV DUO LOOP CATHETER
ApplicantBiosense Webster, Inc.
Product CodeDRF · Cardiovascular
Decision DateAug 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Intended Use

The LASSO® NAV Duo Loop eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart. The LASSO® NAV Duo Loop eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.2.)

Device Story

The LASSO® NAV Duo Loop eco Catheter is a diagnostic electrophysiology catheter used to map atrial cardiac structures. It features a fixed spiral loop distal tip with 20 platinum/iridium electrodes for recording intracardiac signals and cardiac stimulation. The device is inserted into the right or left atrium via an 8F guiding sheath. It incorporates a tri-axial magnetic location sensor for integration with the CARTO® 3 EP Navigation System (v3.2 or later) to provide real-time location data. The catheter includes a bi-directional deflection mechanism for maneuverability. Physicians use the device to obtain electrograms, which are transmitted to external recording equipment and the navigation system for display and analysis. This information assists clinicians in identifying and diagnosing heart arrhythmias. The device benefits patients by facilitating precise anatomical and electrical mapping of the heart during electrophysiological procedures.

Clinical Evidence

Bench testing only. Validation included integrity and performance testing after ethylene oxide sterilization, thermal cycling, accelerated aging, and simulated transportation. Compatibility with the CARTO® 3 EP Navigation System was verified. No clinical data provided.

Technological Characteristics

7 Fr diameter catheter with 20 platinum/iridium ring electrodes in a fixed spiral loop configuration. Features tri-axial magnetic location sensor for navigation. Bi-directional deflection mechanism. Connects to CARTO® 3 EP Navigation System and standard recording equipment via accessory cables. Sterilized via ethylene oxide.

Indications for Use

Indicated for multiple electrode electrophysiological mapping (recording or stimulation) of cardiac structures, specifically atrial regions, in patients requiring cardiac arrhythmia diagnosis.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 2. 510(K) SUMMARY # K130602 This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT | Biosense Webster, Inc.<br>3333 Diamond Canyon Road<br>Diamond Bar, CA 91765 | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------| | OFFICIAL<br>CORRESPONDENT | John Jimenez<br>Senior Specialist, Regulatory Affairs<br>Telephone: 909-839-8534<br>Fax: 909-839-8804<br>Email: jjimene7@its.jnj.com<br>AUG 1 4 2013 | | TRADE NAME | LASSO® NAV Duo Loop eco Catheter | | COMMON NAME | Electrophysiological Mapping Catheter | | CLASSIFICATION NAME | Electrode recording catheter or electrode recording<br>probe | | DEVICE CLASSIFICATION | Class II, 21 CFR §870.1220<br>Product Code: DRF | | PRODUCT CODES | D-1354-01-S, D-1354-02-S, D-1354-03-S<br>D-1354-04-S, D-1354-05-S, D-1354-06-S | | PREDICATE DEVICE | Biosense Webster LASSO® 2515 NAV eco Variable<br>Catheter (510(k) K113213) | | REFERENCE DEVICE | Biosense Webster Webster® CS Catheter with EZ Steer<br>Technology (510(k) K101345) | #### SUBSTANTIALLY EQUIVALENT TO: The Biosense Webster LASSO®NA V Duo Loop eco Catheter is substantially equivalent to the Biosense Webster LASSO® 2515 NAV eco Variable Catheter [510(k) K113213, cleared December 5, 2011] and the Biosense Webster® CS Catheter with EZ Steer Technology [510(k) K101345, cleared May 13, 2010]. Like the predicate device, the LASSO®NAV Duo Loop eco Catheter features a tri-axial magnetic location sensor, allowing the ability to provide location information when used with the CARTO® 3 EP Navigation System, Version 3.2. The proposed catheter also has a similar circular loop of 20 electrodes at the distal tip of a 7 Fr diameter catheter, the same patient contact materials, and the same catheter connectivity as the predicate device. The original {1}------------------------------------------------ intended use of the predicate device for diagnosis of heart arrhythmias remains the same in the proposed device: however, the proposed device will now have the added ability of deflecting in two directions (bi-directional). This latter capability is substantially equivalent to the Reference Device, the Webster® CS Catheter with EZ Steer Technology. The Webster® CS Catheter with EZ Steer Technology went through this same transition with FDA when FDA cleared the update of the uni-directional version to the bi-directional version. # DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: The LASSO*NAV Duo Loop eco Catheter has been designed to facilitate electrophysiological mapping of the atria of the heart with the CARTO® 3 EP Navigation System and a reference device. The catheter is deployed in the right or left atrium through an 8F guiding sheath. The bi-directional catheter consists of a circular spine on the distal tip, with platinum/iridium electrodes that can be used for stimulation and recording. The purpose of this Premarket Notification is to create a bi-directional fixed multi-loop high density circular diagnostic catheter for recording intracardiac signals and cardiac stimulation. The modifications consist of changing from a single variable loop to a fixed spiral loop, including an insertion tool with atraumatic tip to the catheter, and incorporating a bi-directional deflection in two directions) feature to the distal end of the catheter. The new catheter will retain the extended barrel ("pig tail") connector and will connect to the system cable by way of an accessory cable. The LASSO® NAV Duo Loop eco Catheter features a fixed spiral loop design available in two diameters, 20mm diameter and 25mm diameter. Each catheter with 20 electrodes spaced along its multiple loops in a uni-polar configuration for the 20mm and either a uni-polar or bi-polar configuration for the 25mm. The proposed catheters will continue to interface with standard recording equipment via interface cables and appropriate connectors. # INDICATIONS FOR USE: The LASSO® NAV Duo Loop eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart. The LASSO® NAV Duo Loop eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO 3 EP Navigation Systems prior to Version 3.2.) # TECHNICAL CHARACTERISTICS: The LASSO® NAV Duo Loop eco Catheter is a typical electrophysiological catheter that is unique only in its geometrical arrangement of 20 ring electrodes spaced along its {2}------------------------------------------------ multiple loops. Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation System and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias. ### PERFORMANCE DATA: The safety and performance of the LASSO® NAV Duo Loop eco Catheter have been validated through Bench Testing, Biocompatibility Testing, GLP Animal Study, and System Level Testing. Bench testing evaluated the integrity and performance of the LASSO® NAV Duo Loop eco Catheter following three times ethylene oxide (EtO) sterilization, thermal cycling, accelerated aging (equivalent to one year real-time aging), and simulated transportation conditions. After thorough investigations and mitigations where appropriate, the device passed all intended criteria in accordance with appropriate test criteria and standards. In addition, the proposed catheter was tested with the CARTO® 3 EP Navigation System to demonstrate compatibility. Based on the results of the risk analysis and the bench testing performed, it is concluded that the minor differences in design and use that exist between the predicate device and the LASSO® NAV Duo Loop eco Catheter do not compromise the safety and efficacy of the catheter. This testing demonstrates that the subject device is at least as safe and effective as the legally marketed predicate devices. #### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE: The technological comparison and nonclinical studies demonstrate that the LASSO® NAV Duo Loop eco Catheter is safe and effective for anatomic mapping of the heart and establish equivalence to its predicate and reference devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around the perimeter. At the center of the seal is a stylized emblem, consisting of a white abstract design resembling an eagle or bird-like figure, set against a dark background. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 August 14, 2013 Biosense Webster, Inc. C/O John Jimenez 3333 Diamond Canyon Rd Diamond Bar, CA 91765 US Re: K130602 > Trade/Device Name: Lasso nav duo loop catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Regulatory Class: Class II Product Code: DRF Dated: July 1, 2013 Received: July 2, 2013 Dear John Jimenez: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading, If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ # Page 2 - John Jimenez or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Owen Farris -S for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 1. INDICATIONS FOR USE STATEMENT 510(k) No (if known): K130602 Device Name: LASSO® NAV Duo Loop eco Catheter Indication for Use: The LASSO® NAV Duo Loop eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart. The LASSO® NAV Duo Loop eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.2.) Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Digitally signed by Owen P. Faris -5 Date: 2013.08.14 08:30:52 -04'00' Page of of age
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...